Tender For Cardiac Cath Lab With Ep Facility Model B
Tender Overview
Project Description
Tender for Cardiac Cath Lab With Ep Facility Model B , Cathlab Machine As Per The Specification Except The Following , Medical Grade Large High Definition Fhd Display 55 , Four Medical Grade High Definition Display Lcd/ Led Flat 19 Or Higher Monitors With Wide Viewing Angle, High Luminance High Contrast, Flicker Free, Distortion Free , Electrophysiology Application , Workstation As Per Spec , High Pressure Injector With 200 Syringes , Online Ups For Cathlab & Accessories , Console, Workstation, Lights And Computer , All In One Computer With Laser Printer , 12 Channel Ecg With Pc Connectivity , Full Wrap Around Lead Aprons , Skirt (With Fastening Belt)-Vest Type Lead Aprons , Front Full Type Lead Apron , Light Weight Leadless Single Piece Aprons With Front And Back Coverage , Light Weight Leadless Skirt And Vest With Fastening Belt , Thyroid Guard Lead Less , Leadless Head Gear Radiation Protection And Light Weight Lead Goggles , Lead Stockings , Lead Spectacles , Lead Gloves , Heavy Duty Stand With Hangers To Hold Atleast 5 Lead Aprons (Stainless Steel 304 Construction) , Wall Mounted Hangers To Hold Atleast 5Lead Aprons , Intercom Between Exam And Control Room , Instrument Table , Accessory Table , 15.2.1:Demolishing Cement Concrete Manually / By Mechanical Means Including Disposal Of Material Within 50 Metres Lead As Per Direction Of Engineer - In-Charge.Nominal Concrete 1:3:6 Or Richer Mix (I/C Equivalent Design Mix) , 15.12.1:Dismantling Doors, Windows And Clerestory Windows (Steel Or Wood ) Shutter Including Chowkhats, Architrave, Holdfasts Etc. Complete And Stacking Within 50 Metres Lead:Of Area 3 Sq. Metres And Below , 15.12.2:Dismantling Doors, Windows And Clerestory Windows (Steel Or Wood ) Shutter Including Chowkhats, Architrave, Holdfasts Etc. Complete And Stacking Within 50 Metres Lead:Of Area Beyond 3 Sq. Metres , 15.23.1:Dismantling Tile Work In Floors And Roofs Laid In Cement Mortar Including Stacking Material Within 50 Metres Lead.For Thickness Of Tiles 10 Mm To 25 Mm , 15.3:Demolishing Reinforced Cement Concrete Work Manually / By Mechanical Means Including Stacking Of Steel Bars And Disposal Of Unserviceable Material With In 50 Metres Lead As Per Direction Of Engineer -In-Charge. , 15.57:Dismantling Aluminium / Gypsum Partitions, Doors, Windows, Fixed Glazing And False Ceiling Including Disposal Of Unserviceable Material And Stacking Of Serviceable Material With In 50 Metres Lead As Directed By Engineer - In-Charge , 15.7.4:Demolishing Brick Work Manually / By Mechanical Means Including Stacking Of Serviceable Material And Disposal Of Unserviceable Material Within 50 Metres Lead As Per Direction Of Engineer-In-Charge.In Cement Mortar , 15.60:Disposal Of Building Rubbish/ Malba/ Similar Unserviceable, Dismantled Or Waste Materials By Mechanical Means, Including Loading, Transporting, Unloading To Approved Municipal Dumping Ground Or As Approved By Engineer-In-Charge, Beyond 50 M Initial Lead, For All Leads Including All Lifts Involved , 2.1.1:Earth Work In Surface Excavation Not Exceeding 30 Cm In Depth But Exceeding 1.5M In Width As Well As 10 Sqm On Plan Including Disposal Of Excavated Earth Up To 50 M And Lift Up To 1.5 M, Disposed Soil To Be Levelled And Neatly Dressed:All Kinds Of Soil , 2.6.1:Earth Work In Excavation By Mechanical Means (Hydraulic Excavator)/Manual Means Over Areas (Exceeding 30 Cm In Depth, 1.5 M In Width As Well As 10 Sqm On Plan) Including Getting Out And Disposal Of Excavated Earth Lead Upto 50 M And Lift Upto 1.5 M, As Directed By Engineer-In-Charge.All Kinds Of Soil , 50.2.25.1:Filling With Contractor'S Own Earth ( Excluding Rock) In Trenches, Plinth, Sides Of Foundations Etc. In Layers Not Exceeding 20 Cm In Depth, Consolidating Each Deposited Layer By Ramming And Watering, Lead Up To 50 M And Lift Up To 1.5 M As Per Direction Of Site Engineer-In-Charge , 2.34.1:Supplying Chemical Emulsion In Sealed Containers Including Delivery As Specified.Chlorpyriphos / Lindane Emulsifiable Concentrate Of 20% , 2.35.1.1:Diluting And Injecting Chemical Emulsion For Post -Constructional Anti-Termite Treatment (Excluding The Cost Of Chemical Emulsion): Along External Wall Where The Apron Is Not Provided Using Chemical Emulsion @ 7.5 Litres / Sqm Of The Vertical Surface Of The Substructure To A Depth Of 300Mm Including Excavation Channel Along The Wall & Rodding Etc. Complete:With Chlorpyriphos /Lindane E.C. 20% With 1% Concentration , 6.26.2:Brick Work With Common Burnt Clay Selected F.P.S (Non Modular ) Bricks Of Class Designation 7.5 In Exposed Brick Work Including Making Horizontal And Vertical Grooves 10 Mm Wide 12 Mm Deep Complete In Cement Mortar 1:6 ( 1 Cement : 6 Coarse Sand)Above Plinth Level And Upto Floor V Level , 50.6.1.2:Solid Block Masonry Using Pre Cast Solid Blocks (Factory Made) Of Size 40X20x20cm Or Nearest Available Size Confirming To Is 2185 Part I Of 1979 For Super Structure Up To Floor Two Level Thickness 20Cm And Above In: Cm 1:6 ( 1 Cement: 6 Coarse Sand) Etc Complete. , 50.6.3.2:Solid Block Masonry Using Pre Cast Solid Blocks (Factory Made) Of Size 40X20x10 Cm Or Nearest Available Size Confirming To Is 2185 Part I Of 1979 For Super Structure Up To Floor Two Level For 10 Cm Thick Wall In : Cm 1:6 ( 1 Cement : 6 Coarse Sand) Including Cost Of Scaffolding Complete , 50.6.1.8:Solid Block Masonry Using Pre Cast Solid Blocks (Factory Made) Of Size 40X20x15cm Or Nearest Available Size Confirming To Is 2185 Part I Of 1979 For Super Structure Up To Floor Two Level With Thickness 15Cm In: Cm 1:6 ( 1 Cement : 6 Coarse Sand Etc Complete , 4.1.3:Providing And Laying In Position Cement Concrete Of Specified Grade Excluding The Cost Of Centering And Shuttering - All Work Up To Plinth Level:1:2:4 (Cement : 2 Coarse Sand (Zone-Iii)Derived From Natural Sources : 4 Graded Stone Aggregate 20 Mm Nominal Size Derived From Natural Sources) , 4.10A: Providing And Laying Damp-Proof Course 40Mm Thick With Cement Concrete 1:2:4 (1 Cement : 2 Coarse Sand (Zone-Iii) Derived From Natural Sources : 4 Graded Stone Aggregate 12.5Mm Nominal Size Derived From Natural Sources).Providing And Laying Damp-Proof Course 40Mm Thick With Cement Concrete 1:2:4 (1 Cement : 2 Coarse Sand (Zone-Iii) Including Manufactured Sand Derived From Recycled Concrete Aggregate (Rca) Upto 25% : 4 Graded Stone Aggregate 12.5Mm Nominal Size Including (Recycled Concrete Aggregate (Rca) Upto 25%)) , 5.3:Reinforced Cement Concrete Work In Beams, Suspended Floors, Roofs, Having Slope Up To 15° Landings, Balconies, Shelves, Chajjas, Lintels, Bands, Plain Window Sills, Staircases And Spiral Stair Cases Up To Floor Five Level Excluding The Cost Of Centering, Shuttering, Finishing And Reinforcement, With1:1.5:3 (1 Cement : 1.5 Coarse Sand (Zone Iii) Derived From Natural Sources: 3 Graded Stone Aggregate 20 Mm Nominal Size Derived From Natural Sources). , 5.22.1:Steel Reinforcement For Reinforced Cement Concrete Work Including Straightening, Cutting, Bending, Placing In Position And Binding All Complete Upto Plinth Levelmild Steel And Medium Tensile Steel Bars , 5.9.5:Centering And Shuttering Including Strutting, Etc. And Removal Of Form For:Lintels, Beams, Plinth Beams, Girders Bressumers And Cantilevers , 50.4.1.1:Providing And Laying In Position Cement Concrete 1:2:4 ( 1 Cement : 2 Coarse Sand : 4 Graded Stone Aggregate 12.5 Mm Nominal Size) Up To Floor Five Level Excluding The Cost Of Centering, Shuttering, Finishing For Leveling Coarse Of Floor And Roof Etc. , 50.5.33.2:Providing And Laying In Position Machine Batched And Machine Mixed Design Mix M-20 Grade Cement Concrete For Reinforced Cement Concrete Work, Using Cement Content As Per Approved Design Mix, Including Pumping Of Concrete To Site Of Laying But Excluding The Cost Of Centering, Shuttering, Finishing And Reinforcement, Including Admixtures In Recommended Proportions As Per Is: 9103 To Accelerate, Retard Setting Of Concrete, Improve Workability Without Impairing Strength And Durability As Per Direction Of Engineer-In-Charge. Note:- Cement Content Considered In This Item Is @ 330 Kg/Cum. Excess Or Less Cement Used As Per Design Mix Is Payable Or Recoverable Separately. All Work Above Plinth Level Upto Floor V Level , 50.5.5:Reinforced Cement Concrete Work In Arches, Archribs, Domes, Vaults, Shells, Folded Plate And Roof Having Slope More Than 15 Degree Up To Floor Five Level Excluding The Cost Of Centering, Shuttering, Finishing And Reinforcement With 1:1.5:3 ( 1 Cement : 1.5 Coarse Sand: 3 Graded Stone Aggregate 20 Mm Nominal Size). , 5.9.2:Centering And Shuttering Including Strutting, Etc. And Removal Of Form For:Walls (Any Thickness) Including Attached Pilasters, Butteresses, Plinth And String Courses Etc. , 5.9.3:Centering And Shuttering Including Strutting, Etc. And Removal Of Form For:Suspended Floors, Roofs, Landings, Balconies And Access Platform , 5.9.6:Centering And Shuttering Including Strutting, Etc. And Removal Of Form For:Columns, Pillars, Piers, Abutments, Posts And Struts , 13.1.1:12 Mm Cement Plaster Of Mix:1:4 ( 1 Cement : 4 Fine Sand) , 13.7.1:12 Mm Cement Plaster Finished With A Floating Coat Of Neat Cement Of Mix:1:3 ( 1 Cement : 3 Fine Sand) , 9.141:Providing And Fixing Pvc Door Frame Of Size 50X47 Mm With A Wall Thickness Of 5 Mm (±0.2Mm), Made Out Of Single Piece Extruded Pvc Profile, With Mitred Cut Joints And With 2 Nos Of Pvc Bracket Of Size 190 Mm X 100 Mm Long Arms Of Cross Section Size 35 X 15 Mm & Self Driven Self Taping Screws, The Vertical Door Profiles To Be Reinforced With 40X20 Mm M.S. Rectangular Tube Of 0.8 Mm, Including Providing Epdm Rubber Gasket Weather Seal Throughout The Frame, Including Jointing 5 Mm Pvc Frame Strip With Pvc Solvent Cement On The Back Of The Profile. The Door Frame To Be Fixed To The Wall Using 8X100 Mm Long Anchor Fasteners Complete, All As Per Manufacturers Specification And Direction Of Engineer-In--Charge. , 9.142.2:35 Mm Thick Factory Made Solid Panel Pvc Door Shutter, Made Out Of Singledecorative Finish (Wood Grained Finish) , 21.1.1.2:Providing And Fixing Aluminium Work For Doors, Windows, Ventilators And Partitions With Extruded Built Up Standard Tubular Sections/ Appropriate Z Sections And Other Sections Of Approved Make Conforming To Is : 733 And Is: 1285, Fixing With Dash Fasteners Of Required Dia And Size, Including Necessary Filling Up The Gaps At Junctions, I.E. At Top, Bottom And Sides With Required Epdm Rubber/ Neoprene Gasket Etc. Aluminium Sections Shall Be Smooth, Rust Free, Straight, Mitred And Jointed Mechanically Wherever Required Including Cleat Angle, Aluminnium Snap Beading For Glazing /Paneling, C.P. Brass/ Stainless Steel Screws, All Complete As Per Architectural Drawings And The Directions Of Engineer-In-Charge.(Glazing, Paneling And Dash Fasteners To Be Paid For Separately): For Fixed Portionpowder Coated Aluminium (Minimum Thickness Of Powder Coating 50 Micron) , 21.1.2.2:For Shutters Of Doors, Windows & Ventilators Including Providing And Fixing Hinges / Pivots And Making Provision For Fixing Of Fittings Wherever Required Including The Cost Of Epdm Rubber/ Neoprene Gasket Required (Fittings Shall Be Paid For Separately)Powder Coated Aluminium (Minimum Thickness Of Powder Coating 50 Micron) , 21.13:Providing And Fixing Brass100 Mm Mortice Latch And Lock With 6 Levers Without Pair Of Handles ( Best Make Of Approved Quality ) For Aluminum Doors Including Necessary Cutting And Making Good Etc. Complete , 21.2.2:Providing And Fixing 12 Mm Thick Prelaminated Particle Board Flat Pressed Three Layer Or Graded Wood Particle Board Conforming To Is : 12823 Grade I Type Ii, In Panelling Fixed In Aluminum Doors, Windows Shutters And Partition Frames With C.P. Brass / Stainless Steel Screws Etc. Complete As Per Architectural Drawings And Directions Of Engineer - In- Charge.Pre- Laminated Particle Board With Decorative Lamination On Both Sides , 21.3.1:Providing And Fixing Glazing In Aluminium Door, Window, Ventilator Shutters And Partitions Etc. With Epdm Rubber / Neoprene Gasket Etc. Complete As Per The Architectural Drawings And The Directions Of Engineer - In -Charge. ( Cost Of Aluminium Snap Beading Shall Be Paid In Basic Item):With Float Glass Panes Of 4.0 Mm Thickness (Weight Not Less Than 10Kg/Sqm) , 9.83:Providing And Fixing Aluminium Die Cast Body Tubular Type Universal Hydraulic Door Closer (Having Brand Logo With Isi, Is : 3564, Embossed On The Body, Door Weight Upto 35 Kg And Door Width Upto 700 Mm), With Necessary Accessories And Screws Etc. Complete. , 12.45.2:Providing And Fixing False Ceiling At All Height Including Providing And Fixing Of Frame Work Made Of Special Sections, Power Pressed From M.S. Sheets And Galvanized With Zinc Coating Of 120 Gms/ Sqm (Both Side Inclusive) As Per Is : 277 And Consisting Of Angle Cleats Of Size 25 Mm Wide X 1.6 Mm Thick With Flanges Of 27 Mm And 37 Mm, At 1200 Mm Centre To Centre, One Flange Fixed To The Ceiling With Dash Fastener 12.5 Mm Dia X 50 Mm Long With 6 Mm Dia Bolts, Other Flange Of Cleat Fixed To The Angle Hangers Of 25X10x0.50 Mm Of Required Length With Nuts & Bolts Of Required Size And Other End Of Angle Hanger Fixed With Intermediate G.I. Channels 45X15x0.9 Mm Running At The Spacing Of 1200 Mm Centre To Centre, To Which The Ceiling Section 0.5 Mm Thick Bottom Wedge Of 80 Mm With Tapered Flanges Of 26 Mm Each Having Lips Of 10.5 Mm, At 450 Mm Centre To Centre, Shall Be Fixed In A Direction Perpendicular To G.I. Intermediate Channel With Connecting Clips Made Out Of 2.64 Mm Dia X 230 Mm Long G.I. Wire At Every Junction, Including Fixing Perimeter Channels 0.5 Mm Thick 27 Mm High Having Flanges Of 20 Mm And 30 Mm Long, The Perimeter Of Ceiling Fixed To Wall/Partition With The Help Of Rawl Plugs At 450 Mm Centre, With 25 Mm Long Dry Wall Screws @ 230 Mm Interval, Including Fixing Of Gypsum Board To Ceiling Section And Perimeter Channel With The Help Of Dry Wall Screws Of Size 3.5X25 Mm At 230 Mm C/C, Including Jointing And Finishing To A Flush Finish Of Tapered And Square Edges Of The Board With Recommended Jointing Compound, Jointing Tapes, Finishing With Jointing Compound In 3 Layers Covering Upto 150 Mm On Both Sides Of Joint And Two Coats Of Primer Suitable For Board, All As Per Manufacturers Specification And Also Including The Cost Of Making Openings For Light Fittings, Grills, Diffusers, Cutouts Made With Frame Of Perimeter Channels Suitably Fixed, All Complete As Per Drawings, Specification And Direction Of The Engineer In Charge But Excluding The Cost Of Painting With:12.5 Mm Thick Tapered Edge Glass Reinforced Gypsum (Grg) Board Conforming To Is: 2095- (Part 3):1996 (Boards With Bis Certification Marks) , 12.54.1:Providing And Fixing Gi Clip In Metal Ceiling System Of 600X600 Mm Module Which Includes Providing And Fixing C Wall Angle Of Size 20X30x20 Mm Made Of 0.5 Mm Thick Pre Painted Steel Along The Perimeter Of The Room With Help Of Nylon Sleeves And Wooden Screws At 300 Mm Center To Centre, Suspending The Main C Carrier Of Size 10X38x10 Mm Made Of G.I Steel 0.7 Mm Thick From The Soffit With Help Of Soffit Cleat 37X27x25x1.6 Mm, Rawl Plugs Of Sixe38x12 Mm And C Carrier Suspension Clip And Main Carrier Bracket At 1000 Mm C/C. Inverted Triangle Shaped Spring Tee Having Height Of 24 Mm And Width Of 34 Mm Made Of Gi Steel 0.45 Mm Thick Is Then Fixed To The Main C Carrier And In Direction Perpendicular To It At 600 Mm Centers With Help Of Suspension Brackets. Wherever The Main C Carrier And Spring T Have To Join, C Carrier And Spring T Connectors Have To Be Used. All Sections To Be Galvanized @ 120 Gms/Sqm (Both Side Inclusive), Fixing With Clip In Tiles Into Spring T With :Gi Metal Ceiling Clip In Plain Beveled Edge Global White Color Tiles Of Size 600X600 And 0.5Mm Thick With 25 Mm Height, ,Made Of G.I Sheet Having Galvanizing Of 100 Gms/ Sqm (Both Sides Inclusive) And Electro Statically Polyester Powder Coated Of Thickness 60 Microns (Minimum), Including Factory Painted After Bending . , 12.58.1:Providing And Fixing Tiled False Ceiling Of Approved Materials Of Size 595X595 Mm In True Horizontal Level, Suspended On Interlocking Metal Grid Of Hot Dipped Galvanized Steel Sections (Galvanized @ 120 Grams/ Sqm, Both Side Inclusive) Consisting Of Main T Runner With Suitably Spaced Joints To Get Required Length And Of Size 24X38 Mm Made From 0.30 Mm Thick (Minimum) Sheet, Spaced At 1200 Mm Center To Center And Cross T Of Size 24X25 Mm Made Of 0.30 Mm Thick (Minimum) Sheet, 1200 Mm Long Space Between Main T At 600 Mm Center To Center To Form A Grid Of 1200X600 Mm And Secondary Cross T Of Length 600 Mm And Size 24X25 Mm Made Of 0.30 Mm Thick (Minimum) Sheet To Be Interlocked At Middle Of The 1200X600 Mm Panel To Form Grids Of 600X600 Mm And Wall Angle Of Size 24X24x0.3 Mm And Laying False Ceiling Tiles Of Approved Texture In The Grid Including, Required Cutting/Making, Opening Of Services Like Diffusers, Grills, Light Fittings, Fixtures, Smoke Detectors Etc. Main T Runners To Be Suspended From Ceiling Using Gi Slotted Cleats Of Size 27 X 37 X 25 X 1.6 Mm Fixed To Ceiling With 12.5 Mm Dia And 50 Mm Long Dash Fasteners, 4 Mm Gi Adjustable Rods With Galvanized Butterfly Level Clips Of Size 85 X 30 X 0.8 Mm Spaced At 1200 Mm Center To Center Along Main T, Bottom Exposed Width Of 24 Mm Of All T-Sections Shall Be Pre-Painted With Polyester Paint, All Complete For All Heights As Per Specifications, Drawings And As Directed By Engineer-In-Charge.8 Mm Thick Fully Perforated Calcium Silicate Board Made With Calcareous & Siliceous Materials Reinforced With Cellulose Fiber Manufactured Through Autoclaving Process To Give Stable Crystalline Structure With Minimum Compressive Strength 225 Kg/ Sq. Cm, Bending Strength 100 Kg/Sq. Cm, Of Size 595X595 Mm, Having Perforation Of Dia, 10 Mm With Minimum Perforated Area 18% With Non Woven Tissue On The Back Side, Having An Nrc (Noise Reduction Coefficient) Of 0.85, With 50 Mm Thick Rockwool Of 48 Kg/Cum Backing. , Od155583/2026-2027:Providing And Fixing Trap Doors For Access To The Areas Above False Ceiling. The Trap Door Shall Have An External Frame Made Of Powder Coated Aluminium Channel On Which The Shutter Is Hinged. Shutter Shall Be Made 4 Mm Thick Acp Sheet Shutter External Frame Of Shutter Made Of Powder Coated Aluminium Channel. Concealed Heavy Duty Hinges Shall Be Used To Mount The Shutters And Locking Shall Be Provided. The Shutter Shall Be Moulded In Level With The False Ceiling Forming A Groove All Around. Sufficient Number Of Hinges And Locks Shall Be Provided To Avoid Sagging Of The Shutter. Key Also Shall Be Supplied. No Extra Rate Shall Be Provided For Suspenders Or Frame Works All Round For Linking The Trap Door To Gyp Board /Any Other False Ceiling. The Rate Including All Labour Charges ,Material Charges Ect Complete.Size 1.20 M X 0.6 M , Od155584/2026-2027:Providing And Fixing Factory-Made Sandwich Panels (Puf Core) For Iso Wall With 40 Mm Puf Double Skin With All Necessary Support Channels And Accessories Etc Wall And Partition Work, Made Of 0.4 Mm Thick Precoated Galvanized Iron Sheets On Top Side And 0.3Mm Sheet At Bottom Side With 40 Mm Thick Polyurethane Foam Core (Density 40 Kg/M³), Having Tongue And Groove Joints With Cam-Lock Mechanism, Including Necessary Framing, Anchor Fasteners, Joint Sealing With Silicone Sealant, And Complete In All Respects As Per ManufacturerS Specifications. , Od155581/2026-2027:Providing And Fixing Sun Controlled Film On Glass Surface For Door/Windows,After Thoroughly Cleaning And Dust Free The Surface To Get Smooth Finish,To Get Air Tight Clean Surface, Ect,Complete As Directed By Engineer-In -Charge , Od155582/2026-2027: Providing And Fixing Vertical Blinds As Per The Direction Of Engineer In Charge , Od155569/2026-2027:Providing And Laying 2 Mm Thick Anti Static Poly Vinyl Flooring Of Laid Over 25 Mm Thick Leveling Layer Of Cement Concrete 1:2:4 (1 Cement : 2 Coarse Sand : 4 Graded Stone Aggregate 10 Mm Nominal Size) To Be Paid Separately, Coated With A Thin Layer Of Hot Bitumen Penetration 80/25 (Blown Type) @ 2.45 Kg Per Sqm, Including Fixing Blocks In Position After Dipping In Hot Bitumen (Blown Type) Up To Half Depth, Planed, Levelled Smooth And Finished Complete. , 8.2.2.2:Providing