Tender For Supply Installation Testing and Commissioning of Medical Gas Pipe Line System And Modular Operation Theater Work In The Construction Of District Hospital Hindaun Distt Karauli

Issued by Government Departments · Karauli, Rajasthan
Tender Value
3.08 Crore
Estimated cost
Bid Submission
10 Aug 2026
13 days left
EMD
6.17 Lakhs
Bank guarantee accepted
Document Fee
2000
Non-refundable
Tender Type
Online

Tender Overview

Competition Type
NCB
Bidding Type
Tender
Location / State
Karauli → Rajasthan
EMD Exemption
Not Available
Quantity
Not Available

Project Description

Tender for Supply Installation Testing and Commissioning of Medical Gas Pipe Line System And Modular Operation Theater Work In The Construction Of District Hospital Hindaun Distt Karauli

BOQ

Name of Work:- SITC of Medical Gas Pipe Line System and Modular Operation Theater Work in the construction of District Hospital Hindaun distt karauli
Sl. No. Item Description
1PART-A (MODULER-OT))
2Supply , Installation, Testing and Comissioning of site fabricated ducting made from aluminium sheet conforming to IS:737, Grade 31000 fabricated as per IS:655 and as per approved drawings & designs including the cost of hangers, supports , angle , expansion fasteners , closed cell neoprene gasket T&P etc. as required of following thickness.
30.80 mm
41.00mm
5Supply, installation, Testing & commissioning of VRV/VRF compatible OEM make AHU KIT' including expansion valve and controller for connecting AHU with VRF.
6Supply and fixing of acoustic lining of ducts with chemically cross link Oxide accetate foam open cell structure Acoustic Insulation with density 30 Kg / Sqm, UV protected and Class O Fire propperty Water Vapour absorption 0.002 gm / Sqcm, including cost of adhesive etc. complete in all respect. All as per pre approved by Engineer in charge. All as per pre approved by Engineer in charge.
715 mm thick
8Supply & fixing of external thermal closed-cell Nitrile rubber Insulation material FM/UL approved with high water vapor diffusion resistance and low thermal conductivity should be certified 'Class O' in Fire Propagation as per BS 476 Part 6 in form of sheet on duct including cost of adhesive (CPRX Compound) omplete as required. All as per pre approved by Engineer in charge.
919mm thick for Outdoor Application with 26 SWG Aluminium Cladding
10Supply, Installation, Testing & Commissioning of modular type Variable Refrigerant Flow/Variable Refrigerant Volume air cooled Outdoor units suitable for cooling and heating, having 100% hermetically sealed inverter type Scroll Compressor(s) with and minimum two compressors along with two Condensing fan, above 14/18 HP modules, microprocessor based Controller, Side / top discharge type condensing unit(s), with R-410 A Refrigerant, vibration isolators, with suitable foundation etc. complete as required. The unit shall deliver the rated capacity at AHRI Conditions and work even at 50°C ambient temperature without tripping. The unit shall be suitable to work on 400V +/- 10%, 3 Phase, 50Hz AC power supply. The unit shall be filled with first charge of the refrigerant and ready for use as required. The EER at 100% load with AHRI conditions shall not be less than 3.5 and Including OEM supplied Piping connection Kit etc. as required. All as per pre approved by Engineer in charge. For additional technical parameters of products/ work , refer Annexure \\"A\\" attached with this BSR .
11SITC of VFD OEM fabricated VFD cum Bypass starter panel Confirming to IEC60721-3-3, IEC60068-2-60 & IEC60068-2 for variable torque applications, 3C3 avg., 200% torque producing capability at min freq of 0.5 Hz, integrated with 3 to 5% of input reactor or built in DC reactor, Max output frequency 400Hz, built in braking chopper till 30KW, built in RS485 Modbus RTU, Min 7DI, 2DO, 2AI, 1AO, onboard control supply of 24V -150 mA for external operations, RJ45 port, Monitoring parameters – V, I, Torque, Freq, RPM, Watt Hour, DC bus voltage, Analog input V or mA, I/O Status, PID loop monitoring, Torque speed, speed limit, Cooling fan, Protection functions – Ground fault, Over current, Overload, UV/OV, Under/over torque, I/P & O/P phase loss, DBR O/L, inverter over hit alarm, DC bus charge LED indication, heatsink thermistor open detection, stall prevention, fault history of min 4 last faults. Bypass starter shall be provided as per type 2 coordination etc. as required. All as per pre approved by Engineer in charge. For additional technical parameters of products/ work , refer Annexure \\"A\\" attached with this BSR . 11.0 KW
12Wall Panelling System: Supply , Installation, Testing and Comissioning of the wall panels should be manufactured with a 0.8mm thick Galvanized sheet. The wall panels should be reinforced by MR grade Soaked plasterboard with a thickness of 15 mm or above thickness. the wall panels should be manufactured with a 0.8mm thick Galvanized sheet in accordance with IS277:2018. The wall paneling/partition systems of the Modular Operation Theatre shall be tested and certified according to BS:476 Part (20 & 22)-1987 from authorized Fire Research Laboratory under the Government of India to ensure that the wall paneling/partition meets the fire resistance rating criteria of one twenty minutes on thermal insulation, integrity, and stability parameters. Valid certificate/test report shall be submitted along with the bid for technical bid approval. The grid section shall be made of three elements: - Lower wall element (approx. 900mm high). Upper wall element (approx. 