| 1 | SITC of MGPS of medically grade as per BSEN-13348 with third party certification from Lloyds/TUV degreased (ie free from oil, grease and partivulate matters) seamless copper pipe confirming to ASTM B88 and all copper joining done by brazing with AWSA5.8, classification BCuP-5 or BS 1845 following procedure in accordance with ASTM B828 after purging with nitrogen. All pipe cutting shall be carried out with a radial pipecutter and copper pipe fitting's shal comply with B4 alloy designation. Adequate support shall be provided while laying pipeline testing shall be done at 17 bar and after testing pipe shall be painted with two coat of enamel paint as per colour codification as per ISO 7396-Part-1 2016.Third party Test report shall be submitted. All as per pre approved by Engineer in charge. For additional technical parameters of products/ work , refer Annexure \\"A\\" attached with P.W.D. Electrical B.S.R. 2022. |
| 2 | SITC of non lubricated ball type Nickel Coated Isolation Valve with adapter & PTFE Seat having quarter turn / Butterfly handle opening of following sizes suitable for flow capacity of required LPM as per the size of valve at 4.7 bar with maximum allowable pressure of 30 bar and suitable for working temperature ranging - 20°C to 130°C. Appropriate Test Certificate/ report of product/accessories from OEM shall be submitted. All as per pre approved by Engineer in charge. For additional technical parameters of products/ work , refer Annexure \\"A\\" attached with P.W.D. Electrical B.S.R. 2022. |
| 3 | For 12 MM OD pipe |
| 4 | For 15 MM OD pipe |
| 5 | For 22 MM OD pipe |
| 6 | For 28 MM OD pipe |
| 7 | SITC of Gas outlet points Oxy, N2O, (AR-4), (AR-7) and Vaccum (8.7 psig) for dispensing medical gases (oxygen/air/Vaccum) at point of use confirming to BSEN ISO 9170-1:2008. The gas outlet are quick connect type It should have color coding for different gas source and should accept only respective adapters as required. Appropriate Test Certificate/ report of product/accessories from OEM shall be submitted. The Product should be CE Certified / CE Marked. All as per pre approved by Engineer in charge. For additional technical parameters of products/ work , refer Annexure \\"A\\" attached with P.W.D. Electrical B.S.R. 2022. |
| 8 | SITC of BPC (back pressure compensated) flow meter with humidifer made of nickel plated body , double jacketed with flow gradations CE Marked on it . Desired flow will be setteled through rotary knob . Flow range should be between 0 to 15 lpm & work should be executed as per direction of engineer in charge .Appropriate Test Certificate/ report of product/accessories from OEM shall be submitted. The Product should be CE Certified / CE Marked. All as per pre approved by Engineer in charge. For additional technical parameters of products/ work , refer Annexure \\"A\\" attached with P.W.D. Electrical B.S.R. 2022. |
| 9 | SITC of Ward suction unit (Suction Controller and Collection Jar - 600 ml capacity) with posibility to configure required capacity collection Jar and should meets the requirements as per BSEN ISO 10079-3:2009. The unit must designed to operate with the vaccum line of MGPS. The collection Jar must accumpained with Vaccum regulator suitable for 0-14.7 psig with flow rate 0-55 Lmp (min) for operating temperature - 18°C to 50°C with ON/OFF facility. The collection Jar should be transparent, made of high quality polycarbonate that can be autoclaved at 121°C for 20 minutes for reuse and provided with auto stop to avoid over spill and contamination of the suction source when the jar collection reached optimized level. The complete unit supplied with filter to protect the source and pipe line from bacterial contamination. Appropriate Test Certificate/ report of product/accessories from OEM shall be submitted. The Product should be CE Certified / CE Marked. All as per pre approved by Engineer in charge. For additional technical parameters of products/ work , refer Annexure \\"A\\" attached with P.W.D. Electrical B.S.R. 2022. |
| 10 | SITC of Bed Head Panel of 1200 MM long made out from high strength extruded aluminium with powder coating with at least 4 nos 6/16 Amp switch socket and provision for required number of gas outlet points as suggested by the client. Appropriate Test Certificate/ report of product/accessories from OEM shall be submitted. The Product should be CE Certified / CE Marked. All as per pre approved by Engineer in charge. For additional technical parameters of products/ work , refer Annexure \\"A\\" attached with P.W.D. Electrical B.S.R. 2022. |
| 11 | P&F of Bed Panel accessories like Monitor Tray, Utility Basket, Vacuum Slider, Infusion Pump Pole with IV hooks, with proper nut bolts and screw and with proper finishing. Appropriate Test Certificate/ report of product/accessories from OEM shall be submitted. The Product should be CE Certified / CE Marked. All as per pre approved by Engineer in charge. For additional technical parameters of products/ work , refer Annexure \\"A\\" attached with P.W.D. Electrical B.S.R. 2022. |
| 12 | Monitor Tray |
| 13 | Utility Basket |
| 14 | Vacuum Slider |
| 15 | Infusion Pump Pole |
