| 1 | SITC of MGPS of medically grade as per BSEN-13348 with third party certification from Lloyds/TUV degreased [ie free from oil, grease and particulate matters) seamless copper pipe confirming to ASTM BBB and all copper joining done by brazing with AWSAS.B, classification BCuP-S or BS 1845 following procedure in accordance with ASTM BB2B after purging with nitrogen. All pipe cutting shall be carried out with a radial pipe cutter and copper pipe fitting\"s shall comply with 84 alloy designation. Adequate support shall be provided while laying pipeline testing shall be done at L7 bar and after testing pipe shall be painted with two coat of enamel paint as per colour codification as per ISO 7396-Part-\2076. |
| 2 | SITC of non lubricated ball type GM Isolation Valve with adapter & PTFE Seat having quarter turn handle opening of following sizes suitable for flow capacity 1500 LPM at 4.7 bar with maximum allowable pressure of 30 bar and suitable for working temperature ranging - 20\"C to L30\"C.Valve should beCE approved/ CE certified. Annexure \"A\" attached with this BSR . |
| 3 | For 12 MM OD pipe |
| 4 | For 15 MM OD pipe |
| 5 | For 22 MM OD pipe |
| 6 | For 28 MM OD pipe |
| 7 | For 42 MM OD pipe |
| 8 | For 54 MM OD pipe |
| 9 | SITC of Gas outlet point (AR -4), (AR - 7) and vaccum (8.7 Psig) for disPensing medical gases [oxygen/air/VaccumJ at point of use confirming to BSEN ISO 9170-1:2008. The gas outlet are quick connect type. it should have colour coding for dieffrentgas source and should accept only respective adapters as required. |
| 10 | SITC of BPC (back pressure compensated) flow meter with Humidifire made of nickei plated body. Double jacket with flow gradations CE Marked on it. Deired flow will be setteided through ritary knob. Flow range shuld be between 0 to 15 lpm & work should be executed as per direction of engineer in charg. Appreoprite Test Certificate / report with liable warranty of product /accessories from OEM shall be submitted. The Product should be CE Certified/ CE Marked. |
| 11 | Ward vacuum unit with 600 ml Capacity collection jar |
| 12 | Suction troly |
| 13 | SITC of Valve box made of CRCA sheet duly powder coated with transparent window which is installed in the MGPS for oxygen and vaccum line and should comply with ENISOS211.2001 complete with ball type valves and gauses as required for oxygencompressed air and vaccum line fthree gas) |
| 14 | For Three gas |
| 15 | For Four gas |
| 16 | SITC of microprocessor based auto alarm system (digital type) suitable for medical gas service provided with audible and visual alarm for MGPS Zoning. It should monitors the current status of oxygen /compressed air pressure in distribution pipeline of central medical supply system and trigger the alarm to alert. The system should comply BSEN IS0: 73 96- 1 :2007 |
| 17 | For Three Gas Services |
| 18 | For Four Gas Services |
| 19 | SITC of Bed Head Panel of 1200 -1500 MM long made out from high strength extruded aluminium with powder coating with 4 nos 6/L6 Amp switch socket and provision forfour Nos gas outlet point work should be executed as per directions of engineer in charge and of CE certified makes. |
| 20 | Providing and fixing of Rail clamp for mountin ward vaccum unit on bed head panel com Plete as required\" |
| 21 | SITC of High pressure Rubber Tube for Oxygen/Vacuum work should be executed as per direction of Engineer incharse of CE Certified. |
| 22 | P&F of B.H.Panel accessories like Monitor Tray, Utility Basket, Vacuum Slider, Infusion PumP Pole with IV hooks, with proper nut bolts and screw and with Proper finishing & work should be executed as per directions of engineer in charge of CE certified/ CE Approved. |
| 23 | Utility Basket |
| 24 | Vacuum Slider |
| 25 | Infusion Pump Pole |
| 26 | Moniter Tray |
| 27 | SITC of Automatic Gas Control Panel for Each oxygen/Nitrous 0xide - Control Panel .out made for M.S. duly Powder coated. Compact and fully automatic manifold control system for 0xygen services. Easy to maintain and use. Designed with both safety and continuity of flow supply at a constant pressure via two banks of bottled gas cylinders with following status monitoring. A -Bank in use [Green Indicationi f - EmPW bank [Red Indication). Control Panel should have built in LED display to indicate Normal, High & Low line pressure. It.should have facility to test all above indicators working perfectly. the control panel should include pressure gauge to indicate the gas pressure of each header and pipe line distribution pressure. Capable to provide a distribution flow rate in excess of 1200 LPM. The control panel should incbrporate manual bypass system in case of failure of shifting mechanism. It should be of CE certified product. |
| 28 | SITC of Oxygen/ Nitrous Oxide Manifold System (12+12) with tail pipes NRV - manifold system comprise of bank of 1,2+12 cylinders (Bulk cylinder of D type) capacity manifold high pressure copper pipe of size 1/2 \" x14 swg (approx. 2.0320 mm) & high pressure fitting made of high tensile strength brass be used to connect the cylinders to the header assemblies with 24 no,s non return valves & 24 no\"s high pressure flexible tail pipe is made of copper pipe of 1f3\"x1-6 swg (approx. L.6256 mm) the manifold is hydraulically tested at 225kg/cm2 pressure it is also degreased and is to be offered duly wrapped in a polythene bag make CE Approved/CE certified Note : Providing of Cylinders is in the scope of Hospital / End Client. |