And Fixing 18 Mm Thick Gang Saw Cut, Mirror, Polished, Premoulded And Prepolished, Machine Cut For Kitchen Platforms, Vanity Counters, Window Sills, Facias And Similar Locations, Of Required Size, Approved Shade, Colour And Texture Laid Over 20 Mm Thick Base Cement Mortar 1:4 ( 1 Cement : 4 Coarse Sand), Joints Treated With White Cement, Mixed With Matching Pigment, Epoxy Touch Ups, Including Rubbing, Curing, Moulding And Polishing To Edges To Give High Gloss Finish Etc. Complete At All Levels.Area Of Slab Over 0.50 Sqm , 11.37:Providing And Laying Ceramic Glazed Floor Tiles Of Size 300X300 Mm (Thickness To Be Specified By The Manufacturer) Of 1St Quality Conforming To Is : 15622 Of Approved Make In Colours Such As White, Ivory, Grey, Fume Red Brown, Laid On 20 Mm Thick Cement Mortar 1:4 (1 Cement : 4 Coarse Sand), Jointing With Grey Cement Slurry @ 3.3 Kg/Sqm Including Pointing The Joints With White Cement And Matching Pigment Etc., Complete , 11.47.2:Providing And Laying Vitrified Tiles In Different Sizes (Thickness To Be Specified By The Manufacturer), With Water Absorption Less Than 0.08% And Conforming To Is : 15622, Of Approved Brand & Manufacturer,In All Colours And Shade, In Skirting, Riser Of Steps, Laid With Cement Based High Polymer Modified Quick Set Tile Adhesive (Water Based) Conforming To Is : 15477, In Average 6 Mm Thickness, Including Grouting Of Joints (Payment For Grouting Of Joints To Be Made Separately).Size Of Tile 600X600 Mm , 11.49.2:Providing And Laying Vitrified Tiles In Floor With Different Sizes (Thickness To Be Specified By The Manufacturer), With Water Absorption Less Than 0.08% And Conforming To Is : 15622, Of Approved Brand & Manufacturer, In All Colours And Shade, Laid With Cement Based High Polymer Modified Quick Set Tile Adhesive (Water Based) Conforming To Is : 15477, In Average 6 Mm Thickness, Including Grouting Of Joints (Payment For Grouting Of Joints To Be Made Separately).Size Of Tiles 600X600 Mm , 11.48.2:Grouting The Joints Of Flooring Tiles Having Joints Of 3 Mm Width, Using Epoxy Grout Mix 0.70 Kg Of Organic Coated Filler Of Desired Shade (0.10 Kg Of Hardener And 0.20 Kg Of Resin Per Kg). Including Filling /Grouting And Finishing Complete As Per Direction Of Engineer-In-Charge.Size Of Tile 600X600 Mm , 13.43.1:Applying One Coat Of Water Thinnable Cement Primer Of Approved Brand And Manufacture On Wall Surface:Water Thinnable Cement Primer , 13.80:Providing And Applying White Cement Based Putty Of Average Thickness 1 Mm, Of Approved Brand And Manufacturer, Over The Plastered Wall Surface To Prepare The Surface Even And Smooth Complete. , 13.82.2:Wall Painting With Acrylic Emulsion Paint, Having Voc (Volatile Organic Compound) Content Less Than 50 Grams/ Litre, Of Approved Brand And Manufacture Including Applying Additional Coats Wherever Required, To Achieve Even Shade And Colour.Two Coats , 13.46.1:Finishing Walls With Acrylic Smooth Exterior Paint Of Required Shade:New Work (Two Or More Coat Applied @ 1.67 Ltr/10 Sqm Over And Including Priming Coat Of Exterior Primer Applied @ 2.20 Kg/10 Sqm) , 13.61.1:Painting With Synthetic Enamel Paint Of Approved Brand And Manufacture To Give An Even Shade:Two Or More Coats On New Work , 13.69.1:Polishing On Wood Work With Ready Mixed Wax Polish Of Approved Brand And Manufacture:New Work , 17.11.1:Providing And Fixing White Vitreous China Laboratory Sink With C.I Brackets, C.P. Brass Chain With Rubber Plug 40 Mm C.P Brass Waste And 40 Mm C.P. Brass Trap With Necessary C.P. Brass Unions Complete, Including Painting Of Fittings And Brackets, Cutting And Making Good The Wall Wherever Required:Size 450X300x150 Mm , 17.2.2:Providing And Fixing White Vitreous China Pedestal Type Water Closet (European Type W.C. Pan) With Seat And Lid, 10 Litre Low Level White P.V.C. Flushing Cistern, Including Flush Pipe, With Manually Controlled Device (Handle Lever), Conforming To Is : 7231, With All Fittings And Fixtures Complete, Including Cutting And Making Good The Walls And Floors Wherever Required:W.C. Pan With Isi Marked Black Solid Plastic Seat And Lid , 17.28.1.1:Providing And Fixing P.V.C. Waste Pipe For Sink Or Wash Basin Including Pvc. Waste Fittings Complete. Semi Rigid Pipe32 Mm Dia , 17.32.1:Providing And Fixing Mirror Of Superior Glass ( Of Approved Quality) And Of Required Shape And Size With Plastic Moulded Frame Of Approved Make And Shade With 6 Mm Thick Hard Board Backing :Circular Shape 450 Mm Dia , 17.7.1:Providing And Fixing Wash Basin With C.I. Brackets, 15 Mm C.P. Brass Pillar Taps, 32 Mm C.P. Brass Waste Of Standard Pattern, Including Painting Of Fittings And Brackets, Cutting And Making Good The Walls Wherever Require:White Vitreous China Wash Basin Size 630X450 Mm With A Pair Of 15 Mm C.P. Brass Pillar Taps , 17.73.2:Providing And Fixing Ptmt Towel Rail Complete With Brackets Fixed To Wooden Cleats With Cp Brass Screws With Concealed Fitting Arrangement Of Approved Quality And Colour600 Mm Long Towel Rail With Total Length Of 645 Mm, Width 78 Mm And Effective Height Of 88 Mm, Weighting Not Less Than 190 Gms , 50.17.1.5:Supplying And Fixing Cp Health Faucet Superior Quality ( Jagur Or Equvalent Make ) Including Cost Of Materials And Labour Charges Etc Complete As Per The Direction Of Site Engineer-In-Charge. , Od155570/2026-2027:Providing And Fixing 15 Mm C.P. Brass Tap With Elbow Operation Lever Of Approved Quality Conforming To Is Standards And Weighing Not Less Than 690 Gms. 15 Mm Nominal Bore , 18.65:Providing And Fixing Ptmt Soap Dish Holder Having Length Of 138 Mm, Breadth 102 Mm, Height Of 75 Mm With Concealed Fittings Arrangements, Weighing Not Less Than 106 Gms. , 18.15.1:Providing And Fixing Brass Bib Cock Of Approved Quality :15 Mm Nominal Bore , 50.18.7.1.1:Providing And Fixing Pvc Pipes, Fittings Including Fixing The Pipe With Clamps At 1.00 M Spacing. This Includes Jointing Of Pipes & Fittings With One Step Pvc Solvent Cement And Testing Of Joints Complete As Per Direction Of Engineer-In-Charge 15Mm Dia 12Kgf/Cm2 -Internal Work- Exposed On Wall , 50.18.7.4.1:Providing And Fixing Pvc Pipes, Fittings Including Fixing The Pipe With Clamps At 1.00 M Spacing. This Includes Jointing Of Pipes & Fittings With One Step Pvc Solvent Cement And Testing Of Joints Complete As Per Direction Of Engineer-In-Charge 32 Mm Dia 10Kgf/Cm2- Internal Work - Exposed On Wall , 50.18.8.1.1:Providing And Fixing Pvc Pipes, Fittings Including Fixing The Pipe With Clamps At 1.00 M Spacing. This Includes Jointing Of Pipes & Fittings With One Step Pvc Solvent Cement And Testing Of Joints Complete As Per Direction Of Engineer- In- Charge. Concealed Work, Including Cutting Chases And Making Good The Wall Etc. 15 Mm Pipe 12 Kgf/ Cm2 , 50.18.8.4.1:Providing And Fixing Pvc Pipes, Fittings Including Fixing The Pipe With Clamps At 1.00 M Spacing. This Includes Jointing Of Pipes & Fittings With One Step Pvc Solvent Cement And Testing Of Joints Complete As Per Direction Of Engineer-In-Charge. Concealed Work, Including Cutting Chases And Making Good The Wall Etc. 32 Mm Pipe 10Kgf/Cm2 , 50.18.8.8.2:Providing And Fixing Pvc Pipes, Fittings, Including Fixing The Pipe With Clamps At 1.00 M Spacing. This Includes Jointing Of Pipes With One Step Pvc Solvent Cement And Testing Of Joints Complete As Per The Direction Of Engineer-In-Charge. Concealed Work, Including Cutting Chases And Making Good The Wall Etc. 75 Mm Pipe 4 Kgf/Cm2 , 50.18.8.9.1:Providing And Fixing Pvc Pipes, Fittings Including Fixing The Pipe With Clamps At 1.00 M Spacing. This Includes Jointing Of Pipes With One Step Pvc Solvent Cement And Testing Of Joints Complete As Per Direction Of Engineer-In-Charge. Concealed Work, Including Cutting Chased And Making Good The Wall Etc. 110 Mm Pipe 6Kgf/Cm2 , 50.9.1.1:Providing Wood Work In Frames Of Doors, Windows, Clerestory Windows And Other Frames, Wrought Framed And Fixed In Position With Hold Fast Lugs Or With Dash Fasteners Of Required Dia & Length (Hold Fast Lugs Or Dash Fastener Shall Be Paid For Separately), Using Good Quality Anjili Wood /Jack Wood , 9.21.2:Providing And Fixing Isi Marked Flush Door Shutters Conforming To Is : 2202 (Part I) Non- Decorative Type, Core Of Block Board Construction With Frame Of 1St Class Hard Wood And Well Matched Commercial 3 Ply Veneering With Vertical Grains Or Cross Bands And Face Veneers On Both Faces Of Shutters:30 Mm Thick Including Isi Marked Stainless Steel Butt Hinges With Necessary Screws , Od155577/2026-2027:Supplying And Lining 2 Mm Thick Radiation Shielding Lead Sheet Conforming To Is 2112 Or Equivalent International Standards, For Doors And Frames, Ensuring Overlap Between Sheets, Joints Staggered And Sealed To Prevent Radiation Leakage. Lead Sheet To Be Properly Glued With Suitable Contact Adhesive, Fixed On Dooe Frame/Shutter As Per Approved Design. Door Frame And Door Shutter Paid Separately .All Work To Be Done As Per Regulatory Body Requirements (Like Aerb Guidelines). , 9.127.2:Providing & Fixing Decorative High Pressure Laminated Sheet Of Plain / Wood Grain In Gloss/ Matt/Suede Finish With High Density Protective Surface Layer And Reverse Side Of Adhesive Bonding Quality Conforming To Is: 2046 Types S, Including Cost Of Adhesive Of Approved Quality.1.0 Mm Thick , 9.140.1:Providing And Fixing Plain Lining With Necessary Screws/Nuts/& Bolts/Nails, Including A Coat Of Approved Primer On One Face, And Fixed On Wooden /Steel Frame Work, Complete As Per Direction Of Engineer-In-Charge ( Frame Work Shall Be Paid For Separately)12 Mm Thick Commercial Ply Conforming To Is: 1328 Bwr Type , 9.169:Providing And Fixing 19Mm Thick Both Side Balancing Lamination Factory Pressed Bwp Grade Marine Ply As Per Is 710 Of Approved Brand Boxes, Shelves, Racks, Almirah, Cupboard And Drawer Etc. Including Necessary Nails, Screws Etc. Complete As Per Direction Of Engineer-In-Charge. , 9.170.1:Providing And Fixing Stainless Steel Fancy Handle Of Approved Make Fixed With Ss Screws Etc. Complete As Per Direction Of Engineer-In-Charge.200 Mm , 9.171:Providing And Fixing Stainless Steel Soft Closing Spring Hinges At 0 Degree Hinges (Hydraulic Type) Of Approved Make/Brand To Cupboard Shutters With Full Threaded Steel Screws Including Making Necessary Recess In Board And Finished Etc. Complete As Per Direction Of Engineer-In-Charge. , 14.34:Providing And Fixing 150 Mm Bright Finished Floor Brass Door Stopper With Rubber Cushion, Necessary Brass Screws Etc. To Suit Shutter Thickness Complete , 9.72.3:Providing And Fixing Bright Finished Brass Butt Hinges With Necessary Screws Etc. Complete:100X85x5.5 Mm (Heavy Type) , 9.74.3:Providing And Fixing Bright Finished Brass Tower Bolts (Barrel Type) With Necessary Screws Etc. Complete:150X10 Mm , Od155580/2026-2027:Supply And Fixing Solid Steel Under Desk Keyboard Tray Slide Out With 360° Rotating Mouse Platform, Keyboard Shelf Sliding , Od155576/2026-2027:File Cabinet 465 W X 620 D X 1350 H(4 Cabinet Type Godrej Or Equivalent) , Od155573/2026-2027:Shoe Rack Three Shelves : Pretreated And Powder Coated, Three Shelves, Made Of Cold Rolled Ms Sheet Of 16 Guages, Six Cold Rolled Ms Tube Frame Levels As Support For 6 Pairs Of Shoes In One Shelf, Abrasion Resistant, Non-Discoloring, Corrosion Resistant , Od155575/2026-2027:Computer Table 1500*600 With Keyboard Tray, Cpu Storage At Base In A Side And Storage Drawers At Other Side., To Be Made Out Of 25Mm Thick Top & Whereas Body, Back & Shutter In 18Mm Prelaminated Particle Board With Pvc Edge Band On All Sides & . Handle &Quot;C&Quot; Type Ss Brush Finish (For Workstation And Reports) Reputed Brand Godrej,Wipro Or Equivalent , Od155574/2026-2027:Supplying Of Steel Cupboardfor Storage 6Ft Tall 3 Feet Width, Door With Glass Transparent Doors With Godrej Make Or Equivalent , Od155572/2026-2027:Supplying Of Steel Cupboard For Storage 6Ft Tall 3 Feet Width, Ordinary Door With Godrej Or Equivalent Make , Od155578/2026-2027:Supplying And Installation Of Medium Back Revolving Chairs For Workstation, Conference Hall And Training Room (With Synchronized Tilt Mechanism, Zero Position Tilt Lock, Pneumatic Height Adjustments, Abs Net Back Rest & Seat With Abs , Adjustable Armrest, Nylon Base With Castors) (Godrej, Wipro Make Or Equivalent) , Od155571/2026-2027:Supplying And Installation Of High Back Revolving Chair For Cabin (With Synchronized Tilt Mechanism, Zero Position Tilt Lock, Pneumatic Height Adjustment, Abs Net Back Rest, Seat With Abs, Contoured Seat Mold, Adjustable Armrest, Aluminium Dia Cast Base With Castors) (Godrej, Wipro Make Or Equivalent). , Od155579/2026-2027:Supplying And Installation Of Visitors Chair, Adjustable Armrest, Non Revolving, Abs Net Back Rest & Seat With Abs, Countoured Seat Mould, (Godrej,Wipro Make Or Equivalent) , Od158443/2026-2027:Providing And Fixing 3Inch Chrome Finish Handle In Cub Board , Od158444/2026-2027:Providing And Fixing Iauto Hinges With Normal Close , 50.10.1:Steel Work In Built Up G I Tubular ( Round, Square Or Rectangular Hollow Tubes Etc.) Trusses Etc., Including Cutting, Hoisting,Fixing In Position And Applying A Priming Coat Of Approved Steel Primer, Including Welding And Bolted With Special Shaped Washers Etc. Complete , Od155565/2026-2027:Supplying And Fixing Skylight Of Size 1.2Mx 1.8M With Designed Acrylic Sheet Including 4 Numbers Led Tube As Per The Direction Of Engineer In Charge , Od155563/2026-2027: Supplying And Fixing Inaugration Board Using Granite Slab Including Lettering As Per Direction Of Departmental Officers Including Cost Of Granite, Lettering And Fixing Charges Etc Complete. Size: 100X84cm , Od155560/2026-2027:Supplying And Fixing Name Board Of Size 1.20 X 0.60 Using Acrylic Board Including Lettering As Per The Direction Of Departmental Officers Including All Cost , Conveyance And Labour Charges Etc Complete. , Od155564/2026-2027:Design And Placement Of Wall-Mounted Acrylic Name Board Material: High-Quality, Ultra-Durable Acrylic Dimensions: 0.45 M (L) X 0.10 M (H) Design: Sleek, Wall-Mounted Name Board With Clear And Bold Lettering Installation: Suitable For Indoor Wall Mounting Finish: Smooth Edges With Polished Surface For A Clean, Professional Look Durability: Weather-Resistant And Long-Lasting , Od155561/2026-2027:Cardiac Catheter Wash Station : Completely Made Of Ss 304 G Satin Finish, All Curved Joints Should Be As Per Htm Standards To Ensure Perfect Hygiene And Ease Of Cleaning, Provision For Connecting The Catheter Using 3 Way Adapter, Provision To Connect Air (High Pressure) And Water Through Separate Discharge Pipe Lines To Clean The Catheter, Wall Mounted Type, Approximate Dimension 33Lx 24W X 57H, Provisions To Connect And Clean At Least 6 Catheters Simultaneously. , Od155562/2026-2027:2 Bay Scrub Station : Providing, Fixing And Installation Of Floor Mounted 2 Bay Scrub Station Made Of Ss 304 1.6Mm Thick. Sheet With Timeout Facilities, Foot Operated With Hot And Cold Water Mixture Complete With All Fittings Etc. Satin Finish And Most Optimal Ra Value. Elbow Operated Water And Hand Scrub. Curved Joints To Avoid Microbial Deposition And Growth, Ensuring Perfect Hygiene And Ease Of Cleaning. The Station Shall Be Equipped With Soap Dispensers. Easily Openable For Refilling The Hand Scrub/ Routine Maintenance. Two Bay 63X2944 , Od155567/2026-2027:Wall Mounted Lead Apron Rack (Ss 304 Grade) For Lead Aprons : Stainless Steel 304 Grade, The Wall Mounted Foldaway Lead Apron Rack Is Designed As A Simple Solution To Store Six Lead Aprons Effectively. The Arms Of The Hanger Fold Away Against The Wall To Save Valuable Space, You Also Get The Option To Choose Whether They Swing To The Left Or Swing To The Right Side. Wall Mounted Lead Apron Rack Are Made From High Quality Stainless Steel. Six Movable Hangers For Lead Apron. Easy To Install & Scratch Resistant. Overall Frame Size Approx. : 56Cm Length X 28Cm Width (With Hanging Bracket) Depth Of Hanger Is Approx. 52Cm At 90 Degree Angle From Racks, , Od155568/2026-2027:Catheer Stouage Unit :Full-Height Cabinets With Specialized, Heavy-Duty Sliding Catheter Hooks (Cath Managers), Tempered Glass Or Solid Doors With Locks. Material: 18-Gauge, 304 Stainless Steel (Grade 304 Or Higher Is Standard For Hygiene) With Polished Finish. Structure: Freestanding, Mobile (With Lockable 5 Casters), Or Recessed Options. Dimensions: Typical Full-Height Cabinets Are ~2100 Mm (H) X 1000 Mm (W) X 600 Mm (D). , Od155566/2026-2027:Supplying And Laying Polyester, Washable And Stain Free Cloth(Bio-Active Fabric Washable Germ Free Suitable For Sterile Area) Curtain System With Maximum Curtain Cloth Height 230 Cm(Top 30 Cm White Colour Net And 200Cm Light Blue Color As Per Approval) Stitching With Stifiner Tape, Pleats, Ironing And Laying Around Each Bed Using Pvc Runner With Hooks & Stopper For Hanging/Sliding Etc. As Per The Direction Of Engineer In Charge Including Conveyance To Site, Laying In Line And Level, Sealing All Cutting In False Ceiling With Powder Coated Ms Cup, All Leads And Lifts Including All Labour Charges Etc. Complete. (Unpleated Length Of The Cloth Measurement Will Be Considered For Payment Also The Bottom And Top Foldings Will Not Be Considered In Cloth Measurement)/ Supplying And Fixing Powder Coated 20Mm Dia. Gi Round Tube Hanging Support Of Required Length(App. Length 1.10M) One End Fixing From Slab By Using 6Mm Dia Expansion Bolt At A Distance Not More Than 2M, C/C Down End Fixed To Powder Coated Aluminium Channel Frame 45 X 20Mm X 1Mm Thick Section By Using Ss Screw As Per The Direction Of Engineer In Charge Including Conveyance. , Od158451/2026-2027:Providing And Fixing 30 Mm Thick Puf Insulated Corrugated Panel 30 Mm Thick (Colour As Required By Engineer-In-Charge) With Crest Of 28 Mm And Required Pitch With 0.4Mm Ppgl Sheet At Top Side And 0.3Mm Ppgl Sheet At Bottom Side In 40±2 Kg Density, The Pu Panel Shall Be Cfc Free Using N-Pentane As A Blowing Agent. Panel Shall Have Ppgl Sheet On Both Side Of Polyurethane Foam Confirming To Is 12436:1988.The Pre Coated Sheet Shall Be Of Minimum 550 Mpa Steel Grade Confirming To Is 14246:1995 And Shall Have Zinc Coating Of Minimum 120 Gsm As Per Is :513 : 5-7 Microns Epoxy Primer On Both Side Of The Sheet And Polyester Top Coat 15-18 Micron. The Ppgl Sheet Shall Have Plastic Protective Guard Film Of Minimum 25 Microns To Avoid Scratches During Transportation. The Pu Foam Shall Be Self Extinguishing, Fire Retardant Type Having Minimum Density Of 40 Kg/Cu.Mt (+, -2 Kgs) The Roofing Panels Shall Be Fixed To Purlins Using Stsd Screws Having Suitable Washers & Length Complete As Per Direction Of Engineer-In- Charge , 24.2:Providing And Fixing Double Scaffolding System (Cup Lock Type ) On The Exterior Side Of Building/Structure, Upto 25 Metre Height, Above Ground Level, Including Additional Rows Of Scaffolding In Stepped Manner As Per Requirement Of Site, Made With 40 Mm Dia M.S. Tube, Placed 1.5 Metre Centre To Centre, Horizontal & Vertical Tubes Joint With Cup & Lock System With M.S. Tubes, M.S. Tube Challis, M.S. Clamps And Staircase System In The Scaffolding For Working Platform Etc. And Maintaining It In A Serviceable Condition For Execution Of Work Of Cleaning And / Or Pointing And / Or Applying Chemical And Removing It Thereafter. The Scaffolding System Shall Be Stiffened With Bracings, Runners, Connecting With The Building Etc, Wherever Required, If Feasible, For Inspection Of Work At Required Locations With Essential Safety Features For The Workmen Etc., Complete As Per Directions And Approval Of Engineer-In-Charge. Note: - ( 1) The Evaluational Area Of The Scaffolding Shall Be Measured For Payment Purpose, (2) The Payment Will Be Made Once Only For Execution Of All Items Of Such Works. , 1.54.2:Wiring For Group Controlled (Looped) Light Point/Fan Point/Exhaust Fan Point/ Call Bell Point (Without Independent Switch Etc.) With 1.5 Sq. Mm Frls Pvc Insulated Copper Conductor Single Core Cable In Surface/ Recessed Steel Conduit, And Earthing The Point With 1.5 Sq. Mm Frls Pvc Insulated Copper Conductor Single Core Cable Etc. As Required.Group B , 1.10.3:Wiring For Light Point/ Fan Point/ Exhaust Fan Point/ Call Bell Point With 1.5 Sq.Mm Frls Pvc Insulated Copper Conductor Single Core Cable In Surface / Recessed Medium Class Pvc Conduit, With Modular Switch, Modular Plate, Suitable Gi Box And Earthing The Point