| 1 | SUB HEAD - A : GAS PLANT EQUIPMENT |
| 2 | Supply, Installation Testing, Commissioning of Duplex MEDICAL AIR COMPRESSOR SYSTEM, multi drive control mode, Oil injected Screw air compressor with foundation, Filteration system, receiver, electrical connection from pump to control panel and all as required as per specification. entire system shall be installed as a package and compatable to fulfill the demand of system. Product should be as per specification.. Make Compressor - Garden Denver/ Hitachi / Anestiwata /Atlas copco/ Ingersol Rand/ Kirloskar / Becom Meades/ Technoex/ Medepha (Italy)/ Reanimed (Turkey)/ CMS (UK)Make Air Drier:- Trident/ Summits/ Delair |
| 3 | Main component of system are as below |
| 4 | Air compressor with capacity 49 CFM(+/- 5%) with minimum 15 HP Motor as required. Working pressure 10 Bar. |
| 5 | Air receiver tank 1500 Ltr. |
| 6 | Air filteration Set as per specification |
| 7 | Air Bacteria filter suitable capacity |
| 8 | Air dryer suitable capacity |
| 9 | Air Pr. Reducing Station 4-Bar (Duplex) |
| 10 | Supply, Installation Testing, Commissioning of Duplex MEDICAL VACUUM SYSTEM. The package shall consist of Reciprocating Vacuum pumps (including civil foundation and electrical wiring of suitable capacity from Pump to electrical panel and non return valve complete) with suitable HP Motor. The vacuum pump, motor and associated accessories shall be installed as a package by the manufacturer/installer and not as discrete items. complete system are as per specification. Make- Becker /Becom Meades/ Atlas copco/ Technoex /GD /Ingersol Rand/ Anestiwata/ Hitachi/ Busch/ Garden Denver/ Edward/ Reanimed (Turkey)/ CMS (UK) |
| 11 | Main component of system are as below |
| 12 | Reciprocating Vacuum Pump With capacity 3000 Ltr FAD (+/- 5%) with suitable Motor (7.5 H.P.) |
| 13 | Pressure Switch |
| 14 | Air receiver tank 1500 Ltr. |
| 15 | Silencer muffler type 100 Cfm |
| 16 | Vacuum Bacterial Filter |
| 17 | SITC of OXYGEN MANIFOLD minimum of 2 x 10 cylinders as size given in the drawings having top frame comprising of high pressure copper pipe of size 16 mm O.D. 2 mm thick with high pressure brass fitting made of high tensile brass, NRV and high pressure copper tailpiece made of high pressure copper pipe. 1x10 bank shall be a reserve source of oxygen supply. It shall automatically supply the pipeline when primary source becomes exhausted or fails due to any reason. The system shall have source shut-off valve for easy changing and positioning, without hook. The cylinder should be placed with the help of cylinder brackets and fixing chains which should be zinc plated appropriate. Make- Sigma/ Aktiv/ HMP/ DTPL/ MPS/ Linde/ Prime Technomed/ Synergic/ MEC |
| 18 | S.I.T.C of Notified CE Marked, compliant with HTM/IS/ NFPA Oxygen Change-Over Reduction Box (Automatic Control Panel) of Digital fully Automatic type, capable of switching from “Bank in Use” to “Reserve Bank” without fluctuation in delivery line pressure and without the need for external electrical power. The Changeover panel shall be rated for min. 120 QMH at 400 kPA, and shall include the following accessories. Panel shall comply with HTM/NFPA as per specification. Make - Medicop/ Beacon Medeas/ Drager/ Allied Health Care USA/ Power EX USA/ Phoenix Pipeline UK/ Tri Tech USA/ Precision UK/ Ohio/ Reanimed (Turkey)/ CMS (UK) |
| 19 | Main Component/ Feature is as below |
| 20 | Touch Panel LCD Display |
| 21 | Programmble microprocesor based controller |
| 22 | Able for Data Logging |
| 23 | able to show flow of gas on display with help of optional flow Sensor |
| 24 | Double Stage Rugulation |
| 25 | Overflow Safety Valve |
| 26 | Audio Visual alarm |
| 27 | CE Certified with Noyified no. on it |
| 28 | Oxygen Emergency System (4+4):- The oxygen emergency system contains 8 nos Gas cylinder manifold along with copper header, Brass Non Return valve & High pressure copper tail pipe for each cylinder and one number of semi auto matic control panel complete. Make- Sigma/ Aktiv/ HMP/ DTPL/ MPS/ Linde/ Synergic/ MDD/ NSP/ MR/ MEC |
| 29 | SITC of NITROUS OXIDE MANIFOLD SYSTEM Minimum of 2 x 2 cylinders and reserve source of nitrous oxide will be 2x1 cylinders. Cylinder size shall be as given in the drawings having top frame comprising of high pressure copper pipe of size 16mm ODx 2 mm wall thickness and with high pressure brass fitting made of high tensile brass, NRV, high pressure copper tailpiece made of high pressure copper. Reserve source of N2O shall automatically supply the pipeline when primary source of supply become exhausted or fails due to any reason. (Nitrous Oxide gas cylinders are not in the scope of this work) Make- Sigma/ Aktiv/ HMP/ DTPL/ MPS/ Linde/ Synergic/ MDD/ NSP/ MR/ MEC |
