Supply Of Dental Flash Autoclave Medical Instruments & Equipments To Yalburga Tq Under Dho Koppal
Tender Overview
Project Description
Supply Of Dental Flash Autoclave Medical Instruments & Equipments To Yalburga Tq Under Dho Koppal 1. Dental Flash Autoclave 1.Should Be A Table Top Autoclave For Dental And Ophthalmic Applications 2. Two Automatic Programmes Approx. At 2.2 Bar At 134 Degrees C And 1.1 Bar At 121 Degree C. The Equipment Should Have Automatic Pressure Control Switch Or Automatic Water Control Device To Ensure That The Equipment Does Not Run Dry. 3. Should Have Flash Cycle For Rapid Sterilization And Should Have An Option For Liquid Cycle. 4. Should Have Air Pump For Closed Door Drying. 5. Should Have Rapid Warm Up Facility. Built In Reservoir To Store Water Required To Produce Steam And Used Water Separately For Easy Decantation. 6. The System Should Be Equipped With Required Safety Features. The Door Should Have Double Locking Safety Feature And Should Open Only With Atmospheric Pressure In The Chamber. 7. Should Have Automatic Cut-Off To Prevent Overheating And Cut-Off For Insufficient Water The Machine Should Not Start Without Sufficient Water. 8. Should Have A Minimum Chamber Capacity Of 19 Litres Or Above. 9. Should Have Pressure Display And Temperature Display. 10.Unit Should Function With 200-240Vac 50 Or 60 Hz Input Power Supply. 11.The System Should Comply With National Quality Certification Or International Standards For Sterilization Safety. 12.Following Accessories Should Be Supplied Along With The Equipment. A. 1 Set Of 3 Removable Shelves Stainless Steel. B. 1 Instrument Basket Stainless Steel. C. 1 Set Of 2 Drum For Sterilization Stainless Steel. D. 1 Roll Of Sterilization Indicator. E. 1 Box Paper Sheet 100 Nos Crepe For Sterilization Packs. F. 2 Spare Silicone Gaskets. G. 1 Sets Of Spare Fuses. H. Equipment Should Be Provided With A Line Cord Power Cord Of Acceptable Durability Quality Length And Current Carrying Capacity And Should Be Compatible With Indian Standard Power Socket. 14.Controls Should Be Visible And Clearly Defined. 15.Labels And Markings Should Be Clear And Visible. 16.Should Have Safety Certificate From A Competent Authority Ce Or Fda Us Or Stqc Cb Certificate Or Stqc S Certificate Or Valid Detailed Electrical And Functional Safety Test Report From Ertl. Copy Of The Certificate Or Test Report Shall Be Produced Along With The Technical Bid. 17.Should Have Air Filters. 18.Gaskets Should Be Replaced At Free Of Cost Whenever Required In The Comprehensive Warranty And Cmc Period.,
BOQ
| Sl No | Item Code | Item Name | Item Description | Unit | Quantity |
|---|---|---|---|---|---|
| 1 | DHO 09 | DENTAL FLASH AUTOCLAVE 1.Should be a table top autoclave for Dental and ophthalmic applications 2. Two automatic programmes approx. at 2.2 bar at 134 degrees C and 1.1 bar at 121 degree C. The equipment should have automatic pressure control switch OR automatic water control device to ensure that the equipment does not run dry. 3. Should have flash cycle for rapid sterilization and should have an option for liquid cycle. 4. Should have Air Pump for closed door drying. 5. Should have rapid warm up facility. Built in reservoir to store water required to produce steam and used water separately for easy decantation. 6. The system should be equipped with required safety features. The door should have double locking safety feature and should open only with atmospheric pressure in the chamber. 7. Should have automatic cut-off to prevent overheating and cut-off for insufficient water the machine should not start without sufficient water. 8. Should have a minimum chamber capacity of 19 litres or above. 9. Should have pressure display and temperature display. 10.Unit should function with 200-240Vac 50 OR 60 Hz input power supply. 11.The system should comply with National quality certification or International standards for sterilization safety. 12.Following accessories should be supplied along with the equipment. A. 1 set of 3 removable shelves stainless steel. B. 1 instrument basket stainless steel. C. 1 set of 2 Drum for sterilization stainless steel. D. 1 Roll of sterilization indicator. E. 1 box paper sheet 100 nos crepe for sterilization packs. F. 2 spare silicone gaskets. G. 1 sets of spare fuses. H. Equipment should be provided with a line cord power cord of acceptable durability quality length and current carrying capacity and should be compatible with Indian standard power socket. 14.Controls should be visible and clearly defined. 15.Labels and markings should be clear and visible. 16.Should have safety certificate from a competent authority CE OR FDA US OR STQC CB certificate OR STQC S certificate or valid detailed electrical and functional safety test report from ERTL. Copy of the certificate OR test report shall be produced along with the technical bid. 17.Should have air filters. 18.Gaskets should be replaced at free of cost whenever required in the comprehensive Warranty and CMC period. | Each | 1 |
AI Tender Summary
Tender Timeline
Tender Published
Tender notice published.
CompletedBid Submission Deadline
Online submission via eProcurement portal.
CompletedBid Opening Date
Technical bids will be opened and evaluated.
CompletedTender Documents
Download All (ZIP) ↓We takes all possible care for accurate & authentic tender information. However users are requested to refer Original source of Tender Notice / Tender Document published by Tender Issuing Agency before taking any call regarding this tender