1800mm high). A customizable installation module of 200 mm between the lower and upper wall element. The facory made antimicrobial powder coating should provide 24-hour protection against bacteria, fungi and mold. This should be confirmed by third party certificate as per ISO 22196:2011 for minimum 9 different bacteria and 2 different fungus. Valid certificate/test report shall be submitted along with the bid for technical bid approval. Any kind of cutting, chipping, spraying and painting at site shall be deemed unacceptable. Panels should be mounted on a structure made of galvanized steel / iron of a thickness of 1.5 to 2 mm. The structure should be of square tube of minimum 38x38 mm. It should enable the distribution of medical gases, electricity and all required fittings etc inside the wall. The joints between the panels should be between 6 -10 mm and should be provided with a press fit food grade antimicrobial (as per ISO 22196:2011) silicone gasket to close the gap. Sound proof up to 47 dB or better. Valid certificate/test report shall be submitted along with the bid for technical bid approval. The system shall be 100% modular to ensure quick and easy installation, maintenance, retrofitting and upgradation including Free-standing sub frames, perimeter wall panels and removable sealing gaskets. The system shall be installed at site without any welding.The Wall panels should be CE compliance/DIN 1.4301 certified.OEM shall submit the material verification quality report at the time of pre-dispatch of the material.
13Ceiling System: Supply , Installation, Testing and Comissioning of the ceiling panels should be manufactured with high-quality 0.8 mm thick galvanized sheet with Soaked plasterboard combined with glued with a thickness of 10-12 mm thickness. OR The state-of-the-art standard for modular OT ceilings is the 0.8mm antibacterial steel clip-in tile system at 1200×600mm. This system provides: • Individual tile access in under a minute. • Zero civil works for above-ceiling maintenance. • Consistent antibacterial surface with no joints, cracks, or plasterboard edges. • Reduced dead load on the ceiling substructure. • Each Individual Panels can be removed without affecting other ceiling tiles.and should be factory made antimicrobial powder-coated as per ISO 22196:2011 in any RAL color. Ceiling tiles shall be tested and certified for ASTM E84 to ensure that the material does not provoke fire and does not generate smoke. Valid certificate/test report shall be submitted along with the bid for technical bid approval. Ceiling tiles should have a standard module sized of 1200 x 600 mm. Installation Ceiling should create a tight space without any leakage. The ceiling panels mounted to the structure can be removed individually to access the area above ceiling. The Ceiling system should be CE compliance/DIN 1.4301 certified. OEM shall submit the material verification quality report at the time of pre-dispatch of the material. 3.3.2 along with suitable supports/substructure frame.
14Supply , Installation, Testing and Comissioning of Storage Unit
15The storage unit should be made with at least 0.8 mm thick Galvanized iron panels
16The shelves should be of galvanized & removable for cleaning.
17Storage shall have two-point locking system with Chrome finished D-Handles. Storage shall have adjustable shelves and there shall be 4 loading levels. Each Shelf shall have loading capacity of 35 Kg uniformly distribute load. Storage rack shall have 4 set of soft closing hinges for noiseless operation. Door shall have feature of opening angle of 135-165 degree for complete fumigation inside the corners.
18The overall size should be approx. 200 cm X 100 cm X 35 cm
19Supply , Installation, Testing and Comissioning of Scrub Station (min size 1500mm )
20Compact surgical scrub sink should be designed for use in OT complex providing for pre procedural scrub up.( Double sink combination as suitable)
21Each fixture should be fabricated from heavy gauge type 304 stainless steel (minimum thickness 1.5mm) and should be seamless welded construction, polished to a satin finish with all required fittingshigh and rich specification.
22The scrub sink should be provided with a front access panel which should be easily removed for access to the water controlled valve, waste connections, stoppers and strainers.
23Hands free operation should include infra red sensors with programmable adjustment.
24Thermostatic mixing, valve control should be located behind the access panel and maintain constant water temperature.
25Timing should be adjustable to meet individual application requirements.
26Provided with infrared sensors, thermostatic control taps with fail safe temperature controls.
27All units should have reduced anti- splash fronts.
28Supply , Installation, Testing and Comissioning of PVC Flooring It should be with 2mm antistatic seamless PVC flooring Floor should be smooth, non-slip, impervious homogeneos material conductive enough to dissipate static electricity but not conductive enough to endanger personnel from electric shock.Electrostatic charge dissipation combat PVC seamless self leveling.flooring of very high quality should be provided Thickness not less than 2 mm. Continuous roll should be used and joints should be welded by special PVC thermal welding units using PVC welding bars of same colour The sheets should be highly durable with resistance to shock and indentation. It should be scratchproof also. The conductive material should be homogeneos grains.It should be inert to body fluids, chemicals and disinfectants. Should not be affected by temperature variation within theOT.The floor should efficiently discharge electric charges up to 2 kV Flooring should be done by skilled workers of accredited agencies authorized by the supplier of PVC sheets. The electrical resistance (point to ground) should be within2.5x10^4 to 5x10^6 ohms. The floor should not allow build up of electrical charge beyond 100 volts due to antistatic effect. The corners should not be terminated sharply and concealed cove- former (aluminium) should be used to overlap the wall panel to a height of approx.100mm and sealed perfectly and uniformly. Self leveling compounds should be used. The conductive copper grid laid underneath the PVC sheet should be supported by liquid epoxy compounds allowed to set as a uniform and level surface. The copper strips to be made visible by grinding and no copper strip should project more than 0.5mm above level surface to avoid damage to the PVC sheet. One earthing lead should be brought out from every 150sq.ft area and attaching it to the main earthing strip/ground. Copper grounding strips (0.05 mm thick, 50 mm width) should be laid flat on the floor in the conductive adhesive and connected to copper strip of grounding. The connection from copper grid should be brought out uniformly at places to form equipotential grid.Flooring should be mechanically shock proof, scratch proof, fire resistance should be not less then or equal to 8kw/sqmantibectrial corners should be uniformly curved and required sound insulation.
