Bids Are invited for Integrated Digital Colposcope Thermocoagulator - Its A Compact Handheld Device With One Button Turn On/Offfunction That Runs On Batteries, Featuring Large 5 Inch Lcd Touch Display And An Optical Camera To Aid In Diagnosis And Tre
Tender Overview
Project Description
Bids Are invited for Integrated Digital Colposcope Thermocoagulator - Its A Compact Handheld Device With One Button Turn On/Offfunction That Runs On Batteries, Featuring Large 5 Inch Lcd Touch Display And An Optical Camera To Aid In Diagnosis And Treatment Of Cervical Cancer. , It Can Be Used The Colposcopy And Thermal Ablation Modes Independently. , It Can Be Used For Digital Colposcopy, Thermal Ablation, Or Both. , The Optical Camera Should Have Resolution Of 2880*2016 Px , Material: Polycarbonate , Method Of Visualization: Led Illumination With Direct Observation With Unaided Eye And/Or Camera Assisted Visualization With Display , Camera Features: , It Should Have Single Touch Green Light Filter Feature. , The Capture Button Should Have 8Ms Shutter Speed To Avoid Motion Blur. , Should Support Hand Gesture To Capture The Image. , It Shall Be Stand-Alone Unit Consisting Of Battery Back-Up With Thermos Probes. , It Shall Be A Fully Mobile System With Interchangeable Probes Designed For The Use Of Ablative Procedures. , The Unit Shall Be Light Weight And Easy To Carry. , Unit Shall Support Different Diameter Probe Consisting Two Small And Two Big Probes , Therapy Probe Must Be Made Of Durable Robust Material That Can Be Sterilized And Constructed With Anti-Sticks Surface And Be Biocompatible. , Treatment Temperature Range Must Be Selectable Between 1000C To 1200C And Treatment Time 20 To 60 Sec. , Must Have Inbuilt Wi-Fi System For Transferring Of The Data , Must Have Bluetooth Features , The System Should Have Future Software Updatability System With Cloud And Ai Facility . , Should Have Hdmi Facility To Connect With Exterior Display Devices For Wide Viewing. , Bidder Shall Indicate The Details Of Probes To Be Supplied , System Should Be Supplied With Transport Case For Easy Transportation. , Instruction For Use Should Be Embedded On The Software For Easy To Use. , Accessories, Spares And Consumables , Must Submit Iso 9001 Or Iso 13485:2003/Ac:2007 For Medical Devices And Us Fda 510(K) Clearance Must Be There , The Device Should Be Battery Operated And Hand Held Type Device. , It Should Consist Of Main Unit With Removable Probe And Removal Battery Pack. , It Should Have Safety Feature For Guiding Led Light, Auto Shutoff, Low Battery Indicator Etc. , The Device Must Have Inbuilt Timer Facility For Treatment. , The Probe Tip Used In The Device Should Produce At Least 100 Degree Centigradefor Destruction Of Human Tissue. , The Device Should Have A Safety Mechanism, It Should Be Ablestop The Procedure Anytime During The Treatment And The Probe Should Cool Down To A Safe Temperature (Contactable With Bare Hands) Within 20Sec , The Temperature Of Probe Shaft Should Not Reach Beyond 43 Degree Centigrade For Safety Of The Patient. Only Tip Of Probe Should Be Maintained Target Temperature. , Probe Shaft Should Be Usable Without Any Extra/External Protecting Covers. , Probe Shaft Should Be Usable Without Any Extra/External Protecting Covers. , The Probe Should Be Sterilized With Sterilizers Like Cidex. , Should Be Able To Produce Heat After The Application On Tissue, No Preheating Should Be Required. , Should Consist Of Only Following Reusable Accessories: , Quoted Model Should Be Usfda Or Ce Certified. , Quoted Model Should Be Iso13485 Certified.
