| 1 | Supply and installation of Medical Equipments |
| 2 | Ventilator |
| 3 | Specification:Operational Requirements: The portable ventilator should be light weight( < 10 kg) Should be microprocessor controlled, portable, light weight. Should operate with main electric supply as well as with battery. Should be able to work both with cylinders and pipeline, connectors and high pressure tubing of appropriate length to be supplied. Demonstration of the equipment is a must. Technical Specifications Should have turbine technology for supplying air- oxygen mixture. Should have following modes of ventilation:CMV, Assist-control, SIMV, PS-PEEP Audio-visual alarms for a. Low supply pressure b. High/low airway pressure c. Leakage/disconnection d. Power failure e. Apnea f. Low battery Should have following settings a. TV 30 – 2200ml b. PEEP/CPAP – 0-40 cm H20 , PS 0-60 cm H2O c. RR up to 40bpm d. I: E ratio 1:99 to 3:1 e. FiO2 : 21 – 100% Battery backup for minimum 4 hour. Should fix, on rails of transport trolley and on stand with wheels |
| 4 | System Configuration Accessories, spares and consumables Portable Ventilator-01 Adult Reusable /Autoclavable Silicon Patient Circuit-01 Oxygen Hose-01 Air Hose-01 Rechargeable Batteries- 01 set Environmental factors The unit shall be capable of being stored continuously in ambient temperature of 0-500 C and relative humidity of 15-90%.The unit shall be capable of operating continuously in ambient temperature of 10 -40deg C and relative humidity of 15-90%.Shall meet IEC-60601-1-2:200(Or Equivalent BIS) General Requirements of Safety for Electromagnetic Compatibility. Power Supply Power input to be 220-240VAC, 50Hz Standards, Safety and Training Electrical safety conforms to standards for electrical safety IEC-60601 / IS-13450 Product should be US FDA and European CE Manufacturer should have ISO certification for quality standards. Comprehensive warranty as per bid. Compliance Report to be submitted in a tabulated and point wise manner clearly mentioning the page/para number of original catalogue/data sheet. Any point ,if not substantiated with authenticated catalogue /manual, will not be considered |
| 5 | Documentation User Manual in English. Service manual in English. Certificate of calibration and inspection from factory List of Equipment’s available for providing calibration and routine maintenance support as per manufacturer documentation in service / technical manual List of important spare parts and accessories with their part number and costing Log book with instruction for daily, weekly, monthly and quarterly maintenance checklist. The job description of the hospital technician and company service engineer should be clearly spelt out Must submit user list and performance report as per bid. |
| 6 | Suction apparatus |
| 7 | Specification The Suction Machine should be compact.Lightweight and capable of providing suciton not less than 700±10 mm Hg at 30 ltrs/min.The unit should be supplied with two Autoclavable collection Jar of 2 Ltrs. each.The pump used should be ISI Mark. |
| 8 | Blood gas analyzer ABG |
| 9 | Specification The machine can measure following parameters in the same single cartridge: PH, PO2, PCO2, NA+, K+, Cl-, Ca++, Hb%, Hct and lactate It can give 23 calculated parameters also: CH+, CH+(T), pH(T), pCO2(T), pO2(T), HCO3-atd, HCO3-std, BB(B), BE(B), BE(ecf), ct CO2, Ca++(7.4), AnGap, tHb(est), sO2(est), pO2(A-a), pO2(A-a) (T), pO2(a/A), pO2(a/A)(T), RI, RI(T), pO2/FI O2, pO2(T)/FI O2 User interface: 8 inches Touch Screen LCD Display. Sample volume in micro liters: 80µl. It is based on individual blood sample cartridge system. It has the acid base interpretation either graphically or descriptive wise. Accept samples automatically by aspiration and be syringe. Should have long life sensors of at least 1 year. Calibration and measurement reagents should be available in a easy reagent packs/bottles/cartridges. Cartridge/ reagent to be used can be stored mostly at room temperature between 2-30 Degree C. Shall have patient Bar Code facility, Ability to run a wide range of patient panels on one system. Weight in kg: 6.5 kg. Throughput time in seconds: no more than 70 sec. It should auto calibrate with facility for individual electrode calibration through calibration pack. The cartridge should have a shelf life of at last 06 months. Test results should be displayed on the LCD screen along with print out. It should have data storage facility. System should be CE certified. The system should be based on technology of the year 2015 or after. |
