Fourier Transform Infrared Spectroscopy (Ftir)• The Instrument Spectral Resolution Should Be 0.8 Cm-1• The Standard Spectral Operating Range Should Be No Less Than 6,000 - 500Cm-1 For Znse Optics.• Signal To Noise Ratio Typically >50,000:1 (1Min Measurem
Tender Overview
Project Description
Fourier Transform Infrared Spectroscopy (Ftir)• The Instrument Spectral Resolution Should Be 0.8 Cm-1• The Standard Spectral Operating Range Should Be No Less Than 6,000 - 500Cm-1 For Znse Optics.• Signal To Noise Ratio Typically >50,000:1 (1Min Measurement Time,Spectral Resolution4cm-1)• It Should Be Portable.• Spectrometer Components Like Source, Laser, Detector, Andinterferometer Must Be Continuously Monitored For Operation Within