With 1.5 Sq.Mm. Frls Pvc Insulated Copper Conductor Single Core Cable Etc As Required.Group C , 1.14.1:Wiring For Circuit/ Submain Wiring Alongwith Earth Wire With The Following Sizes Of Frls Pvc Insulated Copper Conductor, Single Core Cable In Surface / Recessed Medium Class Pvc Conduit As Required2x1.5 Sq.Mm + 1X1.5 Sq.Mm Earth Wire , 1.14.2:Wiring For Circuit/ Submain Wiring Alongwith Earth Wire With The Following Sizes Of Frls Pvc Insulated Copper Conductor, Single Core Cable In Surface / Recessed Medium Class Pvc Conduit As Required2x 2.5 Sq. Mm + 1 X 2.5 Sq. Mm Earth Wire , 1.14.3:Wiring For Circuit/ Submain Wiring Alongwith Earth Wire With The Following Sizes Of Frls Pvc Insulated Copper Conductor, Single Core Cable In Surface / Recessed Medium Class Pvc Conduit As Required2 X 4 Sq. Mm + 1 X 4 Sq. Mm Earth Wire , 1.17.31:Supplying And Drawing Following Sizes Of Frls Pvc Insulated Copper Conductor, Single Core Cable In The Existing Surface / Recessed Steel/ Pvc Conduit As Required.4 X 6 Sq. Mm , 1.17.3:Supplying And Drawing Following Sizes Of Frls Pvc Insulated Copper Conductor, Single Core Cable In The Existing Surface / Recessed Steel/ Pvc Conduit As Required.3 X 1.5 Sq.Mm , 1.17.12:Supplying And Drawing Following Sizes Of Frls Pvc Insulated Copper Conductor, Single Core Cable In The Existing Surface / Recessed Steel/ Pvc Conduit As Required.3 X 2.5 Sq. Mm , 1.24.1:Supplying And Fixing Following Modular Switch/ Socket On The Existing Modular Plate & Switch Box Including Connections But Excluding Modular Plate Etc. As Required.5/6 Amps Switch , 1.24.3:Supplying And Fixing Following Modular Switch/ Socket On The Existing Modular Plate & Switch Box Including Connections But Excluding Modular Plate Etc. As Required.15/16 annual periodical maintenance Switch , 1.24.4:Supplying And Fixing Following Modular Switch/ Socket On The Existing Modular Plate & Switch Box Including Connections But Excluding Modular Plate Etc. As Required.3 Pin 5/6 annual periodical maintenance Socket Outlet , 1.24.5:Supplying And Fixing Following Modular Switch/ Socket On The Existing Modular Plate & Switch Box Including Connections But Excluding Modular Plate Etc. As Required.6 Pin 15/16 annual periodical maintenance Socket Outlet , 1.24.8:Supplying And Fixing Following Modular Switch/ Socket On The Existing Modular Plate & Switch Box Including Connections But Excluding Modular Plate Etc. As Required.Bell Push , 1.26:Supplying And Fixing Modular Blanking Plate On The Existing Modular Plate & Switch Box Excluding Modular Plate As Required. , 1.27.1:Supplying And Fixing Following Size/ Modules, Gi Box Alongwith Modular Base & Cover Plate For Modular Switches In Recess Etc As Required.1 Or 2 Module (75Mmx75mm) , 1.27.2:Supplying And Fixing Following Size/ Modules, Gi Box Alongwith Modular Base & Cover Plate For Modular Switches In Recess Etc As Required.3 Module (100Mmx75mm) , 1.27.4:Supplying And Fixing Following Size/ Modules, Gi Box Alongwith Modular Base & Cover Plate For Modular Switches In Recess Etc As Required.6 Module (200Mmx75mm) , 1.27.5:Supplying And Fixing Following Size/ Modules, Gi Box Alongwith Modular Base & Cover Plate For Modular Switches In Recess Etc As Required.8 Module (125Mmx125mm) , 1.38:Supplying And Fixing Call Bell/ Buzzer Suitable For Single Phase, 230 Volts, Complete As Required. , 90.4.8:Charges For Cutting Holes Suitable For Accommodating Exhaust Fans Up To 305Mm Sweep Including Plastering Colour Washing Etc. As Required. , 1.50.1:Installation Of Exhaust Fan In The Existing Opening, Including Making Good The Damage Connection Testing, Commissioning Etc. As Required.Upto 450 Mm Sweep , 90.11.1.10:Supply And Installation Of Sheet Steel, Phosphatised And Painted, Dust And Vermin Proof Enclosure Of Mcb Db Including Copper /Brass Bus Bar, Neutral Link, Earth Bus And Din Rail Suitable For Fixing Mcb/ Isolator Etc. Fixed On Wall Using Suitable Anchor Bolts Or Fixed In Recess Including Cutting Hole On The Wall , Making Good The Damages, Colour Washing Etc. As Required4 Way (8+12) - Three Phase Double Cover (Ip 42/43) , 90.11.1.11:Supply And Installation Of Sheet Steel, Phosphatised And Painted, Dust And Vermin Proof Enclosure Of Mcb Db Including Copper /Brass Bus Bar, Neutral Link, Earth Bus And Din Rail Suitable For Fixing Mcb/ Isolator Etc. Fixed On Wall Using Suitable Anchor Bolts Or Fixed In Recess Including Cutting Hole On The Wall , Making Good The Damages, Colour Washing Etc. As Required6 Way (8+18) - Three Phase Double Cover (Ip 42/43) , 90.11.1.20:Supply And Installation Of Sheet Steel, Phosphatised And Painted, Dust And Vermin Proof Enclosure Of Mcb Db Including Copper /Brass Bus Bar, Neutral Link, Earth Bus And Din Rail Suitable For Fixing Mcb/ Isolator Etc. Fixed On Wall Using Suitable Anchor Bolts Or Fixed In Recess Including Cutting Hole On The Wall , Making Good The Damages, Colour Washing Etc. As Required8 Way (8+12) - Double Cover Tpn Vertical Db With Provision For Fixing Mccb As Incomer And Sp/ Tp Mcb As Outgoing (Ip 42/43) , 2.12.1:Supplying And Fixing Following Rating, Double Pole, 240 Volts, Isolator In The Existing Mcb Db Complete With Connections, Testing And Commissioning Etc. As Required.40 Amps , 2.14.2:Supplying And Fixing Following Rating, Double Pole, (Single Phase And Neutral),240 Volts, Residual Current Circuit Breaker (Rccb), Having A Sensitivity Current 30 Milli Amperes In The Existing Mcb Db Complete With Connections, Testing And Commissioning Etc. As Required. 40 Amps , 2.15.3:Supplying And Fixing Following Rating, Four Pole, (Three Phase And Neutral), 415 Volts, Residual Current Circuit Breaker (Rccb), Having A Sensitivity Current 30 Milli Amperes In The Existing Mcb Db Complete With Connections, Testing And Commissioning Etc. As Required. 63 Amps , 2.15.2:Supplying And Fixing Following Rating, Four Pole, (Three Phase And Neutral), 415 Volts, Residual Current Circuit Breaker (Rccb), Having A Sensitivity Current 30 Milli Amperes In The Existing Mcb Db Complete With Connections, Testing And Commissioning Etc. As Required. 40 Amps , 90.12.7.49:Supply, Laying And Clamping Of 1 No. Pvc Insulated And Pvc Sheathed Armoured Aluminium Power Cable, 1.1 Kv Grade Of The Following Sizes Using Clamps Noted Along With The Cables, Spacing Of Clamps Not Exceeding 60Cms, Making Good The Damages , Colour Washing Etc. As Required.4 Core 185 Sq Mm , 90.12.41.34:Supply, Laying And Clamping Of 1 No. Pvc Insulated And Pvc Sheathed Armoured Copper Power Cable, 1.1 Kv Grade Of The Following Sizes Using Clamps Noted Along With The Cables, Spacing Of Clamps Not Exceeding 60Cms, Making Good The Damages , Colour Washing Etc. As Required.4 Core 16 Sq Mm With Factory Made Clamp , 90.12.1.45:Supply & Laying Of One Number Pvc Insulated And Pvc Sheathed Armoured Aluminium Power Cable Of 1.1Kv Grade Of The Following Sizes In Ground Including Excavation Of Trench Of Size 35 X 75 Cm, Refilling The Trench Etc. As Required But Excluding Sand Cushioning And Protective Covering (In Ordinary Soil).4 Core 70 Sq Mm , 9.1.23:Supplying And Making End Termination With Brass Compression Gland And Aluminium Lugs For Following Size Of Pvc Insulated And Pvc Sheathed / Xlpe Aluminium Conductor Cable Of 1.1 Kv Grade As Required.3 1/2X 70 Sq. Mm (38Mm) , 9.1.32:Supplying And Making End Termination With Brass Compression Gland And Aluminium Lugs For Following Size Of Pvc Insulated And Pvc Sheathed / Xlpe Aluminium Conductor Cable Of 1.1 Kv Grade As Required.4 X 10 Sq. Mm (25Mm) , 9.1.33:Supplying And Making End Termination With Brass Compression Gland And Aluminium Lugs For Following Size Of Pvc Insulated And Pvc Sheathed / Xlpe Aluminium Conductor Cable Of 1.1 Kv Grade As Required.4 X 16 Sq. Mm (28Mm) , 90.12.24.7:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.35 Sq Mm , 90.12.24.5:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.16 Sq Mm , 90.12.24.4:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.10 Sq Mm , 90.16.1.3:Supply And Drawing Bare Earthing Conductors Of The Following Sizes Along With Wiring/ Cables And Giving Connection As Required3.15 Mm Copper Conductor ( 10 Swg) , 90.16.5.3:Supply Of Superior Quality Copper Earth Socket For The Following Size Of Earth Conductor Including Crimping Etc. As Required.3.15 Mm (10 Swg) , 5.6:Earthing With Copper Earth Plate 600 Mm X 600 Mm X 3 Mm Thick Including Accessories, And Providing Masonry Enclosure With Cover Plate Having Locking Arrangement And Watering Pipe Of 2.7 Metre Long Etc. With Charcoal/ Coke And Salt As Required. , 90.16.3.9:Supply And Clamping The Following Size Of Strips On Surface Of Wall/ Parapet/ Existing Cable Tray Using Clamps Fabricated From 20 X 3 Mm Gi Flat Duly Painted Or Heavy Duty Gi Spacer Saddles Spacing Of Clamps Not Exceeding 1 M, Making Good The Damages, Colour Washing Etc. As Required ( For Horizontal Run)25 X 3 Mm Copper Strip , 90.16.2.4:Supply And Laying Of The Following Sizes Of B Class Gi Pipe Laid 0.6 M Below Ground Level Including Excavating Trench Of Suitable Size And Refilling Etc, As Required.40 Mm Dia G.I. Pipe ( Medium Class) , Od156110/2026-2027:Recess Mounted Led Luminaire With Full Diffuser Type Upto 36 W (Max), 2X2 Led, Cool White, 5700K, White Powder Coated Crca Housing, Wipro Crco10r038hp57gl1 Or Bajaj Bzrsql 36W Gzi Cw Or Equivalent , Od156113/2026-2027:Recess Mounted Circular Led Retrofit Downlighter, 15W, Powder Coated White Finish, Bpslr 15W Wh Bajaj Or Equivalent , Od156114/2026-2027:6 Kg Fire Extinguisher Abc Powder Type With Nabl Approval , Od156095/2026-2027:5 Kg Fire Extinguisher Carbon Di-Oxide Type With Nabl Approval , 17.2.14:Intelligent Fire Alarm Systemsupplying, Installation, Testing & Commissioning Of Addressable Manual Call Point Complete As Required. , 17.1.2:Fire Detection And Alarm System (Conventional)Supplying, Installation, Testing & Commissioning Of Smoke Detector With Built-In Led And Mounting Base Complete With All Connections Etc. As Required. , 17.2.5:Intelligent Fire Alarm Systemsupplying, Installation, Testing & Commissioning Of Response Indicator On Surface/ Recessed Ms Box Having Two Led, Metallic Cover Complete With All Connections Etc As Required. , 17.1.9.3:Supplying, Installation, Testing & Commissioning Sector Panel Suitable For Following Zones, Complete With Visual Indications For Short Circuit Fault, Open Circuit Fault, Fire Condition And All Other Standard Facilities As Per Is:2189 With Mimic Diagram For All Area/Zone Covered, Complete With All Connections, Interconnections As Required.10 Zone , 17.1.6:Fire Detection And Alarm System (Conventional)Supplying, Installation, Testing & Commissioning Fire Alarm Sounder With Facility To Make Announcement, Mounted In M.S. Box (16 Swg) With Hinged Cover Plate & Suitable For Operation With Amplifier I/C Line Matching Transformer Etc. Complete As Required. , 17.5.3.3:Supplying And Drawing Of Cable Fire Retardant Pvc Insulated Copper Conductor Cable In The Existing Surface / Recessed Steel Conduit Of Following Pairs, Cores And Size Including Connections And Interconnections Etc. As Required.Speaker Cable Three Pair, 2-Core, 1.5 Sqmm , 17.1.5:Fire Detection And Alarm System (Conventional)Supplying, Installation, Testing & Commissioning Response Indicator On Surface/ Recess Ms Box Having Two Leds Metallic Cover Complete With All Connections Etc. As Required. , 17.1.9.1:Supplying, Installation, Testing & Commissioning Sector Panel Suitable For Following Zones, Complete With Visual Indications For Short Circuit Fault, Open Circuit Fault, Fire Condition And All Other Standard Facilities As Per Is:2189 With Mimic Diagram For All Area/Zone Covered, Complete With All Connections, Interconnections As Required.4 Zone , 90.14.10.6:Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required.125A, 35/36 Ka (Ics=100%Icu), 3 Pole, Current Limiting Type Mccb With Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc , Od156102/2026-2027:Preparation Of Electrical Drawing, Submission Of Drawings To Electrical Inspectorate, Kseb And Obtaining Approval, Arranging Inspection For Entire Installation And Other Utilities. Obtaining Road Cutting Permission Obtaining Sanction Orders For The Electrical Installation, Testing, Commissioning & As-Built Documentation,Power Allocation, Energisation Of The Installation Complete. Including All Incidental Expenses Except Statutory Fees Payable To The Statutory Bodies. The Statutory Fees Will Be Reimbursed On Production Of Valid Receipts. , 1.19:Supplying And Drawing Co-Axial Tv Cable Rg-6 Grade, 0.7 Mm Solid Copper Conductor Pe Insulated, Shielded With Fine Tinned Copper Braid And Protected With Pvc Sheath In The Existing Surface/ Recessed Steel/ Pvc Conduit As Required. , 1.21.1:Supplying And Fixing Of Following Sizes Of Medium Class Pvc Conduit Along With Accessories In Surface/ Recess Including Cutting The Wall And Making Good The Same In Case Of Recessed Conduit As Required.20 Mm , 1.24.6:Supplying And Fixing Following Modular Switch/ Socket On The Existing Modular Plate & Switch Box Including Connections But Excluding Modular Plate Etc. As Required.Telephone Socket Outlet , 1.24.7:Supplying And Fixing Following Modular Switch/ Socket On The Existing Modular Plate & Switch Box Including Connections But Excluding Modular Plate Etc. As Required.Tv Antenna Socket Outlet , 1.21.2:Supplying And Fixing Of Following Sizes Of Medium Class Pvc Conduit Along With Accessories In Surface/ Recess Including Cutting The Wall And Making Good The Same In Case Of Recessed Conduit As Required.25 Mm , 1.21.3:Supplying And Fixing Of Following Sizes Of Medium Class Pvc Conduit Along With Accessories In Surface/ Recess Including Cutting The Wall And Making Good The Same In Case Of Recessed Conduit As Required.32 Mm , 90.6.9.5:Supplying & Drawing Unarmored Pvc Insulated And & Pvc Sheathed Copper Conductor Telephone Pair Cable Conforming To Itd Specification For Indoor Application Through The Existing Conduit5 Pair , 1.53.1:Supplying And Drawn Of Utp 4 Pair Cat 6 Lan Cable In The Existing Surface/Recessed Steel/Pvc Conduit As Required1 Run Of Cable , Od156096/2026-2027:Cat 6 Rj 45 Computer Jack Modular Socket Outlet , Od156111/2026-2027:Supply Of Branded 2 Meter Patch Cords With Cat6 Cable , Od156115/2026-2027:Patch Panel- 24-Port Patch Panel With Cable Tray-Rack Mountable , Od156109/2026-2027:24-Port Gigabit Manageable Smart Network Switch With 2 Sfps.Rack Mountable , Od158671/2026-2027:Supply Of Lt Switch Board Cubicle Type, Made Out Of 2 Mm Thick Crca Sheet, Totally Enclosed, Ip42, Free Standing, Wall/Floor Mounting, Dust And Vermin Proof, Powder Coated, Indoor, Compartmentalized, Suitable For Operation On 3 Phase And Neutral, 415 V, 50Hz Ac System With Bus-Bars Extendable On Both Sides, Including All The Necessary Internal Wiring , Suitable Spreader Etc. The Panel Shall Be Fabricated From A Firm Having Cpri Certification With Test Certificates For Similar Panel(Required For Short Circuit Rating, Temperature Rise And Ip Classification) And With Following Switchgears, Metering Instruments And Accessories Etc. Complete As Required As Per Drawing & Specification.Ryb Indicating Lamps Led Type4a Sp Mcb (0-500) V Class 1 With 90 Deg Defl. Scale.(0-200)A Class 1 With 90 Deg Defl. Scale, Voltmeter Selector Switch 200/5 A 5 Va Class .5S Cts On/Off Indication Lamp, Digital Multi-Function Meter To Read V,A.Hz,Pf,Kwh,Incomer 200A Fp Moulded Case Circuit Breaker With Thermal Release With Ol And Sc Protection Neutral Link (Ics =100%Icu= 25Ka) -1 No&Rsquo;S200a Copper Tpn Busbar Chamber With Copper Busbar-Out Goings 100 A Fp Moulded Case Circuit Breaker With Thermal Release With Ol And Sc Protection Neutral Link (Ics =100%Icu= 25Ka) - 3 No&Rsquo;S .63 A Fp Moulded Case Circuit Breaker With Thermal Based Release With Ol And Sc Protection Neutral Link (Ics =100%Icu= 25Ka) - 3 No&Rsquo;Sthe Panel Shall Be Top Entry/Exit And Suitable For Termination Of Cable System As Per Sld And Technical Specification. , Od158672/2026-2027:Supply Of Lt Switch Board Cubicle Type, Made Out Of 2 Mm Thick Crca Sheet, Totally Enclosed, Ip42, Free Standing, Wall/Floor Mounting, Dust And Vermin Proof, Powder Coated, Indoor, Compartmentalized, Suitable For Operation On 3 Phase And Neutral, 415 V, 50Hz Ac System With Bus-Bars Extendable On Both Sides, Including All The Necessary Internal Wiring , Suitable Spreader Etc. The Panel Shall Be Fabricated From A Firm Having Cpri Certification With Test Certificates For Similar Panel (Required For Short Circuit Rating, Temperature Rise And Ip Classification) And With Following Switchgears, Metering Instruments And Accessories Etc. Complete As Required As Per Drawing & Specification. Ryb Indicating Lamps Led Type, 4A Sp Mcb, Digital Multi-Function Meter To Read V,A,Hz,Pf,Kwh, Surge Protection Device, Efr, 400/5 A ,5 Va Class .5S Cts, On/Off Indication Lamp. Incomer -Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required. 250A, 25 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc With Copper Tpn Busbar Chamber With Copper Busbar. Outgoings-1)Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required.200A, 25 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Thermal/Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc-1No 2) Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required.100A, 16 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Thermal Based Release With Overload Setting Of Having Adjustable Ol & Sc-2 Nos 3)63 A 4Pole Mccb With Thermal Release With Ol & Sc Protection Nuetral Link (Ics=100%, Icu 16Ka)-3 Nos 4) Spare-1No , Od158673/2026-2027:Supply Of Lt Switch Board Cubicle Type, Made Out Of 2 Mm Thick Crca Sheet, Totally Enclosed, Ip42, Free Standing, Wall/Floor Mounting, Dust And Vermin Proof, Powder Coated, Indoor, Compartmentalized, Suitable For Operation On 3 Phase And Neutral, 415 V, 50Hz Ac System With Bus-Bars Extendable On Both Sides, Including All The Necessary Internal Wiring , Suitable Spreader Etc. The Panel Shall Be Fabricated From A Firm Having Cpri Certification With Test Certificates For Similar Panel (Required For Short Circuit Rating, Temperature Rise And Ip Classification) And With Following Switchgears, Metering Instruments And Accessories Etc. Complete As Required As Per Drawing & Specification. Ryb Indicating Lamps Led Type, 4A Sp Mcb, Digital Multi-Function Meter To Read V,A,Hz,Pf,Kwh, Surge Protection Device, Efr, 400/5 A ,5 Va Class .5S Cts, On/Off Indication Lamp, Incomer -Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required. 320A/400A, 50/55 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc Outgoings-1)Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required.160A, 35/36 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc-1No 2) Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required.16A -100A, 35/36 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc-3 Nos 3)Spare- 2Nos , 9.1.3:Supplying And Making End Termination With Brass Compression Gland And Aluminium Lugs For Following Size Of Pvc Insulated And Pvc Sheathed / Xlpe Aluminium Conductor Cable Of 1.1 Kv Grade As Required.2 X 16 Sq. Mm (22Mm) , 4.4.3:Supplying And Installing Following Size Of Perforated Painted With Powder Coating M.S. Cable Trays Cross Member With Perforation Not More Than 17.5%, Joined With Connectors, Suspended From The Ceiling With M.S. Suspenders Including Bolts & Nuts, Painting Suspenders Etc As Required. 