| 1 | Should be fully automatic double door high pressure steam sterilizer with rectangular chamber and can accommodate 10 STU in one cycle. | Should be fully automatic double door high pressure steam sterilizer with rectangular chamber and can accommodate 10 STU in one cycle. |
| 2 | Should have Integrated steam generator should be made of 316 L steel and two level water ring vacuum pump with water saving mechanism and silent operation. | Should have Integrated steam generator should be made of 316 L steel and two level water ring vacuum pump with water saving mechanism and silent operation. |
| 3 | The system should have inbuilt thermal degasification. Auto steam generator blow down systems should be inbuilt. | The system should have inbuilt thermal degasification. Auto steam generator blow down systems should be inbuilt. |
| 4 | Sterilization chamber- doors should be made of 316 L and chamber jacket should be made up of high quality AISI 316 L steel. | Sterilization chamber- doors should be made of 316 L and chamber jacket should be made up of high quality AISI 316 L steel. |
| 5 | System must have vertical sliding door (pneumatic/ electrical operated) with safety locks, manual actuation of doors. | System must have vertical sliding door (pneumatic/ electrical operated) with safety locks, manual actuation of doors. |
| 6 | should have automatic door gasket seal and should have safety feature if accidental steam is released. | should have automatic door gasket seal and should have safety feature if accidental steam is released. |
| 7 | All pipes in contact with steam should be made of AISI 316 L. | All pipes in contact with steam should be made of AISI 316 L. |
| 8 | Should have integrated thermal /Ink printer for reports. | Should have integrated thermal /Ink printer for reports. |
| 9 | Should be complete auto diagnostic system providing error codes on display in case of failure. | Should be complete auto diagnostic system providing error codes on display in case of failure. |
| 10 | Should have Dual microprocessor control system/ PLC for fully independent checking and control of the sterilization system. | Should have Dual microprocessor control system/ PLC for fully independent checking and control of the sterilization system. |
| 11 | Manometer gauge / digitally display on screen should be provided for the chamber & steam generator pressure. | Manometer gauge / digitally display on screen should be provided for the chamber & steam generator pressure. |
| 12 | Should have Graphic indication of curves of chamber temp and pressure on display. | Should have Graphic indication of curves of chamber temp and pressure on display. |
| 13 | Display of remaining cycle time, display of temperature and pressure in chamber & jacket should be provided on the touch screen . | Display of remaining cycle time, display of temperature and pressure in chamber & jacket should be provided on the touch screen . |
| 14 | Absolute pressure sensors for the precise recording and control of vacuum and pressure in the chamber and additionally in the jacket. | Absolute pressure sensors for the precise recording and control of vacuum and pressure in the chamber and additionally in the jacket. |
| 15 | Should have alarm systems for compressed air deficiency, water deficiency. | Should have alarm systems for compressed air deficiency, water deficiency. |
| 16 | should have Integrated water saving device, Touch screen display, Chip card reader/USB and RS 232 interface, should be capable of remote service assistance. | should have Integrated water saving device, Touch screen display, Chip card reader/USB and RS 232 interface, should be capable of remote service assistance. |
| 17 | Sterilization to be achieved at 134°C/121°C degrees. The sterilizer should have at least 4 preloaded programs for standard sterilization, containers,Textile and rubbers. It should also have preloaded test programs for Bowie-Dick test and Vacuum test. Should have the option to load customer specific programs in the sterilizer. | Sterilization to be achieved at 134°C/121°C degrees. The sterilizer should have at least 4 preloaded programs for standard sterilization, containers,Textile and rubbers. It should also have preloaded test programs for Bowie-Dick test and Vacuum test. Should have the option to load customer specific programs in the sterilizer. |
| 18 | Water consumption should not be more than 220- 280 L per cycle. | Water consumption should not be more than 220- 280 L per cycle. |
| 19 | The sterilizer should come with standard accessories like carriage with shelf & side railings, one set of transport and loading trolley with each sterilizer, etc. The sterilizer should be complete with side and top paneling. | The sterilizer should come with standard accessories like carriage with shelf & side railings, one set of transport and loading trolley with each sterilizer, etc. The sterilizer should be complete with side and top paneling. |