| 10 | Over Head warmer |
| 11 | Specification Movable Warmer StandBest in class highly reliable micro-controller and CE marked temperature sensor ensures safe and durable micro temperature controlled environment for an InfantControl Modes: Skin, Manual & PrewarmCalibration free thermistor based sensorsDisplay for Actual Baby’s Temp (°C), Set Temp., Heater Power % and LCD Screen for display of Current Control Mode, Continuous display of Skin Temp in °F, Apgar (Up) & Down Timer and Alarm Messages.Alarms messages: Sensor Failure, Heating Power Failure, Heater Power Drive Failure, Over/Under Temp., and High/Low Temp.APGAR/Down TimerFR Grade Swiveling overhead heater box with CE marked ceramic heater, reflector and safety grillLED examination lamp for extra illumination during procedureFacility to attach multiple syringe/infusion pump, monitor shelf and SS multi hook IV stand |
| 12 | Multi para cardiac monitor |
| 13 | Specification Should have a minimum 12 inches or more color CRT/TFT display Should be suitable for the monitoring of wide variety of patients in Recovery Rooms, HDUs and step down ICUs etc Monitor should have port for connectivity to central nursing station & hospital local networking system interfacing facility Should have a neonatal to adult usage Should be able to display 6 or more waveforms along with related numerical parameters on a single screen. The monitor should give choice of selection of the parameters to be displayed as waveforms Should have single and multiple parameter modular system. Should have space saving design with minimum cable clutter Should display following parameters ECG/Respiration Should have heart rate range from 20-250 or more beats/mt Should have Auto, ECG or Plethysmograph as source for heart rate Should have adjustable heart rate tone Should have sweep speed of 12.5/25/50mm/sec Should have auto as well as manual gain selection Should have protection against electrosurgical interference and defibrillation Should have ST segment analysis Should have software standards for lethal arrhythmias, Vfib, Vtach, Asystole, advanced arrhythmia and alarms Should display audio, visual alarms Should have impedance as a method to know respiration Should display a range of 0-99 breath pm for respiration |
| 14 | NIBP Should be based upon Oscillometric method Should have manual/auto/stat modes Should display systolic, diastolic and mean arterial pressures Should have a scale in mmhg Should have a range of 1- 120mts for repeat cycle Should have a cuff pressure range of 0-300 mmHg for adults and 0-150 mmHg for neonates Should have a range of systolic BP from 40-250; diastolic BP from 20-200 and MAP from 25-220 mmHg Pulse Oximetry Should have saturation range of 0-100% for adult, child and neonate Should have a accuracy of +- 2% at 70-100% range and +- 3% Should have a HR range of 30-250bpm Plethysmographic waveform display LED pulse oximetry Should be capable of measuring oxygen saturation even in case of motion artifact (patient motion or movement), low perfusion & intensive ambient light Temperature Should have a range of 0-50 degree C Should have a resolution of 0.1 degree C Should have a accuracy of +- 0.2 degree C or less Should provide skin probe Should have future upgradability of following parameters: ECTO2 & 2-IBP. Trends Should have atleast 48 hrs trend storage of all parameters in graphic and chart form Should provide user configurable graphical and numerical trend pages for providing complete historical view to patient status Should have trend cursor in both graphical and numeric trends Should have waveform snapshot to store and recall waveform samples for historical review. Should have a facility to printout trend data on a printer |
