Tender For Supply And Installation 01 No. Electrochemical Workstation At Chemistry Department- 1 Electrochemical Workstation 2 No. of channel: 1Interface: USBMeasured current/Potential resolution: 0.0033% full scale/bitApplied Potential Resolution: 150 MicroV or BetterIMP Amplitude Range: 0.00001 V- 0.7 V RMSElectrode Configuration, Terminals: 2,3,4 and 5Potentiostat Rise time: 400ns or betterReference Electrode Input Impedance: > 10TOhmCurrent Range: 100nA - 1A or BetterCompliance Voltage: +/- 12 V or BetterScan Rate: 10 microV/s to 10000 V/s or Better-0.1 mV/sPotentiostat Bandwidth: 10Mhz or BetterReference Electrode Input Bias Current: 1 pA or BetterApplied Current Resolution: 0.03% of Current Range or BetterMaximum Current: +/-1 A or BetterBandwidth of Electrometer: 10MHz or BetterIMP Frequency Range10µHZ to 2MHzMaximum Potential: +/- 10 V or BetterAdditional InformationFeature: Licensed and full version of software provision for measurement of I-V measurements, Imax, Pmax fill factor etc. upgradation of software should be made as and whenever required, computer with latest version of windows OS compatible with instrument software.Additional Requirements1. Instrument should come with multichannel (8channel) dummy cell for external cell calibration & also with EMI shielding & GND option.2. Instrument should be having interface for spectro-electrochemistry analysis.3. Instrument must be having floating/GND option.4. Instrument software must be able to support multiple channels with Multichannel add on facility and also should be having custom experiment design option with multiple voltage/current waveforms.Electrochemical Cell System & Accessories: Ag/Agcl reference electrode(B10) -1qty, Platinum mesh electrode (1x1cm) - 1qty, Working electrode holder (clip type)-1qty, Gastight cell with stand (20ml)-1qtyWarranty Summary: 2 YearsUpgradable option: There should be a provision that in future the system will use in Spectro-electrochemistry Setup. System should be upgradable to Bipotentiostat in futureTechniques: Cyclic Voltammetry, Rapid alternating Polarity, Linear Sweep Voltammetry, Tafel Plot (TAFEL) potentiodynamic deactivation, pitting corrosion, corrosion rate, linear Polarization, Corrosion current etc, Chrono Amperometry, Chrono Coulometry, Differential Pulse Voltammetry, Amperometric i-t Curve, Bulk Electrolysis with Coulometry, AC Impedance, Impedance - Time, Impedance - Potential, Chrono Potentiometry, Chronopotentiometry with Current Ramp, Multi-Current Steps, Potentiometric Stripping Analysis, Open Circuit Potential - Time, Galvanostat, IR Compensation, Bode and Nyquist & Mott-Schottky plot, Chronoamperometry, Pulsed Voltammetry (SCP, NPV, DPV & SWV), OCP Measurement, Tafel Analysis (Corrosion Measurement),Linear Polarization (Corrosion Measurement), Chronopotentiometry, Galvanostatic Charge-Discharge