Procurement Brightness Tester , Equipments Details , Brighntess / Opacity Tester Spectro Photometer With Eric / Ink Removal Value Measurement, Pc / Touch Screen, With Printer, Inclusive= Color-Brightness Software, Working Standards, Apertus, Black Cavity, Cables. Brightness Measuring Standard Tiles & Essentials Spares To Measure Above Parameters, Photometric Range / Measurement Range= 0-200%, Repeatability=More Than 0.01 E Cielab Units On White Creamic Tiles, Reproducibility= More Than 0.25 E Cielab Units ( Average ) On Ncs Colour Samples , Instrument Required For Above , Spectral Range / Wave Length=360-700, Reporting Range / =400-700 Nm Adjustable To 360- 700 Mm, Measurement Geometry=D / 0 According To Iso 2469, Measuring Principle = Sp 2000 Dual- Beam Monochromator With 2X 256 Diodes In 2 Nm Intervals, Illuminant / Observer =D65 / 10 , C / 2 , D50 / 2 , T11 / 10 , Eric / Ink Removal Value, Bandwidth / Band Pass / Wave Length Pitch = 10 Nm, Light Source =Pulsed Xenon Lamp D 65 Filtered, = 4 -20 S, Measurement Time =Depending On The Type Of Measurement , Software , Operating System= Microsoft Windows, Graphical Reports= Reflectance Factor, Opacity, Scattering, Absorption, Trend, L* -A * -B*, Properties= X , Y, Z, L * , A* , B * , X , Y, Rx, Ry, Rz, Cie Whiteness / Tint - D65 Resp. C, Fluorescene, Iso Brightness, 65- Brightness, Yellowness, Opacity, Transparency, K / N - Value, Ie , Residual Ink, Dw , Pe, S, K, Sr457, Kr457, Metamerism, E And Others , Installation Required , Dimensions= Printer 0.3X0.3X0.3M / 12X12x12in Volume 0.2M2 / 7 Ft3, Pc / Screen 0.6X0.6X0.5 M / 24X24x20 In Volume 0.43 / 14 Ft3 , Applicable Standards= Iso Standards 12625—7, 12625-16, 2469, 2470, 2471, 5631, Tappi Optical Methods T519, T525, T527, T560, T567