29HERMETICALLY SEALED DOOR Door sizes should be as per below option and quantities will be as per BOQ of respective institute Door Size 2.1 M(H) X 1.5M(W) /Size 2.1 M(H) X 1.8M(W) Supply , Installation, Testing and Comissioning of the door shall be designed to open and close by waving a hand at a distance of 10 cm from the sensor (touchless sensors provided on both sides) and shall also be equipped with a foot-operated switch. The doors shall additionally have the feature of effortless manual operation in case of failure of the automatic mechanism. The track should be of stainless steel/Aluminum and the running surface for the top rollers should be suitably angled to reduce resistance to movement The door leaf should be hung by means of hard plastic rollers of high quality with double bearing at the top. Rollers should be provided under the stainless steel/Aluminium track to enable smooth and noiseless movement. Opening and closing of the door should be microprocessor controlled electromechanical movement. The door material should be of same as wall panelling material Color should match the interior and care should be taken to make the leaf strong and light weight. One should be able to open and close the door effortlessly incase of failure of automatic mechanism. Door opening handle should be strong and sturdy. Material should be of SS (gloss/matte finish). Door leaf should have high quality synthetic rubber gasket with long life to ensure hermetic sealing (to maintain air pressure differential). Air tightness 99.99% at a pressure of minimum 75Pa The finished floor on either side of the door should be perfectly level (maximum permissible difference +1mm). The door is designed and constructed in accordance with the fire-resistance requirements of BS 476 (Parts 20 & 22): 1987 and IS 3614 (Part 2): 2021. The overall thickness of the finished door should be 50- 60mm. The inner part of the door should be filled with CFC free polyurethane foam / Honey Comb / Mineral wool thickness of 48mm or nearby. (Sealed airtight to prevent further ingress of any microbial organism). The door and controls should comply with IEE regulation. All motors used should be DC brushless/PMDC motors with essential isolation from mains.Door should be with vision window 300 mm x 300 mm with double glazed panels and hermetically sealed.
30CIVIL WORK Such as scaffolding, wall breaking ,Core Cutting , Cement board covering, gap filling, platform dismentaling, AHU and outdoor foundation etc will be in the scope of contractor.
31NON-BSR ITEMS ELECTRICAL)
32SITC of Laminar Flow Units (LFU) comprising of Al frame 18 guage with HEPA filter and HEPA filter box with powder coated single skin casing made out 18 guage sheet complete with HЕРА EPA filters .with aluminium perforated laminar diffuser 2400 x2400 x600 mm.
33Touch Surgeon Control Panel: Supply , Installation, Testing and Comissioning of the Control Panel Should be a touch screen .this control panel should work as the central control panel for light control, gas alarm,etc the touch screen shold be wall mounted. Stationed in the visiblity line of surgeon and ot staff the access Hieght Should be convenient for the nurse to opreate and help/assistant when in need. the panel should accomodate digital clock and the elipsed time indicator .the medical gas alarm should indicate high and low gas pressure for each gas servicein the ot and normal low indication for vacuum this should be supported by audible alarm also the panel should have an alarm mute facility the sensor should be at nearst isolation valve control for general lighting on-off to be provided. the control opreating light to be provided hand free telephone should be located one side. terprature and humidity set points/display. digital room pressure in cm of H2O or equivalent. Hepa filter Bank Differential pressure indicator.Operation capability while wearing surgical latex gloves. The system shall include an OT Occupancy Control feature within the Touch Control Panel. When a surgery is in progress, this feature will prevent unauthorized door opening through the hand-wave sensor. During OT occupancy, the door shall remain locked from the sensor control and can only be operated through the foot switch or authorized control panel command. This prevents accidental door opening due to hand-wave sensor activation in high-traffic areas such as corridors during trolley or patient movement. it should have CE compliance 1.8.1 The control panel shall be touch screen pane This control panel shall work as the central control panel for the HVAC controls, instruction board, touch screen and ON/OFF of MOT light. It shall be CE Compliance/US FDA/ETL/UL/BIS certified/ listed.
34Supply , Installation, Testing and Comissioning of Pressure Relief Damper
35Pressure relief dampers or over flow ports should be provided in each room to prevent contamination of air from clean and dirty areas.
36Suitably sized air pressure relief damper should be strategically placed, enabling differential room pressure to be maintained and ensure that when doors are opened betweenclean and dirty areas.
37Counter- weight balancing system should be provided in the PRD to maintain positive pressure inside the operation room.
38Air pressure stabilizers should have unique capability of controlling differential pressure to close tolerance. The PRD should remain closed at pressure below the set pressure and should open fully at a pressure only fractionally above the threshold pressure.