| 15 | Alarms Should have alarms display and adjustment facility. There should be facility for Setting and suspending alarms Should have adjustable alarm limits ( high and low) Should have both audio and visual alarms. Power supply/ Battery Should be able to operate according to the Indian standards of electricity supply as well as on battery of 3hrs after fully charging Should display battery status indicator Should have a battery charge time between 2-3 hrs. ACCESSORIES Should provide following accessories with each monitor ECG cables 5 lead-----------1 pieces with at least 5 stick electrodes NIBP cuffs------------ 1 for adult SpO2 probes…………. 1 each for Adult Temperature Probe………..1 skin probes Rolling stand for mounting the monitor with attached baskets to keep its accessories should be quoted separately as optional Operator & service manuals in English |
| 16 | Mega o2 cylinder |
| 17 | Specification Other Known Name :- Big Oxygen Cylinder, D Type Cylinder, Jumbo Oxygen Cylinder, 7000ltr Cylinder and 7 Cum Cylinder Cylinder Water Capacity :- 46.7 Ltr Cylinder Oxygen Capacity :- 7000 ltrs (7 Cum) Weight of Cylinder :- 51kg(Approx) Height of Cylinder with valve :- 1445mm(Approx) Diameter of Cylinder (Outer) :- 232mm (Approx) Cylinder working duration/Last :- 48hrs @ 2 lpm Cylinder Gas Type :- Medical Oxygen Ideal use of Cylinder :- Hosptial, Ambulance, Clinic & Home Use Working pressure of Cylinder :- 150kg/cm2 at 15 degree Celsius Cylinder Test Pressure :- 250kg/cm2 at 15 degree Celsius Cylinder Wall thickness :- 5.5 mm Standard :- IS 7825 part II Cylinder Valve :- IS 3224 Certified by Dept of Explosives :- GOI (PESO) |
| 18 | Overhead Table |
| 19 | Specification Overbed table with PE table top Size: 920mm X 400mm × 740 / 1115 mm. Framework is made of Epoxy steel. Aluminium column lift system, Adjustable height with gas spring 4 Swivel castor all with brakes. |
| 20 | Defibrillator |
| 21 | Specification It should be CE Approved. Should have facility for: a) ECG Monitoring b) Defibrillation c) Recorder Should be of Rectilinear Biphasic Technology having energy selection up to 1-200 Joules Must be capable of monitoring ECG through ECG cables, Multi function electrodes and paddles Should have the facility for future up gradation of AED, Pacing, SPO2 & NIBP for both adults and pediatric patients Should have ECG waveform display on bright high resolution display Should be compact and portable with inbuilt rechargeable battery Should have an inbuilt recorder printing ECG trace and stored information Should not be more than 7 kgs with battery and paddles Should have charging time less than 6 seconds. Should have user selectable alarm settings Should work on mains as well as on rechargeable battery Should be supplied with Adult and inbuilt Pediatric external paddles Should have facility to increase/decrease Energy selection on paddles preferred. Should have battery capacity up to 50-70 shock delivery on single charge. |
| 22 | Syringe pump |
| 23 | Specification It is user friendly & easy to operate pump It can perform Many Functions like it has Anti-bolus & power lost memory Specifications and functions: Syringe Size : 5, 10, 20, 30, 50/60ml Infusion Modes: to satisfy different infusion requirements (must to have rate mode, time mode, body weight mode). Flow Rate: 0.1-1200ml/h Preset Volume (VTBI) : 0.1~ 9999.99 ml Accumulated Volume: 0.1~ 9999.99 ml Increment: 0.1 ml/h. Accuracy: ±2%. KVO: 0.1-3.0 ml/h. Bolus: Manual bolus and Programmable bolus. Occlusion Level: It has 3 levels of occlusion pressure. Screen: 3.5\" colored screen The machine have the facility of auto recognition of Syringe size |
| 24 | It should have following alarms: VTBI Done VTBI Near Done KVO Finish Syringe Empty Syringe Disengaged Syringe Near Empty Occlusion System Malfunction Battery Low Battery Empty AC Power Disconnection Syringe not Correctly inserted Battery backup of min 8 hrs. Machine has high protection level IP24. The machine is European CE/ISO certified. |
| 25 | Revolving stool |
| 26 | Specification Usage/Application:- Clinic And Hospital Brand:- Accure Minimum Order Quantity:- 10Adjustable:- No Type:- Hospital Equipment Design Type:- Customize Color:- Satinless Steel No Of Leg:- 4 |
| 27 | Crash cart |
