Strengthening And Improvement Of Laboratory Facilities Of Awrmi ( Procurement Of Highly Equipped Laboratory Equipments / Machineries For Soil Laboratory, Concrete And Material Testing And Geo-Textile Laboratory ) Under Sopd-G Strengthening And Improve, guwahati-Assam

Water Resources Department-WRD has published Strengthening And Improvement Of Laboratory Facilities Of Awrmi ( Procurement Of Highly Equipped Laboratory Equipments / Machineries For Soil Laboratory, Concrete And Material Testing And Geo-Textile Laboratory ) Under Sopd-G Strengthening And Improve. Submission Date for this Tender is 15-06-2022. Clamps Tenders in guwahati Assam. Bidders can get complete Tender details and download the document.




Tender Notice

32399643
Strengthening And Improvement Of Laboratory Facilities Of Awrmi ( Procurement Of Highly Equipped Laboratory Equipments / Machineries For Soil Laboratory, Concrete And Material Testing And Geo-Textile Laboratory ) Under Sopd-G Strengthening And Improve
Open Tender
Indian
Assam
guwahati
15-06-2022

Tender Details

Strengthening And Improvement Of Laboratory Facilities Of Awrmi ( Procurement Of Highly Equipped Laboratory Equipments / Machineries For Soil Laboratory, Concrete And Material Testing And Geo-Textile Laboratory ) Under Sopd-G Strengthening And Improvement Of Laboratory Facilities Of Awrmi ( Procurement Of Highly Equipped Laboratory Equipments / Machineries For Soil Laboratory, Concrete And Material Testing And Geo-Textile Laboratory ) Under Sopd-G , Supply And Installation Melt Flow Index Tester For Testing Of Polypropylene Sample As Per Astm D1238. Band Heater For Precision Temperature Control ( + / - 0.1°C From Set Point ) Along The Entire Testing Area Of The Bore, Meeting The New Requirements Specified In Iso 1133-2 Must Be Offered. The Furnace Must Have A Quick Action Die Release Mechanism For Easy Removal Of The Die For Cleaning After A Test. The Basic Melt Indexer Should Be Used For Procedure A - Manual Operation Or For Procedure B - Automatically Timed Flow Rate Measurement. ( Detailed Specification Is Mentioned In The Next Sheet Please Look Into That Carefully. ) , Supply And Installation Consolidation Apparatus, Three Gang Bench Type Model, Analogue The Equipment Meets The General Requirements Of Is:2720 ( Part 15 ) And Is Designed For A Capacity Of 20 Kgf / Cm². Fixed Ring Consolidation Cell For 60Mm Dia Specimen Is Used In The Outfit. ( Detailed Specification Is Mentioned In The Next Sheet Please Look Into That Carefully. ) , Supply And Installation Laboratory Permeability Apparatus Ref:- Is -2720 ( Part-17 ) , This Standard ( Part 17 ) Covers The Methods For Laboratory Determination Of Coefficient Of Permeability Of Soils Using Falling Head And The Constant Head Methods. This Test Is Recommended For Soils With Coefficient Of Permeability In The Range 10^-3 To 10^-7 Cm / S And Maximum Particle Size Of 9.5 Mm. The Equipment Consist Of The Following Replaceable Parts:- Stand With Three Glass Tubes Of 6 Mm, 10 Mm And 20 Mm Dia Approx. Metallic Mould 100 Mm Dia X 127.3 Mm Height, 1000 Ml Volume With Metal Thickness 5Mm As Per Is :-11209-1985 