39The frame, body and blade should be of grade SS 304 stainless steel.
40Supply , Installation, Testing and Comissioning of X RAY Film Viewer
41LED type flat panel X-ray viewing panel should be flicker free
42This should comply with relevant electrical safely codes.
43Mounting should be flush with the wall to avoid dust accumulation and growth or organisms between wall and panel.
44No external frame shall be installed on the modular wall, and the surface shall be free from protrusions or irregularities that may cause interruptions. Outer peripheral facia of X-ray viewer shall be in similar material of wall panel and also front peripheral facia finish shall have same powder cating color theme as wall panel.
45The diffuser on the front panel should be a uniformly lit screen.
46Dimming electronic control should be enclosed at the bottom of the cabinet.
47Proper spring loaded film clip with rollers should be provided to hold the films firmly and to remove the film without scratches.
48This should be 2 Numbers of dual panel viewing screen (14\\"x17\\" each), it may be on one wall panel or adjacent.
49Supply , Installation, Testing and Comissioning of Operating List Board
50One operating list board should be provided in each operating theater.
51It should be made of ceramic having magnetic properties and should be flushed to the wall of the operating room.No external frame shall be installed on the modular wall, and the surface shall be free from protrusions or irregularities that may cause interruptions. Outer peripheral facia of writing board shall be in similar material of wall panel and also front peripheral facia finish shall have same powder cating color theme as wall panel.
52Supply, Installation, Testing and Commissioning (SITC) of peripheral OT LED lights, recessed-mounted, IP66-rated, non-hygroscopic luminaires suitable for cleanroom applications, providing a minimum illumination level of 500–700 lux on the operating table. The lights shall be equipped with dimming control through a touch control panel.Luminaries cover shall be made of highly resistant, disinfectant proof laminated acrylic sheet with stylish fine-grained surface, light box shall have laminated transparent acrylic screwed. The dimensions shall be: 2 ft(L) X 2ft(W). Light cover shall have provision to open from bottom side when necessary for maintenance and placed it back after maintenace to form hermetic sealing. The reflectors shall be of high quality, cleanable and non-deteriorating. Recess frames shall be gas tight. The fitting shall remain flushed with the ceiling and shall be removable when required. The light fitting shall be uniformly and aesthetically distributed on the ceiling to provide uniform illumination in the modular operation theater. Light shall not interfere when green mode endoscopy is performed. The light box / body shall be factory made antibacterial powder coated.
53SINGLE ARM PENDANT FOR ANAESTHETIST ( Indigenious)
54Supply , Installation, Testing and Comissioning of the Pendants made from Special High Strength Aluminium Alloy and finished with Powder Coated paint should comply with NFPA 99C/HTM 2022/02-01. or Pendant arms are made from powder coated from aluminium Extruded. The support arms should be extremely robust and revolve on high quality bearings, so that the pendant head glides smoothly and quickly to any desired position. Single arm moveable arms with coverage of 1000mm and 330 deg. Horizontal movements for each arm. Weight carrying capacity of the arm should not be less than 150 Kg. .
55Each arm should be capable of 300-340 degrees of rotation, which can be easily adjusted to suit the desired mode of operation. The Pendant Service Heads should be modular with minimum 800mm head. The heads should be capable of accepting a range of shelves, infusion poles, electrical switch/sokets, gas outlets other accessories as asked in tender.The Pendant Service Heads should had provision for medical gas terminal units and 6/16 Amps universal Sockets with switches. Each 6/16 Amp circuit to be connected to a separate circuit, and should not be looped together.
56Each pendant should have provision for outlets and probes as mentioned below – Oxygen Outlets - 2 nos., Vacuum Outlets - 2 nos., Nitrous oxide - 2 nos., Medical Air(4 bar) Outlets -1 No.
57Each pendant should be supplied following accessories as mentioned below –Electrical sockets - 4 nos - 6/16A, Data socket cat 6 -1 no. RJ45 socket -1 no Adjustable shelves shall be with two rails one on each side
58Manufacturer and products should also be registered under Indian CDSCO regulatory authority.
59DOUBLE ARM PENDANTS FOR SURGEON ( Indigenious)
60Supply , Installation, Testing and Comissioning of the Pendants The support arms should be extremely robust and revolve on high quality bearings, so that the pendant head glides smoothly and quickly to any desired position
61Double moveable arms (any combination) with total coverage of min 1000 mm and 330 to 340 deg. Horizontal movements for each arm.
62Weight carrying capacity of the arm should not be less than 150 Kg.
63The Pendant Service Heads should be modular with minimum 800 mm head. The heads should be capable of accepting a range of shelves, infusion poles, electrical switch/sokets, gas outlets other accessories asasked in tender.
64The Pendant Service Heads should be supplied with medical gas terminal units and 5/15 or 6/16 Amps hybridSockets with switches.
65Each pendant should be supplied with outlets and probes as mentioned below –
66O2 & Nitrous & MA4 -1Nos each
67Vacuum Outlets – 2nos,
68Electrical sockets - 4 nos.
69Adjustable Shelf with two rails one on each side – 1 no.
70Data socket RJ-45 -1 no.
71Manufacturer and products should also be registered under Indian CDSCO regulatory authority.
72The Pendants shall comply with NFPA 99C/HTM 02-01/DIN/EN/ISO 7396-1 as amended up to the last date of receipt of tenders. The support arms shall be extremely robust and revolve on high quality bearings, so that the pendant head glides smoothly and quickly to any desired position.