| 29 | SITC of Oxygen/Nitrous Oxide Manifold System [8+8) with tail pipes NRV - manifold system comprise of bank of 8+8 cylinders (Bulk cylinder of D type) capacity manifold high pressure copper pipe of size 1/2 \" x14 swg (approx. 2.0320 mm) & high pressure fitting made of high tensile strength brass be used to connect the cylinders to the header assemblies with 24 no,s non return valves & 24 no\"s high pressure flexible tail pipe is made of copper pipe of 1f3\"x1-6 swg (approx. L.6256 mm) the manifold is hydraulically tested at 225kg/cm2 pressure it is also degreased and is to be offered duly wrapped in a polythene bag make CE Approved/CE certified Note : Providing of Cylinders is in the scope of Hospital / End Client. |
| 30 | SITC of Dessicant Air Dryer 0utput Capacity - 60 cfm with UL Certified Microprocessor Controller with dewpoint based changeover option, Pre-filter rating 0.3 microns , Shuttle valve for Low Pressure drop, AII Aluminium resist oxidation & scale formation, Dewpoint better than -40 degree C including fixing on wall with required T & P making connections etc complete in all respect. |
| 31 | SITC of Compressed Air System should consist of 1 nos. Air Compressors with Electric Motors, Electrical Switch Gear and Automatic Control SYstem of following Specifications:a)- Air Compressor L(0ne) Nos; Type Non-lubricated / Oil-free; Capacity 57 cfm at 8.5 kg/cm2; Cooling Air cooled b) Drive- Electric Motor 2 [two) Nos; Capacity 15 HP; Speed L440 rpm (syn.); Drive Belt drive; Electricity supply 440 volt, 50 Hz. 3 Phasec) Starter Star/Delta Control & Switchgear with overload protection.including making PCC foundation and fixing withbolts , anti vibration pad and making connections complete as required. |
| 32 | SITC of Vacuum Pump of Single stage, capacity 4235 lpm / L49cfm(Piston displacementJ. The pumP should be suited for maximum working pressure of 29 inch mercury [730 mm mercury) The pump should be fitted with MS channel frame comPlete having vibration isolators with 1 Nos.silencers at vacuum pump discharge end. The pump should be air cooled.l sets of isolation valve, on return valve and filter should be provided with pumps. Drive Assembly. The electrical drive for vacuum pump should consist of suitable TEFC induction motor operating at 1440 RPM and connected through a V-belt. The motor shouldbe suited for 415v + 10% 50 HZ .3 Phase AC supply. Capacity : 10 HP of a rePuted make Vacuum Switch Vacuum switch of 0-760 mm range should be provided, it should have both NO/NC contracts and suitable for 230 V 50 Hz.AC supply. The vacuum switch I should be housed in weather Prool I casing with 3/B\"BSP gas service I connection at the bottom of the Assembly. vaccum filter 1 Nos vacuum filters with drain valve. This filter should be non-ferrous and in line. including making PCC foundation and fixing with bolts , anti vibration pad and making connections complete as reouired. |
| 33 | SITCof Compressed air filtration system:One set of 3 - stage filter The filters should be made of die cast aluminums housing with epoxy power paint on the outside and anodized surface treatment inside to prevent corrosion and ensure extra long life. The filters should have maximum contaminant removal efficiency with minimum pressure drop. Total 3-stage of filters should be used as mentioned below:-Stage 1 : coalescing filters for generalpurpose protection, removing liquid water and oil aerosol to 0.1 mg/cum. [0.1 pm) and particles down to 5 micron.Stage 2 : particular filter for dust protection, removing particles down to 1 micron. Stage 3 : high efficiency coalescing filters, removing liquid water and oilaerosol to 0.01 mg/cu.m. (0.01 pip) and particles down to 0.0L micron. |
| 34 | SITC of Hot dipped galvanized ( inside and outside )Air receiver tank fas per lS:7938-L97 6 and upto date amendment) for compressed air system to dampen out pulsation and condense and trap the moisture in the compressed air and also to let oil and other impurities settle down before air passes into the piping with safety valve , Drain Valve , Pressure gauge and fusible plugs of 1000 litres capacity complete as required including making PCC foundation and fixing with bolts and making connections complete as required. |
| 35 | SITC of PLC Based Cascading Duplex electrical control panel suitable for 2 nos vacuum pump(1O /7.5 / 5 HP) and 2nos compressor [15 HP).Electrical Control panel with duplexing and cascading facility of air compressor- Zrios and vacuum pumps- 2 nos. powder quoted Electrical panel made out from CRCA sheet with independent starting units, with 250 Amp MCCB with spreader links of 250 Amp as incomer including rotary handle, Amp/Volt/Freq meter with required CT\"s , Indication lamp, push button, programmable logic controller, SMPS 1.5 A to 2.1A , Control relay,Contactors of required ratings ,timer,TP and SP MCB\"s as required , single phase preventer, over load protection relay, with 2 nos Star delta & 2 nos DOL starter including control wiring etc. complete with testing etc as required.(Brand of components used in Cascding panel should be as per Group I of Electrical BSR). |