Bids Are invited for Fully Automatic Biochemistry Analyzer (V2) (Q2) - System Type: Fully Automated Random Access(For Routine And Stat S Le) Open System (Should Be Compatible To Analytical Grade Fully Automated Biochemistry Reagents) , Type Of Bioche

Issued by Statutory Bodies & Commissions/Committees · goa, Goa
Tender Value
Ref. Documents
Estimated cost
Bid Submission
28 Jul 2026
17 days left
EMD
2.07 Lakhs
Bank guarantee accepted
Document Fee
Ref. Documents
Non-refundable
Tender Type
Gem

Tender Overview

Competition Type
NCB
Bidding Type
Tender
Location / State
goa → Goa
EMD Exemption
Available
Quantity
2

Project Description

Bids Are invited for Fully Automatic Biochemistry Analyzer (V2) (Q2) - System Type: Fully Automated Random Access(For Routine And Stat S Le) Open System (Should Be Compatible To Analytical Grade Fully Automated Biochemistry Reagents) , Type Of Biochemistry: Floor Model Type/Bench Model Type , S Le Type: Serum,Plasma,Whole Blood,Urine,Csf S Les/Apecimens & Ise (Sodium,Pottasium & Chloride) , Assay Type: End-Point, Rate, Kinetic & Two-Point Assays/Methodologies & Optional Ise , Throughput Tests / Hour Photometric Tests With Ise Module: 400 Or More , Throughput Tests/Hour Photometric Tests Without Ise Module: 560 Or More , Calibration Facility: Factor, Linear (One, Two And Multi-Point), Exponential, Logit-Log, Spline, Non-Linear , Equipment Must Have Self Diagnostic Tests With Error Message And Online Display , Equipment Must Be Programmable For All Test Menus And State Of The Art Work Station Should Be Up-Gradable By Addition Of Ise Module , S Le Loading Type: S Le Loading Capacity Should Be 125 Or More And Should Have Liquid Level Detection , Equipment Should Have Both Internal And External Probe Cleaning/Washing Capacity To Avoid Carry Over , Repeat Facility Should Be Available For Calibrator And Control , Should Have Reagent Refill Message And Monitoring Facility , Minimum Number Of Test Parameters On Board At A Time: ≥ 30 , Minimum S Le Capacity At A Time: 50 To 99 , Minimum Number Of Stat S Les At A Time: ≥ 10 , Should Have Pre Post And Auto Dilution Of S Les And Return Capability For Out Of Range S Les , Probe Dispensers Must Have Level Detectors And Must Typically Use Between 2-35 Μl Of S Le And Have The Ability To Aspirate Liquid From Micro S Le Viales With Liquid Level Detection , Should Have Clot Detection Capability For S Le And Should Be A Critical Alert , Should Have Vertical Collision Protection For The Integrated Reagent (R1 & R2& S Le Probe) , Should Have Non-Stop 24 Hour Cooling With Peltier Element , Spectoral Range In Nm: 340 To 670 Nm , Wave Length Selection Method: Diffraction Grating System , Light Source: Led Light Source , Life Of L In Hrs: ≥ 1000 , Power Consumption Of L In Va: ≤ 1000 , Type Of S Le Cups: Reusable , Number Of S Le Cups Provided With Machine: 1000 , Qc Program :Qc Programme With L-J Graphs, Print Out Of Reports/ Westgard Rules , Reaction Cuvette: Reusable And Is Of Borosilicate Glass/And The Analyzer Should Have Onboard Cuvette Integrity Check And Rotating With Atleast 120 Cuvette Available For Analyser , System Should Have On Board Cooling System: Refrigerated At 6 Degree C - 11 Degree C , Provide Di Water Plant For The Washing Of The Cuvettes/ To Avoid Carry Over And Contamination (Optional) , Should Provide Compatible On Line Ups: 5Kva , Back Up Time: 1 Hour , Data Management Software: Equipment To Be Provided With Compatible Programmable Windows Based Comprehensive Data Processing And Management System Graphical User Interface Software, Lm Capability Complete Back Up Of Data Base For Calibration Control And Patient S Le Results, Capability For Bi-Directional Interfacing