73Supply , Installation, Testing and Comissioning of Double Dome OT Light
741. The Dual Dome LED Operating Lights provide advanced illumination, ensuring ample light for all surgical fields, including critical cases and deep cavities. 2. Single-color, dual-color LEDs in white and yellow with circular light beam types 3. Adaptable light field size and intensity with controls for brightness, color temperature, and light field size via a control panel.4. 360 Degree rotation is present for the light heads.5. Specifically designed to improve the lighting depth and range of view, the tampered glass is scratch resistant and has strong laminar flow qualities.6. True-color, cool light ensures precise visibility of nerves and vessels while enhancing surgeon comfort during long procedures. 7. Endoscopy mode – yes 8. Total irradiance at max. illuminance - W/m2 – 545 Parameter Specification Number of Light Heads2 Light Heads No. of LEDs per Light Head 48 LEDs / Dome Light Head Diameter 500 mm Lux Intensity 60,000 Lux / Dome Illuminated Field Diameter 150–250 mm CRI (RA / R9) 95 / 95 Colour Temperature 3000–4500 K (Adjustable) Life Span of LEDs 60,000 Hours Dimming Range 30–100% Quality Certificates USFDA Registered, CE, ISO 9001, ISO 13485 Standards Compliance Conformity to IEC 60601 Standards Manufacturer and products should also be registered under Indian CDSCO regulatory authority.
75PASSBOX - 600 MM* 600 MM*600 MM size should be provided in the Operation Theatre as specified in the scope of the work to remove waste materials from the Operation Theatre to Dirty linen Area just adjacent to Operation Theatre. Each Hatch should be equipped with two doors and the door should be operated electronically. The Hatch should be designed in such a way that only one door should be opened at one time. The UV light should be so installed that it is kept on while both the doors are closed, this UV light has to be automatically turned off in case of opening of either of the doors. There should be indicators on both sides of the OT so that the door open/close status can be monitored from both ends. Hatch Box should be manufactured from SS 304 Stainless Steel
76MGPS WORK INSIDE OT such as pendant , piping installation, ( Inside OT only) etc. Complete as per site requirement and pre approved by engineer incharge
77Supplying, installing, testing and commissioning of double skin floor mounted type AHU with 45 mm thick panel PUF CFC FREE DENSITY 40 kg/M³ with thermal break profile. the AHU inner skin shall be 0.6 mm pre coated Gl,outer skin shall be 0.6 mm pre coated GI with colour.complete with fan section centrifugal backword curved DIDW type fan(s), belt drive, vibration isolators, prefilter, micro filter section made of ALUMINIUM (with out micro filter) mixing box for return air, fresh air. SS drain pan.6 row deep 2 nos (one working + one standby) DX cooling coil of copper tube & aluminium fins, 12-13 FPI, bypass filter not more then 0.12.fan motor shall be suitable for 415±10%volts, 50Hz, 3 phase AC supply and shall be high efficiency(IE-3 Class).including fresh air, return air, supply air damper with air volume control.AHU CAPACITY : 3000 CFM
78Supply , Installation, Testing and Comissioning of internal electrification work including switch,socket,point and Circuit wiring with proper size of coper wire. Distribution Board, RCCB, MCB, 8 Way 3 phase PPI DB, Switch socket 6 Amp and 16 Amp both etc. required for the Modular OT, complete in all respects as per the requirements of EIC/MOIC, shall be provided and wired by the MOT vendor. Need to veryfied quantity as per bsr rates.
79SITC of 3-Phase Input, 3-Phase Output Online double conversion VFI-SS-111 IGBT based Rectifier & Inverter UPS with 0.99 Input power factor & unity Output factor (KVA=KW), having Inbuilt Isolation Transformer at Input, Phase balance of 120° ±1° (100% unbalanced load), input voltage & frequency range of 320-480V & 45-55 Hz and output voltage of 415V (+/- 1% for balanced & +/- 2% for unbalanced load) & output frequency of 50Hz +/- 0.1 Hz. UPS should have Inbuilt phase sequence corrector & protector both (UPS should correct the Input Phase reversal condition & shall not go to battery mode) and Input, output & Bypass breaker UPS should have Inbuilt Static & maintenance bypass, Parallel kit, EPO, dust filter & conformal coated boards. Overall efficiency without transformer loss (93% for 10KVA, 94% for 20KVA & 30KVA, 95% for 40KVA & 60KVA). Overloading on mains mode: ≤110% for 60 mins ≤125% for 10 mins ≤150% for 1 min UPS should have LED Panel with minimum 5-inch Intuitive Touch Screen LCD Display showing UPS status with other parameters like Input, Output, Bypass - Voltage/Frequency, load%, Battery%. UPS should have protection from Input High/Low, Over Load, Short Circuit, Over Temperature and give audible alarms on Battery Mode, Low Battery, Fault, Over Load & Bypass Mode. UPS communication through RS485. SMF Batteries of 12V VRLA for 30 min backup with MS Rack having battery breaker with enclosure and necessary cable for UPS to Batteries Installation of 20 Mtrs. UPS should have feature to connect similar model & capacity with common battery bank. Operating Temperature of UPS 40-degree, relative humidity 0-95% (Non condensing). 10KVA UPS with 10080 VAH Battery
80PART-B (MGPS))
81NON-BSR ITEMS)
82SITC of MGPS pipe of medically grade as per BSEN-13348 with third party certification from Lloyds/TUV degreased [ie free from oil, grease and particulate matters) seamless copper pipe confirming to ASTM B88 and all copper joining done by brazing with AWSA5.8, classification BCuP-5 or BS 1845 following procedure in accordance with ASTM B828 after purging with nitrogen. All pipe cutting shall be carried out with a radial pipe cutter and copper pipe fitting's shall comply with B4 alloy designation. Adequate support shall be provided while laying pipeline testing shall be done at 17 bar and after testing pipe shall be painted with two coat of enamel paint as per colour codification as per ISO 7396-Part-1 2016.Thisd partyTest reportshall be submitted. all as per approved by Engineer in charge.