Extension Collar 100 Mm Dia X 60 Mm Height. Drainage Base Plate With A Recess For Porous Stone And An Outlet Valve. Metallic Clamping Ring Drainage Cap With Recess For A Porous Stone And Fitted With Inlet Valve And Air Release Valve. Dummy Plate To Serve As False Bottom During Compaction. ( Detailed Specification Is Mentioned In The Next Sheet Please Look Into That Carefully. ) , Supply And Installation Overhead Tank, Made Of Steel, Approx, 37.5Cm Dia X 1M High , Supply And Installation Rammer 2.6Kg X 310Mm Controlled Drop , Supply And Installation Rammer 4.9 Kg X 450 Mm Controlled Drop. ( Aim-11102 ) , Supply And Installation Universal Testing Machine Computerized. Capacity, 1000 Kn, 6 Pillar Type With Hydraulic Jaws With Related Softwareandcomputer And Printer Machine Specification:- • Load Resolution :- 0.01Kn • Elongation Scale Resolution :- 0.1Mm • Extensometer Resolution:- 0.001Mm • Max Clearance For Tensile Testing :-50-650Mm • Max Clearance For Compression Testing :- 0-600Mm • Clearance Between Columns:- 650Mm • Ram Stroke :- 250Mm • Straining / Piston Speed At No Load:- 0-100Mm / Min • For Tension Test:- Clamping Jaws For Round Specimen :- 8-16Mm / 20-40Mm / 40-60Mm Clamping Jaws For Flat Specimen :- 0-40Mm Flat Specimen Width:- 70Mm • For Compression Test :- Pair Of Compression Plates Dia Of Circular Plates:- 233Mm • For Transverse Test:-Table With Adjustable Rollers Width Of Roller – 160Mm ; Diameter Of Roller -50Mm ( Detailed Specification Is Mentioned In The Next Sheet Please Look Into That Carefully. ) , Supply And Installation Digital Compression Testing Machine, Capacity 2000Kn , Compression Testing Machines Conform To Is: 14858 ( 2000 ) And Calibrated With An Accuracy Of ± 1% As Per The Requirement Of 1828 ( Class1 ) . Capacity Of Machine ( Kn ) :2000Kn Least Count ( Kn ) : 0.1Kn, Maximum Clearance Between Platens ( Mm ) :- 370Mm Maximum Distance Between Side Plates ( Mm ) :- 340Mm Platen Size ( Mm ) :-222.0Mm Dia Pistondia ( Mm ) :222.2Mm Piston Stroke ( Mm ) :-50Mm ( Detailed Specification Is Mentioned In The Next Sheet Please Look Into That Carefully. ) , Supply And Installation Universal Automatic Compactor With Mild Steel Compaction Mould As Per Ref Standard Is:2720 ( Part 7 & 8 ) , Supplied With ( Mould, 100Mm Dia X 127.3Mm Height 1, 000 Ml Volume ) :- The Compaction Mould Shall Be Made Ofbrass. It Shall Be Cylindrical In Shape And Finished Smooth The Mould Or Weided, Shall Have Two Eyes Either Cast Integral With The Body It Shall Have Suitable Seatings At The Top End For Positioning The Collar, Dimension:-100 Mm Id, 127.3 Mm Height , 1, 000 Ml Volume ) ( Detailed Specification Is Mentioned In The Next Sheet Please Look Into That Carefully. ) , Supply And Installation Jaw Faces For Tensile Grips: Grips For 100Kn Utm , Supply And Installation Hand Held Raman Analyzer Geo-Polymer Identification Through Hand Held Raman Spectrometer : Geo-Synthetic Identification Through Hand-Held Raman Spectroscopy Forlab And Field Application, No Prior Sample Preparation And Instant Result. ( Detailed Specification Is Mentioned In The Next Sheet Please Look Into That Carefully. )

