Tender For Supply, Installation, Testing & Commissioning Of Mgps System In 150 Bedded Cchb At Aiims Bhopal.
Tender Value
4.91 Crore
Estimated cost
Bid Submission
12 Sep 2026
17 days left
EMD
9.83 Lakhs
Bank guarantee accepted
Document Fee
Ref. Documents
Non-refundable
Tender Type
Online
Tender Overview
Organization
Tender ID
Competition Type
NCB
Bidding Type
Tender
Location / State
Bhopal → Madhya Pradesh
EMD Exemption
Not Available
Quantity
Not Available
Website
Contact Person
Address
Project Description
Tender for Supply, Installation, Testing & Commissioning Of Mgps System In 150 Bedded Cchb At Aiims Bhopal.
BOQ
| Sl. No. | Item Description |
| 1 | Supply, Installation, Testing and Commissioning of Medical Grade BSI Kite Mark & CE certified Copper Pipes/Tubes CDSCO: It shall meet with BS:EN: 13348:2016 standards, Solid Drawn, seamless, deoxidized, non-arsenical, half hard, tempered and degreased copper pipe confirming to the standard. All copper pipes should be medical grade degreased &delivered capped at both ends. The Copper pipes should be accompanied with Manufactured Test certificate for the physical properties & chemical composition. Degreasing of pipe shall be such that there is less than 20mg/m2 (0.002mg/cm2) of hydrocarbons on the degreased surface when tested by the method specified. Valid BSI Kite Mark and CE certificate must be submitted. Installation & testing: Installation of piping be carried out with utmost cleanliness. Only pipes, fittings and valves that have been degreased and fittings shall be used at site. Pipe fixing clamps shall be of nonferrous or non-deteriorating plastic suitable for the diameter of the pipe. Inert gas welding technique should be used by passing oxygen Free Nitrogen Gas inside the copper pipes during silver brazing, in order to avoid carbon deposition inside the copper pipes. Only copper-to-copper joints are permitted on site except threaded or flanged joints may be made where pipelines are connected to items such as valves and control equipment. No flux used for joining Copper to Copper joints and on for joints made on site. Copper to copper joints be brazed using a 5% silver-copper phosphorous brazing alloy CP104. Suitable sleeves shall be provided wherever pipes cross through walls / slabs. All pipe clamps be non-reactive to copper. After erection, the pipes are to be flushed with dry nitrogen gas and then pressure tested with dry nitrogen at a pressure equal to twice. All the pipes from manifold/plant upto the outlets painted with two coats of synthetic enamel paint and colour codification should be as per standards. |
| 2 | Supply, Installation, Testing and Commissioning of Medical Grade Copper Pipe size 108 mm OD x 1.5mm thk |
| 3 | Supply, Installation, Testing and Commissioning of Medical Grade Copper Pipe size 76 mm OD x 1.5mm thk |
| 4 | Supply, Installation, Testing and Commissioning of Medical Grade Copper Pipe size 54 mm OD x 1.2mm thk |
| 5 | Supply, Installation, Testing and Commissioning of Medical Grade Copper Pipe size 42 mm OD x 1.2mm thk |
| 6 | Supply, Installation, Testing and Commissioning of Medical Grade Copper Pipe size 35 mm OD x 1.2mm thk |
| 7 | Supply, Installation, Testing and Commissioning of Medical Grade Copper Pipe size 22 mm OD x 1mm thk |
| 8 | Imported Medical Gas Isolation Valve with Pipe line assembly Pre Piped Degreased Ball Valve with pipe extension. It should be full-port design. It should be blow-out proof stem. It should be lockable valve. The Valve shall be 3 piece ball-type design with a bronze body and chrome plated brass ball for sizes 1/2” to 4”. Seats shall be Teflon (TFE) and seals Viton for 1/2” - 4” valves. A blow-out proof stem shall be used and the valve shall have a maximum pressure rating of 600 psi [4,137 kPa]. Valves shall be operated by a lever-type handle requiring only a quarter turn from a fully open position to a fully closed position. Valves shall be designed in such a manner that it can be “swung-out” during installation so as to prevent damage due to heat transfer during the brazing operation. Each valve assembly shall be washed and degreased for medical gas service. The valve shall be supplied in a sealed plastic bag to prevent contamination prior to installation. |
| 9 | Imported ISO VALVE - SINGLE PORT/LOCKING 1-1/4\\" (35mm) |
| 10 | Imported ISO VALVE - SINGLE PORT/LOCKING 1-1/2\\" (42mm) |
| 11 | Imported ISO VALVE - SINGLE PORT/LOCKING 2\\" (54mm) |
| 12 | Imported ISO VALVE – DUAL PORT/LOCKING 3\\"(76mm)) |
| 13 | Imported ISO VALVE – DUAL PORT/LOCKING 4\\"(108mm) |
| 14 | Supply, Installation, Testing and Commissioning of Medical Gas AVSU (Area Valve Service Unit) CDSCO Licence : Single-service Aluminum body AVSU with NIST connection for Oxygen. It should be manufactured from Virgin Aluminium Alloy 6063-T6 extrusion profiles. Scrap aluminium shall not be used. All aluminium sections shall be anodized except the top cover, which shall be high-quality powder-coated aluminium. |