With Lis , Patient Result Storage Capacity: 1000000 Or More , Onboard 7 Step Cuvette Washing & 2 Step Drying Facility Should Be Provided , Should Provide Photometeric Checking Of Cuvettes Before Next Reaction , System Should Have Facility For Reading Results On Monitor And Print Out Facility , Reports: Prinout Of Reports And Full Patient Demographics Should Be Available , Should Have Availability Of Rs 232 Port/Usb For Data Transfer , Power Supply: 220-230V Ac, 50Hz-60Hz Single Phase , Reagents Position: 75 Or More , Should Provide With Bar Code Reader For Reagents , Should Provide Bar Code Reader For S Les , Seller Must Provide Original Documentary Proof Of The Date And Place Of Manufacturing Of Provided Equipment , Service Manual And Operating Manual To Be Provided , System Shall Be Provided With A Compatible Desktop Pc , Speed Of Processor In Ghz: Greater Than 3 Ghz , Ram In Mb: ≥ 512 Mb Or Compatible , Hard Disk Drive Capacity In Gb: ≥ 80 Gb , Monitor: Led/ Digital Color Monitor , Monitor Size (Inches): 17 Inches Or More , Compatible Operating System Should Be Provided In Pc , Should Provide With Laser Printer/Inkjet Printer , The Unit Shall Be Capable Of Operating Continuously In Ambient Temperature And Relative Humidity: Temp Of 10° To 25° C , Startup Kit, Calibrators, Consumables Required In Training The Staff/Technician/Doctors Should Be Provided At The Time Of Installation &Training , The Reaction Cuvettes Should Be Replaced At Free Of Cost When Ever Required During The Warranty Period , The Analyser Should Atleast Have Pre-Dilusion Ratio Of 1:200 For S Les Oh High Concentration , Waste Tank Should Have Sensing Capability To Alert The End User For Liquid Fill , Water Consumption Should Be Less Then 14 Litres Per Hour And Have Sensors To Alert The End User For Liquid Insufficiency/Exhaust , Certification And Reports , Should Have European Ce/Fda(Us)/Stqc Cb Certificate/Stqc/Iec Certificate Or Valid Electrical And Functioinal Safety Test Reports , Manufacturing Unit Certification Should Be Iso:13485 (Latest) , Should Provide With Calibration Certificates Issued By Manufacturer At The Time Of Installation , Submission Of All Necessary Certifications, Licenses And Test Reports To The Buyer At The Time Of Bid Submission And/Or Along With Supplies As Per Buyer Requirement , Analyzer Should Have Integritated Mixer Systems With Standardise Mixing Procedures For Optimal Homogenisation Of The Reagent And S Le , Miscellaneous Requirements , Principal Manufacturer Shall Ensure Uninterrupted Availability Of All Spares For Atleast 10 Years , Availability Of Toll Free Facility For Technical Support Maintened By Principal Manufacturer Or Authorized Agencies , User/Technical/Maintenance Manuals To Be Provided In English In Hard And Soft Copy , List Of Important Spares And Accessories, With Their Part Numbers To Be Provided To The Buyer At The Time Of Supplying The Equipment , Installation And Demonstration Of Equipment And Training To Be Provided After Completing Supplies Before Acceptance , The Principal Manufacturer Must Have Direct Presence/Approved Service Center In India , Calibration Certificates As Per Nabh Requirement , Warranty , Ø Warranty In Years(Including Machine, Di Water Plant If Provided , Ups, Computer And Printer) Should Be Of 2 Years , Additional Requirement , Breakdown Calls Should Be Attended Within 24 Hours Of Brakdown , Preventive Maintenance To Be Carried Out Atleast Twice A Year , Should Be Supplied Complete With Probes/Accessories, Consumables, User & Service Manuals, Certificates At Time Of Installation , Delivery: The Item To Be Delivered At 2 X North Goa District Hospital (Asilo), Mapusa, Goa Total Quantity : 2