| 30 | Repair Installation testing & Commissioning of Notified CE Marked, Nitrous Oxide Change-Over Reduction Box (Automatic Control Panel) of Digital fully Automatic type, capable of switching from “Bank in Use” to “Reserve Bank” without fluctuation in delivery line pressure and without the need for external electrical powe.the changeover panel is rated for 50 QMH at 400 kPA. old panel requird repair for service/replace of High Pressure, line pressure regulators and required sensor as per requirement ot restore all functioning .Make- Medicop/ Allied Health Care USA/ Power EX USA/ Phoenix Pipeline UK/ Tri Tech USA/ Precision UK/ Ohio/ Medepha (Italy)/ Reanimed (Turkey)/ CMS (UK) |
| 31 | Nitrous Oxide Emergency Manifold:- The Nitrous Oxide Emergency system contains 2 Nos Gas cylinder manifold along with copper header, Brass Non Return valve & High pressure copper tail pipe for each cylinder and one number of high pressure regulator. Make- Aktiv/ HMP/ DTPL/ Sigma / MPS/ Linde/ Synergic/ MDD/ NSP/ MR/ MEC/ Prime Technologies |
| 32 | Supply, Installation Testing, Commissioning of gas lockable AVSU/CONTROL-CLOSING VALVE BOXES (INTEGRATED WITH AREA ALARM) with integrated alarm system as per specifications, and as required. Product should be as per specification. As per approved make attached. Make: Drager/ Ohio/ Medicop/Becon-Medas/ Pneumatech (UK), Tritech Medical (USA), Allied Healthcare (USA), Delta P (Italy), Medepha Italy/ Reanimed (Turkey)/ CMS (UK) |
| 33 | Must have Gas Pressure alarm facility LED digital display, audio visual feature |
| 34 | must have data logging feature |
| 35 | Must have emergency feed facility in case of shutdown for intrepted supply |
| 36 | SS sheet construction, CE certified/UL Listed with relevant notified no. on it as required, entry/ exist of gases from upward direction |
| 37 | 3 Services (Oxygen, MA-4, Vacuum) |
| 38 | 4 Services (Oxygen, MA-4, Nitrous Oxide,Vacuum) |
| 39 | 6 Services |
| 40 | Master Main Alarm including all Gases |
| 41 | SUB HEAD B : ZONAL ACESSORIES |
| 42 | SITC of copper pipes for Medical Gases, complete with elbows, flanges, specials, others fittings, and steel suppots . The copper pipes should be of solid drawn seamless deoxidized, un arsenical, half hard, tempered and of degreased material confirming to BS-EN 13348.2001 Part 1 Table X of copper pipe. All copper pipe shall be inspected and certified by third party inspection agency for medical use before dispatch and plugged / capped at both end. Make- Maxflow/ Rajco Tubes / Mandev/ Janya |
| 43 | 54mm OD x 1.2 mm thk |
| 44 | 42 mm OD x 1.2 mm thk |
| 45 | 28 mm OD x 0.9 mm thk |
| 46 | 22 mm OD x 0.9 mm thk |
| 47 | 15 mm OD x 0.9 mm thk |
| 48 | 12 mm OD x 0.6 mm thk |
| 49 | S.I.T.C of Isolation Ball Valves suitable for Medical Gas use, for Installation for the Purpose of Isolation of Medical Gas Mains distribution/control system on Respective Floor/plant/area. It must have quarter turn open and close operation. It must be suitable for 10 bar working pressure. Make: RB, RN, Matrix, L&T, Legris/ Audco/ Cimberio (Cim) |
| 50 | 54mm φ |
| 51 | 42mm φ |
| 52 | 28mm φ |
| 53 | 22mm φ |
| 54 | 15mm φ |
| 55 | 12mm φ |
| 56 | Sub Head C : (MEDICAL GAS PIPELINE SYSTEM ACCESSORIES) |
| 57 | SITC Oxygen Flow Meter - Made from brass, chrome plated. It shall be back pressure comphensated type having housing of Polycorbonate autoclayble resuable flow range 0- 15 LPM. Humidifier shall be compatible with flowmeter and reusable and as per specifications and as required. Product should be make as per specification. Make- Aktiv/ Medimatic/ HMP/ Sigma/ DTPL/ Perfect/ Reanimed (Turkey)/ MDD MPS/ NSP/ MR/ MEC |
| 58 | SITC Ward Vacuum Unit - Regulator capable of controlling vacuum between 0 to 760 mm Hg and Polycorbonate autoclayble, resuable type receiver jar of not less 600 ml. Vacuum regulator shall have a dial gauge, on/off switch, valve control knob and as per specifications as required. Product should be make as per specification. Make: Aktiv/ HMP/Sigma/ DTPL/ Medimatic/ NSP/ MDD/ MPS/ Linde/ Prime Technologies/ Synergic/ MR/ MEC |