| 15 | Supply, Installation, Testing and Commissioning of Medical Gas AVSU (Area Valve Service Unit) CDSCO Licence : Single-service Aluminum body AVSU with NIST connection for MA4 Air. It should be manufactured from Virgin Aluminium Alloy 6063-T6 extrusion profiles. Scrap aluminium shall not be used. All aluminium sections shall be anodized except the top cover, which shall be high-quality powder-coated aluminium. |
| 16 | Supply, Installation, Testing and Commissioning of Medical Gas AVSU (Area Valve Service Unit) CDSCO Licence : Single-service Aluminum body AVSU with NIST connection for Vacuum. It should be manufactured from Virgin Aluminium Alloy 6063-T6 extrusion profiles. Scrap aluminium shall not be used. All aluminium sections shall be anodized except the top cover, which shall be high-quality powder-coated aluminium. |
| 17 | Supply, Installation, Testing and Commissioning of Imported Medical Gas LCD Alarm (Pressure Monitor) It should be UL or ETL Listed. (Alarm of 3 Gas Service Oxygen, Air 4 & Vacuum)LCD Alarm shall be microprocessor based with a 10\\" (25.4 cm) screen and capable of monitoring medical gases. The digital read-out shall provide a constant indication of each gas being measured, indicating a green “NORM4L” and a red “HIGH” or “LOW” alarm condition. A mute button silences the alarm, through the red status remains until the issue is resolved. !t should have customizable Alarm set points : The default set-points shall be +/- 20% variation from normal condition. It should support wired and wireless (Wifi) Ethernet connections for seamless integration with facility networks. It must have remote monitoring and notifications : Alarm should sends real time email and text notifications for critical alarm conditions. It should also provides a remote display that mirrors the alarm screen, viewable on desktops, laptops or tablets connected to facility network in same connection. The alarm should be connected with Mobile APP ( Mobile ECo system APP should be available on Apple APP store or Google play to download) |
| 18 | Supply, Installation, Testing and Commissioning of Imported Pressure Sensors : It shall be mounted Iocally (in the rough-in box) by installing the copper pipe provided or mounted remotely. |
| 19 | Supply, Installation, Testing and Commissioning of Imported Gas Outllet Points Terminal Unit for OxygenMedical Gas Terminal Units (Gas Outlet Points) : All Medical Gas Terminal Units shall be of 100% metal construction. The terminal unit front fascia shall be metallic. Plastic front fascias or plastic structural components shall not be accepted. Gas-specific second-fix components shall be manufactured from die-cast zinc alloy. Plastic components are not acceptable. Terminal unit socket castings shall be permanently coated with a low-friction fluoropolymer to ensure maximum reliability, smooth operation, and extended service life. The terminal unit socket die-casting shall incorporate a permanent gas indexing pin to eliminate the risk of loosening due to rough handling or misuse. |
| 20 | Supply, Installation, Testing and Commissioning of Imported Gas Outllet Points Terminal Unit for MA4 AirMedical Gas Terminal Units (Gas Outlet Points) : All Medical Gas Terminal Units shall be of 100% metal construction. The terminal unit front fascia shall be metallic. Plastic front fascias or plastic structural components shall not be accepted. Gas-specific second-fix components shall be manufactured from die-cast zinc alloy. Plastic components are not acceptable. Terminal unit socket castings shall be permanently coated with a low-friction fluoropolymer to ensure maximum reliability, smooth operation, and extended service life. The terminal unit socket die-casting shall incorporate a permanent gas indexing pin to eliminate the risk of loosening due to rough handling or misuse. |
| 21 | Supply, Installation, Testing and Commissioning of Imported Gas Outllet Points Terminal Unit for VacuumMedical Gas Terminal Units (Gas Outlet Points) : All Medical Gas Terminal Units shall be of 100% metal construction. The terminal unit front fascia shall be metallic. Plastic front fascias or plastic structural components shall not be accepted. Gas-specific second-fix components shall be manufactured from die-cast zinc alloy. Plastic components are not acceptable. Terminal unit socket castings shall be permanently coated with a low-friction fluoropolymer to ensure maximum reliability, smooth operation, and extended service life. The terminal unit socket die-casting shall incorporate a permanent gas indexing pin to eliminate the risk of loosening due to rough handling or misuse. |
| 22 | Supply, Installation, Testing and Commissioning of Bed Head Panel (Horizontal / Vertical) : Minimum length: 1500 mm. BHP Unit shall be manufactured from Virgin Aluminium Alloy 6063-T6 extrusion profiles. Scrap aluminium shall not be used. All aluminium sections shall be anodized except the top cover, which shall be high-quality powder-coated aluminium. The Bed Head Unit shall consist of two independent service channels: Channel-1: Dedicated Medical Gas Channel with factory pre-piped, factory-tested gas outlets (minimum 4/5/6 gas outlet points).Pre piped minimum 4/5/6 gas outlet points as per design matrix (*medical gas terminal outlet are seperate line item as mentioned in theBOQ) Channel-2: Dedicated Electrical Channel with factory pre-wired and factory-tested electrical accessories comprising:(Minimum six electrical sockets) Two data sockets Integrated LED Reading Light and on/off switch Built-in lower aluminium equipment rails. One stainless steel basket with rail clamp. |
