| 1 | Copper Pipeline: - Supplying and Distribution of copper piping. The scope of work shall cover all distribution piping and terminal units for oxygen, nitrous oxide, vacuum and compressed air. Solid drawn, seamless, de-oxidized, non-arsenical, half-hard, tempered and de-greased copper pipe conforming to EN 13348: 2008. All copper pipes will be de-greased & delivered capped at both ends. The pipes will be accompanied with manufacturers test certificate for the physical properties & chemical composition. The Pipe Sizes to be used are from among as under: Copper fittings shall be made of copper and suitable for a steam working Pressure of 17 bar and especially made for brazed socket type connections. All copper fittings will be conforming to EN 1254-1, should be factory de-greased, certified, and individually packed and identified for medical use. Fittings should be kite marked up to 54 mm size. Etc. as per requirement(i) 12 mm (OD) x 0.90 mm Thick.MAKE: - MEXFLOW OR MANDEV OR PAPRIWAL (MEDICAL CERTIFIED.) |
| 2 | Copper Pipeline: - Supplying and Distribution of copper piping. The scope of work shall cover all distribution piping and terminal units for oxygen, nitrous oxide, vacuum and compressed air. Solid drawn, seamless, de-oxidized, non-arsenical, half-hard, tempered and de-greased copper pipe conforming to EN 13348: 2008. All copper pipes will be de-greased & delivered capped at both ends. The pipes will be accompanied with manufacturers test certificate for the physical properties & chemical composition. The Pipe Sizes to be used are from among as under: Copper fittings shall be made of copper and suitable for a steam working Pressure of 17 bar and especially made for brazed socket type connections. All copper fittings will be conforming to EN 1254-1, should be factory de-greased, certified, and individually packed and identified for medical use. Fittings should be kite marked up to 54 mm size. Etc. as per requirementi) 15 mm (OD) x 0.90 mm Thick.MAKE: - MEXFLOW OR MANDEV OR PAPRIWAL (MEDICAL CERTIFIED.) |
| 3 | Floor Isolation Valve :-Supplying and Installation of Isolation valve of following sizes and isolation valves will be Non Lubricated, 900 turn level, Ball type, suitable for oxygen service. All valves shall be pneumatically tested for twice the working pressure and factory degreased for medical gas service before supply etc. as per requirement. (MAKE- JK/MR/RN)i)Isolation Valve-12 mm. (MAKE- JK/MR/RN). |
| 4 | Floor Isolation Valve :-Supplying and Installation of Isolation valve of following sizes and isolation valves will be Non Lubricated, 900 turn level, Ball type, suitable for oxygen service. All valves shall be pneumatically tested for twice the working pressure and factory degreased for medical gas service before supply etc. as per requirement. (MAKE- JK/MR/RN)i)Isolation Valve-15 mm. (MAKE- JK/MR/RN). |
| 5 | Supplying and Installation for heavy duty regulator IOX-14 type for Oxygen with following specification:-• IOX regulator Control will have first stage regulators it’s capable of delivering 100-200 PSI outlet pressure. It should comply international standards: HTM 02-01/NFPA 99/SO- 7396-1.• Both the first stage regulator in the oxygen IOX regulator will have non-halogenated polymer in the high pressure side toensure that there will be no ignition due to adiabatic compression furthermore, 40 micro filters should be provided at the inlet of each high pressure regulators of the oxygen regulator.• The iox heavy duty regulators will be delivering an outlet pressure of 60 psig.• The iox heavy duty regulators will be covered with aesthetically suitable cover for safe operation indicating the respective services. The iox heavy duty regulators will have built in transformer to ensure safe operation by low voltage etc. as per requirement.(MAKE- JK/MR/AKTIVE) |
| 6 | Supplying and Installation of Digital alarm system with single channel microprocessor controlled alarm and services should have the following features.• Digital Display of Line Pressure for all the services with factory calibrated pressure sensors.• Color coded LED Display of Line pressure status (High – Caution – Normal – Caution–Low). Audible Alarm for High & Low pressure condition.• Test and Alarm Acknowledge (Mute) facility. (Alarm acknowledge (Mute) time span is programmable from 1 to 60 min).• Programming facility of alarm limits from front panel.• Facility to connect to remote alarm box by potential free contacts provided in the alarm box. Small and compact design. Light Weight (3 kg)• Imported highly sensitive gas pressure sensors & USA/CE marked power supply. Mounted on a powder coated MS box.• Nut & Nipples are to be provided for connection.• Low voltage internal operation with input power supply of 220V AC.• Easy wall mounting facility etc. as per requirement.(MAKE- JK/MR/AKTIVE) |
| 7 | Supplying and Installation of double lock Oxygen outlet point. (O2).(MAKE- JK/MR/AKTIVE) |
| 8 | Supplying and Installation of outlet probe or connector for BPC flow meter. (SS-304 standard) Oxygen (O2) probe / connector.(MAKE- JK/MR/AKTIVE) |
| 9 | Supplying and Installation of BPC Flow meter with Humidifier with following specification: -Back Pressure Compensated flow meter will be of accurate gas flow measurement with following features:• Control within a range of 0 – 15 lpm (calibration within +/- 10%).• It will meet strict precision and durability standard.• The flow meter body shall be made of brass chrome plated materials.• The flow tube and shroud components shall be made of clear, impact resistant polycarbonate.• Flow Tube shall have large and expanded 0 – 5 lpm range for improved readability at low flows.• Inlet filter of stainless steel wire mesh to prevent entry of foreign particles.• The humidifier bottle shall be made of unbreakable polycarbonate material and auto cleavable at 121 degree Centigrade temperature etc. as per requirement.(MAKE- JK/MR/AKTIVE) |
| 10 | Supplying and Installation of Bed head panel Horizontal/Vertical it should have the following features- Minimum length 5 feet. Efficient, safe & Robust design in extruded 19luminium section. Smooth curved surfaces, with acceptable color choice Should have an integrated rail system to mount accessories Segregation of services i.e. low voltage supplies, high voltage supplies, and medical gases should be maintained throughout. Entire pipe line should run in continuous horizontal panels with no break for each unit & length as per area where it has to be installed. Provision for outlet cut off:-Oxygen -1 point, Electrical sockets- 6. Combined 15/5 etc. as per requirement. (MAKE- JK/MR/AKTIVE) i) Bed head panel Size-2 feet |
| 11 | Supplying and Installation of Bed head panel Horizontal/Vertical it should have the following features- Minimum length 5 feet. Efficient, safe & Robust design in extruded 19luminium section. Smooth curved surfaces, with acceptable color choice Should have an integrated rail system to mount accessories Segregation of services i.e. low voltage supplies, high voltage supplies, and medical gases should be maintained throughout. Entire pipe line should run in continuous horizontal panels with no break for each unit & length as per area where it has to be installed. Provision for outlet cut off:-Oxygen -1 point, Electrical sockets- 6. Combined 15/5 etc. as per requirement. (MAKE- JK/MR/AKTIVE) i) Bed head panel Size-3 feet |