225 Mm Width X 50 Mm Depth X 1.6 Mm Thickness , 16.16:Supplying & Fixing Of Powder Coated Extruded Aluminum Return Air Grills With Louvers But Without Volume Control Dampers Complete As Required. , 90.12.7.35:Supply, Laying And Clamping Of 1 No. Pvc Insulated And Pvc Sheathed Armoured Aluminium Power Cable, 1.1 Kv Grade Of The Following Sizes Using Clamps Noted Along With The Cables, Spacing Of Clamps Not Exceeding 60Cms, Making Good The Damages , Colour Washing Etc. As Required.3.5 Core 240 Sq Mm , 1.27.6:Supplying And Fixing Following Size/ Modules, Gi Box Alongwith Modular Base & Cover Plate For Modular Switches In Recess Etc As Required.12 Module (200Mmx150mm) , 9.1.29:Supplying And Making End Termination With Brass Compression Gland And Aluminium Lugs For Following Size Of Pvc Insulated And Pvc Sheathed / Xlpe Aluminium Conductor Cable Of 1.1 Kv Grade As Required.3 1/2X 240 Sq. Mm (62Mm) , 90.12.41.36:Supply, Laying And Clamping Of 1 No. Pvc Insulated And Pvc Sheathed Armoured Copper Power Cable, 1.1 Kv Grade Of The Following Sizes Using Clamps Noted Along With The Cables, Spacing Of Clamps Not Exceeding 60Cms, Making Good The Damages , Colour Washing Etc. As Required.4 Core 35 Sq Mm With Factory Made Clamp , 90.12.38.7:Supply And Laying Of One Number Additional Pvc Insulated And Pvc Sheathed Armoured Copper Power Cable Of 1.1Kv Grade Of The Following Sizes In Ground In The Same Trench In One Tier Horizontal Formation Including Widening The Trench By 20 Cm, Refilling The Trench Etc. As Required But Excluding Sand Cushioning And Protective Covering (In Ordinary Soil)3 Core 10 Sq Mm , 1.49:Numbering Of Ceiling Fan / Exhaust Fan/ Fluorescent Fittings As Required. , 4.1.3:Supplying And Installing Following Size Of Perforated Painted With Powder Coating M.S. Cable Trays With Perforation Not More Than 17.5%, In Convenient Sections, Joined With Connectors, Suspended From The Ceiling With M.S. Suspenders Including Bolts & Nuts, Painting Suspenders Etc As Required. 225 Mm Width X 50 Mm Depth X 1.6 Mm Thickness , 4.3.2:Supplying And Installing Following Size Of Perforated Painted With Powder Coating M.S. Cable Trays Tee With Perforation Not More Than 17.5%, Joined With Connectors, Suspended From The Ceiling With M.S. Suspenders Including Bolts & Nuts, Painting Suspenders Etc As Required. 150 Mm Width X 50 Mm Depth X 1.6 Mm Thickness , 4.3.3:Supplying And Installing Following Size Of Perforated Painted With Powder Coating M.S. Cable Trays Tee With Perforation Not More Than 17.5%, Joined With Connectors, Suspended From The Ceiling With M.S. Suspenders Including Bolts & Nuts, Painting Suspenders Etc As Required. 225Mm Width X 50 Mm Depth X 1.6 Mm Thickness , 90.12.24.10:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.95 Sq Mm , 90.12.24.11:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.120 Sq Mm , 90.12.24.12:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.150 Sq Mm , 90.12.24.13:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.185 Sq Mm , 90.12.24.14:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.240 Sq Mm , 90.12.24.6:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.25 Sq Mm , 90.12.24.8:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.50 Sq Mm , 90.12.24.9:Supply Of Superior Quality Copper Tubular Cable Socket And Making Core Termination Of Cables Of The Following Sizes Including Crimping Etc. As Required.70 Sq Mm , Od158671/2026-2027:Supply Of Lt Switch Board Cubicle Type, Made Out Of 2 Mm Thick Crca Sheet, Totally Enclosed, Ip42, Free Standing, Wall/Floor Mounting, Dust And Vermin Proof, Powder Coated, Indoor, Compartmentalized, Suitable For Operation On 3 Phase And Neutral, 415 V, 50Hz Ac System With Bus-Bars Extendable On Both Sides, Including All The Necessary Internal Wiring , Suitable Spreader Etc. The Panel Shall Be Fabricated From A Firm Having Cpri Certification With Test Certificates For Similar Panel(Required For Short Circuit Rating, Temperature Rise And Ip Classification) And With Following Switchgears, Metering Instruments And Accessories Etc. Complete As Required As Per Drawing & Specification.Ryb Indicating Lamps Led Type4a Sp Mcb (0-500) V Class 1 With 90 Deg Defl. Scale.(0-200)A Class 1 With 90 Deg Defl. Scale, Voltmeter Selector Switch 200/5 A 5 Va Class .5S Cts On/Off Indication Lamp, Digital Multi-Function Meter To Read V,A.Hz,Pf,Kwh,Incomer 200A Fp Moulded Case Circuit Breaker With Thermal Release With Ol And Sc Protection Neutral Link (Ics =100%Icu= 25Ka) -1 No&Rsquo;S200a Copper Tpn Busbar Chamber With Copper Busbar-Out Goings 100 A Fp Moulded Case Circuit Breaker With Thermal Release With Ol And Sc Protection Neutral Link (Ics =100%Icu= 25Ka) - 3 No&Rsquo;S .63 A Fp Moulded Case Circuit Breaker With Thermal Based Release With Ol And Sc Protection Neutral Link (Ics =100%Icu= 25Ka) - 3 No&Rsquo;Sthe Panel Shall Be Top Entry/Exit And Suitable For Termination Of Cable System As Per Sld And Technical Specification. , Od158672/2026-2027:Supply Of Lt Switch Board Cubicle Type, Made Out Of 2 Mm Thick Crca Sheet, Totally Enclosed, Ip42, Free Standing, Wall/Floor Mounting, Dust And Vermin Proof, Powder Coated, Indoor, Compartmentalized, Suitable For Operation On 3 Phase And Neutral, 415 V, 50Hz Ac System With Bus-Bars Extendable On Both Sides, Including All The Necessary Internal Wiring , Suitable Spreader Etc. The Panel Shall Be Fabricated From A Firm Having Cpri Certification With Test Certificates For Similar Panel (Required For Short Circuit Rating, Temperature Rise And Ip Classification) And With Following Switchgears, Metering Instruments And Accessories Etc. Complete As Required As Per Drawing & Specification. Ryb Indicating Lamps Led Type, 4A Sp Mcb, Digital Multi-Function Meter To Read V,A,Hz,Pf,Kwh, Surge Protection Device, Efr, 400/5 A ,5 Va Class .5S Cts, On/Off Indication Lamp. Incomer -Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required. 250A, 25 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc With Copper Tpn Busbar Chamber With Copper Busbar. Outgoings-1)Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required.200A, 25 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Thermal/Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc-1No 2) Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required.100A, 16 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Thermal Based Release With Overload Setting Of Having Adjustable Ol & Sc-2 Nos 3)63 A 4Pole Mccb With Thermal Release With Ol & Sc Protection Nuetral Link (Ics=100%, Icu 16Ka)-3 Nos 4) Spare-1No , Od158673/2026-2027:Supply Of Lt Switch Board Cubicle Type, Made Out Of 2 Mm Thick Crca Sheet, Totally Enclosed, Ip42, Free Standing, Wall/Floor Mounting, Dust And Vermin Proof, Powder Coated, Indoor, Compartmentalized, Suitable For Operation On 3 Phase And Neutral, 415 V, 50Hz Ac System With Bus-Bars Extendable On Both Sides, Including All The Necessary Internal Wiring , Suitable Spreader Etc. The Panel Shall Be Fabricated From A Firm Having Cpri Certification With Test Certificates For Similar Panel (Required For Short Circuit Rating, Temperature Rise And Ip Classification) And With Following Switchgears, Metering Instruments And Accessories Etc. Complete As Required As Per Drawing & Specification. Ryb Indicating Lamps Led Type, 4A Sp Mcb, Digital Multi-Function Meter To Read V,A,Hz,Pf,Kwh, Surge Protection Device, Efr, 400/5 A ,5 Va Class .5S Cts, On/Off Indication Lamp, Incomer -Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required. 320A/400A, 50/55 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc Outgoings-1)Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required.160A, 35/36 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc-1No 2) Supply, Conveyance And Fixing The Following Types & Current Rated Control Gears & Switchgears Conforming To Is 13947 Suitable For 440 V, 50 Hz, Ac Supply In The Existing Panel Assembly As Required.16A -100A, 35/36 Ka (Ics=100%Icu), 4 Pole, Current Limiting Type Mccb With Microprocessor Based Release With Overload Setting Of 50 - 100% Having Adjustable Ol & Sc-3 Nos 3)Spare- 2Nos , Od158710/2026-2027:Supply, Laying And Clamping Of 1 No. Pvc Insulated And Pvc Sheathed Unarmoured Copper Power Cable, 70 Sqmm Isi Marked Fr Ls Pvc Insulated Single Core Copper Conductor Cable Using Clamps Noted Along With The Cables, Spacing Of Clamps Not Exceeding 60Cms, Making Good The Damages , Colour Washing Etc. As Required. , Od158711/2026-2027:Supply, Laying And Clamping Of 1 No. Pvc Insulated And Pvc Sheathed Unarmoured Copper Power Cable, 1.1 Kv Grade Of The Following Sizes Using Clamps Noted Along With The Cables, Spacing Of Clamps Not Exceeding 60Cms, Making Good The Damages , Colour Washing Etc. As Required. 3.5 Core 50 Sq Mm With Factory Made Clamp , Od158712/2026-2027:Supply, Laying And Clamping Of 1 No. Pvc Insulated And Pvc Sheathed Unarmoured Copper Power Cable, 1.1 Kv Grade Of The Following Sizes Using Clamps Noted Along With The Cables, Spacing Of Clamps Not Exceeding 60Cms, Making Good The Damages , Colour Washing Etc. As Required. 3.5 Core 35 Sq Mm With Factory Made Clamp , Od158743/2026-2027:Supply, Laying And Clamping Of 1 No. Pvc Insulated And Pvc Sheathed Unarmoured Copper Power Cable, 25 Sqmm Isi Marked Fr Ls Pvc Insulated Single Core Copper Conductor Cable Using Clamps Noted Along With The Cables, Spacing Of Clamps Not Exceeding 60Cms, Making Good The Damages , Colour Washing Etc. As Required. , 16.5.2:Supplying, Laying,Fixing, Testing And Commissioning Of Following Thicknessinsulation With Nitrile Rubber Of Thickness 19Mm On Existing Pipe , Od156296/2026-2027:Charges For Cutting Holes Suitable For Accommodating The Supply Ducts Including Plastering, Colour Washing Etc. As Required. , Od156414/2026-2027:Supply,Conveyance Testing And Commissioning Of 2 Ton 3 Star Inverter Type,Split Ac Unit With Copper Tube Cooling Coils , Bluestar/Lg/Voltas/Mitsubishi(Dual Inverter Type ) Or Superior Brands,Cooling Capacity 6000Kw/Hr With Rotary/Scroll Type Compressors, Microprocessor Controlled Codeless Remote,Moisture Removal Above 1.8, Cfm Above 500/450/350,Led/Lcd Display. , Od156415/2026-2027:Supply,Conveyance Testing And Commissioning Of 5Kva Voltage Stabilizer Wall Mount Type With Time Delay,High& Low Voltage& Thermal Overload Protection,Input Voltage 170-220V ,Output Voltage 200-240 V,High Voltage Cut Off 240,Efficiency: ≫95%,Frequency : 50Hz , Od156416/2026-2027:Supply, Installation, Testing And Commissioning Upvc Drain Pipe 6Kg/Sq.Cmof Suitable Dia Connecting The Drain Pan Of The Indoor Units To Nearest Drain Point, With 12Mm Thick Closed Cell Elastomeric Insulation As Per Specification , Od156417/2026-2027:Sitc Of 19 Mm Thick Nitrile Rubber For Under Deck Insulation , Od156453/2026-2027:Motorized Volume Control Damper For Individual Operation Of Ductable Split Air Conditioners , Od156454/2026-2027:Supplying And Fixing Outdoor Unit Stand For 1 Ton To 2 Ton Split Type Air Conditioner, Fabricated From Ms Angle Of Suitable Size Duly Welded, Provided With Suitable Stiffeners And Cross Members, Painted With Two Coats Of Anti-Corrosive Primer And Powder Coated, Complete With Necessary Anchor Fasteners/Bolts And Rubber Vibration Pads For Mounting The Concerned Ac Outdoor Unit, Including Transportation, Labour And Materials As Required , Od156455/2026-2027:Additional Copper Pipe With Superlon Insulation For Suction And Discharge Suitable For 1/1.5/2 Tr Split Ac Units , Od156456/2026-2027:Electrical Cabling Between Indoor And Outdoor Units, With Required Size Cables/Wires(Havells/Rr Kabel/V-Guard Frls Wires) , Od156457/2026-2027:Providing And Fixing Of Thermal Insulation Lining With 13Mm Thick Nitrile Rubber Closed Cell Type With Adhesive, On Supply Duct , Od156458/2026-2027:Providing And Fixing Of Acoustic Duct Lining With 10Mm Thick Nitrile Rubber Open Cell Type Inside The Ducts , Od156463/2026-2027:Supply, Installation, Testing And Commissioning (Sitc) Of 5.5 Tr X 2 No'S Or Higher Capacity Air Cooled Ductable Split Air Conditioner With Ceiling Suspended Idu And All Standard Fittings, Idu Ss Drain Pan, As Per Specifications Complete With Suitable Capacity Drive Motor, Air Cooled Condenser With All Accessories, Hermetically Sealed Scroll Compressor, Electronic Microprocessor Pendant Controller With Digital Display, First Charge Refrigerent Gas, Vibration Isolator Pads, Structural Supports For Fixing Idu/Odu As Shown In The Drawing, Safety Devices Including Thermal Protector, Pressure Relief Valve, Over Load Relays Single Phase Preventor Etc, Power Wiring(Max 3M From Isolator) And Control Cabling(Indoor To Outdoor),Thermostat Wiring, High-Low Pressure Cut Out, Suitable For Three Phase 440V Supply Etc Complete As Required , Od156464/2026-2027:Supply, Installation, Testing And Commissioning Of Volume Control Duct Damper Made Of Extruded Aluminium For Ducts/Collars To Be Provided With Suitable Links, Lebers And Quadrants For Manual Control Of Volume Of Air Flow And For Proper Balancing Of The Air Distribution In The Main Duct Complete With Neoprene Rubber Gaskets, Nuts, Bolts, Screws Linkages, Flanges Etc., As Per Specifications.The Material Shall Be Paramagnetic , Od156465/2026-2027:Supply, Installation, Testing And Commissioning Of Suitable Sized Hard Drawn 18 Swg Copper Refrigerant Piping(Both Suction And Liquid Piping) Including 25Mm Thick Closed Cell Elastomeric Insulation As Per Specification For 11/8.75/ 5.5 Tr Machine Between Indoor & Outdoor Units As Per Site Conditions All Complete. , Od156466/2026-2027:Supply, Installation, Testing And Commissioning (Sitc) Of 8.75 Tr X 2 No'S Or Higher Capacity Air Cooled Ductable Split Air Conditioner With Ceiling Suspended Idu And All Standard Fittings, Idu Ss Drain Pan, As Per Specifications Complete With Suitable Capacity Drive Motor, Air Cooled Condenser With All Accessories, Hermetically Sealed Scroll Compressor, Electronic Microprocessor Pendant Controller With Digital Display, First Charge Refrigerent Gas, Vibration Isolator Pads, Structural Supports For Fixing Idu/Odu As Shown In The Drawing, Safety Devices Including Thermal Protector, Pressure Relief Valve, Over Load Relays Single Phase Preventor Etc, Power Wiring(Max 3M From Isolator) And Control Cabling(Indoor To Outdoor),Thermostat Wiring, High-Low Pressure Cut Out, Suitable For Three Phase 440V Supply Etc Complete As Required , Od156467/2026-2027:Supply, Installation, Testing And Commissioning (Sitc) Of 11 Tr X 2 No'S Or Higher Capacity Air Cooled Ductable Split Air Conditioner With Ceiling Suspended Idu And All Standard Fittings, Idu Ss Drain Pan, As Per Specifications Complete With Suitable Capacity Drive Motor, Air Cooled Condenser With All Accessories, Hermetically Sealed Scroll Compressor, Electronic Microprocessor Pendant Controller With Digital Display, First Charge Refrigerent Gas, Vibration Isolator Pads, Structural Supports For Fixing Idu/Odu As Shown In The Drawing, Safety Devices Including Thermal Protector, Pressure Relief Valve, Over Load Relays Single Phase Preventor Etc, Power Wiring(Max 3M From Isolator) And Control Cabling(Indoor To Outdoor),Thermostat Wiring, High-Low Pressure Cut Out, Suitable For Three Phase 440V Supply Etc Complete As Required , Od156469/2026-2027:Suitable De-Humidifier Units-Extraction Atleast 35Litres/Day,Compressor-Rotary,Airflow Atleast 220Mcube/Hr,Reputed Brand ,Input Voltage 230V Ac , Od156470/2026-2027:Supply, Installation, Testing And Commissioning Of Powder Coated Extruded Aluminimum Continuous Grilles 150 Mm High For Supply Air With Aluminium Damper As Per Spec , Od156471/2026-2027:Supplying, Fixing Testing Commissioning Of Supply Of Air Diffusers Of Powder Coated Aluminium Without Volume Control Dampers With Anti Smudge Ring & Removable Core. , Od156472/2026-2027:Supplying, Fixing Testing Commissioning Of Return Air Diffusers Of Powder Coated Aluminium Without Volume Control Dampers With Anti Smudge Ring & Removable Core , Od156473/2026-2027: Supply, Installation, Testing And Commissioning Of Powder Coated Extruded Aluminium Continuous Grilles 150Mm High For Return Air As Per Spec , 16.12.1.1:Supply, Installation, Balancing And Commissioning Of Factory Fabricated Gss Sheet Metal Rectangular/Round Ducting Complete With Neoprene Rubber Gaskets, Elbows, Splitter Dampers, Vanes, Hangers, Supports Etc. As Per Approved Drawings And Specifications Of Following Sheet Thickness Complete As Required.Factory Fabricated Gss Sheet Metal Rectangular Ducting As Per Is 655-2006 Thickness 0.63 Mm Sheet , 16.12.1.2:Supply, Installation, Balancing And Commissioning Of Factory Fabricated Gss Sheet Metal Rectangular/Round Ducting Complete With Neoprene Rubber Gaskets, Elbows, Splitter Dampers, Vanes, Hangers, Supports Etc. As Per Approved Drawings And Specifications Of Following Sheet Thickness Complete As Required.Factory Fabricated Gss Sheet Metal Rectangular Ducting As Per Is 655-2006 Thickness 0.80 Mm Sheet , Od158751/2026-2027:Supplying And Fixing Powder-Coated Ms Outdoor Unit Stand Suitable For 5.5 Tr Ductable Split Ac Unit, Fabricated From Heavy-Duty Angle Sections With Adequate Stiffeners, Complete With Ribbed Rubber Vibration Isolation Pads, Anchoring Bolts, And Anti-Corrosive Paint Finish As Required , Od158752/2026-2027:Supplying And Fixing Powder-Coated Ms Outdoor Unit Stand Suitable For 11/8.75 Tr Ductable Split Ac Unit, Fabricated From Heavy-Duty Angle Sections With Adequate Stiffeners, Complete With Ribbed Rubber Vibration Isolation Pads, Anchoring Bolts, And Anti-Corrosive Paint Finish As Required , Od158753/2026-2027:Supply, Installation, Testing And Commissioning Of 8.5 Tr Inverter Type Air-Cooled Ductable Split Air-Conditioning Unit, Comprising Ceiling-Suspended Indoor Unit And Air-Cooled Outdoor Unit(Both Copper Coils), Complete With All Standard Accessories And Safety Devices, Conforming To Relevant Is/Iec Standards And ManufacturerS Specifications. The Unit Shall Be Provided With Inverter-Driven Hermetically Sealed Scroll Compressor And Shall Operate On R-410A / R-32 Refrigerant With Factory First Charge. The Indoor Unit Shall Be Complete With Stainless Steel Drain Pan, Inverter-Compatible Blower Motor And Factory Insulation, While The Outdoor Unit Shall Be Provided With Blue-Fin / Gold-Fin Coated Condenser Coils And Corrosion-Resistant Powder-Coated Enclosure, Suitable For Operation At Ambient Temperatures Up To 50°C. The System Shall Be Suitable For 3-Phase, 415-440 V, 50 Hz Power Supply With Wide Voltage Operating Range And Shall Include Microprocessor-Based Wired Controller With Digital Display. Safety Features Such As High And Low Pressure Cut-Outs, Overload Relays, Phase Sequence / Single-Phasing Preventer And Thermal Protection For Compressor And Motors Shall Be Provided. The Work Shall Include Vibration Isolation Pads For Indoor And Outdoor Units, Structural Supports For Mounting As Per Site Requirements, Power Wiring Up To 3 Metres From Isolator, Control Cabling Between Indoor And Outdoor Units, Thermostat Wiring, Testing, Commissioning And Handing Over In Working Condition, Complete In All Respects, As Per ManufacturerS Recommendations And Directions Of Engineer-In-Charge. , The Scrub Station Shall Be Made Of High-Grade Stainless Steel 304 With Stain Polish Finish And Wired To The Current Iee Regulations & Bs 7971. The Scrub Stations Control System Shall Be Based On Advanced Microprocessor. The Scrub Station Shall Be Sensor-Activated 24Vac. The Scrub Station Shall Be Provided With User Selectable Time Out Control. The Supply Shall Not Be Interrupted In Case Of Power Failure Or Maintenance. The Station Shall Be Equipped With Thermostatic Mixing Valve To Controls The Temperature Of Hot / Cold Water Supply. The Washing Faucet Shall Be Made Of Chrome Plated Brass Hand. The Station Shall Be Equipped With Soap Dispensers. Two Bay 63X2431 (Not Taken For Evaluation) , Eto Sterlizer, Sealing Machine & Accessories ( Not Taken For Evaluation ) , Latest Dose Reduction / Image Quality Improvement Packages Like Autoright / Clarity / Optiq. ( Not Taken For Evaluation ) , Iabp As Per The Specification ( Not Taken For Evaluation)
BOQ
| Sl. No. | Item Description |
| 1 | Cathlab machine as per the specification except the following |
| 2 | Medical Grade large high definition FHD display 55\\" |
| 3 | Four Medical grade high definition display LCD/ LED flat 19\\" or higher monitors with wide viewing angle, high luminance high contrast, flicker free, distortion free |
| 4 | Electrophysiology Application |
| 5 | Workstation as per spec |
| 6 | High Pressure Injector with 200 Syringes |
| 7 | Online UPS for Cathlab & accessories , Console, Workstation, lights and computer |
| 8 | All in one Computer with Laser Printer |
| 9 | 12 channel ECG with PC Connectivity |
| 10 | Full Wrap Around Lead Aprons |
| 11 | Skirt (with fastening belt)-Vest type Lead Aprons |
| 12 | Front full type Lead Apron |
| 13 | Light weight leadless single piece aprons with front and back coverage |
| 14 | Light weight leadless skirt and vest with fastening belt |
| 15 | Thyroid Guard lead less |
| 16 | Leadless Head Gear radiation protection and light weight Lead Goggles |
| 17 | Lead Stockings |
| 18 | lead Spectacles |
| 19 | Lead Gloves |
