Supply, Installation And Commissioning Of Portable Biogas Analyzer Supply, Installation And Commissioning Of Portable Biogas Analyzer , Supply, Installation And Commissioning Of Portable Biogas Analyzer , Portable Biogas Analyzer • Type: Portable Typ, chennai-Tamil Nadu

Council Of Scientific And Industrial Research-CSIR has published Supply, Installation And Commissioning Of Portable Biogas Analyzer Supply, Installation And Commissioning Of Portable Biogas Analyzer , Supply, Installation And Commissioning Of Portable Biogas Analyzer , Portable Biogas Analyzer • Type: Portable Typ. Submission Date for this Tender is 12-05-2022. Gas analyzer Tenders in chennai Tamil Nadu. Bidders can get complete Tender details and download the document.




Tender Notice

31943627
Supply, Installation And Commissioning Of Portable Biogas Analyzer Supply, Installation And Commissioning Of Portable Biogas Analyzer , Supply, Installation And Commissioning Of Portable Biogas Analyzer , Portable Biogas Analyzer • Type: Portable Typ
Open Tender
Indian
Tamil Nadu
chennai
12-05-2022

Tender Details

Supply, Installation And Commissioning Of Portable Biogas Analyzer Supply, Installation And Commissioning Of Portable Biogas Analyzer , Supply, Installation And Commissioning Of Portable Biogas Analyzer , Portable Biogas Analyzer • Type: Portable Type Gas Analyzer Gases To Be Measured : Ch4 , Co2 , Nh3 , O2 , H2s Ch4 Measurement Range : 0-100%; Typical Accuracy: 0-70% = ±0.5% ( Vol ) ; 70-100% = ±1.5% Co2 Measurement Range : 0-100%; Typical Accuracy: 0-60% = ±0.5% ( Vol ) ; 70-100% = ±1.5% O2 Measurement Range : 0-25%; Typical Accuracy: ±1.0% ( Vol ) Nh3 Measurement Range : 0-1000 Ppm; Typical Accuracy: ±10.0% Fs H2s Measurement Range : 0-10, 000 Ppm; Typical Accuracy: ±5.0% Fs Power Supply: - Battery Type: Rechargeable Nickel Metal Hydride Battery Pack - Battery Life: Atleast 7-8 Hours After Fully Charging - Battery Charger: Separate Charger Required With Charging Time Of Less Than 4-5 Hours From Complete Discharge Operating Temperature Range: -10°C To +50°C Atmospheric Pressure Range: 700-1200 Mbar Relative Humidity: 0-95% Non-Condensing Case Seal: Ip65 Case Material: Abs / Polypropylene With Rubber Over-Moulding Keys: Alpha-Numeric Keypad With Tactile Membrane Display: Ultra-Clear High Resolution 4.3 Full Colour Tft Connections: Colour Coded Gas Inlet, Pressure And Outlet Ports. Waterproof Usb Port, Anemometer And Charger / Temperature Probe Connections. Gas Sample Filters: External User Changeable 2.0?M Ptfe Water Traps Sensor: Ch4 & Co2 - By Dual Wavelength Infrared Sensor With Reference Channel ; O2 / H2s / Nh3 : By Internal Electrochemical Cell , Packing & Forwarding, Insurance And Transportation Charges , Installation And Training Charges

Key Value

Document Fees
Refer document
EMD
INR 13750.0 /-
Tender Value
Refer document

BOQ Items

Name of Work: Supply, installation and commissioning of Portable biogas analyzer
Sl. No. Item Description
1Supply, installation and commissioning of Portable biogas analyzer
2Portable Biogas analyzer • Type: Portable type gas analyzer Gases to be measured : CH4 , CO2 , NH3 , O2 , H2S CH4 measurement range : 0-100%; Typical accuracy: 0-70% = ±0.5% (vol); 70-100% = ±1.5% CO2 measurement range : 0-100%; Typical accuracy: 0-60% = ±0.5% (vol); 70-100% = ±1.5% O2 measurement range : 0-25%; Typical accuracy: ±1.0% (vol) NH3 measurement range : 0-1000 ppm; Typical accuracy: ±10.0% FS H2S measurement range : 0-10,000 ppm; Typical accuracy: ±5.0% FS Power supply: - Battery type: Rechargeable nickel metal hydride battery pack - Battery life: Atleast 7-8 hours after fully charging - Battery charger: Separate charger required with charging time of less than 4-5 hours from complete discharge Operating temperature range: -10°C to +50°C Atmospheric pressure range: 700-1200 mbar Relative humidity: 0-95% Non-condensing Case seal: IP65 Case material: ABS / polypropylene with rubber over-moulding Keys: Alpha-numeric keypad with tactile membrane Display: Ultra-clear high resolution 4.3" full colour TFT Connections: Colour coded gas inlet, pressure and outlet ports. Waterproof USB port, anemometer and charger / temperature probe connections. Gas sample filters: External user changeable 2.0µm ptfe water traps Sensor: CH4 & CO2 - By dual wavelength infrared sensor with reference channel ; O2 / H2S/ NH3 : By internal electrochemical cell
3Packing & Forwarding, Insurance and Transportation charges
4Installation and training charges