| 59 | Supply, Installation, Testing and commissioning of High pressure Hose antistatic duly colour coded for oxygen Nitrous oxide, Compressed Air and Vacuum as per technical specifications. Make- Sigma/ Aktiv/ HMP |
| 60 | Supply, Installation, Testing and commissioning of Low pressure Hose antistatic flexiable for Vacuum as per technical specifications. Make- Sigma/ Aktiv/ HMP |
| 61 | SUB HEAD D : EARTH , ELECTRICAL & MISCELLANEOUS WORK |
| 62 | Electrical Panel for air and vacuum system as per drawings, specifications and as required, electrical panel should be made of sheet metal/ epoxy painted and having all essential electrical gadgets, phase preventer complete with comprising of MCCBs, voltmeter, amperemeter, gauges, Hour meter, fuses, and main switches and wiring individual starter units, for motors of all packages like Vacuum Pump System air compressor system, single phase circuits shall be provided for connection of various accessories.Only 415V, 50 Cycles, 3 Phase supply shall be provided by the department. Termination of the same shall be done by the contractor including lugs, crimping and any other accessories, appropriate earthing as required. Make: Tricolite/ KEPL/ International Switch Gears/ L&T/ Schematic |
| 63 | Earthing with copper earth plate 600 mm X 600 mm X 3 mm thick including accessories, and providing masonry enclosure with cover plate having locking arrangement and watering pipe of 2.7 metre long etc. with charcoal/ coke and salt as required. |
| 64 | Providing and fixing 25 mm X 5 mm G.I. strip in 40 mm dia G.I. pipe from earth electrode including connection with G.I. nut, bolt, spring, washer excavation and re-filling etc. as required. |
| 65 | Providing and fixing 25 mm X 5 mm G.I. strip on surface or in recess for connections etc. as required. |
| 66 | Providing and fixing 6 SWG dia G.I. wire on surface or in recess for loop earthing as required. |
| 67 | Providing and fixing 8 SWG dia GI Wire on surface or in recessfor loop earthing complete as required. |
| 68 | SUB HEAD E : CIVIL WORK |
| 69 | Providing and laying in position cement concrete of specified grade excluding the cost of centering and shuttering - All work up to plinth level : |
| 70 | 1:2:4 (1 cement : 2 coarse sand (zone-III) derived from natural sources : 4 graded stone aggregate 20 mm nominal size derived from natural sources) |
| 71 | 1:5:10 (1 cement : 5 coarse sand (zone-III) derived from natural sources : 10 graded stone aggregate 40 mm nominal size derived from natural sources) |
| 72 | Providing and laying in position specified grade of reinforced cement concrete, excluding the cost of centering, shuttering, finishing and reinforcement - All work up to plinth level : |
| 73 | 1:1.5:3 (1 cement : 1.5 coarse sand (zone-III) derived from natural sources : 3 graded stone aggregate 20 mm nominal size derived from natural sources) |
| 74 | Centering and shuttering including strutting, propping etc. and removal of form for : |
| 75 | Suspended floors, roofs, landings, balconies and access platform |
| 76 | Steel reinforcement for R.C.C. work including straightening, cutting, bending, placing in position and binding all complete upto plinth level. |
| 77 | Thermo-Mechanically Treated bars of grade Fe-500D or more. |
| 78 | Brick work with common burnt clay F.P.S. (non modular) bricks of class designation 7.5 in foundation and plinth in: |
| 79 | Cement mortar 1:4 (1 cement : 4 coarse sand) |
| 80 | Stone work (machine cut edges in Veneer work) for wall lining upto 10 m height, backing filled with a grout of 12 mm thick cement mortar 1:3 (1 Cement : 3 coarse sand) and jointed with Cement mortar 1:2 (1 cement : 2 stone dust), including grinding area hand grinder on joints rubbing and polishing complete. (To be secured to the backing and the sides by means of cramps and pins which shall be paid for separately). |
| 81 | Polished Kota stone slabs on exposed face. |
| 82 | 25 mm thick |
| 83 | Providing and fixing Ist quality ceramic glazed wall tiles conforming to IS: 15622 (thickness to be specified by the manufacturer), of approved make, in all colours, shades except burgundy, bottle green, black of any size as approved by Engineer-in-Charge, in skirting, risers of steps and dados, over 12 mm thick bed of cement mortar 1:3 (1 cement : 3 coarse sand) and jointing with grey cement slurry @ 3.3kg per sqm, including pointing in white cement mixed with pigment of matching shade complete. (Size 300 x 600 mm) |
