Material Bio medically compliant Polyurethane PU polymer no pyrogenic and latex free Axial extension shall be approximate 70 cm plus minus 5 percent Open ended Distal configuration should consist of a meticulously formed atraumatic profile Should have full length intrinsic radiopacity must be integrated within the shaft matrix and circumferential graduation markings at standardized intervals and engineered torque stability to mitigate deformation or kinking during fluoroscopic or cystoscopy manipulation Secure Connector to allow leak free instillation of contrast via syringe to be provided to be supplied as sterile individually sealed units designated for single use clinical application only Secondary packaging must indicate the manufacturers name product details LOT number Mfg and Exp Dates and sterilization method The product must be ethylene oxide sterilized with visual confirmation to be provided Manufacturer shall possess a valid ISO 13485 2016 certification and the device must be CDSCO registered
2
5 Fr Ureteric Catheter 70 cm
Material Bio medically compliant Polyurethane PU polymer no pyrogenic and latex free Axial extension shall be approximate 70 cm plus minus 5 percent Open ended Distal configuration should consist of a meticulously formed atraumatic profile Should have full length intrinsic radiopacity must be integrated within the shaft matrix and circumferential graduation markings at standardized intervals and engineered torque stability to mitigate deformation or kinking during fluoroscopic or cystoscopy manipulation Secure Connector to allow leak free instillation of contrast via syringe to be provided to be supplied as sterile individually sealed units designated for single use clinical application only Secondary packaging must indicate the manufacturers name product details LOT number Mfg and Exp Dates and sterilization method The product must be ethylene oxide sterilized with visual confirmation to be provided Manufacturer shall possess a valid ISO 13485 2016 certification and the device must be CDSCO registered
3
6 Fr Ureteric Catheter 70 cm
Material Bio medically compliant Polyurethane PU polymer no pyrogenic and latex free Axial extension shall be approximate 70 cm plus minus 5 percent Open ended Distal configuration should consist of a meticulously formed atraumatic profile Should have full length intrinsic radiopacity must be integrated within the shaft matrix and circumferential graduation markings at standardized intervals and engineered torque stability to mitigate deformation or kinking during fluoroscopic or cystoscopy manipulation Secure Connector to allow leak free instillation of contrast via syringe to be provided to be supplied as sterile individually sealed units designated for single use clinical application only Secondary packaging must indicate the manufacturers name product details LOT number Mfg and Exp Dates and sterilization method The product must be ethylene oxide sterilized with visual confirmation to be provided Manufacturer shall possess a valid ISO 13485 2016 certification and the device must be CDSCO registered
4
Smoke Evacuator Filter
Should retain solid particles down to nano size at highest efficiencies remove bio hazardous contaminants retain aerosol droplets in liquid phase and retain volatile substances in the gas phase Should be able to retain inorganic particles bacteria viruses and cells efficiently Should possess a hydrophobic filtration medium retaining any contaminants in the liquid phase It should contain a medium for the retention of volatile chemicals Filtration efficiency Airborne particle removal efficiency less than 99 point 99 percent in accordance with EN1822 Airborne cellular bacterial removal efficiency 99 point 99 percent Airborne viral removal efficiency 99 point 99 percent Organic volatile reduction significant reduction of index substances Effective filtration area Approximately 50 sq cm connector ISO male luer connector with a rotating locking collar
5
Catheter valve
Ergonomically designed slide button to pass urine when no urine bag is attached valve should open automatically when urine bag to the valve is connected and should close once bag tube is removed should be suitable for use with both urethral and suprapubic catheters
6
120 cm Veress needle
Needle should be sproing loaded blunt style mechanism should have sharp bevelled tip for quick and safe insertion outer diameter size 2 mm working length 120 mm bis certification
7
150 cm Veress needle
Needle should be sproing loaded blunt style mechanism should have sharp bevelled tip for quick and safe insertion outer diameter size 2 mm working length 150 mm bis certification