| 1 | Oxygen System | Mani fold- Oxygen Manifold 4 + 4 cylinder banks complete with NRV, tail Pipes, Middle frame, Bottom Frame with CE marked, |
| 2 | Oxygen System 2 | Semi Automatic Control Panel for Oxygen- 4 bar line pressure, 500lpm as per as per Medical Device Directives |
| 3 | Oxygen System 3 | Emergency Oxygen Manifold- 01 + 01 Cylinder Emergency Manifold with complete regulator, NRV, tail pipes, Middle frame & Bottom frame, |
| 4 | Oxygen System 4 | Oxygen Flowmeter with humidifier Bottle, CE Marked, |
| 5 | Oxygen System 5 | Oxygen Gas outlet front loaded DIN /BS STD. |
| 6 | Oxygen System 6 | DIN/British standard quick connect adapter(safety probe) |
| 7 | Oxygen System 7 | Oxygen H.P. Rubber tube white color, |
| 8 | AREA ALARM | Area line pressure alarm-1 Gases |
| 9 | Area Valve Box, | 1 gases |
| 10 | Isolation Valve, | 42mm |
| 11 | Isolation Valve 2 | 28mm |
| 12 | Isolation Valve 3 | 22mm |
| 13 | Isolation Valve 4 | 15mm |
| 14 | Isolation Valve 5 | 12mm |
| 15 | Medical Grade Copper Pipeline (In Mtrs.) as per EN 13348, | 12 mm (OD) x 0.7 mm Thickness |
| 16 | Medical Grade Copper Pipeline (In Mtrs.) as per EN 13348,(2) | 15 mm (OD) X 0.9 mm Thickness |
| 17 | Medical Grade Copper Pipeline (In Mtrs.) as per EN 13348,(3) | 22 mm (OD) X 0.9 mm Thickness |
| 18 | Medical Grade Copper Pipeline (In Mtrs.) as per EN 13348,(4) | 28 mm (OD) X 0.9 mm Thickness |
| 19 | Medical Grade Copper Pipeline (In Mtrs.) as per EN 13348,(5) | 42 mm (OD) X 1.2 mm Thickness |
| 20 | Manifold room | Construction of manifold room of 100-120 sq.ft. with brick work, plaster, distempering, Painting, GI sheet, IPC flooring, Door, window, ventilator, grill, light, fan, exhaust etc. all complete at suitable location in premises. |
| 21 | Structural steel work | Structural steel work bolted in built up sections, trusess andframed work, including cutting,hoisting,fixing in position and applying a priming coat of approved ateef primer all complete. |
| 22 | Electrical work | Electrical work for Manifold, Alarm Panel, Light etc , Wiring for light point/ fan point/ exhaust fan point/ callbell point with 1.5 sq.mm FR PVC insulated copperconductor single core cable in surface / recessedmedium class PVC conduit with modular switch ,modular plate, suitable size G.I. box etc as required.Group C |
| 23 | Electrical work 2 | Wiring for circuit/ submain wiring alongwith earth wirewith the following sizes of FR PVC insulated copperconductor, single core cable in surface/ recessedmedium class PVC conduit as required. 2 X 2.5 sq. mm + 1 X 2.5 sq. mm earth wire |
| 24 | Electrical work 3 | Wiring for circuit/ submain wiring alongwith earth wirewith the following sizes of FR PVC insulated copperconductor, single core cable in surface/ recessedmedium class PVC conduit as required.2 X 4 sq. mm + 1 X 4 sq. mm earth wire |
| 25 | Electrical work 4 | Wiring for circuit/ submain wiring alongwith earth wirewith the following sizes of FR PVC insulated copperconductor, single core cable in surface/ recessedmedium class PVC conduit as required.2 X 6 sq. mm + 1 X 6 sq. mm earth wire |