Key Value

Document Fees
INR 2000 /-
EMD
INR 156323.0 /-
Tender Value
INR 78.16 Lakhs /-

BOQ Items

Name of Work: Strengthening and improvement of laboratory facilities of AWRMI (Procurement of highly equipped laboratory equipments/ machineries for Soil Laboratory, Concrete and Material Testing and GeoTextile Laboratory) under SOPD-G
Sl. No. Item Description
1SUPPLY AND INSTALLATION Melt Flow Index Tester for Testing of Polypropylene sample as per ASTM D1238. Band heater for precision temperature control (+/- 0.1°C from set point) along the entire testing area of the bore, meeting the new requirements specified in ISO 1133-2 must be offered. The furnace must have a quick action die release mechanism for easy removal of the die for cleaning after a test. The basic Melt Indexer should be used for Procedure A - Manual Operation or for Procedure B - Automatically Timed Flow Rate Measurement. (Detailed Specification is mentioned in the next sheet Please look into that carefully.)
2SUPPLY AND INSTALLATION Consolidation Apparatus, Three Gang Bench type Model, Analogue The equipment meets the general requirements of IS:2720 (Part 15) and is designed for a capacity of 20 kgf/cm². Fixed ring consolidation cell for 60mm dia specimen is used in the outfit. (Detailed Specification is mentioned in the next sheet Please look into that carefully.)
3SUPPLY AND INSTALLATION Laboratory Permeability Apparatus Ref:- IS -2720(Part-17),This standard ( Part 17 ) covers the methods for laboratory determination of coefficient of permeability of soils using falling head and the constant head methods. This test is recommended for soils with coefficient of permeability in the range 10^-3 to 10^-7 cm/s and maximum particle size of 9.5 mm. The equipment consist of the following replaceable parts:- Stand with three glass tubes of 6 mm,10 mm and 20 mm dia approx. Metallic Mould 100 mm dia x 127.3 mm height, 1000 ml volume with metal thickness 5mm as per IS :-11209-1985 Extension Collar 100 mm dia x 60 mm height. Drainage Base Plate with a recess for Porous Stone and an Outlet Valve. Metallic Clamping Ring Drainage Cap with recess for a Porous Stone and fitted with Inlet Valve and Air Release Valve. Dummy Plate to serve as False Bottom during compaction. (Detailed Specification is mentioned in the next sheet Please look into that carefully.)
4SUPPLY AND INSTALLATION Overhead Tank, made of steel, approx, 37.5cm dia x 1m high
5SUPPLY AND INSTALLATION Rammer 2.6kg x 310mm controlled drop
6SUPPLY AND INSTALLATION Rammer 4.9 kg x 450 mm controlled drop. (AIM-11102)
7SUPPLY AND INSTALLATION Universal Testing Machine Computerized. Capacity, 1000 kN, 6 pillar type with Hydraulic Jaws with related SoftwareandComputer and Printer Machine Specification:- • Load Resolution :- 0.01kN • Elongation scale Resolution :- 0.1mm • Extensometer Resolution:- 0.001mm • Max Clearance for Tensile Testing :-50-650mm • Max Clearance for Compression Testing :- 0-600mm • Clearance Between columns:- 650mm • Ram Stroke :- 250mm • Straining/Piston speed at No Load:- 0-100mm/min • For Tension Test:- Clamping Jaws for round Specimen :- 8-16mm/20-40mm/40-60mm Clamping jaws for flat specimen :- 0-40mm Flat Specimen Width:- 70mm • For Compression Test :- Pair of compression plates Dia of Circular plates:- 233mm • For Transverse Test:-Table with Adjustable Rollers Width of Roller – 160mm ; Diameter of Roller -50mm (Detailed Specification is mentioned in the next sheet Please look into that carefully.)
8SUPPLY AND INSTALLATION Digital Compression Testing Machine, Capacity 2000kN , Compression Testing Machines conform to IS: 14858(2000) and calibrated with an accuracy of ± 1% as per the requirement of 1828(Class1) . Capacity Of Machine(kN):2000kN Least Count(kN): 0.1kN, Maximum clearance between Platens(mm):- 370mm Maximum distance between side plates(mm):- 340mm Platen Size(mm):-222.0mm Dia PistonDia(mm):222.2mm Piston Stroke(mm):-50mm (Detailed Specification is mentioned in the next sheet Please look into that carefully.)
9SUPPLY AND INSTALLATION Universal Automatic Compactor with Mild Steel compaction mould as per Ref Standard IS:2720 (Part 7 & 8), supplied with (Mould, 100mm dia x 127.3mm height 1,000 ml volume):- The compaction mould shall be made ofBrass. It shall be cylindrical in shape and finished smooth The mould or weided, shall have two eyes either cast integral with the body It shall have suitable seatings at the top end for positioning the collar, Dimension:-100 mm ID, 127.3 mm height ,1,000 ml volume) (Detailed Specification is mentioned in the next sheet Please look into that carefully.)
10SUPPLY AND INSTALLATION Jaw faces for Tensile grips: Grips for 100kN UTM
11SUPPLY AND INSTALLATION Hand held Raman Analyzer Geo-polymer Identification through Hand Held RAMAN Spectrometer : Geo-synthetic identification through hand-held RAMAN Spectroscopy forlab and field application, no prior sample preparation and instant result. (Detailed Specification is mentioned in the next sheet Please look into that carefully.)