| 23 | Supply, Installation, Testing and Commissioning of Ceiling Mounted Vertical Column, Vertical column 1200mm length with ceiling pole. Vertical tube + Column joint + Flange Swivel Movement 330 degree CE Compliance under class -1 It should be made up of Aluminum extrusion profiles with aluminum alloy 6063 temper T6. It should be manufactured from Virgin Aluminum’ no scrap added in aluminum). All sections are anodized aluminum except top plate. The top plate will be aluminum powder coating (high quality). Built in high strength anodized aluminum rail of full length without any joint Prewired minimum 8nos 6/16amp multpin electrical sockets. Pre piped minimum 3/4/5/6 gas outlet points as per design matrix (*medical gas terminal outlet are separate line item as mentioned in the BOQ). 2 data sockets Valid CDSCO Manufacturing License submission is mandatory at the time of bidding.\\" |
| 24 | Supply, Installation, Testing and Commissioning of Pressure Gauge (*Pressure range 0-16, Dial size 4inch) : Pressure Range: 0 – 16 Kg/cm2 with psi measuring system and Movement: Copper alloy, Case: Black steel. Window: Press fit plastic. Dial: Aluminium white with stop pin. Pointer: Black plastic, Accuracy: 2.5% as per EN 837-1. |
| 25 | Supply, Installation, Testing and Commissioning of Imported O2 Flow Meter 70 L/Min , 1/2 BS 1/4 G M INLET : Pressure compensated to prevent back pressure build up on flow indicator. It should have cartridge knob, extremely reliable, easy-to-clean and ensuring an optimized adjustment. Perfect tightness of the knob thanks to its polyamide seat. Expanded scale providing higher reading accuracy. Durable polycarbonate flow tube with cover. Monoblock scale cover made of polyamide for a higher sturdiness and a greater safety. The inopportune unscrewing of the monoblock scale cover is not possible. Body made of nickel-plated brass, very strong. Flow meter should be placed in the vertical position. It should be imported. It should be light weight. The flow meters should be of 0-70lpm for oxygen and with inlet pressure 50-60psi. (4.5 bar). The closing of the knob should be without any leakage. |
| 26 | Supply, Installation, Testing and Commissioning of Imported O2 Low Flow Meter 15 L.Min , 1/2 with stainless steel BS Direct Probe 1/4 GFPressure compensated to prevent back pressure build up on flow indicator. It should have cartridge knob, extremely reliable, easy-to-clean and ensuring an optimized adjustment. Perfect tightness of the knob thanks to its polyamide seat. Expanded scale providing higher reading accuracy. Durable polycarbonate flow tube with cover. Monoblock scale cover made of polyamide for a higher sturdiness and a greater safety. The inopportune unscrewing of the monoblock scale cover is not possible. Body made of nickel-plated brass, very strong. Flow meter should be placed in the vertical position. It should be imported. It should be light weight. The flow meters should be of 0-15lpm for oxygen and with inlet pressure 50-60psi. (4.5 bar). The closing of the knob. |
| 27 | Supply, Installation, Testing and Commissioning of Imported Polysulphone 250cc Humidifier bottle should be unbreakable, reusable to disinfectants and complements. |
| 28 | Supply, Installation, Testing and Commissioning of Medical Gas Vacuum Regulator CDSCO License |
| 29 | Supply, Installation, Testing and Commissioning of Imported Emergency Oxygen Inlet It should have an inlet to connect a temporary, auxiliary source of supply for emergency or maintenance situations. This connection should include necessary valves and to allow for an emergency supply of oxygen and Isolation of the normal source of supply. The inlet connection should be a 1\\"\\" female npt connection and be provided with a plug to prevent entry of foreign matter into the oxygen system when not in use. 1\\"\\" brass ball valve controls access to the Emergency supply, and the gauge indicates the pressure being introduced into oxygen pipeline. The enclosure door should be labelled “Emergency High Pressure Oxygen Inlet Station” and should be equipped with a staple for padlocking to allow entry only by authorized personnel. The interior of the enclosure should be clearly labeled with instructions for connection and operation of the emergency oxygen inlet. The relief valve should be supplied factory cleaned for oxygen service and should automatically reset after discharging to provide a positive seal. Please Enable Macros to View BoQ information |
AI Tender Summary
Tender Timeline
Aug 22, 2026
11:30 IST
Tender Published
Tender notice published.
CompletedSep 12, 2026
17:00 IST
Bid Submission Deadline
Online submission via eProcurement portal.
Upcoming · 17 daysSep 14, 2026
Bid Opening Date
Technical bids will be opened and evaluated.
Upcoming · 19 daysTender Documents
Download All (ZIP) ↓