83SITC of non lubricated ball type nicel Coated Isolation Valve with adapter & PTFE Seat having quarter turn / Butterfly handle opening of following sizes suitable fo flow capacity of required LPMas per the size of valve at 4.7 bar with maximum allowable pressure of 30 bars and suitable for working temperature renging - 20 °C to 130 °C. Appropriate Test Certificate/ report of product/ accessories from OEM shall be submitted. All a per approved by Engineer in charge.
84For 12 MM OD pipe
85For 15 MM OD pipe
86For 22 MM OD pipe
87For 28 MM OD pipe
88For 42 MM OD pipe
89For 54 MM OD pipe
90SITC of Gas outlet point Oxy, (AR-4), (AR-7) and Vaccum (8.7 psig) for dispeensing medical pgases (oxygen/air/Vaccum) at point of use confirming to BSEN ISO 9 170-1:2008.. The gas outlet are quick connect type. it should have colour coding for dieffrent gas source and should accept only respective adapters as required.appropriat Test Certificate/ report of product/accessories from OEM shall be submitted. The Product should be CE Certified /CE marked.
91SITC of BPC (back pressure compensated) flow meter with Humidifire made of nickei plated body. Double jacketed with flow gradations CE Marked on it. Desired flow will be setteided through rotary knob. Flow range shuld be between 0 to 15 lpm & work should be executed as per direction of engineer in charg. Appreoprite Test Certificate / report of product /accessories from OEM shall be submitted. The Product should be CE Certified/ CE Marked.
92SITC of ward suction unit (Sution Controller and Collevtion Jar - 600 ml caapicity) with posibility to configure required capcity collection Jar and should meets the reuirements as per BSEN ISO 10079-3:2009. The unit must designed to operate with the vaccum line of MGPS. The collection jar must accumpained with Vaccum regulator suitable for 0-14.7 psig with flow rate 0-55 lmp (min)for operating tempurature - 18 °C to 50 °C with ON/OFF facility. The collection Jar should be transparent, made of hogh quality polycarbonate that can be autoclaved at 121 °C for 20 minuts for reuse and provided with auto stop to avoid over spill and contamination of the suction source when the jar collection reached optimized level. The completer unit supplied with filter to protect the source and pipe line from baceterial contamination. Appropriate Test Certificate / report of product/ accessories from OEM shall be submitted . The Product shpuld be CE Certificate / CE marked
93Suction troly
94SITC of Valve box made of CRCA sheet duly powder coated with transparent window which is installed in the MGPS for oxygen and vaccum line and should comply with BSEN ISO 5211:2001 complete with ball type valves and gauses as required for oxygen compressed air and vaccum line. Appropriate Test Certificate / report of product/ accessories from OEM shall be submitted . The Product shpuld be CE Certificate / CE marked
95For Three gas
96For Four gas
97SITC of microprocessor based auto alarm system (digital type) suitable for medical gas service provided with audible and visual alarm for MGPS Zoning. It should monitors the current status of oxygen /compressed air pressure in distribution pipeline of central medical supply system and trigger the alarm to alert. The system should comply BSEN IS0: 73 96- 1 :2007
98For Three Gas Services
99For Four Gas Services
100SITC of Bed Head Panel of 900 -1500 MM long made out from high strength extruded aluminium with powder coating with at least 4 nos 6/16 Amp switch socket and provision for required number of gas outlet points as suggested by the client. Appropriate Test Certificate/report o product/accessories from OEM shall be submitted. The Product shpuld be CE Certificate / CE marked. All as per approved engineer incharge.
101Providing and fixing of Rail clamp for mountin ward vaccum unit on bed head panel complete as required
102SITC of High pressure Rubber Tube for Oxygen/Vacuum work should be executed as per direction of Engineer incharse of CE Certified.
103P&F of Bed Panel accessories like Monitor Tray, Utility Basket, Vacuum Slider, Infusion PumP Pole with IV hooks, with proper nut bolts and screw and with Proper finishing . Appropriate Test Certificate/report o product/accessories from OEM shall be submitted. The Product shpuld be CE Certificate / CE marked. All as per approved engineer incharge.
104Utility Basket
105Vacuum Slider
106Infusion Pump Pole
107Monitor Tray
108SITC of Automatic Gas Control Panel for Each oxygen/Nitrous 0xide - Control Panel .out made for M.S. duly Powder coated. Compact and fully automatic manifold control Panel cover made of M.S. duly power coated. Compact and ffully automatic manifold control system for Oxygen service. Easy to maintain and use. Designed with both safety and continuity of flow supply at a constant pressure via two banks of bottled gas cylinders with following status monitoring. A -Bank in use [Green Indicationi f - EmPW bank [Red Indication). Control Panel should have built in LED display to indicate Normal, High & Low line pressure. It.should have facility to test all above indicators working perfectly. the control panel should include pressure gauge to indicate the gas pressure of each header and pipe line distribution pressure. Capable to provide a distribution flow rate in excess of 1200 LPM. The control panel should incbrporate manual bypass system in case of failure of shifting mechanism. It should be of CE Certificate / CE marked. All as per approved engineer incharge.