BOQ
| Sl. No. | Item Title | Item Description |
| 1 | It’s a compact handheld device with one button turn on/offfunction that runs on batteries, featuring large 5 inch LCD touch display and an optical camera to aid in diagnosis and treatment of cervical cancer. | It’s a compact handheld device with one button turn on/offfunction that runs on batteries, featuring large 5 inch LCD touch display and an optical camera to aid in diagnosis and treatment of cervical cancer. |
| 2 | It can be used the colposcopy and thermal ablation modes independently. | It can be used the colposcopy and thermal ablation modes independently. |
| 3 | It can be used for digital colposcopy, thermal ablation, or both. | It can be used for digital colposcopy, thermal ablation, or both. |
| 4 | The optical camera should have resolution of 2880*2016 PX | The optical camera should have resolution of 2880*2016 PX |
| 5 | Material: Polycarbonate | Material: Polycarbonate |
| 6 | Method of Visualization: LED Illumination with Direct Observation with Unaided Eye and/or Camera Assisted visualization with Display | Method of Visualization: LED Illumination with Direct Observation with Unaided Eye and/or Camera Assisted visualization with Display |
| 7 | Camera Features: | Camera Features: |
| 8 | It should have single touch green light filter feature. | It should have single touch green light filter feature. |
| 9 | The capture button should have 8ms shutter speed to avoid motion blur. | The capture button should have 8ms shutter speed to avoid motion blur. |
| 10 | Should support hand gesture to capture the image. | Should support hand gesture to capture the image. |
| 11 | It shall be stand-alone unit consisting of battery back-up with thermos probes. | It shall be stand-alone unit consisting of battery back-up with thermos probes. |
| 12 | It shall be a fully mobile system with interchangeable probes designed for the use of ablative procedures. | It shall be a fully mobile system with interchangeable probes designed for the use of ablative procedures. |
| 13 | The unit shall be light weight and easy to carry. | The unit shall be light weight and easy to carry. |
| 14 | Unit shall support different diameter probe consisting two small and two big probes | Unit shall support different diameter probe consisting two small and two big probes |
| 15 | Therapy probe must be made of durable robust material that can be sterilized and constructed with anti-sticks surface and be biocompatible. | Therapy probe must be made of durable robust material that can be sterilized and constructed with anti-sticks surface and be biocompatible. |
| 16 | Treatment temperature range must be selectable between 1000C to 1200C and treatment time 20 to 60 sec. | Treatment temperature range must be selectable between 1000C to 1200C and treatment time 20 to 60 sec. |
| 17 | Must have inbuilt Wi-Fi system for transferring of the data | Must have inbuilt Wi-Fi system for transferring of the data |
| 18 | Must have bluetooth features | Must have bluetooth features |
| 19 | The system should have future software updatability system with cloud and AI facility . | The system should have future software updatability system with cloud and AI facility . |
| 20 | Should have HDMI facility to connect with exterior display devices for wide viewing. | Should have HDMI facility to connect with exterior display devices for wide viewing. |
| 21 | Bidder shall indicate the details of probes to be supplied | Bidder shall indicate the details of probes to be supplied |
| 22 | System should be supplied with transport case for easy transportation. | System should be supplied with transport case for easy transportation. |
| 23 | Instruction for use should be embedded on the software for easy to use. | Instruction for use should be embedded on the software for easy to use. |
| 24 | Accessories, spares and consumables | Accessories, spares and consumables |
| 25 | Must submit ISO 9001 or ISO 13485:2003/AC:2007 for medical devices and US FDA 510(k) clearance must be there | Must submit ISO 9001 or ISO 13485:2003/AC:2007 for medical devices and US FDA 510(k) clearance must be there |
| 26 | The device should be battery operated and hand held type device. | The device should be battery operated and hand held type device. |
| 27 | It should consist of Main unit with removable probe and removal battery pack. | It should consist of Main unit with removable probe and removal battery pack. |
| 28 | It should have safety feature for guiding LED light, auto shutoff, low battery indicator etc. | It should have safety feature for guiding LED light, auto shutoff, low battery indicator etc. |
| 29 | The device must have Inbuilt timer facility for treatment. | The device must have Inbuilt timer facility for treatment. |
| 30 | The probe tip used in the device should produce at least 100 Degree Centigradefor destruction of human tissue. | The probe tip used in the device should produce at least 100 Degree Centigradefor destruction of human tissue. |
| 31 | The device should have a safety mechanism, it should be ablestop the procedure anytime during the treatment and the probe should cool down to a safe temperature (contactable with bare hands) within 20sec | The device should have a safety mechanism, it should be ablestop the procedure anytime during the treatment and the probe should cool down to a safe temperature (contactable with bare hands) within 20sec |
| 32 | The temperature of probe shaft should not reach beyond 43 degree centigrade for safety of the patient. Only tip of probe should be maintained target temperature. | The temperature of probe shaft should not reach beyond 43 degree centigrade for safety of the patient. Only tip of probe should be maintained target temperature. |
| 33 | Probe Shaft should be usable without any extra/external protecting covers. | Probe Shaft should be usable without any extra/external protecting covers. |
| 34 | Probe Shaft should be usable without any extra/external protecting covers. | Probe Shaft should be usable without any extra/external protecting covers. |
| 35 | The probe should be sterilized with sterilizers like CIDEX. | The probe should be sterilized with sterilizers like CIDEX. |
| 36 | Should be able to produce heat after the application on tissue, no preheating should be required. | Should be able to produce heat after the application on tissue, no preheating should be required. |
| 37 | Should consist of only following reusable Accessories: | Should consist of only following reusable Accessories: |
| 38 | Quoted model should be USFDA or CE certified. | Quoted model should be USFDA or CE certified. |
| 39 | Quoted model should be ISO13485 certified. | Quoted model should be ISO13485 certified. |
AI Tender Summary
Tender Timeline
Tender Published
Tender notice published.
CompletedBid Submission Deadline
Online submission via eProcurement portal.
Upcoming · 17 daysBid Opening Date
Technical bids will be opened and evaluated.
Upcoming · 18 daysTender Documents
Download All (ZIP) ↓We takes all possible care for accurate & authentic tender information. However users are requested to refer Original source of Tender Notice / Tender Document published by Tender Issuing Agency before taking any call regarding this tender