| 20 | Heavy Duty Stand with Hangers to hold atleast 5 Lead Aprons (Stainless steel 304 construction) |
| 21 | Wall mounted hangers to hold atleast 5Lead Aprons |
| 22 | Intercom between exam and control room |
| 23 | Instrument table |
| 24 | Accessory table |
| 25 | 15.2.1:Demolishing cement concrete manually / by mechanical means including disposal of material within 50 metres lead as per direction of Engineer - in-Charge.Nominal concrete 1:3:6 or richer mix (i/c equivalent design mix) |
| 26 | 15.12.1:Dismantling doors, windows and clerestory windows (steel or wood ) shutter including chowkhats, architrave, holdfasts etc. complete and stacking within 50 metres lead:Of area 3 sq. metres and below |
| 27 | 15.12.2:Dismantling doors, windows and clerestory windows (steel or wood ) shutter including chowkhats, architrave, holdfasts etc. complete and stacking within 50 metres lead:Of area beyond 3 sq. metres |
| 28 | 15.23.1:Dismantling tile work in floors and roofs laid in cement mortar including stacking material within 50 metres lead.For thickness of tiles 10 mm to 25 mm |
| 29 | 15.3:Demolishing R.C.C. work manually / by mechanical means including stacking of steel bars and disposal of unserviceable material with in 50 metres lead as per direction of Engineer -in-Charge. |
| 30 | 15.57:Dismantling aluminium / Gypsum partitions, doors, windows, fixed glazing and false ceiling including disposal of unserviceable material and stacking of serviceable material with in 50 metres lead as directed by Engineer - in-Charge |
| 31 | 15.7.4:Demolishing brick work manually / by mechanical means including stacking of serviceable material and disposal of unserviceable material within 50 metres lead as per direction of Engineer-in-Charge.In cement mortar |
| 32 | 15.60:Disposal of building rubbish/ malba/ similar unserviceable, dismantled or waste materials by mechanical means, including loading, transporting, unloading to approved municipal dumping ground or as approved by Engineer-in-charge, beyond 50 m initial lead, for all leads including all lifts involved |
| 33 | 2.1.1:Earth work in surface excavation not exceeding 30 cm in depth but exceeding 1.5m in width as well as 10 sqm on plan including disposal of excavated earth up to 50 m and lift up to 1.5 m, disposed soil to be levelled and neatly dressed:All Kinds of soil |
| 34 | 2.6.1:Earth work in excavation by mechanical means (Hydraulic excavator)/manual means over areas (exceeding 30 cm in depth, 1.5 m in width as well as 10 sqm on plan) including getting out and disposal of excavated earth lead upto 50 m and lift upto 1.5 m, as directed by Engineer-in-charge.All kinds of soil |
| 35 | 50.2.25.1:Filling with contractor's own earth ( excluding rock) in trenches, plinth, sides of foundations etc. in layers not exceeding 20 cm in depth, consolidating each deposited layer by ramming and watering, lead up to 50 m and lift up to 1.5 m as per direction of site Engineer-in-charge |
| 36 | 2.34.1:Supplying chemical emulsion in sealed containers including delivery as specified.Chlorpyriphos / Lindane emulsifiable concentrate of 20% |
| 37 | 2.35.1.1:Diluting and injecting chemical emulsion for POST -CONSTRUCTIONAL anti-termite treatment (excluding the cost of chemical emulsion):Along external wall where the apron is not provided using chemical emulsion @ 7.5 litres / sqm of the vertical surface of the substructure to a depth of 300mm including excavation channel along the wall & rodding etc. complete:With Chlorpyriphos /Lindane E.C. 20% with 1% concentration |
| 38 | 6.26.2:Brick work with common burnt clay selected F.P.S (non modular ) bricks of class designation 7.5 in exposed brick work including making horizontal and vertical grooves 10 mm wide 12 mm deep complete in cement mortar 1:6 ( 1 cement : 6 coarse sand)Above plinth level and upto floor V level |
| 39 | 50.6.1.2:Solid block masonry using pre cast solid blocks (Factory made) of size 40x20x20cm or nearest available size confirming to IS 2185 part I of 1979 for super structure up to floor two level thickness 20cm and above in: CM 1:6 ( 1 cement: 6 coarse sand) etc complete. |
| 40 | 50.6.3.2:Solid block masonry using pre cast solid blocks (Factory made) of size 40x20x10 cm or nearest available size confirming to IS 2185 part I of 1979 for super structure up to floor two level for 10 cm thick wall in : CM 1:6 ( 1 cement : 6 coarse sand) including cost of scaffolding complete |
| 41 | 50.6.1.8:Solid block masonry using pre cast solid blocks (Factory made) of size 40x20x15cm or nearest available size confirming to IS 2185 part I of 1979 for super structure up to floor two level with thickness 15cm in: CM 1:6 ( 1 cement : 6 coarse sand etc complete |
| 42 | 4.1.3:Providing and laying in position cement concrete of specified grade excluding the cost of centering and shuttering - All work up to plinth level:1:2:4 (cement : 2 coarse sand (zone-III)derived from natural sources : 4 graded stone aggregate 20 mm nominal size derived from natural sources) |
| 43 | 4.10A: Providing and laying damp-proof course 40mm thick with cement concrete 1:2:4 (1 cement : 2 coarse sand (zone-III) derived from natural sources : 4 graded stone aggregate 12.5mm nominal size derived from natural sources).Providing and laying damp-proof course 40mm thick with cement concrete 1:2:4(1 cement : 2 coarse sand (zone-III) including manufactured sand derived fromRecycled Concrete Aggregate (RCA) upto 25% : 4 graded stone aggregate 12.5mmnominal size including (Recycled Concrete Aggregate (RCA) upto 25%)) |
| 44 | 5.3:Reinforced cement concrete work in beams, suspended floors, roofs, having slope up to 15° landings, balconies, shelves, chajjas, lintels, bands, plain window sills, staircases and spiral stair cases up to floor five level excluding the cost of centering, shuttering, finishing and reinforcement, with1:1.5:3 (1 cement : 1.5 coarse sand (Zone III) derived from natural sources: 3 graded stone aggregate 20 mm nominal size derived from natural sources). |
| 45 | 5.22.1:Steel reinforcement for R.C.C work including straightening, cutting, bending, placing in position and binding all complete upto plinth levelMild steel and Medium Tensile steel bars |
| 46 | 5.9.5:Centering and shuttering including strutting, etc. and removal of form for:Lintels, beams, plinth beams, girders bressumers and cantilevers |
| 47 | 50.4.1.1:Providing and laying in position cement concrete 1:2:4 ( 1 cement : 2 coarse sand : 4 graded stone aggregate 12.5 mm nominal size) up to floor five level excluding the cost of centering, shuttering, finishing for leveling coarse of floor and roof etc. |
| 48 | 50.5.33.2:Providing and laying in position machine batched and machine mixed design mix M-20 grade cement concrete for reinforced cement concrete work, using cement content as per approved design mix, including pumping of concrete to site of laying but excluding the cost of centering, shuttering, finishing and reinforcement, including admixtures in recommended proportions as per IS: 9103 to accelerate, retard setting of concrete, improve workability without impairing strength and durability as per direction of Engineer-in-charge. Note:- Cement content considered in this item is @ 330 kg/cum. Excess or less cement used as per design mix is payable or recoverable separately. All work above plinth level upto floor V level |
| 49 | 50.5.5:Reinforced cement concrete work in arches, archribs, domes, vaults, shells, folded plate and roof having slope more than 15 degree up to floor five level excluding the cost of centering, shuttering, finishing and reinforcement with 1:1.5:3 ( 1 cement : 1.5 coarse sand: 3 graded stone aggregate 20 mm nominal size). |
| 50 | 5.9.2:Centering and shuttering including strutting, etc. and removal of form for:Walls (any thickness) including attached pilasters, butteresses, plinth and string courses etc. |
| 51 | 5.9.3:Centering and shuttering including strutting, etc. and removal of form for:Suspended floors, roofs, landings, balconies and access platform |
| 52 | 5.9.6:Centering and shuttering including strutting, etc. and removal of form for:Columns, Pillars, Piers, Abutments, Posts and Struts |
| 53 | 13.1.1:12 mm cement plaster of mix:1:4 ( 1 cement : 4 fine sand) |
| 54 | 13.7.1:12 mm cement plaster finished with a floating coat of neat cement of mix:1:3 ( 1 cement : 3 fine sand) |
| 55 | 9.141:Providing and fixing PVC Door Frame of size 50x47 mm with a wall thickness of 5 mm (±0.2mm), made out of single piece extruded PVC profile, with mitred cut joints and with 2 nos of PVC bracket of size 190 mm x 100 mm long arms of cross section size 35 x 15 mm & self driven self taping screws, the vertical door profiles to be reinforced with 40x20 mm M.S. rectangular tube of 0.8 mm, including providing EPDM rubber gasket weather seal throughout the frame, including jointing 5 mm PVC frame strip with PVC solvent cement on the back of the profile. The door frame to be fixed to the wall using 8x100 mm long anchor fasteners complete, all as per manufacturer's specification and direction of Engineer-in--charge. |
| 56 | 9.142.2:35 mm thick factory made Solid panel PVC Door shutter, made out of singleDecorative finish (wood grained finish) |
| 57 | 21.1.1.2:Providing and fixing aluminium work for doors, windows, ventilators and partitions with extruded built up standard tubular sections/ appropriate Z sections and other sections of approved make conforming to IS : 733 and IS: 1285, fixing with dash fasteners of required dia and size, including necessary filling up the gaps at junctions, i.e. at top, bottom and sides with required EPDM rubber/ neoprene gasket etc. Aluminium sections shall be smooth, rust free, straight, mitred and jointed mechanically wherever required including cleat angle, Aluminnium snap beading for glazing /paneling, C.P. brass/ stainless steel screws, all complete as per architectural drawings and the directions of Engineer-in-charge.(Glazing, paneling and dash fasteners to be paid for separately):For fixed portionPowder coated aluminium (minimum thickness of powder coating 50 micron) |
| 58 | 21.1.2.2:For shutters of doors, windows & ventilators including providing and fixing hinges / pivots and making provision for fixing of fittings wherever required including the cost of EPDM rubber/ neoprene gasket required (Fittings shall be paid for separately)Powder coated aluminium (minimum thickness of powder coating 50 micron) |
| 59 | 21.13:Providing and fixing Brass100 mm mortice latch and lock with 6 levers without pair of handles ( best make of approved quality ) for aluminum doors including necessary cutting and making good etc. complete |
| 60 | 21.2.2:Providing and fixing 12 mm thick prelaminated particle board flat pressed three layer or graded wood particle board conforming to IS : 12823 Grade I Type II, in panelling fixed in aluminum doors, windows shutters and partition frames with C.P. brass / stainless steel screws etc. complete as per architectural drawings and directions of Engineer - in- Charge.Pre- laminated particle board with decorative lamination on both sides |
| 61 | 21.3.1:Providing and fixing glazing in aluminium door, window, ventilator shutters and partitions etc. with EPDM rubber / neoprene gasket etc. complete as per the architectural drawings and the directions of Engineer - in -Charge. ( Cost of aluminium snap beading shall be paid in basic item):With float glass panes of 4.0 mm thickness (weight not less than 10kg/sqm) |
| 62 | 9.83:Providing and fixing aluminium die cast body tubular type universal hydraulic door closer (having brand logo with ISI, IS : 3564, embossed on the body, door weight upto 35 kg and door width upto 700 mm), with necessary accessories and screws etc. complete. |
| 63 | 12.45.2:Providing and fixing false ceiling at all height including providing and fixing of frame work made of special sections, power pressed from M.S. sheets and galvanized with zinc coating of 120 gms/ sqm (both side inclusive) as per IS : 277 and consisting of angle cleats of size 25 mm wide x 1.6 mm thick with flanges of 27 mm and 37 mm, at 1200 mm centre to centre, one flange fixed to the ceiling with dash fastener 12.5 mm dia x 50 mm long with 6 mm dia bolts, other flange of cleat fixed to the angle hangers of 25x10x0.50 mm of required length with nuts & bolts of required size and other end of angle hanger fixed with intermediate G.I. channels 45x15x0.9 mm running at the spacing of 1200 mm centre to centre, to which the ceiling section 0.5 mm thick bottom wedge of 80 mm with tapered flanges of 26 mm each having lips of 10.5 mm, at 450 mm centre to centre, shall be fixed in a direction perpendicular to G.I. intermediate channel with connecting clips made out of 2.64 mm dia x 230 mm long G.I. wire at every junction, including fixing perimeter channels 0.5 mm thick 27 mm high having flanges of 20 mm and 30 mm long, the perimeter of ceiling fixed to wall/partition with the help of rawl plugs at 450 mm centre, with 25 mm long dry wall screws @ 230 mm interval, including fixing of gypsum board to ceiling section and perimeter channel with the help of dry wall screws of size 3.5x25 mm at 230 mm c/c, including jointing and finishing to a flush finish of tapered and square edges of the board with recommended jointing compound, jointing tapes, finishing with jointing compound in 3 layers covering upto 150 mm on both sides of joint and two coats of primer suitable for board, all as per manufacturer's specification and also including the cost of making openings for light fittings, grills, diffusers, cutouts made with frame of perimeter channels suitably fixed, all complete as per drawings, specification and direction of the Engineer in Charge but excluding the cost of painting with:12.5 mm thick tapered edge Glass Reinforced Gypsum (GRG) board conformingto IS: 2095- (Part 3):1996 (Boards with BIS certification marks) |
| 64 | 12.54.1:Providing and fixing GI Clip in Metal Ceiling System of 600x600 mm module which includes providing and fixing 'C' wall angle of size 20x30x20 mm made of 0.5 mm thick pre painted steel along the perimeter of the room with help of nylon sleeves and wooden screws at 300 mm center to centre, suspending the main C carrier of size 10x38x10 mm made of G.I steel 0.7 mm thick from the soffit with help of soffit cleat 37x27x25x1.6 mm, rawl plugs of sixe38x12 mm and C carrier suspension clip and main carrier bracket at 1000 mm c/c. Inverted triangle shaped Spring Tee having height of 24 mm and width of 34 mm made of GI steel 0.45 mm thick is then fixed to the main C carrier and in direction perpendicular to it at 600 mm centers with help of suspension brackets. Wherever the main C carrier and spring T have to join, C carrier and spring T connectors have to be used. All sections to be galvanized @ 120 gms/sqm (both side inclusive), fixing with clip in tiles into spring 'T' with :GI Metal Ceiling Clip in plain Beveled edge global white color tiles of size 600x600 and 0.5mm thick with 25 mm height, ,made of G.I sheet having galvanizing of 100 gms/ sqm (both sides inclusive) and electro statically polyester powder coated of thickness 60 microns (minimum), including factory painted after bending . |
| 65 | 12.58.1:Providing and fixing tiled false ceiling of approved materials of size 595x595 mm in true horizontal level, suspended on interlocking metal grid of hot dipped galvanized steel sections (galvanized @ 120 grams/ sqm, both side inclusive) consisting of main \\"T\\" runner with suitably spaced joints to get required length and of size 24x38 mm made from 0.30 mm thick (minimum) sheet, spaced at 1200 mm center to center and cross \\"T\\" of size 24x25 mm made of 0.30 mm thick (minimum) sheet, 1200 mm long space between main \\"T\\" at 600 mm center to center to form a grid of 1200x600 mm and secondary cross \\"T\\" of length 600 mm and size 24x25 mm made of 0.30 mm thick (minimum) sheet to be interlocked at middle of the 1200x600 mm panel to form grids of 600x600 mm and wall angle of size 24x24x0.3 mm and laying false ceiling tiles of approved texture in the grid including, required cutting/making, opening of services like diffusers, grills, light fittings, fixtures, smoke detectors etc. Main \\"T\\" runners to be suspended from ceiling using GI slotted cleats of size 27 x 37 x 25 x 1.6 mm fixed to ceiling with 12.5 mm dia and 50 mm long dash fasteners, 4 mm GI adjustable rods with galvanized butterfly level clips of size 85 x 30 x 0.8 mm spaced at 1200 mm center to center along main T, bottom exposed width of 24 mm of all T-sections shall be pre-painted with polyester paint, all complete for all heights as per specifications, drawings and as directed by Engineer-in-charge.8 mm thick fully perforated calcium silicate board made with Calcareous & Siliceous materials reinforced with cellulose fiber manufactured through autoclaving process to give stable crystalline structure with minimum compressive strength 225 Kg/ sq. cm, bending strength 100 Kg/sq. cm, of size 595x595 mm, having perforation of dia, 10 mm with minimum perforated area 18% with non woven tissue on the back side, having an NRC (Noise Reduction Coefficient) of 0.85, with 50 mm thick rockwool of 48 Kg/cum backing. |
| 66 | OD155583/2026-2027:Providing and fixing trap doors for access to the areas above false ceiling. The trap door shall have an external frame made of Powder coated aluminium channel on which the shutter is hinged. Shutter shall be made 4 mm thick ACP sheet shutter external frame of shutter made of Powder coated aluminium channel. Concealed heavy duty hinges shall be used to mount the shutters and locking shall be provided. The shutter shall be moulded in level with the false ceiling forming a groove all around. Sufficient number of hinges and locks shall be provided to avoid sagging of the shutter. Key also shall be supplied. No extra rate shall be provided for suspenders or frame works all round for linking the trap door to gyp board /any other false ceiling. The rate including all labour charges ,material charges ect complete.Size 1.20 m x 0.6 m |
| 67 | OD155584/2026-2027:Providing and fixing factory-made sandwich panels (PUF core) for ISO WALL with 40 mm PUF double skin with all necessary support channels and accessories etc wall and partition work, made of 0.4 mm thick precoated galvanized iron sheets on top side and 0.3mm sheet at bottom side with 40 mm thick polyurethane foam core (density 40 kg/m³), having tongue and groove joints with cam-lock mechanism, including necessary framing, anchor fasteners, joint sealing with silicone sealant, and complete in all respects as per manufacturer’s specifications. |
| 68 | OD155581/2026-2027:Providing and fixing sun controlled film on glass surface for door/windows,after thoroughly cleaning and dust free the surface to get smooth finish,to get air tight clean surface, ect,complete as directed by Engineer-in -charge |
| 69 | OD155582/2026-2027: Providing and fixing Vertical blinds as per the direction of Engineer in charge |
| 70 | OD155569/2026-2027:Providing and laying 2 mm thick anti static Poly vinyl flooring of laid over 25 mm thick leveling layer of cement concrete 1:2:4 (1 cement : 2 coarse sand : 4 graded stone aggregate 10 mm nominal size) to be paid separately, coated with a thin layer of hot bitumen penetration 80/25 (blown type) @ 2.45 kg per sqm, including fixing blocks in position after dipping in hot bitumen (blown type) up to half depth, planed, levelled smooth and finished complete. |
| 71 | 8.2.2.2:Providing and fixing 18 mm thick gang saw cut, mirror, polished, premoulded and prepolished, machine cut for kitchen platforms, vanity counters, window sills, facias and similar locations, of required size, approved shade, colour and texture laid over 20 mm thick base cement mortar 1:4 ( 1 cement : 4 coarse sand), joints treated with white cement, mixed with matching pigment, epoxy touch ups, including rubbing, curing, moulding and polishing to edges to give high gloss finish etc. complete at all levels.Area of slab over 0.50 sqm |