BOQ

Sl. No. Item TitleItem Description
1System type: Fully Automated Random Access(for Routine and STAT sample) open System (Should be compatible to analytical grade fully automated biochemistry reagents)System type: Fully Automated Random Access(for Routine and STAT sample) open System (Should be compatible to analytical grade fully automated biochemistry reagents)
2Type of Biochemistry: Floor Model type/Bench model typeType of Biochemistry: Floor Model type/Bench model type
3Sample type: Serum,Plasma,whole blood,Urine,CSF samples/apecimens & ISE (sodium,pottasium & chloride)Sample type: Serum,Plasma,whole blood,Urine,CSF samples/apecimens & ISE (sodium,pottasium & chloride)
4Assay type: End-point, Rate, Kinetic & two-point assays/methodologies & optional ISEAssay type: End-point, Rate, Kinetic & two-point assays/methodologies & optional ISE
5Throughput tests / hour Photometric Tests with ISE module: 400 or moreThroughput tests / hour Photometric Tests with ISE module: 400 or more
6Throughput tests/hour Photometric Tests without ISE Module: 560 or moreThroughput tests/hour Photometric Tests without ISE Module: 560 or more
7Calibration Facility: Factor, Linear (one, two and multi-point), Exponential, Logit-Log, Spline, Non-LinearCalibration Facility: Factor, Linear (one, two and multi-point), Exponential, Logit-Log, Spline, Non-Linear
8Equipment must have self diagnostic tests with error message and online displayEquipment must have self diagnostic tests with error message and online display
9Equipment must be programmable for all test menus and state of the art work station should be up-gradable by addition of ISE moduleEquipment must be programmable for all test menus and state of the art work station should be up-gradable by addition of ISE module
10Sample loading type: Sample Loading capacity should be 125 or more and should have liquid level detectionSample loading type: Sample Loading capacity should be 125 or more and should have liquid level detection
11Equipment should have both internal and external probe cleaning/washing capacity to avoid carry overEquipment should have both internal and external probe cleaning/washing capacity to avoid carry over
12Repeat facility should be available for calibrator and controlRepeat facility should be available for calibrator and control
13Should have reagent refill message and monitoring facilityShould have reagent refill message and monitoring facility
14Minimum number of test parameters on board at a time: ≥ 30Minimum number of test parameters on board at a time: ≥ 30
15Minimum sample capacity at a time: 50 to 99Minimum sample capacity at a time: 50 to 99
16Minimum number of STAT samples at a time: ≥ 10Minimum number of STAT samples at a time: ≥ 10
17Should have pre post and auto dilution of samples and return capability for out of range samplesShould have pre post and auto dilution of samples and return capability for out of range samples
18Probe dispensers must have level detectors and must typically use between 2-35 μl of sample and have the ability to aspirate liquid from micro sample viales with liquid level detectionProbe dispensers must have level detectors and must typically use between 2-35 μl of sample and have the ability to aspirate liquid from micro sample viales with liquid level detection
19Should have clot detection capability for sample and should be a critical alertShould have clot detection capability for sample and should be a critical alert
20Should have vertical collision protection for the integrated reagent (R1 & R2& sample probe)Should have vertical collision protection for the integrated reagent (R1 & R2& sample probe)
21Should have non-stop 24 hour cooling with Peltier elementShould have non-stop 24 hour cooling with Peltier element
22Spectoral range in nm: 340 to 670 nmSpectoral range in nm: 340 to 670 nm
23Wave length selection method: Diffraction Grating systemWave length selection method: Diffraction Grating system
24Light source: LED light sourceLight source: LED light source
25Life of lamp in hrs: ≥ 1000Life of lamp in hrs: ≥ 1000
26Power consumption of lamp in VA: ≤ 1000Power consumption of lamp in VA: ≤ 1000
27Type of sample cups: reusableType of sample cups: reusable
28Number of sample cups provided with machine: 1000Number of sample cups provided with machine: 1000