109SITC of Oxygen/ Nitrous Oxide Manifold System (12+12) with tail pipes NRV - manifold system comprise of bank of 1,2+12 cylinders (Bulk cylinder of D type) capacity manifold high pressure copper pipe of size 1/2 \\" x14 swg (approx. 2.0320 mm) & high pressure fitting made of high tensile strength brass be used to connect the cylinders to the header assemblies with 24 no,s non return valves & 24 no's high pressure flexible tail pipe is made of copper pipe of 1f3\\"x1-6 swg (approx. L.6256 mm) the manifold is hydraulically tested at 225kg/cm2 pressure it is also degreased and is to be offered duly wrapped in a polythene bag make CE Approved/CE certified Note : Providing of Cylinders is in the scope of Hospital / End Client.
110SITC of Oxygen/Nitrous Oxide Manifold System [8+8) with tail pipes NRV - manifold system comprise of bank of 8+8 cylinders (Bulk cylinder of D type) capacity manifold high pressure copper pipe of size 1/2 \\" x14 swg (approx. 2.0320 mm) & high pressure fitting made of high tensile strength brass be used to connect the cylinders to the header assemblies with 16 no,s non return valves & 16 no's high pressure flexible tail pipe is made of copper pipe of 1/3\\"x16 swg (approx. L.6256 mm) the manifold is hydraulically tested at 225kg/cm2 pressure it is also degreased and is to be offered duly wrapped in a polythene bag. Appropriate Test Certificate/report of product/accessories from OEM shall be submitted. The Product shpuld be CE Certificate / CE marked. All as per approved engineer incharge.
111SITC of Dessicant Air Dryer 0utput Capacity - 60 cfm with UL Certified Microprocessor Controller with dewpoint based changeover option, Pre-filter rating 0.3 microns , Shuttle valve for Low Pressure drop, AII Aluminium resist oxidation & scale formation, Dewpoint better than -40 degree C including fixing on wall with required T & P making connections etc complete in all respect.
112SITC of Compressed Air System should consist of 1 nos. Air Compressors with Electric Motors, Electrical Switch Gear and Automatic Control SYstem of following Specifications:a)- Air Compressor L(0ne) Nos; Type Non-lubricated / Oil-free; Capacity 57 cfm at 8.5 kg/cm2; Cooling Air cooled b) Drive- Electric Motor 2 [two) Nos; Capacity 15 HP; Speed L440 rpm (syn.); Drive Belt drive; Electricity supply 440 volt, 50 Hz. 3 Phase c) Starter Star/Delta Control & Switchgear with overload protection.including making PCC foundation and fixing withbolts , anti vibration pad and making connections complete as required.
113SITC of Vacuum Pump of Single stage, capacity 4235 lpm / L49cfm(Piston displacementJ. The pumP should be suited for maximum working pressure of 29 inch mercury [730 mm mercury) The pump should be fitted with MS channel frame comPlete having vibration isolators with 1 Nos.silencers at vacuum pump discharge end. The pump should be air cooled.l sets of isolation valve, on return valve and filter should be provided with pumps. Drive Assembly. The electrical drive for vacuum pump should consist of suitable TEFC induction motor operating at 1440 RPM and connected through a V-belt. The motor shouldbe suited for 415v + 10% 50 HZ .3 Phase AC supply. Capacity : 10 HP of a rePuted make Vacuum Switch Vacuum switch of 0-760 mm range should be provided, it should have both NO/NC contracts and suitable for 230 V 50 Hz.AC supply. The vacuum switch I should be housed in weather Prool I casing with 3/B\\"BSP gas service I connection at the bottom of the Assembly. vaccum filter 1 Nos vacuum filters with drain valve. This filter should be non-ferrous and in line. including making PCC foundation and fixing with bolts , anti vibration pad and making connections complete as reouired.
114SITC of Compressed air filtration system and regulating systum :One set of 3 - stage filter The filters should be made of die cast aluminums housing with epoxy power paint on the outside and anodized surface treatment inside to prevent corrosion and ensure extra long life. The filters should have maximum contaminant removal efficiency with minimum pressure drop. Total 3-stage of filters should be used as mentioned below:-Stage 1 : coalescing filters for generalpurpose protection, removing liquid water and oil aerosol to 0.1 mg/cum. [0.1 pm) and particles down to 5 micron.Stage 2 : particular filter for dust protection, removing particles down to 1 micron. Stage 3 : high efficiency coalescing filters, removing liquid water and oilaerosol to 0.01 mg/cu.m. (0.01 pip) and particles down to 0.0L micron.Janatics regulator 1/2\\" or 3/4\\" to be installde for regulating the Output air pressure. . Appropriate Test Certificate/report of product/accessories from OEM shall be submitted. The Product shpuld be CE Certificate / CE marked. All as per approved engineer incharge.
115SITC of Hot dipped galvanized ( inside and outside ) and powder coated . Air/ Vacuum receiver tank fas per lS:7938-L97 6 and upto date amendment) for compressed air system to dampen out pulsation and condense and trap the moisture in the compressed air and also to let oil and other impurities settle down before air passes into the piping with safety valve , Drain Valve , Pressure gauge and fusible plugs of 1000 litres capacity complete as required including fixing with bolts and making cinnections complete as required.Appropriat Test Certificate/report of product/accessories from OEM shall be submitted.
116SITC of PLC Based Cascading Duplex electrical control panel suitable for 2 nos vacuum pump(1O /7.5 / 5 HP) and 2nos compressor [15 HP).Electrical Control panel with duplexing and cascading facility of air compressor- Zrios and vacuum pumps- 2 nos. powder quoted Electrical panel made out from CRCA sheet with independent starting units, with 250 Amp MCCB with spreader links of 250 Amp as incomer including rotary handle, Amp/Volt/Freq meter with required CT's , Indication lamp, push button, programmable logic controller, SMPS 1.5 A to 2.1A , Control relay,Contactors of required ratings ,timer,TP and SP MCB's as required , single phase preventer, over load protection relay, with 2 nos Star delta & 2 nos DOL starter including control wiring etc. complete with testing etc as required.(Brand of components used in Cascding panel should be as per Group I of Electrical BSR). Please Enable Macros to View BoQ information

