| 2 | Dynamic Viscosity range: 15 mPa.s to 6M mPa.s, Torque range (10% to 100%): 0.00673 mNm to 0.0673 mNm, Torque resolution: ± 0.1 %of FSR (i.e. 0.0673 µNm), Accuracy: ±1.0 % of FSR (i.e 0.673µNm), Repeatability: ±0.2%, Speed range: 0.01 rpm to 250 rpm, Number of speeds: Speed list with 18 standard speeds - 1 fixed custom speed C0 (200 rpm) - Freely selectable speeds within the speed range - Speed resolution: 0.01 rpm to 9.99 rpm: 0.01 rpm - 10 rpm to 59.9 rpm: 0.1 rpm - > 60 rpm: 1 rpm, Spindle Coupling, Magnetic Spindle Coupling, Spindle Recognition, Instrument should automatically recognise Spindle, MOC of Spindle, SS316 Measurement Modes: Manual Mode, Stop at Time, Stop at Torque/Temperature/Viscosity Speed/Time Scan, All methods with: QC-limit function, User Instructions, User Interface: 7” color touchscreen, Data Storage: Data memory (upto 999 measurements) Print/Export from data storage, Data Transfer: PDF Export, CSV Table, Page Printer (USB or network), Temperature Measurement: Pt-100 Temp probe, Speed/Spindle Selection Mode: In stand-alone mode, the instrument should automatically run at different speeds to determine ideal speed needed to reach 80% torque. If torque cannot be reached with the attached spindle, the instrument should make spindle suggestions. Compatibility for Keyboard and Mouse: It should be possible to connect wireless keyboard and mouse directly to the viscometer, No of Ports: There should be minimum three USB ports and One Ethernet port in instrument, Measuring System SC4-18, Sample Volume : 6.7 ml, Viscosity range: 1.5 to 30K, DIN Adaptor, Required for measurement with a concentric cylinder (CC). Double-gap (DG) measuring system, Visco check VC 500, Viscosity standard for calibration |