| 1 | 12 mm lockable line valves with nut/nipple adapter |
| 2 | 15 mm lockable line valves with nut/nipple adapter |
| 3 | 22 mm lockable line valves with nut/nipple adapter |
| 4 | 28 mm lockable line valves with nut/nipple adapter |
| 5 | Suuply,fixing & testing ofdigital medical gas area alarm for line pressureof the following services in compliance with relevant standard and manufactured in compliance with ISO 13485, ISO 45001.Make: AKtive/asha/MPS/PNS or equv. Approved. |
| 6 | 3 services ((Oxygen,MA-4 & Vacuum). |
| 7 | Supply,fixing,testing & commissioning of terminal outlets gas specific with with matching s/s probe/ adaptor for the servicesindicated (viz. Oxygen,Medical Air & vacuum) and to accept only compatible quick connect pin/geometric index type adapters conforming to ISO7396 standard , manufactured in compliance with ISO13485,ISO45001.Make: PNS/Aktive/Asha/MPS or equv. approved. |
| 8 | Oxygen Outlet |
| 9 | Vacuum Outlet |
| 10 | Medical Air-4 Outlet |
| 11 | Suuply,fixing,testing & commissioning of Oxygen flow meter made for brass chrom plated . It shall be back pressure compensated type having housing of polycarbonate autoclavable reusable with flow range of 0-15 LPM fittedwithmatching adopter for oxygen outlet point & humidifer and humidifier shall shall be compatible with flow meter . It shall be manufactured in compliance relevant ISO standards etc as required.Make: PNS/Active/Asha/MPS/PMP or equv. approved. |
| 12 | Supply,installation,testing and commissioning of indigenous ward vacuum unit with regulator capable of controlling vacuuum between 0 to 760 mm of Hg & polycarbonate autoclavable , reusable type jar receiver ,vacuum regulator shall have dial gauge , on/off switch ,valve control knob, hygrophobic filter and additional polycarbonate safety trap bottle as per specifications and as required.The product shall be in compliance with ISO 7396 standard.Make: PMP/Asha/Active or equivalent approved. |
| 13 | 1000 Ml Receiver Jar |
| 14 | Low pressure suction tube for suction outlet adopter as required. |
| 15 | High pressure suction tube for suction outlet adopter as required. |
| 16 | Supply,installation,testing & commissioning of modular bed head panel made out of high grade aluminium with provision of gas outlets. Each panel shall be equiped with 4 pole vacuum slider , utility basket,monitor tray &each panel shall be fitted with with 6 Nos. of 15 A, modular switch & 6 Nos. of 6 pin modular power sockets.The panel shall be fabricated by extrusion method & powder coated as per approved colour . It should have dual compartment facility separately for gases and electrical services as per international standards for safety & service. Unit shall be manufactured as ISO13485 & 45001. Make: PNS/Active/asha/MPS or equv. approved. |
| 17 | Bed Head Panel for 3 gases outlet each of 1200 mm long. |
| 18 | Supply,installation,testing & commissioning of Area zonal valve box (three line) for required service as per ISO7396 standard comprising of isolation valve,pressure gauge,locking arrangement , transparent breakable front to break in case of emergency. The unit shall be manufactured as per ISO 13485 & 45001 etc as required. Make: Speed/Asha/Active or equivalent approved. |
| 19 | 3 services ((Oxygen,MA-4 & Vacuum). |