AI Tender Summary

OUR REF NO 56593612
AUTHORITY Government Departments
TENDER VALUE 3.08 Crore
LAST DATE 10-08-2026
Authority
Authority Name Medical And Health Services
Work Description Tender for Supply Installation Testing and Commissioning of Medical Gas Pipe Line System And Modular Operation Theater Work In The Construction Of District Hospital Hindaun Distt Karauli
Basic Detail
Tender No NIT No.09/2026-27 EE M and H DivBharatpur/
Bidding Type Tender
Location
City Karauli
State Rajasthan
Key Dates
Publish Date 21 Jul 2026
Submission Date 10 Aug 2026
Open Date 11 Aug 2026
Finance
Tender Value 3.08 Crore
Tender Fee 2000
EMD 6.17 Lakhs
Exemption Not Available
Document List
1-FrontPage.pdf
363729f5-5ef4-477c-8e39-bbe7d407275b.html
BOQ_1024710.xls
MGPS_CDSCO.xls
SBD.pdf
Tendernotice_1.pdf
Tendernotice_2 (3).pdf
Tendernotice_3 (2).pdf
Tendernotice_4.pdf
Tendernotice_5.pdf
Tendernotice_6.pdf
Tendernotice_7.pdf
Terms_and_Conditions.pdf

Unlock Full AI Tender Summary

Get instant access to the complete AI-generated analysis — scope, eligibility, timeline & more.

Instant Access Secure Free
Your details are secure and used only for document delivery.

Tender Timeline

Jul 21, 2026
11:30 IST

Tender Published

Tender notice published.

Completed
Aug 10, 2026
17:00 IST

Bid Submission Deadline

Online submission via eProcurement portal.

Upcoming · 13 days
Aug 11, 2026

Bid Opening Date

Technical bids will be opened and evaluated.

Upcoming · 14 days

Tender Documents

Download All (ZIP) ↓
pdf

1-FrontPage.pdf

html

363729f5-5ef4-477c-8e39-bbe7d407275b.html

xls

BOQ_1024710.xls

xls

MGPS_CDSCO.xls

pdf

SBD.pdf

pdf

Tendernotice_1.pdf

pdf

Tendernotice_2 (3).pdf

pdf

Tendernotice_3 (2).pdf

pdf

Tendernotice_4.pdf

pdf

Tendernotice_5.pdf

pdf

Tendernotice_6.pdf

pdf

Tendernotice_7.pdf

pdf

Terms_and_Conditions.pdf

Disclaimer

We takes all possible care for accurate & authentic tender information. However users are requested to refer Original source of Tender Notice / Tender Document published by Tender Issuing Agency before taking any call regarding this tender

TenderDetail
Loading tenders