Bids Are invited for Hybridization Chamber For Fish Test (Hybrite) - Temperature Range: 30°C -99°C. , Lowest Programmable Temperature 30°C Or Ambient Temperature +5°C (Whichever Is Higher). , Processing Time: 00:00 - 99:59 Hr : Min. , Number Of Progr
Tender Overview
Project Description
Bids Are invited for Hybridization Chamber For Fish Test (Hybrite) - Temperature Range: 30°C -99°C. , Lowest Programmable Temperature 30°C Or Ambient Temperature +5°C (Whichever Is Higher). , Processing Time: 00:00 - 99:59 Hr : Min. , Number Of Programs: 40 Approximately , Three Operating Modes (Fixed Temperature, Hybridization Only Or Denaturation And Hybridization). , Slide Separator & Cover Gasket (Qty - 1 Each). , Humidity Cards (Qty - 5 Pairs). , Numeric Keypad For Easy Programming. , Ambient Operating Humidity :20-80% Relative. , Ambient Operating Temperature: 5-40°C (41-104°F) , R Time: 37-95°C In Less Than 3 Minutes , Cooling Time: 95-45°C In Less Than 6 Minutes , Environmental: Temperature 15° C To 40° C. , Ivd Certification Necessary For Use In Vitro Diagnostic Use In Denaturation / Hybridization Of Slide-Based Fish Procedures , System Should Be Verified During Installation By A Calibrated Temperature Verification , Local Service Support: Should Have Local Office And Service Support/Service Engineer For Attending The Breakdown Calls. Details To Be Submitted. , Response Time: Should Not Be More Than 06 Hrs From Lodging A Breakdown Complaint On Toll Free Or By Email. , Warranty- 2 Years , Amc/Cmc- Cost Per Annum Will Be Considered For All Bidders For 8 Years After Warranty
BOQ
| Sl. No. | Item Title | Item Description |
| 1 | Temperature Range: 30°C -99°C. | Temperature Range: 30°C -99°C. |
| 2 | Lowest programmable temperature 30°C or ambient temperature +5°C (whichever is higher). | Lowest programmable temperature 30°C or ambient temperature +5°C (whichever is higher). |
| 3 | Processing Time: 00:00 - 99:59 hr : min. | Processing Time: 00:00 - 99:59 hr : min. |
| 4 | Number of Programs: 40 approximately | Number of Programs: 40 approximately |
| 5 | Three operating modes (Fixed Temperature, Hybridization Only or Denaturation and Hybridization). | Three operating modes (Fixed Temperature, Hybridization Only or Denaturation and Hybridization). |
| 6 | Slide separator & Cover gasket (Qty - 1 each). | Slide separator & Cover gasket (Qty - 1 each). |
| 7 | Humidity cards (Qty - 5 Pairs). | Humidity cards (Qty - 5 Pairs). |
| 8 | Numeric keypad for easy programming. | Numeric keypad for easy programming. |
| 9 | Ambient operating Humidity :20-80% relative. | Ambient operating Humidity :20-80% relative. |
| 10 | Ambient operating Temperature: 5-40°C (41-104°F) | Ambient operating Temperature: 5-40°C (41-104°F) |
| 11 | Ramp Time: 37-95°C in less than 3 minutes | Ramp Time: 37-95°C in less than 3 minutes |
| 12 | Cooling Time: 95-45°C in less than 6 minutes | Cooling Time: 95-45°C in less than 6 minutes |
| 13 | Environmental: Temperature 15° C to 40° C. | Environmental: Temperature 15° C to 40° C. |
| 14 | IVD Certification Necessary for use in vitro diagnostic use in denaturation / hybridization of slide-based FISH procedures | IVD Certification Necessary for use in vitro diagnostic use in denaturation / hybridization of slide-based FISH procedures |
| 15 | System should be verified during installation by a calibrated Temperature Verification | System should be verified during installation by a calibrated Temperature Verification |
| 16 | Local Service Support: Should have local office and service support/service engineer for attending the breakdown calls. Details to be submitted. | Local Service Support: Should have local office and service support/service engineer for attending the breakdown calls. Details to be submitted. |
| 17 | Response Time: Should not be more than 06 hrs from lodging a breakdown complaint on toll free or by email. | Response Time: Should not be more than 06 hrs from lodging a breakdown complaint on toll free or by email. |
| 18 | Warranty- 2 years | Warranty- 2 years |
| 19 | AMC/CMC- cost per annum will be considered for all bidders for 8 years after warranty | AMC/CMC- cost per annum will be considered for all bidders for 8 years after warranty |
AI Tender Summary
Tender Timeline
Tender Published
Tender notice published.
CompletedBid Submission Deadline
Online submission via eProcurement portal.
Upcoming · 9 daysBid Opening Date
Technical bids will be opened and evaluated.
Upcoming · 9 daysTender Documents
Download All (ZIP) ↓We takes all possible care for accurate & authentic tender information. However users are requested to refer Original source of Tender Notice / Tender Document published by Tender Issuing Agency before taking any call regarding this tender