| 72 | 11.37:Providing and laying Ceramic glazed floor tiles of size 300x300 mm (thickness to be specified by the manufacturer) of 1st quality conforming to IS : 15622 of approved make in colours such as White, Ivory, Grey, Fume Red Brown, laid on 20 mm thick cement mortar 1:4 (1 Cement : 4 Coarse sand), Jointing with grey cement slurry @ 3.3 kg/sqm including pointing the joints with white cement and matching pigment etc., complete |
| 73 | 11.47.2:Providing and laying Vitrified tiles in different sizes (thickness to be specified by the manufacturer), with water absorption less than 0.08% and conforming to IS : 15622, of approved brand & manufacturer,in all colours and shade, in skirting, riser of steps, laid with cement based high polymer modified quick set tile adhesive (water based) conforming to IS : 15477, in average 6 mm thickness, including grouting of joints (Payment for grouting of joints to be made separately).Size of Tile 600x600 mm |
| 74 | 11.49.2:Providing and laying Vitrified tiles in floor with different sizes (thickness to be specified by the manufacturer), with water absorption less than 0.08% and conforming to IS : 15622, of approved brand & manufacturer, in all colours and shade, laid with cement based high polymer modified quick set tile adhesive (water based) conforming to IS : 15477, in average 6 mm thickness, including grouting of joints (Payment for grouting of joints to be made separately).Size of Tiles 600x600 mm |
| 75 | 11.48.2:Grouting the joints of flooring tiles having joints of 3 mm width, using epoxy grout mix 0.70 kg of organic coated filler of desired shade (0.10 kg of hardener and 0.20 kg of resin per kg). including filling /grouting and finishing complete as per direction of Engineer-in-charge.Size of Tile 600x600 mm |
| 76 | 13.43.1:Applying one coat of water thinnable cement primer of approved brand and manufacture on wall surface:Water thinnable cement primer |
| 77 | 13.80:Providing and applying white cement based putty of average thickness 1 mm, of approved brand and manufacturer, over the plastered wall surface to prepare the surface even and smooth complete. |
| 78 | 13.82.2:Wall painting with acrylic emulsion paint, having VOC (Volatile Organic Compound) content less than 50 grams/ litre, of approved brand and manufacture including applying additional coats wherever required, to achieve even shade and colour.Two coats |
| 79 | 13.46.1:Finishing walls with Acrylic Smooth exterior paint of required shade:New work (Two or more coat applied @ 1.67 ltr/10 sqm over and including priming coat of exterior primer applied @ 2.20 kg/10 sqm) |
| 80 | 13.61.1:Painting with synthetic enamel paint of approved brand and manufacture to give an even shade:Two or more coats on new work |
| 81 | 13.69.1:Polishing on wood work with ready mixed wax polish of approved brand and manufacture:New work |
| 82 | 17.11.1:Providing and fixing white vitreous china laboratory sink with C.I brackets, C.P. brass chain with rubber plug 40 mm C.P brass waste and 40 mm C.P. brass trap with necessary C.P. brass unions complete, including painting of fittings and brackets, cutting and making good the wall wherever required:Size 450x300x150 mm |
| 83 | 17.2.2:Providing and fixing white vitreous china pedestal type water closet (European type W.C. pan) with seat and lid, 10 litre low level white P.V.C. flushing cistern, including flush pipe, with manually controlled device (handle lever), conforming to IS : 7231, with all fittings and fixtures complete, including cutting and making good the walls and floors wherever required:W.C. pan with ISI marked black solid plastic seat and lid |
| 84 | 17.28.1.1:Providing and fixing P.V.C. waste pipe for sink or wash basin including PVC.waste fittings complete.Semi rigid pipe32 mm dia |
| 85 | 17.32.1:Providing and fixing mirror of superior glass ( of approved quality) and of required shape and size with plastic moulded frame of approved make and shade with 6 mm thick hard board backing :Circular shape 450 mm dia |
| 86 | 17.7.1:Providing and fixing wash basin with C.I. brackets, 15 mm C.P. brass pillar taps, 32 mm C.P. brass waste of standard pattern, including painting of fittings and brackets, cutting and making good the walls wherever require:White Vitreous China Wash basin size 630x450 mm with a pair of 15 mm C.P. brass pillar taps |
| 87 | 17.73.2:Providing and fixing PTMT towel rail complete with brackets fixed to wooden cleats with CP brass screws with concealed fitting arrangement of approved quality and colour600 mm long towel rail with total length of 645 mm, width 78 mm and effective height of 88 mm, weighting not less than 190 gms |
| 88 | 50.17.1.5:Supplying and fixing CP Health Faucet superior quality ( Jagur or equvalent make ) including cost of materials and labour charges etc complete as per the direction of site Engineer-in-charge. |
| 89 | OD155570/2026-2027:Providing and fixing 15 mm C.P. brass tap with elbow operation lever of approved quality conforming to IS standards and weighing not less than 690 gms. 15 mm nominal bore |
| 90 | 18.65:Providing and fixing PTMT Soap Dish Holder having length of 138 mm, breadth 102 mm, height of 75 mm with concealed fittings arrangements, weighing not less than 106 gms. |
| 91 | 18.15.1:Providing and fixing brass bib cock of approved quality :15 mm nominal bore |
| 92 | 50.18.7.1.1:Providing and fixing PVC pipes, fittings including fixing the pipe with clamps at 1.00 m spacing. This includes jointing of pipes & fittings with one step PVC solvent cement and testing of joints complete as per direction of Engineer-in-Charge 15mm dia 12kgf/cm2 -Internal work- Exposed on wall |
| 93 | 50.18.7.4.1:Providing and fixing PVC pipes, fittings including fixing the pipe with clamps at 1.00 m spacing. This includes jointing of pipes & fittings with one step PVC solvent cement and testing of joints complete as per direction of Engineer-in-Charge 32 mm dia 10Kgf/cm2- Internal work - Exposed on wall |
| 94 | 50.18.8.1.1:Providing and fixing PVC pipes, fittings including fixing the pipe with clamps at 1.00 m spacing. This includes jointing of pipes & fittings with one step PVC solvent cement and testing of joints complete as per direction of Engineer- in- Charge. Concealed work, including cutting chases and making good the wall etc. 15 mm pipe 12 Kgf/ cm2 |
| 95 | 50.18.8.4.1:Providing and fixing PVC pipes, fittings including fixing the pipe with clamps at 1.00 m spacing. This includes jointing of pipes & fittings with one step PVC solvent cement and testing of joints complete as per direction of Engineer-in-Charge. Concealed work, including cutting chases and making good the wall etc. 32 mm pipe 10Kgf/cm2 |
| 96 | 50.18.8.8.2:Providing and fixing PVC pipes, fittings, including fixing the pipe with clamps at 1.00 m spacing. this includes jointing of pipes with one step PVC solvent cement and testing of joints complete as per the direction of Engineer-in-Charge. Concealed work, including cutting chases and making good the wall etc. 75 mm pipe 4 kgf/cm2 |
| 97 | 50.18.8.9.1:Providing and fixing PVC pipes, fittings including fixing the pipe with clamps at 1.00 m spacing. This includes jointing of pipes with one step PVC solvent cement and testing of joints complete as per direction of Engineer-in-Charge. Concealed work, including cutting chased and making good the wall etc. 110 mm pipe 6kgf/cm2 |
| 98 | 50.9.1.1:Providing wood work in frames of doors, windows, clerestory windows and other frames, wrought framed and fixed in position with hold fast lugs or with dash fasteners of required dia & length (hold fast lugs or dash fastener shall be paid for separately), using good quality Anjili wood /jack wood |
| 99 | 9.21.2:Providing and fixing ISI marked flush door shutters conforming to IS : 2202 (Part I) non- decorative Type, core of block board construction with frame of 1st class hard wood and well matched commercial 3 ply veneering with vertical grains or cross bands and face veneers on both faces of shutters:30 mm thick including ISI marked Stainless Steel butt hinges with necessary screws |
| 100 | OD155577/2026-2027:Supplying and lining 2 mm thick radiation shielding lead sheet conforming to IS 2112 or equivalent international standards, for doors and frames, ensuring overlap between sheets, joints staggered and sealed to prevent radiation leakage. Lead sheet to be properly glued with suitable contact adhesive, fixed on dooe frame/shutter as per approved design. Door frame and door shutter paid separately .All work to be done as per regulatory body requirements (like AERB guidelines). |
| 101 | 9.127.2:Providing & Fixing decorative high pressure laminated sheet of plain / wood grain in gloss/ matt/suede finish with high density protective surface layer and reverse side of adhesive bonding quality conforming to IS: 2046 Types S, including cost of adhesive of approved quality.1.0 mm thick |
| 102 | 9.140.1:Providing and fixing plain lining with necessary screws/nuts/& bolts/nails, including a coat of approved primer on one face, and fixed on wooden /steel frame work, complete as per direction of Engineer-in-charge ( Frame work shall be paid for separately)12 mm thick commercial ply conforming to IS: 1328 BWR type |
| 103 | 9.169:Providing and fixing 19mm thick both side balancing lamination factory pressed BWP grade marine ply as per IS 710 of approved brand boxes, shelves, racks, almirah, cupboard and drawer etc. including necessary nails, screws etc. complete as per direction of Engineer-in-charge. |
| 104 | 9.170.1:Providing and fixing stainless steel fancy handle of approved make fixed with SS screws etc. complete as per direction of Engineer-in-charge.200 mm |
| 105 | 9.171:Providing and fixing stainless steel soft closing spring hinges at 0 degree hinges (hydraulic type) of approved make/brand to cupboard shutters with full threaded steel screws including making necessary recess in board and finished etc. complete as per direction of Engineer-in-charge. |
| 106 | 14.34:Providing and fixing 150 mm bright finished floor brass door stopper with rubber cushion, necessary brass screws etc. to suit shutter thickness complete |
| 107 | 9.72.3:Providing and fixing bright finished brass butt hinges with necessary screws etc. complete:100x85x5.5 mm (heavy type) |
| 108 | 9.74.3:Providing and fixing bright finished brass tower bolts (barrel type) with necessary screws etc. complete:150x10 mm |
| 109 | OD155580/2026-2027:Supply and fixing Solid Steel Under Desk Keyboard Tray Slide Out with 360° Rotating Mouse Platform, Keyboard Shelf Sliding |
| 110 | OD155576/2026-2027:File cabinet 465 w x 620 d x 1350 h(4 cabinet type Godrej or equivalent) |
| 111 | OD155573/2026-2027:Shoe Rack Three Shelves : Pretreated and powder coated, Three shelves, Made of cold rolled MS sheet of 16 guages, Six cold rolled MS tube frame levels as support for 6 pairs of shoes in one shelf, Abrasion resistant, non-discoloring, corrosion resistant |
| 112 | OD155575/2026-2027:Computer table 1500*600 with keyboard tray, CPU storage at base in a side and storage drawers at other side., To be made out of 25mm Thick top & whereas Body, Back & Shutter in 18mm Prelaminated particle board with PVC edge band on all sides & . Handle "C" Type SS Brush Finish (For workstation and reports) reputed brand Godrej,Wipro or equivalent |
| 113 | OD155574/2026-2027:Supplying of Steel cupboardfor storage 6ft tall 3 feet width, Door with glass transparent doors with Godrej make or equivalent |
| 114 | OD155572/2026-2027:Supplying of Steel cupboard for storage 6ft tall 3 feetwidth, Ordinary door with Godrej or equivalent make |
| 115 | OD155578/2026-2027:Supplying and installation of Medium Back revolving Chairs for workstation, Conference hall and Training room (With Synchronized Tilt Mechanism, Zero position tilt lock, Pneumatic Height Adjustments, ABS Net Back Rest & seat with ABS , Adjustable armrest, Nylon base with castors) (Godrej, Wipro make or equivalent) |
| 116 | OD155571/2026-2027:Supplying and installation of High Back revolving Chair for Cabin (With Synchronized Tilt Mechanism, Zero position tilt lock, Pneumatic Height Adjustment, ABS Net Back Rest, seat with ABS, contoured seat mold, Adjustable armrest, aluminium dia cast Base with castors) (Godrej, Wipro make or equivalent). |
| 117 | OD155579/2026-2027:Supplying and installation of Visitors chair, Adjustable armrest, Non revolving, ABS net Back Rest & seat with ABS, countoured seat mould, (Godrej,Wipro make or equivalent) |
| 118 | OD158443/2026-2027:Providing and fixing 3inch chrome finish handle in cub board |
| 119 | OD158444/2026-2027:Providing and fixing IAuto hinges with normal close |
| 120 | 50.10.1:Steel work in built up G I tubular ( round, square or rectangular hollow tubes etc.) trusses etc., including cutting, hoisting,fixing in position and applying a priming coat of approved steel primer, including welding and bolted with special shaped washers etc. complete |
| 121 | OD155565/2026-2027:Supplying and fixing skylight of size 1.2mx 1.8m with designed acrylic sheet including 4 numbers LED Tube as per the direction of engineer in charge |
| 122 | OD155563/2026-2027: Supplying and fixing Inaugration board using granite slab including lettering as per direction of departmental officers including cost of granite, lettering and fixing charges etc complete. Size: 100x84cm |
| 123 | OD155560/2026-2027:Supplying and fixing name board of size 1.20 x 0.60 using acrylic board including lettering as per the direction of departmental officers including all cost , conveyance and labour charges etc complete. |
| 124 | OD155564/2026-2027:Design and Placement of Wall-Mounted Acrylic Name Board Material: High-quality, ultra-durable acrylic Dimensions: 0.45 m (L) x 0.10 m (H) Design: Sleek, wall-mounted name board with clear and bold lettering Installation: Suitable for indoor wall mounting Finish: Smooth edges with polished surface for a clean, professional look Durability: Weather-resistant and long-lasting |
| 125 | OD155561/2026-2027:Cardiac catheter wash station : Completely made of SS 304 G Satin finish, All curved joints should be as per HTM standards to ensure perfect hygiene and ease of cleaning, Provision for connecting the catheter using 3 way adapter, provision to connect Air (high pressure) and water through separate discharge pipe lines to clean the catheter, wall mounted type, approximate dimension 33”LX 24”W X 57”H, provisions to connect and clean at least 6 catheters simultaneously. |
| 126 | OD155562/2026-2027:2 Bay Scrub Station : Providing, Fixing and Installation of floor mounted 2 Bay Scrub Station made of SS 304 1.6mm thick. Sheet with timeout facilities, foot operated with hot and cold water mixture complete with all fittings etc. Satin finish and most optimal RA value. Elbow operated water and hand scrub. Curved joints to avoid microbial deposition and growth, ensuring perfect hygiene and ease of cleaning. The station shall be equipped with soap dispensers. Easily openable for refilling the hand scrub/ routine maintenance. Two bay 63”x29”44” |
| 127 | OD155567/2026-2027:Wall Mounted Lead Apron Rack (SS 304 Grade) for Lead Aprons : Stainless Steel 304 Grade, The wall mounted foldaway lead apron rack is designed as a simple solution to store six lead aprons effectively. The arms of the hanger fold away against the wall to save valuable space, you also get the option to choose whether they swing to the left or swing to the right side. Wall mounted lead apron rack are made from high quality stainless steel. Six movable Hangers for Lead Apron. Easy to install & scratch resistant.Overall Frame size approx. : 56cm Length x 28cm Width (with Hanging Bracket)Depth of Hanger is approx. 52cm at 90 degree angle from racks, |
| 128 | OD155568/2026-2027:Catheer stouage unit :full-height cabinets with specialized, heavy-duty sliding catheter hooks (cath managers), tempered glass or solid doors with locks.Material: 18-gauge, 304 Stainless Steel (Grade 304 or higher is standard for hygiene) with polished finish.Structure: Freestanding, mobile (with lockable 5\\" casters), or recessed options.Dimensions: Typical full-height cabinets are ~2100 mm (H) x 1000 mm (W) x 600 mm (D). |
| 129 | OD155566/2026-2027:Supplying and laying polyester, washable and stain free cloth(Bio-Active fabric washable germ free suitable for sterile area) curtain system with maximum curtain cloth height 230 cm(top 30 cm white colour net and 200cm light blue color as per approval) stitching with stifiner tape, pleats, ironing and laying around each bed using PVC runner with hooks & stopper for hanging/sliding etc. as per the direction of engineer in charge including conveyance to site, laying in line and level, sealing all cutting in false ceiling with powder coated MS cup, all leads and lifts including all labour charges etc. complete. (unpleated length of the cloth measurement will be considered for payment also the bottom and top foldings will not be considered in cloth measurement)/ supplying and fixing powder coated 20mm dia. GI round tube hanging support of required length(app. length 1.10m) one end fixing from slab by using 6mm dia expansion bolt at a distance not more than 2m, c/c down end fixed to powder coated aluminium channel frame 45 x 20mm x 1mm thick section by using SS screw as per the direction of engineer in charge including conveyance. |
| 130 | OD158451/2026-2027:Providing and fixing 30 mm thick PUF insulated Corrugated Panel 30 mm thick (colour as required by Engineer-in-Charge) with crest of 28 mm and required pitch with 0.4mm PPGL sheet at top side and 0.3mm PPGL sheet at bottom side in 40±2 Kg Density, The PU panel shall be CFC free using n-Pentane as a blowing agent.Panel shall have PPGL sheet on both side of Polyurethane Foam confirming to IS 12436:1988.The pre coated sheet shall be of minimum 550 mpa steel grade confirming to IS 14246:1995 and shall have zinc coating of minimum 120 gsm as per IS :513 : 5-7 microns epoxy primer on both side of the sheet and polyester top coat 15-18 micron. The PPGL sheet shall have plastic protective guard film of minimum 25 microns to avoid scratches during transportation. The PU Foam shall be self extinguishing, fire retardant type having minimum density of 40 Kg/Cu.mt (+, -2 Kgs) The roofing panels shall be fixed to purlins using STSD screws having suitable washers & length complete as per direction of Engineer-in- Charge |
| 131 | 24.2:Providing and fixing double scaffolding system (Cup lock type ) on the exterior side of building/structure, upto 25 metre height, above ground level, including additional rows of scaffolding in stepped manner as per requirement of site, made with 40 mm dia M.S. tube, placed 1.5 metre centre to centre, horizontal & vertical tubes joint with cup & lock system with M.S. Tubes, M.S. tube challis, M.S. clamps and staircase system in the scaffolding for working platform etc. and maintaining it in a serviceable condition for execution of work of cleaning and / or pointing and / or applying chemical and removing it thereafter. The scaffolding system shall be stiffened with bracings, runners, connecting with the building etc, wherever required, if feasible, for inspection of work at required locations with essential safety features for the workmen etc., complete as per directions and approval of Engineer-in-Charge. Note: - ( 1) The evaluational area of the scaffolding shall be measured for payment purpose, (2) The payment will be made once only for execution of all items of such works. |
| 132 | 1.54.2:Wiring for group controlled (looped) light point/fan point/exhaust fan point/ call bell point (without independent switch etc.) with 1.5 sq. mm FRLS PVC insulated copper conductor single core cable in surface/ recessed steel conduit, and earthing the point with 1.5 sq. mm FRLS PVC insulated copper conductor single core cable etc. as required.Group B |
| 133 | 1.10.3:Wiring for light point/ fan point/ exhaust fan point/ call bell point with 1.5 sq.mm FRLS PVCinsulated copper conductor single core cable in surface / recessed medium class PVC conduit,with modular switch, modular plate, suitable GI box and earthing the point with 1.5 sq.mm. FRLSPVC insulated copper conductor single core cable etc as required.Group C |
| 134 | 1.14.1:Wiring for circuit/ submain wiring alongwith earth wire with the following sizes of FRLS PVC insulated copper conductor, single core cable in surface / recessed medium class PVC conduit as required2x1.5 sq.mm + 1x1.5 sq.mm earth wire |