| 28 | Specification Overall approx Size: 940E X 490W X 1535H mm. Six colored removable bins and two polystyrene lockable storage units with three drawers have each. The top of drawers have containers of different size. Four swivel castors,125mm dia, two with brake. Complete with comer buffers, powder coated oxygen cylinder holder, S.S., I. V. rod, S.S. shelves, cardiac massage board and electric lamp S.S. Tubular frame. |
| 29 | Bed side locker |
| 30 | Specification Size : 500mmx 530mmx 753mmH(without Castors).Structure Made of Melamine Particle board and Aluminum columnUpper Drawer Size : 415 x 360 x 100mmBottom Drawer Size : 415 x 360 x 200mmShelf Height is 135mmWith Four plastic castors |
| 31 | PICU Bed |
| 32 | Specification 50mm X 25mm approx. CRCA rectangular tube frame. Two section perforated CRCA sheet top. Two separate screws for backrest and trendelenburg & reverse trendelenburg position by individual SS folding handles. Stainless Steel Head and foot end bows with laminated panels Full length drop type rails Four non-rusting 125 mm dia. Polyurethane Wheels 2 with brakes and 2 without brakes. Pretreated and epoxy powder coated. Dimensions Approximate Length : 1600mm Width : 840mm Height : 600mm Accessories Telescopic I.V. Rod SS Mattress |
| 33 | Oxygen concentrator |
| 34 | Specification Description of Function: Oxygen concentrator to provide oxygen from ambient air Product Quality standards: It is European CE approved model Manufacturer should be ISO certified for quality standards Technical Specification: Oxygen concentrator to provide oxygen from ambient air. Oxygen concentration measured at the flow meter by oxygen sensing device (OSD) Character of service: Continuous Operation. Sound level <60 dB The oxygen concentration does not reduce at 1828 meters above the sea level, and efficiency is less than 90% from 1828 meters to 4000 meters. It should have Digital Purity meter on display for Oxygen Purity measurement. Integrated Timer should be there It has a big touch screen display for easy operation. Should have audio and visual alarm for Low oxygen concentration, low pressure and power failure The unit should be movable. Provided with four wheels with at least two breaks |
| 35 | Oxygen purity: 93-95% Oxygen concentration 95% + 3 at 5LPM flow Oxygen output, approx: 0 - 5 LPM The machine should have built-in nebulizer port. Pressure, approx: 20.75 kpa. Single outlet for oxygen Delivery controlled from a pressure compensated flow meter to ensure accurate flow display Masks : Necessary mask suitable for Adult should be provided –1 sets Supplied with 2 x spare set of internal and external filters (bacterial) 1x humidifier bottles Environmental factors: The unit shall be capable of being stored continuously in ambient temperature of 0-50deg C and relative humidity of 15-90% The unit shall be capable of operating continuously in ambient temperature of 10 -40 deg C and relative humidity of 15-90% Power Supply Power input to be 220-240VAC, 50Hz fitted with Indian plug |
| 36 | Sub-total of Medical Equipments (without GST) |
| 37 | MEDICAL STAFF/ HR - PEDIATRIC INTENSIVE CARE UNIT (PICU) for One Year-(without GST) |
| 38 | MBSS Doctor |
| 39 | Pediatrician |
| 40 | Staff Nurse |
| 41 | Technician |
| 42 | Operator |
| 43 | Lab Technician |
| 44 | Fourth Class |
| 45 | Sub-total of Medical Staff/HR(without GST) |
| 46 | MINOR CIVIL & ELECTRICAL WORKS - PEDIATRIC INTENSIVE CARE UNIT (PICU) |
| 47 | Chipping of existing old PCC flooring to make the surface rough |
| 48 | Dismantling wooden door and Z- section windows |
| 49 | Vitrified tiles flooring |
| 50 | Ceramic titles on wall/skirting upto 5’ height |
| 51 | P/F Aluminum doors and windows |
| 52 | P/F False ceiling with PVC Laminated gypsum ceiling tiles(595*595*8mm) |
| 53 | P/F Aluminum grill in windows |
| 54 | S/F SS fly proof wire gauge to windows |
| 55 | S/F 36 watt LED Ceiling Square light fitting (60cm*60cm) |
| 56 | Electric work (points, wall fans, sub main extra fittings and switches as required) |
| 57 | Painting with plastic emulsion paint |
| 58 | S I T C of window AC (t-2)with stabilizer MCB and necessary power wiring inclusive of all taxes |
| 59 | Sub-total of Minor Works(without GST) |
| 60 | Sub-total of Minor Works(without GST) for 17 PICU units |
| 61 | Supply and installation of Medical Gas Pipeline for 17 PICU Units (excluding GST) |