29QC program :QC programme with L-j graphs, print out of reports/ westgard rulesQC program :QC programme with L-j graphs, print out of reports/ westgard rules
30Reaction Cuvette: reusable and is of borosilicate glass/and the analyzer should have onboard cuvette integrity check and rotating with atleast 120 cuvette available for analyserReaction Cuvette: reusable and is of borosilicate glass/and the analyzer should have onboard cuvette integrity check and rotating with atleast 120 cuvette available for analyser
31System should have on board cooling system: Refrigerated at 6 degree C - 11 degree CSystem should have on board cooling system: Refrigerated at 6 degree C - 11 degree C
32Provide DI water plant for the washing of the cuvettes/ to avoid carry over and contamination (optional)Provide DI water plant for the washing of the cuvettes/ to avoid carry over and contamination (optional)
33Should provide Compatible on line UPS: 5KVAShould provide Compatible on line UPS: 5KVA
34Back up time: 1 hourBack up time: 1 hour
35Data management software: Equipment to be provided with compatible programmable windows based comprehensive data processing and management system graphical user interface software, LM Capability complete back up of data base for calibration control and patient sample results, capability for Bi-directional interfacing with LISData management software: Equipment to be provided with compatible programmable windows based comprehensive data processing and management system graphical user interface software, LM Capability complete back up of data base for calibration control and patient sample results, capability for Bi-directional interfacing with LIS
36Patient result storage capacity: 1000000 or morePatient result storage capacity: 1000000 or more
37Onboard 7 step cuvette washing & 2 step drying facility should be providedOnboard 7 step cuvette washing & 2 step drying facility should be provided
38Should provide Photometeric Checking of cuvettes before next reactionShould provide Photometeric Checking of cuvettes before next reaction
39System should have facility for reading results on monitor and print out facilitySystem should have facility for reading results on monitor and print out facility
40Reports: Prinout of reports and full patient demographics should be availableReports: Prinout of reports and full patient demographics should be available
41Should have availability of RS 232 port/USB for data transferShould have availability of RS 232 port/USB for data transfer
42Power supply: 220-230V AC, 50Hz-60Hz Single phasePower supply: 220-230V AC, 50Hz-60Hz Single phase
43Reagents position: 75 or moreReagents position: 75 or more
44Should provide with Bar code reader for reagentsShould provide with Bar code reader for reagents
45Should provide bar code reader for samplesShould provide bar code reader for samples
46Seller must provide original documentary proof of the date and place of manufacturing of provided equipmentSeller must provide original documentary proof of the date and place of manufacturing of provided equipment
47Service manual and operating manual to be providedService manual and operating manual to be provided
48System shall be provided with a compatible desktop PCSystem shall be provided with a compatible desktop PC
49Speed of processor in GHZ: Greater than 3 GHZSpeed of processor in GHZ: Greater than 3 GHZ
50RAM in MB: ≥ 512 MB or compatibleRAM in MB: ≥ 512 MB or compatible
51Hard disk drive capacity in GB: ≥ 80 GBHard disk drive capacity in GB: ≥ 80 GB
52Monitor: LED/ Digital color monitorMonitor: LED/ Digital color monitor
53Monitor size (inches): 17 inches or moreMonitor size (inches): 17 inches or more
54Compatible operating system should be provided in PCCompatible operating system should be provided in PC
55Should provide with laser printer/inkjet printerShould provide with laser printer/inkjet printer
56The unit shall be capable of operating continuously in ambient temperature and relative humidity: Temp of 10° to 25° CThe unit shall be capable of operating continuously in ambient temperature and relative humidity: Temp of 10° to 25° C
57Startup kit, Calibrators, consumables required in training the staff/Technician/Doctors should be provided at the time of installation &TrainingStartup kit, Calibrators, consumables required in training the staff/Technician/Doctors should be provided at the time of installation &Training