| 84 | Providing and fixing ISI marked flush door shutters conforming to IS : 2202 (Part I) decorative type, core of block board construction withframe of 1st class hard wood and well matched teak 3 ply veneering with vertical grains or cross bands and face veneers on both faces of shutters. |
| 85 | 30 mm thick including ISI marked Stainless Steel butt hinges with necessary screws |
| 86 | Extra for Providing and fixing flush doors with decorative veneering instead of non decorative ISI marked flush door shutters conforming to IS: 2202 (Part I) |
| 87 | On both side (2 x 10.54) |
| 88 | Providing and fixing aluminium sliding door bolts, ISI marked anodised (anodic coating not less than grade AC 10 as per IS : 1868), transparent or dyed to required colour or shade, with nuts and screws etc. complete : |
| 89 | 250x16 mm |
| 90 | Providing and fixing aluminium tower bolts, ISI marked, anodised (anodic coating not less than grade AC 10 as per IS : 1868 ) transparent or dyed to required colour or shade, with necessary screws etc. complete : |
| 91 | 150x10 mm |
| 92 | Providing and fixing aluminium handles, ISI marked, anodised (anodic coating not less than grade AC 10 as per IS : 1868) transparent or dyed to required colour or shade, with necessary screws etc. complete : |
| 93 | 125 mm |
| 94 | Providing and fixing aluminium hanging floor door stopper, ISI marked, anodised (anodic coating not less than grade AC 10 as per IS : 1868) transparent or dyed to required colour and shade, with necessary screws etc. complete. |
| 95 | Twin rubber stopper |
| 96 | Providing and fixing pressed steel door frames conforming to IS: 4351, manufactured from commercial mild steel sheet of 1.60 mm thickness, including hinges, jamb, lock jamb, bead and if required angle threshold of mild steel angle of section 50x25 mm, or base ties of 1.60 mm, pressed mild steel welded or rigidly fixed together by mechanical means, including M.S. pressed butt hinges 2.5 mm thick with mortar guards, lock strike-plate and shock absorbers as specified and applying a coat of approved steel primer after pre-treatment of the surface as directed by Engineer-in-charge: |
| 97 | Profile B |
| 98 | Fixing with adjustable lugs with split end tail to each jamb |
| 99 | 12 mm cement plaster of mix : |
| 100 | 1:4 (1 cement: 4 coarse sand) |
| 101 | Distempering with 1st quality acrylic distemper (ready mixed) having VOC content less than 50 gram/litre, of approved manufacturer and of required shade and colour all complete to achieve even shade and colour : |
| 102 | New work (two or more coats) over and including water thinnable priming coat with cement primer having VOC content less than 50 gram/litre |
| 103 | Dismantling tile work in floors and roofs laid in cement mortar including stacking material within 50 metres lead. |
| 104 | For thickness of tiles 10 mm to 25 mm |
| 105 | Providing and laying water proofing treatment in sunken portion of WCs, bathroom etc., by applying cement slurry mixed with water proofing cement compound consisting of applying : |
| 106 | First layer of slurry of cement @ 0.488 kg/sqm mixed with water proofing cement compound @ 0.253 kg/ sqm. This layer will be allowed to air cure for 4 hours. |
| 107 | Second layer of slurry of cement @ 0.242 kg/sqm mixed with water proofing cement compound @ 0.126 kg/sqm. This layer will be allowed to air cure for 4 hours followed with water curing for 48 hours |
| 108 | The rate includes preparation of surface, treatment and sealing of all joints, corners, junctions of pipes and masonry with polymer mixed slurry. |
| 109 | Dismantling of Wall tiles and scraping the plaster including removing of malba complete in all respect and to the entire satisfaction of EIC. |
| 110 | Providing & fixing IRC encloser jali of mesh size 40 mm x 40 mm of 10 gauge with frame work of MS angle 40x40x5 mm and MS flat 32x5 mm including painting with two coats of synthetic enamel paint over a coat of steel arrangement all as specified. (Size of angle grid should not be more than 12 sft) with two coats of black paint as per site requirement as desired by EIC |
| 111 | Labour for hiring crane for lifting and placing stainless steel tank SS 304 on terrace complete in all respect. |
| 112 | The Buy back for Dismentaling and old unused item installed at site including cleaning of area as per site requirement to the entire satisfaction of EIC. |
| 113 | (Rs.1,18,70,960 - Rate quoted = Tendered Amount - Rs.6,00,000/- = Net Payable Amount) Please Enable Macros to View BoQ information |