| 135 | 1.14.2:Wiring for circuit/ submain wiring alongwith earth wire with the following sizes of FRLS PVC insulated copper conductor, single core cable in surface / recessed medium class PVC conduit as required2X 2.5 sq. mm + 1 X 2.5 sq. mm earth wire |
| 136 | 1.14.3:Wiring for circuit/ submain wiring alongwith earth wire with the following sizes of FRLS PVC insulated copper conductor, single core cable in surface / recessed medium class PVC conduit as required2 X 4 sq. mm + 1 X 4 sq. mm earth wire |
| 137 | 1.17.31:Supplying and drawing following sizes of FRLS PVC insulated copper conductor, single core cable in the existing surface / recessed steel/ PVC conduit as required.4 x 6 sq. mm |
| 138 | 1.17.3:Supplying and drawing following sizes of FRLS PVC insulated copper conductor, single core cable in the existing surface / recessed steel/ PVC conduit as required.3 X 1.5 sq.mm |
| 139 | 1.17.12:Supplying and drawing following sizes of FRLS PVC insulated copper conductor, single core cable in the existing surface / recessed steel/ PVC conduit as required.3 x 2.5 sq. mm |
| 140 | 1.24.1:Supplying and fixing following modular switch/ socket on the existing modular plate & switch box including connections but excluding modular plate etc. as required.5/6 amps switch |
| 141 | 1.24.3:Supplying and fixing following modular switch/ socket on the existing modular plate & switch box including connections but excluding modular plate etc. as required.15/16 amp switch |
| 142 | 1.24.4:Supplying and fixing following modular switch/ socket on the existing modular plate & switch box including connections but excluding modular plate etc. as required.3 pin 5/6 amp socket outlet |
| 143 | 1.24.5:Supplying and fixing following modular switch/ socket on the existing modular plate & switch box including connections but excluding modular plate etc. as required.6 pin 15/16 amp socket outlet |
| 144 | 1.24.8:Supplying and fixing following modular switch/ socket on the existing modular plate & switch box including connections but excluding modular plate etc. as required.Bell push |
| 145 | 1.26:Supplying and fixing modular blanking plate on the existing modular plate & switch box excluding modular plate as required. |
| 146 | 1.27.1:Supplying and fixing following size/ modules, GI box alongwith modular base & cover plate for modular switches in recess etc as required.1 or 2 Module (75mmX75mm) |
| 147 | 1.27.2:Supplying and fixing following size/ modules, GI box alongwith modular base & cover plate for modular switches in recess etc as required.3 Module (100mmX75mm) |
| 148 | 1.27.4:Supplying and fixing following size/ modules, GI box alongwith modular base & cover plate for modular switches in recess etc as required.6 Module (200mmX75mm) |
| 149 | 1.27.5:Supplying and fixing following size/ modules, GI box alongwith modular base & cover plate for modular switches in recess etc as required.8 Module (125mmX125mm) |
| 150 | 1.38:Supplying and fixing call bell/ buzzer suitable for single phase, 230 volts, complete as required. |
| 151 | 90.4.8:Charges for cutting holes suitable for accommodating exhaust fans up to 305mm sweep including plastering colour washing etc. as required. |
| 152 | 1.50.1:Installation of exhaust fan in the existing opening, including making good the damage connection testing, commissioning etc. as required.Upto 450 mm sweep |
| 153 | 90.11.1.10:Supply and installation of sheet steel, phosphatised and painted, dust and vermin proof enclosure of MCB DB including copper /brass bus bar, neutral link, earth bus and DIN rail suitable for fixing MCB/ isolator etc. fixed on wall using suitable anchor bolts or fixed in recess including cutting hole on the wall , making good the damages, colour washing etc. as required4 way (8+12) - three phase double cover (IP 42/43) |
| 154 | 90.11.1.11:Supply and installation of sheet steel, phosphatised and painted, dust and vermin proof enclosure of MCB DB including copper /brass bus bar, neutral link, earth bus and DIN rail suitable for fixing MCB/ isolator etc. fixed on wall using suitable anchor bolts or fixed in recess including cutting hole on the wall , making good the damages, colour washing etc. as required6 way (8+18) - three phase double cover (IP 42/43) |
| 155 | 90.11.1.20:Supply and installation of sheet steel, phosphatised and painted, dust and vermin proof enclosure of MCB DB including copper /brass bus bar, neutral link, earth bus and DIN rail suitable for fixing MCB/ isolator etc. fixed on wall using suitable anchor bolts or fixed in recess including cutting hole on the wall , making good the damages, colour washing etc. as required8 way (8+12) - double cover TPN vertical DB with provision for fixing MCCB as incomer and SP/ TP MCB as outgoing (IP 42/43) |
| 156 | 2.12.1:Supplying and fixing following rating, double pole, 240 volts, isolator in the existing MCB DB complete with connections, testing and commissioning etc. as required.40 amps |
| 157 | 2.14.2:Supplying and fixing following rating, double pole, (single phase and neutral),240 volts, residual current circuit breaker (RCCB), having a sensitivity current 30 milli amperes in the existing MCB DB complete with connections, testing and commissioning etc. as required.40 amps |
| 158 | 2.15.3:Supplying and fixing following rating, four pole, (three phase and neutral), 415volts, residual current circuit breaker (RCCB), having a sensitivity current 30milli Amperes in the existing MCB DB complete with connections, testing andcommissioning etc. as required.63 amps |
| 159 | 2.15.2:Supplying and fixing following rating, four pole, (three phase and neutral), 415volts, residual current circuit breaker (RCCB), having a sensitivity current 30milli Amperes in the existing MCB DB complete with connections, testing andcommissioning etc. as required.40 amps |
| 160 | 90.12.7.49:Supply, laying and clamping of 1 no. PVC insulated and PVC sheathed armoured aluminium power cable, 1.1 KV grade of the following sizes using clamps noted along with the cables, spacing of clamps not exceeding 60cms, making good the damages , colour washing etc. as required.4 core 185 sq mm |
| 161 | 90.12.41.34:Supply, laying and clamping of 1 no. PVC insulated and PVC sheathed armoured copper power cable, 1.1 KV grade of the following sizes using clamps noted along with the cables, spacing of clamps not exceeding 60cms, making good the damages , colour washing etc. as required.4 core 16 sq mm with factory made clamp |
| 162 | 90.12.1.45:Supply & laying of one number PVC insulated and PVC sheathed armoured aluminium power cable of 1.1KV grade of the following sizes in ground including excavation of trench of size 35 x 75 cm, refilling the trench etc. as required but excluding sand cushioning and protective covering (in ordinary soil).4 core 70 sq mm |
| 163 | 9.1.23:Supplying and making end termination with brass compression gland and aluminium lugs for following size of PVC insulated and PVC sheathed / XLPE aluminium conductor cable of 1.1 KV grade as required.3 1/2X 70 sq. mm (38mm) |
| 164 | 9.1.32:Supplying and making end termination with brass compression gland and aluminium lugs for following size of PVC insulated and PVC sheathed / XLPE aluminium conductor cable of 1.1 KV grade as required.4 X 10 sq. mm (25mm) |
| 165 | 9.1.33:Supplying and making end termination with brass compression gland and aluminium lugs for following size of PVC insulated and PVC sheathed / XLPE aluminium conductor cable of 1.1 KV grade as required.4 X 16 sq. mm (28mm) |
| 166 | 90.12.24.7:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.35 sq mm |
| 167 | 90.12.24.5:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.16 sq mm |
| 168 | 90.12.24.4:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.10 sq mm |
| 169 | 90.16.1.3:Supply and drawing bare earthing conductors of the following sizes along with wiring/ cables and giving connection as required3.15 mm copper conductor ( 10 SWG) |
| 170 | 90.16.5.3:Supply of superior quality copper earth socket for the following size of earth conductor including crimping etc. as required.3.15 mm (10 SWG) |
| 171 | 5.6:Earthing with copper earth plate 600 mm X 600 mm X 3 mm thick including accessories, and providing masonry enclosure with cover plate having locking arrangement and watering pipe of 2.7 metre long etc. with charcoal/ coke and salt as required. |
| 172 | 90.16.3.9:Supply and clamping the following size of strips on surface of wall/ parapet/ existing cable tray using clamps fabricated from 20 X 3 mm GI flat duly painted or heavy duty GI spacer saddles spacing of clamps not exceeding 1 m, making good the damages, colour washing etc. as required ( for horizontal run)25 x 3 mm copper strip |
| 173 | 90.16.2.4:Supply and laying of the following sizes of B class GI pipe laid 0.6 m below ground level including excavating trench of suitable size and refilling etc, as required.40 mm dia G.I. pipe ( medium class) |
| 174 | OD156110/2026-2027:Recess mounted LED luminaire with full diffuser type upto 36 W (max), 2x2 LED, Cool White, 5700k, White powder coated CRCA housing, WIPRO CRCO10R038HP57GL1 or Bajaj BZRSQL 36W GZi CW or Equivalent |
| 175 | OD156113/2026-2027:Recess mounted circular LED retrofit downlighter, 15W, powder coated white finish, BPSLR 15W WH Bajaj or Equivalent |
| 176 | OD156114/2026-2027:6 kg Fire Extinguisher ABC Powder Type With NABL Approval |
| 177 | OD156095/2026-2027:5 kg Fire Extinguisher Carbon Di-oxide Type With NABL Approval |
| 178 | 17.2.14:INTELLIGENT FIRE ALARM SYSTEMSupplying, installation, testing & commissioning of addressable manual call point complete as required. |
| 179 | 17.1.2:FIRE DETECTION AND ALARM SYSTEM (CONVENTIONAL)Supplying, installation, testing & commissioning of smoke detector with built-in LED and mounting base complete with all connections etc. as required. |
| 180 | 17.2.5:INTELLIGENT FIRE ALARM SYSTEMSupplying, installation, testing & commissioning of response indicator on surface/ recessed MS Box having two LED, metallic cover complete with all connections etc as required. |
| 181 | 17.1.9.3:Supplying, installation, testing & commissioning sector panel suitable for following zones, complete with visual indications for short circuit fault, open circuit fault, fire condition and all other standard facilities as per IS:2189 with mimic diagram for all area/zone covered, complete with all connections, interconnections as required.10 Zone |
| 182 | 17.1.6:FIRE DETECTION AND ALARM SYSTEM (CONVENTIONAL)Supplying, installation, testing & commissioning fire alarm sounder with facility to make announcement, mounted in M.S. box (16 SWG) with hinged cover plate & suitable for operation with amplifier i/c line matching transformer etc. complete as required. |
| 183 | 17.5.3.3:Supplying and drawing of cable Fire Retardant PVC insulated copper conductor cable in the existing surface / recessed steel conduit of following pairs, cores and size including connections and interconnections etc. as required.speaker cable Three pair, 2-core, 1.5 sqmm |
| 184 | 17.1.5:FIRE DETECTION AND ALARM SYSTEM (CONVENTIONAL)Supplying, installation, testing & commissioning response indicator on surface/ recess MS box having two LEDs metallic cover complete with all connections etc. as required. |
| 185 | 17.1.9.1:Supplying, installation, testing & commissioning sector panel suitable for following zones, complete with visual indications for short circuit fault, open circuit fault, fire condition and all other standard facilities as per IS:2189 with mimic diagram for all area/zone covered, complete with all connections, interconnections as required.4 Zone |
| 186 | 90.14.10.6:Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.125A, 35/36 kA (Ics=100%Icu), 3 pole, current limiting type MCCB with microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC |
| 187 | OD156102/2026-2027:Preparation of electrical drawing, submission of drawings to Electrical Inspectorate, KSEB and obtaining approval, arranging inspection for entire installation and other utilities. Obtaining road cutting permission obtaining sanction orders for the electrical installation, Testing, Commissioning & As-built documentation,power allocation, energisation of the installation complete. Including all incidental expenses except statutory fees payable to the statutory bodies. The statutory fees will be reimbursed on production of valid receipts. |
| 188 | 1.19:Supplying and drawing co-axial TV cable RG-6 grade, 0.7 mm solid copper conductor PE insulated, shielded with fine tinned copper braid and protected with PVC sheath in the existing surface/ recessed steel/ PVC conduit as required. |
| 189 | 1.21.1:Supplying and fixing of following sizes of medium class PVC conduit along with accessories in surface/ recess including cutting the wall and making good the same in case of recessed conduit as required.20 mm |
| 190 | 1.24.6:Supplying and fixing following modular switch/ socket on the existing modular plate & switch box including connections but excluding modular plate etc. as required.Telephone socket outlet |
| 191 | 1.24.7:Supplying and fixing following modular switch/ socket on the existing modular plate & switch box including connections but excluding modular plate etc. as required.TV antenna socket outlet |
| 192 | 1.21.2:Supplying and fixing of following sizes of medium class PVC conduit along with accessories in surface/ recess including cutting the wall and making good the same in case of recessed conduit as required.25 mm |
| 193 | 1.21.3:Supplying and fixing of following sizes of medium class PVC conduit along with accessories in surface/ recess including cutting the wall and making good the same in case of recessed conduit as required.32 mm |
| 194 | 90.6.9.5:Supplying & Drawing unarmored PVC insulated and & PVC sheathed copper conductor Telephone Pair Cable conforming to ITD specification for indoor application through the existing conduit5 pair |
| 195 | 1.53.1:Supplying and drawn of UTP 4 pair CAT 6 LAN cable in the existing surface/recessed steel/PVC conduit as required1 run of Cable |
| 196 | OD156096/2026-2027:CAT 6 RJ 45 Computer jack modular SOCKET OUTLET |
| 197 | OD156111/2026-2027:Supply of branded 2 meter patch cords with cat6 cable |
| 198 | OD156115/2026-2027:Patch panel- 24-Port patch panel with cable tray-Rack mountable |
| 199 | OD156109/2026-2027:24-Port Gigabit manageable smart Network Switch with 2 SFPs.Rack mountable |
| 200 | OD158671/2026-2027:Supply of LT Switch board cubicle type, made out of 2 mm thick CRCA sheet, totally enclosed, IP42, free standing, wall/floor mounting, dust and vermin proof, powder coated, indoor, compartmentalized, suitable for operation on 3 phase and neutral, 415 V, 50Hz AC system with bus-bars extendable on both sides, including all the necessary internal wiring , suitable spreader etc. The panel shall be fabricated from a firm having CPRI certification with test certificates for similar panel(required for short circuit rating, temperature rise and IP classification) and with following switchgears, metering instruments and accessories etc. complete as required as per drawing & specification.RYB indicating lamps LED type4A SP MCB (0-500) V Class 1 with 90 deg defl. Scale.(0-200)A Class 1 with 90 deg defl. Scale, Voltmeter selector switch 200/5 A 5 VA Class .5S CTs ON/OFF INDICATION LAMP, Digital multi-function meter to read V,A.Hz,PF,Kwh,Incomer 200A FP Moulded Case Circuit Breaker with thermal release with OL and SC protection neutral link (Ics =100%Icu= 25kA) -1 No’s200A Copper TPN busbar chamber with Copper busbar-Out Goings 100 A FP Moulded Case Circuit Breaker with thermal release with OL and SC protection neutral link (Ics =100%Icu= 25kA) - 3 No’s .63 A FP Moulded Case Circuit Breaker with thermal based release with OL and SC protection neutral link (Ics =100%Icu= 25kA) - 3 No’sThe panel shall be Top Entry/Exit and suitable for termination of cable system as per SLD and technical specification. |
| 201 | OD158672/2026-2027:Supply of LT Switch board cubicle type, made out of 2 mm thick CRCA sheet, totally enclosed, IP42, free standing, wall/floor mounting, dust and vermin proof, powder coated, indoor, compartmentalized, suitable for operation on 3 phase and neutral, 415 V, 50Hz AC system with bus-bars extendable on both sides, including all the necessary internal wiring , suitable spreader etc. The panel shall be fabricated from a firm having CPRI certification with test certificates for similar panel (required for short circuit rating, temperature rise and IP classification) and with following switchgears, metering instruments and accessories etc. complete as required as per drawing & specification. RYB Indicating lamps LED type, 4A SP MCB, Digital Multi-function Meter to read V,A,Hz,PF,Kwh, Surge Protection Device, EFR, 400/5 A ,5 VA Class .5S CTs, ON/OFF INDICATION LAMP. Incomer -Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required. 250A, 25 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC with copper TPN busbar chamber with copper busbar. Outgoings-1)Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.200A, 25 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with thermal/microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC-1No 2) Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.100A, 16 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with thermal based release with overload setting of having adjustable OL & SC-2 Nos 3)63 A 4pole MCCB with thermal release with OL & SC protection nuetral link (Ics=100%, Icu 16kA)-3 Nos 4) Spare-1no |
| 202 | OD158673/2026-2027:Supply of LT Switch board cubicle type, made out of 2 mm thick CRCA sheet, totally enclosed, IP42, free standing, wall/floor mounting, dust and vermin proof, powder coated, indoor, compartmentalized, suitable for operation on 3 phase and neutral, 415 V, 50Hz AC system with bus-bars extendable on both sides, including all the necessary internal wiring , suitable spreader etc. The panel shall be fabricated from a firm having CPRI certification with test certificates for similar panel (required for short circuit rating, temperature rise and IP classification) and with following switchgears, metering instruments and accessories etc. complete as required as per drawing & specification. RYB Indicating lamps LED type, 4A SP MCB, Digital Multi-function Meter to read V,A,Hz,PF,Kwh, Surge Protection Device, EFR, 400/5 A ,5 VA Class .5S CTs, ON/OFF INDICATION LAMP, Incomer -Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.320A/400A, 50/55 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC Outgoings-1)Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.160A, 35/36 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC-1No2) Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.16A -100A, 35/36 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC-3 Nos3)Spare- 2Nos |
| 203 | 9.1.3:Supplying and making end termination with brass compression gland and aluminium lugs for following size of PVC insulated and PVC sheathed / XLPE aluminium conductor cable of 1.1 KV grade as required.2 X 16 sq. mm (22mm) |
| 204 | 4.4.3:Supplying and installing following size of perforated painted with powder coatingM.S. cable trays Cross Member with perforation not more than 17.5%, joined with connectors, suspended from the ceiling with M.S. suspenders including bolts &nuts, painting suspenders etc as required.225 mm width X 50 mm depth X 1.6 mm thickness |
| 205 | 16.16:Supplying & fixing of powder coated extruded aluminum Return Air Grills with louvers but without volume control dampers complete as required. |
| 206 | 90.12.7.35:Supply, laying and clamping of 1 no. PVC insulated and PVC sheathed armoured aluminium power cable, 1.1 KV grade of the following sizes using clamps noted along with the cables, spacing of clamps not exceeding 60cms, making good the damages , colour washing etc. as required.3.5 core 240 sq mm |
| 207 | 1.27.6:Supplying and fixing following size/ modules, GI box alongwith modular base & cover plate for modular switches in recess etc as required.12 Module (200mmX150mm) |