58The reaction cuvettes should be replaced at free of cost when ever required during the warranty periodThe reaction cuvettes should be replaced at free of cost when ever required during the warranty period
59The analyser should atleast have pre-dilusion ratio of 1:200 for samples oh high concentrationThe analyser should atleast have pre-dilusion ratio of 1:200 for samples oh high concentration
60Waste tank should have sensing capability to alert the end user for liquid fillWaste tank should have sensing capability to alert the end user for liquid fill
61Water consumption should be less then 14 litres per hour and have sensors to alert the end user for liquid insufficiency/ExhaustWater consumption should be less then 14 litres per hour and have sensors to alert the end user for liquid insufficiency/Exhaust
62Certification and reportsCertification and reports
63Should have European CE/FDA(US)/STQC CB certificate/STQC/IEC certificate or valid electrical and functioinal safety test reportsShould have European CE/FDA(US)/STQC CB certificate/STQC/IEC certificate or valid electrical and functioinal safety test reports
64Manufacturing unit certification should be ISO:13485 (Latest)Manufacturing unit certification should be ISO:13485 (Latest)
65Should provide with calibration certificates issued by manufacturer at the time of installationShould provide with calibration certificates issued by manufacturer at the time of installation
66Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission and/or along with supplies as per buyer requirementSubmission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission and/or along with supplies as per buyer requirement
67Analyzer should have integritated mixer systems with standardise mixing procedures for optimal homogenisation of the reagent and sampleAnalyzer should have integritated mixer systems with standardise mixing procedures for optimal homogenisation of the reagent and sample
68MISCELLANEOUS REQUIREMENTSMISCELLANEOUS REQUIREMENTS
69Principal Manufacturer shall ensure uninterrupted availability of all spares for atleast 10 yearsPrincipal Manufacturer shall ensure uninterrupted availability of all spares for atleast 10 years
70Availability of toll free facility for technical support maintened by principal Manufacturer or authorized agenciesAvailability of toll free facility for technical support maintened by principal Manufacturer or authorized agencies
71User/Technical/Maintenance manuals to be provided in English in hard and soft copyUser/Technical/Maintenance manuals to be provided in English in hard and soft copy
72List of important spares and accessories, with their part numbers to be provided to the buyer at the time of supplying the equipmentList of important spares and accessories, with their part numbers to be provided to the buyer at the time of supplying the equipment
73Installation and Demonstration of equipment and training to be provided after completing supplies before acceptanceInstallation and Demonstration of equipment and training to be provided after completing supplies before acceptance
74The Principal Manufacturer must have direct Presence/approved service center In IndiaThe Principal Manufacturer must have direct Presence/approved service center In India
75Calibration certificates as per NABH requirementCalibration certificates as per NABH requirement
76WarrantyWarranty
77Ø Warranty in years(including machine, DI Water Plant if provided , UPS, computer and printer) should be of 2 yearsØ Warranty in years(including machine, DI Water Plant if provided , UPS, computer and printer) should be of 2 years
78Additional RequirementAdditional Requirement
79Breakdown calls should be attended within 24 hours of brakdownBreakdown calls should be attended within 24 hours of brakdown
80Preventive maintenance to be carried out atleast twice a yearPreventive maintenance to be carried out atleast twice a year
81Should be supplied complete with probes/accessories, consumables, user & service manuals, certificates at time of installationShould be supplied complete with probes/accessories, consumables, user & service manuals, certificates at time of installation
82Delivery: The item to be delivered at 2 x North Goa District Hospital (Asilo), Mapusa, GoaDelivery: The item to be delivered at 2 x North Goa District Hospital (Asilo), Mapusa, Goa

