2 DE Exoscope with integrated illuminator , working distance upto 25 to 75 cm in while light, 20 to 30cm for ICG light, Exoscope should be 10mm diameter 0 degree length 11cm, autoclavable, with fiber optic light transmission incorporated andcondenser lenses. Scope shouldbe compatible with white light and ICG light
2
Fiber Optic Light Cable, size 4.8 mm, length 300 cm, heat-resistant.
3
Clamping cylinder folding, for flexible mounting of 10 mm telescopes to telescope sheath, autoclavable.
4
The ClampingCylinder allows vertical movement and rotation of thetelescope. For use with Clamping jaw.
5
Clamping Jaw, matal, for use withTelescopes, clamping range 16.5 up to 23 mm, with fastener
6
L-shaped Articulated Stand especially long operating range with one mechanical central clamp for allfive joint functions, height 45-48 cm, operating range 62-66 cm
7
Socket to clamp on the operating table for use with European and United States Standard rails also suited for rails from 25x10 upto 35x8 mm with lateral clamping element for height adjustment of thearticulated stand
8
Plastic Container for Sterilization and Storage of Variable Instrument Sets, container should be perforated with transparent lid. Container should be with silicone mat and two-level storage. External minimum dimensions (w x d x h): 545 x 260 x 115 mm. The plastic container should be suitable for steam, gas and hydrogen peroxide sterilization.