| 208 | 9.1.29:Supplying and making end termination with brass compression gland and aluminium lugs for following size of PVC insulated and PVC sheathed / XLPE aluminium conductor cable of 1.1 KV grade as required.3 1/2X 240 sq. mm (62mm) |
| 209 | 90.12.41.36:Supply, laying and clamping of 1 no. PVC insulated and PVC sheathed armoured copper power cable, 1.1 KV grade of the following sizes using clamps noted along with the cables, spacing of clamps not exceeding 60cms, making good the damages , colour washing etc. as required.4 core 35 sq mm with factory made clamp |
| 210 | 90.12.38.7:Supply and Laying of one number additional PVC insulated and PVC sheathed armoured copper power cable of 1.1KV grade of the following sizes in ground in the same trench in one tier horizontal formation including widening the trench by 20 cm, refilling the trench etc. as required but excluding sand cushioning and protective covering (in ordinary soil)3 core 10 sq mm |
| 211 | 1.49:Numbering of ceiling fan / exhaust fan/ fluorescent fittings as required. |
| 212 | 4.1.3:Supplying and installing following size of perforated painted with powder coatingM.S. cable trays with perforation not more than 17.5%, in convenient sections,joined with connectors, suspended from the ceiling with M.S. suspendersincluding bolts & nuts, painting suspenders etc as required.225 mm width X 50 mm depth X 1.6 mm thickness |
| 213 | 4.3.2:Supplying and installing following size of perforated painted with powder coatingM.S. cable trays Tee with perforation not more than 17.5%, joined with connectors, suspended from the ceiling with M.S. suspendersincluding bolts & nuts, painting suspenders etc as required.150 mm width X 50 mm depth X 1.6 mm thickness |
| 214 | 4.3.3:Supplying and installing following size of perforated painted with powder coatingM.S. cable trays Tee with perforation not more than 17.5%, joined with connectors, suspended from the ceiling with M.S. suspendersincluding bolts & nuts, painting suspenders etc as required.225mm width X 50 mm depth X 1.6 mm thickness |
| 215 | 90.12.24.10:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.95 sq mm |
| 216 | 90.12.24.11:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.120 sq mm |
| 217 | 90.12.24.12:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.150 sq mm |
| 218 | 90.12.24.13:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.185 sq mm |
| 219 | 90.12.24.14:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.240 sq mm |
| 220 | 90.12.24.6:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.25 sq mm |
| 221 | 90.12.24.8:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.50 sq mm |
| 222 | 90.12.24.9:Supply of superior quality copper tubular cable socket and making core termination of cables of the following sizes including crimping etc. as required.70 sq mm |
| 223 | OD158671/2026-2027:Supply of LT Switch board cubicle type, made out of 2 mm thick CRCA sheet, totally enclosed, IP42, free standing, wall/floor mounting, dust and vermin proof, powder coated, indoor, compartmentalized, suitable for operation on 3 phase and neutral, 415 V, 50Hz AC system with bus-bars extendable on both sides, including all the necessary internal wiring , suitable spreader etc. The panel shall be fabricated from a firm having CPRI certification with test certificates for similar panel(required for short circuit rating, temperature rise and IP classification) and with following switchgears, metering instruments and accessories etc. complete as required as per drawing & specification.RYB indicating lamps LED type4A SP MCB (0-500) V Class 1 with 90 deg defl. Scale.(0-200)A Class 1 with 90 deg defl. Scale, Voltmeter selector switch 200/5 A 5 VA Class .5S CTs ON/OFF INDICATION LAMP, Digital multi-function meter to read V,A.Hz,PF,Kwh,Incomer 200A FP Moulded Case Circuit Breaker with thermal release with OL and SC protection neutral link (Ics =100%Icu= 25kA) -1 No’s200A Copper TPN busbar chamber with Copper busbar-Out Goings 100 A FP Moulded Case Circuit Breaker with thermal release with OL and SC protection neutral link (Ics =100%Icu= 25kA) - 3 No’s .63 A FP Moulded Case Circuit Breaker with thermal based release with OL and SC protection neutral link (Ics =100%Icu= 25kA) - 3 No’sThe panel shall be Top Entry/Exit and suitable for termination of cable system as per SLD and technical specification. |
| 224 | OD158672/2026-2027:Supply of LT Switch board cubicle type, made out of 2 mm thick CRCA sheet, totally enclosed, IP42, free standing, wall/floor mounting, dust and vermin proof, powder coated, indoor, compartmentalized, suitable for operation on 3 phase and neutral, 415 V, 50Hz AC system with bus-bars extendable on both sides, including all the necessary internal wiring , suitable spreader etc. The panel shall be fabricated from a firm having CPRI certification with test certificates for similar panel (required for short circuit rating, temperature rise and IP classification) and with following switchgears, metering instruments and accessories etc. complete as required as per drawing & specification. RYB Indicating lamps LED type, 4A SP MCB, Digital Multi-function Meter to read V,A,Hz,PF,Kwh, Surge Protection Device, EFR, 400/5 A ,5 VA Class .5S CTs, ON/OFF INDICATION LAMP. Incomer -Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required. 250A, 25 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC with copper TPN busbar chamber with copper busbar. Outgoings-1)Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.200A, 25 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with thermal/microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC-1No 2) Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.100A, 16 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with thermal based release with overload setting of having adjustable OL & SC-2 Nos 3)63 A 4pole MCCB with thermal release with OL & SC protection nuetral link (Ics=100%, Icu 16kA)-3 Nos 4) Spare-1no |
| 225 | OD158673/2026-2027:Supply of LT Switch board cubicle type, made out of 2 mm thick CRCA sheet, totally enclosed, IP42, free standing, wall/floor mounting, dust and vermin proof, powder coated, indoor, compartmentalized, suitable for operation on 3 phase and neutral, 415 V, 50Hz AC system with bus-bars extendable on both sides, including all the necessary internal wiring , suitable spreader etc. The panel shall be fabricated from a firm having CPRI certification with test certificates for similar panel (required for short circuit rating, temperature rise and IP classification) and with following switchgears, metering instruments and accessories etc. complete as required as per drawing & specification. RYB Indicating lamps LED type, 4A SP MCB, Digital Multi-function Meter to read V,A,Hz,PF,Kwh, Surge Protection Device, EFR, 400/5 A ,5 VA Class .5S CTs, ON/OFF INDICATION LAMP, Incomer -Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.320A/400A, 50/55 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC Outgoings-1)Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.160A, 35/36 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC-1No2) Supply, conveyance and fixing the following types & current rated control gears & switchgears conforming to IS 13947 suitable for 440 V, 50 Hz, AC supply in the existing panel assembly as required.16A -100A, 35/36 kA (Ics=100%Icu), 4 pole, current limiting type MCCB with microprocessor based release with overload setting of 50 - 100% having adjustable OL & SC-3 Nos3)Spare- 2Nos |
| 226 | OD158710/2026-2027:Supply, laying and clamping of 1 no. PVC insulated and PVC sheathed unarmoured copper power cable, 70 sqmm ISI Marked FR LS PVC insulated single core copper conductor cable using clamps noted along with the cables, spacing of clamps not exceeding 60cms, making good the damages , colour washing etc. as required. |
| 227 | OD158711/2026-2027:Supply, laying and clamping of 1 no. PVC insulated and PVC sheathed unarmoured copper power cable, 1.1 KV grade of the following sizes using clamps noted along with the cables, spacing of clamps not exceeding 60cms, making good the damages , colour washing etc. as required.3.5 core 50 sq mm with factory made clamp |
| 228 | OD158712/2026-2027:Supply, laying and clamping of 1 no. PVC insulated and PVC sheathed unarmoured copper power cable, 1.1 KV grade of the following sizes using clamps noted along with the cables, spacing of clamps not exceeding 60cms, making good the damages , colour washing etc. as required.3.5 core 35 sq mm with factory made clamp |
| 229 | OD158743/2026-2027:Supply, laying and clamping of 1 no. PVC insulated and PVC sheathed unarmoured copper power cable, 25 sqmm ISI Marked FR LS PVC insulated single core copper conductor cable using clamps noted along with the cables, spacing of clamps not exceeding 60cms, making good the damages , colour washing etc. as required. |
| 230 | 16.5.2:Supplying, laying,fixing, testing and commissioning of following thicknessInsulation with nitrile rubber of thickness 19mm on existing pipe |
| 231 | OD156296/2026-2027:Charges for cutting holes suitable for accommodating the supply ducts including plastering, colour washing etc. as required. |
| 232 | OD156414/2026-2027:Supply,Conveyance testing and commissioning of 2 Ton 3 Star InverterType,Split AC unit with Copper Tube cooling coils ,Bluestar/LG/Voltas/Mitsubishi(Dual Inverter Type ) or SuperiorBrands,Cooling Capacity 6000kw/hr with rotary/scroll typeCompressors, Microprocessor controlled codeless remote,MoistureRemoval above 1.8, CFM above 500/450/350,LED/LCD display. |
| 233 | OD156415/2026-2027:Supply,Conveyance testing and commissioning of 5KVA VoltageStabilizer Wall mount type with Time delay,High& lowvoltage& Thermal overload protection,Input voltage 170-220V,Output Voltage 200-240 V,High Voltage Cut Off 240,Efficiency:>95%,Frequency : 50Hz |
| 234 | OD156416/2026-2027:Supply, installation, testing and commissioning UPVC drain pipe 6kg/sq.cmof suitable dia connecting the drain pan of the indoor units to nearest drain point, with 12mm thick closed cell elastomeric insulation as per specification |
| 235 | OD156417/2026-2027:SITC of 19 mm thick nitrile rubber for under deck insulation |
| 236 | OD156453/2026-2027:Motorized volume control damper for individual operation of ductable split air conditioners |
| 237 | OD156454/2026-2027:Supplying and fixing outdoor unit stand for 1 Ton to 2 Ton split type air conditioner, fabricated from MS angle of suitable size duly welded, provided with suitable stiffeners and cross members, painted with two coats of anti-corrosive primer and powder coated, complete with necessary anchor fasteners/bolts and rubber vibration pads for mounting the concerned AC outdoor unit, including transportation, labour and materials as required |
| 238 | OD156455/2026-2027:Additional Copper Pipe with Superlon Insulation for Suction and Discharge suitable for 1/1.5/2 TR Split AC units |
| 239 | OD156456/2026-2027:Electrical Cabling between indoor and outdoor units, with required size cables/wires(Havells/RR Kabel/V-guard FRLS wires) |
| 240 | OD156457/2026-2027:Providing and fixing of Thermal insulation lining with 13mm thick nitrile rubber closed cell type with adhesive, on supply duct |
| 241 | OD156458/2026-2027:Providing and fixing of Acoustic Duct lining with 10mm thick nitrile rubber open cell type inside the ducts |
| 242 | OD156463/2026-2027:Supply, installation, testing and commissioning (SITC) of 5.5 TR x 2 No's or higher capacity air cooled ductable split air conditioner with ceiling suspended IDU and all standard fittings, IDU SS drain pan, as per specifications complete with suitable capacity drive motor, air cooled condenser with all accessories, hermetically sealed scroll compressor, electronic microprocessor pendant controller with digital display, first charge refrigerent gas, vibration isolator pads, structural supports for fixing IDU/ODU as shown in the drawing, safety devices including thermal protector, pressure relief valve, over load relays single phase preventor etc, power wiring(max 3m from isolator) and control cabling(indoor to outdoor),thermostat wiring, high-low pressure cut out, suitable for three phase 440V supply etc complete as required |
| 243 | OD156464/2026-2027:Supply, installation, testing and commissioning of volume control duct damper made of extruded aluminium for ducts/collars to be provided with suitable links, lebers and quadrants for manual control of volume of air flow and for proper balancing of the air distribution in the main duct complete with neoprene rubber gaskets, nuts, bolts, screws linkages, flanges etc., as per specifications.The material shall be paramagnetic |
| 244 | OD156465/2026-2027:Supply, installation, testing and commissioning of suitable sized hard drawn 18 SWG copper refrigerant piping(both suction and liquid piping) including 25mm thick closed cell elastomeric insulation as per specification for 11/8.75/ 5.5 TR machine between indoor & outdoor units as per site conditions all complete. |
| 245 | OD156466/2026-2027:Supply, installation, testing and commissioning (SITC) of 8.75 TR x 2 No's or higher capacity air cooled ductable split air conditioner with ceiling suspended IDU and all standard fittings, IDU SS drain pan, as per specifications complete with suitable capacity drive motor, air cooled condenser with all accessories, hermetically sealed scroll compressor, electronic microprocessor pendant controller with digital display, first charge refrigerent gas, vibration isolator pads, structural supports for fixing IDU/ODU as shown in the drawing, safety devices including thermal protector, pressure relief valve, over load relays single phase preventor etc, power wiring(max 3m from isolator) and control cabling(indoor to outdoor),thermostat wiring, high-low pressure cut out, suitable for three phase 440V supply etc complete as required |
| 246 | OD156467/2026-2027:Supply, installation, testing and commissioning (SITC) of 11 TR x 2 No's or higher capacity air cooled ductable split air conditioner with ceiling suspended IDU and all standard fittings, IDU SS drain pan, as per specifications complete with suitable capacity drive motor, air cooled condenser with all accessories, hermetically sealed scroll compressor, electronic microprocessor pendant controller with digital display, first charge refrigerent gas, vibration isolator pads, structural supports for fixing IDU/ODU as shown in the drawing, safety devices including thermal protector, pressure relief valve, over load relays single phase preventor etc, power wiring(max 3m from isolator) and control cabling(indoor to outdoor),thermostat wiring, high-low pressure cut out, suitable for three phase 440V supply etc complete as required |
| 247 | OD156469/2026-2027:Suitable De-Humidifier units-Extraction atleast 35litres/day,compressor-rotary,Airflow atleast 220Mcube/hr,reputed brand ,input voltage 230V AC |
| 248 | OD156470/2026-2027:Supply, installation, testing and commissioning of Powder coated extruded aluminimum Continuous grilles 150 mm high for supply air with aluminium damper as per spec |
| 249 | OD156471/2026-2027:Supplying, fixing testing commissioning of supply of air Diffusers of powder coated aluminium without volume control dampers with anti smudge ring & removable core. |
| 250 | OD156472/2026-2027:Supplying, fixing testing commissioning of Return air Diffusers of powder coated aluminium without volume control dampers with anti smudge ring & removable core |
| 251 | OD156473/2026-2027:Supply, installation, testing and commissioning of Powder coated extruded aluminium Continuous grilles 150mm high for return air as per spec |
| 252 | 16.12.1.1:Supply, installation, balancing and commissioning of factory fabricated GSS sheet metal rectangular/round ducting complete with neoprene rubber gaskets, elbows, splitter dampers, vanes, hangers, supports etc. as per approved drawings and specifications of following sheet thickness complete as required.Factory fabricated GSS sheet metal rectangular ducting as per IS 655-2006 Thickness 0.63 mm sheet |
| 253 | 16.12.1.2:Supply, installation, balancing and commissioning of factory fabricated GSS sheet metal rectangular/round ducting complete with neoprene rubber gaskets, elbows, splitter dampers, vanes, hangers, supports etc. as per approved drawings and specifications of following sheet thickness complete as required.Factory fabricated GSS sheet metal rectangular ducting as per IS 655-2006 Thickness 0.80 mm sheet |
| 254 | OD158751/2026-2027:Supplying and fixing powder-coated MS outdoor unit stand suitable for 5.5 TR ductable split AC unit, fabricated from heavy-duty angle sections with adequate stiffeners, complete with ribbed rubber vibration isolation pads, anchoring bolts, and anti-corrosive paint finish as required |
| 255 | OD158752/2026-2027:Supplying and fixing powder-coated MS outdoor unit stand suitable for 11/8.75 TR ductable split AC unit, fabricated from heavy-duty angle sections with adequate stiffeners, complete with ribbed rubber vibration isolation pads, anchoring bolts, and anti-corrosive paint finish as required |
| 256 | OD158753/2026-2027:Supply, installation, testing and commissioning of 8.5 TR inverter type air-cooled ductable split air-conditioning unit, comprising ceiling-suspended indoor unit and air-cooled outdoor unit(both copper coils), complete with all standard accessories and safety devices, conforming to relevant IS/IEC standards and manufacturer’s specifications. The unit shall be provided with inverter-driven hermetically sealed scroll compressor and shall operate on R-410A / R-32 refrigerant with factory first charge. The indoor unit shall be complete with stainless steel drain pan, inverter-compatible blower motor and factory insulation, while the outdoor unit shall be provided with blue-fin / gold-fin coated condenser coils and corrosion-resistant powder-coated enclosure, suitable for operation at ambient temperatures up to 50°C. The system shall be suitable for 3-phase, 415–440 V, 50 Hz power supply with wide voltage operating range and shall include microprocessor-based wired controller with digital display. Safety features such as high and low pressure cut-outs, overload relays, phase sequence / single-phasing preventer and thermal protection for compressor and motors shall be provided. The work shall include vibration isolation pads for indoor and outdoor units, structural supports for mounting as per site requirements, power wiring up to 3 metres from isolator, control cabling between indoor and outdoor units, thermostat wiring, testing, commissioning and handing over in working condition, complete in all respects, as per manufacturer’s recommendations and directions of Engineer-in-Charge. |
| 257 | The scrub station shall be made of high-grade stainless steel 304 with stain polish finish and wired to the current IEE regulations & BS 7971. The scrub stations control system shall be based on advanced microprocessor. The scrub station shall be sensor-activated 24VAC. The scrub station shall be provided with user selectable time out control. The supply shall not be interrupted in case of power failure or maintenance. The station shall be equipped with thermostatic mixing valve to controls the temperature of hot / cold water supply. The washing faucet shall be made of Chrome plated brass hand. The station shall be equipped with soap dispensers.Two bay 63”x24”31”(Not taken for evaluation) |
| 258 | ETO Sterlizer, sealing machine & accessories ( Not taken for evaluation ) |
| 259 | Latest Dose reduction / image quality improvement packages like Autoright / Clarity / OPTIQ. ( Not taken for evaluation ) |
| 260 | IABP as per the specification ( Not taken for evaluation) Please Enable Macros to View BoQ information |
AI Tender Summary
Tender Timeline
Tender Published
Tender notice published.
CompletedBid Submission Deadline
Online submission via eProcurement portal.
Upcoming · 19 daysBid Opening Date
Technical bids will be opened and evaluated.
Upcoming · 21 daysTender Documents
Download All (ZIP) ↓We takes all possible care for accurate & authentic tender information. However users are requested to refer Original source of Tender Notice / Tender Document published by Tender Issuing Agency before taking any call regarding this tender