AI Tender Summary

OUR REF NO 56405669
AUTHORITY Statutory Bodies & Commissions/Committees
TENDER VALUE Ref. Documents
LAST DATE 28-07-2026
Authority
Authority Name Directorate Of Health Services
Work Description Bids Are invited for Fully Automatic Biochemistry Analyzer (V2) (Q2) - System Type: Fully Automated Random Access(For Routine And Stat S Le) Open System (Should Be Compatible To Analytical Grade Fully Automated Biochemistry Reagents) , Type Of Biochemistry: Floor Model Type/Bench Model Type , S Le Type: Serum,Plasma,Whole Blood,Urine,Csf S Les/Apecimens & Ise (Sodium,Pottasium & Chloride) , Assay Type: End-Point, Rate, Kinetic & Two-Point Assays/Methodologies & Optional Ise , Throughput Tests / Hour Photometric Tests With Ise Module: 400 Or More , Throughput Tests/Hour Photometric Tests Without Ise Module: 560 Or More , Calibration Facility: Factor, Linear (One, Two And Multi-Point), Exponential, Logit-Log, Spline, Non-Linear , Equipment Must Have Self Diagnostic Tests With Error Message And Online Display , Equipment Must Be Programmable For All Test Menus And State Of The Art Work Station Should Be Up-Gradable By Addition Of Ise Module , S Le Loading Type: S Le Loading Capacity Should Be 125 Or More And Should Have Liquid Level Detection , Equipment Should Have Both Internal And External Probe Cleaning/Washing Capacity To Avoid Carry Over , Repeat Facility Should Be Available For Calibrator And Control , Should Have Reagent Refill Message And Monitoring Facility , Minimum Number Of Test Parameters On Board At A Time: ≥ 30 , Minimum S Le Capacity At A Time: 50 To 99 , Minimum Number Of Stat S Les At A Time: ≥ 10 , Should Have Pre Post And Auto Dilution Of S Les And Return Capability For Out Of Range S Les , Probe Dispensers Must Have Level Detectors And Must Typically Use Between 2-35 Μl Of S Le And Have The Ability To Aspirate Liquid From Micro S Le Viales With Liquid Level Detection , Should Have Clot Detection Capability For S Le And Should Be A Critical Alert , Should Have Vertical Collision Protection For The Integrated Reagent (R1 & R2& S Le Probe) , Should Have Non-Stop 24 Hour Cooling With Peltier Element , Spectoral Range In Nm: 340 To 670 Nm , Wave Length Selection Method: Diffraction Grating System , Light Source: Led Light Source , Life Of L In Hrs: ≥ 1000 , Power Consumption Of L In Va: ≤ 1000 , Type Of S Le Cups: Reusable , Number Of S Le Cups Provided With Machine: 1000 , Qc Program :Qc Programme With L-J Graphs, Print Out Of Reports/ Westgard Rules , Reaction Cuvette: Reusable And Is Of Borosilicate Glass/And The Analyzer Should Have Onboard Cuvette Integrity Check And Rotating With Atleast 120 Cuvette Available For Analyser , System Should Have On Board Cooling System: Refrigerated At 6 Degree C - 11 Degree C , Provide Di Water Plant For The Washing Of The Cuvettes/ To Avoid Carry Over And Contamination (Optional) , Should Provide Compatible On Line Ups: 5Kva , Back Up Time: 1 Hour , Data Management Software: Equipment To Be Provided With Compatible Programmable Windows Based Comprehensive Data Processing And Management System Graphical User Interface Software, Lm Capability Complete Back Up Of Data Base For Calibration Control And Patient S Le Results, Capability For Bi-Directional Interfacing With Lis , Patient Result Storage Capacity: 1000000 Or More , Onboard 7 Step Cuvette Washing & 2 Step Drying Facility Should Be Provided , Should Provide Photometeric Checking Of Cuvettes Before Next Reaction , System Should Have Facility For Reading Results On Monitor And Print Out Facility , Reports: Prinout Of Reports And Full Patient Demographics Should Be Available , Should Have Availability Of Rs 232 Port/Usb For Data Transfer , Power Supply: 220-230V Ac, 50Hz-60Hz Single Phase , Reagents Position: 75 Or More , Should Provide With Bar Code Reader For Reagents , Should Provide Bar Code Reader For S Les , Seller Must Provide Original Documentary Proof Of The Date And Place Of Manufacturing Of Provided Equipment , Service Manual And Operating Manual To Be Provided , System Shall Be Provided With A Compatible Desktop Pc , Speed Of Processor In Ghz: Greater Than 3 Ghz , Ram In Mb: ≥ 512 Mb Or Compatible , Hard Disk Drive Capacity In Gb: ≥ 80 Gb , Monitor: Led/ Digital Color Monitor , Monitor Size (Inches): 17 Inches Or More , Compatible Operating System Should Be Provided In Pc , Should Provide With Laser Printer/Inkjet Printer , The Unit Shall Be Capable Of Operating Continuously In Ambient Temperature And Relative Humidity: Temp Of 10° To 25° C , Startup Kit, Calibrators, Consumables Required In Training The Staff/Technician/Doctors Should Be Provided At The Time Of Installation &Training , The Reaction Cuvettes Should Be Replaced At Free Of Cost When Ever Required During The Warranty Period , The Analyser Should Atleast Have Pre-Dilusion Ratio Of 1:200 For S Les Oh High Concentration , Waste Tank Should Have Sensing Capability To Alert The End User For Liquid Fill , Water Consumption Should Be Less Then 14 Litres Per Hour And Have Sensors To Alert The End User For Liquid Insufficiency/Exhaust , Certification And Reports , Should Have European Ce/Fda(Us)/Stqc Cb Certificate/Stqc/Iec Certificate Or Valid Electrical And Functioinal Safety Test Reports , Manufacturing Unit Certification Should Be Iso:13485 (Latest) , Should Provide With Calibration Certificates Issued By Manufacturer At The Time Of Installation , Submission Of All Necessary Certifications, Licenses And Test Reports To The Buyer At The Time Of Bid Submission And/Or Along With Supplies As Per Buyer Requirement , Analyzer Should Have Integritated Mixer Systems With Standardise Mixing Procedures For Optimal Homogenisation Of The Reagent And S Le , Miscellaneous Requirements , Principal Manufacturer Shall Ensure Uninterrupted Availability Of All Spares For Atleast 10 Years , Availability Of Toll Free Facility For Technical Support Maintened By Principal Manufacturer Or Authorized Agencies , User/Technical/Maintenance Manuals To Be Provided In English In Hard And Soft Copy , List Of Important Spares And Accessories, With Their Part Numbers To Be Provided To The Buyer At The Time Of Supplying The Equipment , Installation And Demonstration Of Equipment And Training To Be Provided After Completing Supplies Before Acceptance , The Principal Manufacturer Must Have Direct Presence/Approved Service Center In India , Calibration Certificates As Per Nabh Requirement , Warranty , Ø Warranty In Years(Including Machine, Di Water Plant If Provided , Ups, Computer And Printer) Should Be Of 2 Years , Additional Requirement , Breakdown Calls Should Be Attended Within 24 Hours Of Brakdown , Preventive Maintenance To Be Carried Out Atleast Twice A Year , Should Be Supplied Complete With Probes/Accessories, Consumables, User & Service Manuals, Certificates At Time Of Installation , Delivery: The Item To Be Delivered At 2 X North Goa District Hospital (Asilo), Mapusa, Goa Total Quantity : 2
Basic Detail
Tender No GEM/2026/B/7635523
Bidding Type Tender
Location
City goa
State Goa
Key Dates
Publish Date 10 Jul 2026
Submission Date 28 Jul 2026
Open Date 28 Jul 2026
Finance
Tender Value Ref. Documents
Tender Fee Ref. Documents
EMD 2.07 Lakhs
Exemption Available
Document List
cca4d80f-7a6e-4e91-abdc-86c67e431fd1.pdf
1780903936.xlsx

Unlock Full AI Tender Summary

Get instant access to the complete AI-generated analysis — scope, eligibility, timeline & more.

Instant Access Secure Free
Your details are secure and used only for document delivery.

Tender Timeline

Jul 10, 2026
11:30 IST

Tender Published

Tender notice published.

Completed
Jul 28, 2026
17:00 IST

Bid Submission Deadline

Online submission via eProcurement portal.

Upcoming · 17 days
Jul 28, 2026

Bid Opening Date

Technical bids will be opened and evaluated.

Upcoming · 17 days

Tender Documents

Download All (ZIP) ↓
pdf

cca4d80f-7a6e-4e91-abdc-86c67e431fd1.pdf

xlsx

1780903936.xlsx

Disclaimer

We takes all possible care for accurate & authentic tender information. However users are requested to refer Original source of Tender Notice / Tender Document published by Tender Issuing Agency before taking any call regarding this tender

TenderDetail
Loading tenders