| 16 | SITC of Valve box made of CRCA sheet duly powder coated with transparent window which is installed in the MGPS for oxygen and vaccum line and should comply with BSEN ISO5211:2001 complete with ball type valves and gauges as required for oxygen compressed air and vaccum line. Appropriate Test Certificate/ report of product/accessories from OEM shall be submitted. The Product should be CE Certified / CE Marked. All as per pre approved by Engineer in charge. For additional technical parameters of products/ work , refer Annexure \\"A\\" attached with P.W.D. Electrical B.S.R. 2022. |
| 17 | Three Gas Service |
| 18 | Five Gas Service |
| 19 | SITC of microprocessor based auto alarm system (digital type) suitable for medical gas service provided with audible and visual alarm for MGPS Zoning it should monitors the current status of oxygen/ compressed air pressure in distrubuton pipeline of central medical supply system and trigger the alarm to alert. The system should comply BSEN ISO:7396-1:2007 Appropriate Test Certificate/report with reliable warranty of product/accessories from OEM shall be submitted. the product should be CE Certified/ CE Marked. All as per pre approved by Engineer in charge. for additional technical parameters of products/work, refer Annexture\\"A\\" attached with P.W.D. Electrical B.S.R. 2022. |
| 20 | For Three Gas Services |
| 21 | Five Gas Service |
| 22 | SITC of Theatre Vacuum unit(2 jar nos capacity 1000 ml each) with posibility to configure required capacity collection Jar and should meets the requirements as per BSEN ISO 10079-3:2009. The unit must designed to operate with the vaccum line of MGPS. |
| 23 | suction troly |
| 24 | S/I/T/C of : Isolation Panel system with insulation monitoring and insulation fault location floor mounted dust and vermin proof (IP30) compartmentlaized cubical type 10 KVA medical isolation panels shall comfirm to following IEC statndard IEC 60364-7-710, IEC 61439-1 and -2, IEC 61557-8, IEC 61558-2-15 (IEC test certificate to be attached with panel) .The panel shall be fabricated out of 14 gauge CRCA sheet steel required hardware duly treated throgh 7 tank process for derusting. The panel door shall have trnsparant inspection window. The panel shall have following 1 no 10 KVA isolation transformer, 1 no Insulation monitor suitable for all outgoings, 1 no 63 Amp 10 KA 2 pole MCB, 12 nos 16/20 Amp 2 pole C curve MCB, all required suitable rating current transformer for incomer & outgoing, 1 no digital voltmeter & ammeter, suiatble for monitoring insulation fault, transfomer fault, insulation fault location. The panel shall be capable to perform insulation test. The Insulation monitor & fault locator should have inbuilt RS485 communication protocol. Panel shall be capble of detecting overheating of cables inside the OT isolation panel and sending alarm as well as alart over BMS. including making necessary connection etc complete in all respect |
| 25 | S/1/T/C of ISO & CE CERTIFIED Twin Plate X-ray viewing Screen designed toprovide a high level of control luminance without flicker. Size:- 1000x700x95 Deep, indigenous, complete set etc as reqd. |
| 26 | S/1/T/C of ISO & CE CERTIFIED writing board, indigenous, complete set etc as required. |
| 27 | SITC of class II IP51, ISO & CE certified single arm pendent made by alloy 6063-T6 extended round aluminium profile thickness not less than 3 mm for vertical & 2.5 mm for body(pendant) rust proof (passivate) trivelent zinc plated(8-10 micron m)steel with 810 mm long lateral arm, functional radious 636 mm , rotating utility console from 0 to 330 degree and console size (L X H X W) 170mm x 160 mm x1224 mm .rotation of 330 degree+ utility console 330 degree , 7Nos pre - piped & customizable combination for GAS outlets, modular type at least six shutter type multi stand socket of 16 amp socket with switch, 1-RJ45,1-RJ11telephone socket.support system for medical rail320 mm-3 nos with clamps, monitor stand with mounting bracket assembly and arm ,vertical round pole for syringe pump cum IV stand with clamp, crimped medical gas tube to connect secondary devices, table top, pole holder extensions, shelves, catheter jar,suction jar, bottle holders,utilities basket, rubbish bag holder, clamp for medical rail, wire basket etc |
| 28 | SITC of class II IP51, ISO & CE certified Double arm pendent made by alloy 6063-T6 extended round aluminium profile thickness not less than 3 mm for vertical & 2.5 mm for body(pendant) rust proof (passivate) trivelent zinc plated(8-10 micron m)steel with 810 mm+610mm long lateral arm, functional radious 636 mm+436mm , rotating utility console from 0 to 330 degree and console size (L X H X W) 170mm x 160 mm x1224 mm .rotation of 330 degree+ utility console 330 degree , 7Nos pre - piped & customizable combination for GAS outlets, modular type at least six shutter type multi stand socket of 16 amp socket with switch, 1-RJ45,1-RJ11telephone socket.support system for medical rail320 mm-3 nos with clamps, monitor stand with mounting bracket assembly and arm ,vertical round pole for syringe pump cum IV stand with clamp, crimped medical gas tube to connect secondary devices, table top, pole holder extensions, shelves, catheter jar,suction jar, bottle holders,utilities basket, rubbish bag holder, clamp for medical rail, wire basket etc Please Enable Macros to View BoQ information |