Uttar Pradesh Housing And Development Board-UPAVP has published Supply , Installation , Testing , Commissioning Of 02 Nos . Modular Operation Theatre , 1 No . Semi Modular Ot And 02 No . Modular Icu/Sncu At 200 Bedded Maternity And Child Hospital Gomti Nagar Lucknow. Submission Date for this Tender is 21-01-2020. Modular Operation Theater Tenders in 200 Bedded Maternity and Child Hospital Gomti Na Uttar Pradesh. Bidders can get complete Tender details and download the document.




Tender Notice

23213073
Supply , Installation , Testing , Commissioning Of 02 Nos . Modular Operation Theatre , 1 No . Semi Modular Ot And 02 No . Modular Icu/Sncu At 200 Bedded Maternity And Child Hospital Gomti Nagar Lucknow
Open Tender
Indian
Uttar Pradesh
200 Bedded Maternity and Child Hospital Gomti Na
21-01-2020

Tender Details

supply , installation , testing , commissioning Of 02 Nos . Modular Operation Theatre , 1 No . Semi Modular Ot And 02 No . Modular Icu/Sncu At 200 Bedded Maternity And Child Hospital Gomti Nagar Lucknow

Key Value

Document Fees
INR 8850 /-
EMD
INR 1600000.0 /-
Tender Value
INR 8 Crore /-

BOQ Items

Name of Work: S/I/T/C of 02 Nos. Modular Operation Theatre, 1 No. Semi Modular OT and 02 No. Modular ICU/SNCU at 200 Bedded Maternity and Child Hospital Gomti Nagar Lucknow
Sl. No. Item Description Item Code / Make Quantity Units
1 1 Pre-fabricated Walls & Ceiling Panel\nPre - fabricated modular construction shall be designed and constructed for exact size The prefabricated Operating Room will be free standing structure from Compasite free standing insulated steel wall panels. The wall will be constructed using 1.0 mm SS 304 sheet panel in the front panels & 0.8mm. SS 304 at the back panel The core between two panels will consist of rigid polyurethane foam. which has been injected under high pressure with a minimum density of 40 kg/m3. The individual wall Panels shall use the tongue and groove technology for joining two panels, no welding will be allowed . The gaps between panels be suitably filed with metal filter/epoxy and sanded flush. BI01010001010000000000000515BI0100001113 165 Sqmr.
2 1.1 The external walls of the room shall be constructed with solid bricks and mortar by the hospital as part of its scope of work. The inner surface walls shall be fixed to the brick wall with essential supports. There Natal be minimum possible cavity/gap in between the masonry and steel walls. The total distance between inside and outside surface of OR shall be variable to Suit architects's layout, out shall be sufficient for flush mounting of the equipments. The cavity between inner and outer walls shell be left with minimum obstructions for possible addition of equipment at a later date and to enable services, pipes, conduits etc; to be run within the cavity. All wall mounted equipment shall be flush mounted and sealed into theatre. The wail panel design and construction shall allow tor installation and support of all equipment and provision of openings required for the installations, without affecting rigidity and strength. Access boxes shall be fitted to the rear of all the wail -mounted equipments to enable maintenance to be carried out from outside the operating room. All the sharp edges and corners shall be smoothened to avoid tactrial contamination. BI01010001010000000000000515BI0100001114
3 1.2 Anti Bacterial/ Anti Fungal Paint\nThe internal surfaces of the SS-304 OT Walls should be sprayed with water based, non-reflective liquid plastic, to a color approved be the architect a minimum dry film thickness of 300 micronas, These plastic coatings should overlap the floor covering Ceiling system and-door frames by 25mm to provide a continuous sealed surface. Shade Coating applied should be wafer resistant. does not support bacteriological or fungal growth and is resistant. to most chemicals commonly used in hospital departments. The sterile coating should remains unaffected by radiation and other ionizing radiation at levels in excess of 1000 mred and is classified to class I when tested in accordance with the requirements specified under BS 478. part 7 1971. Surface spread of flame Test for Materials. The coating system should be easily maintained and can withsland-repeated cleaning with alkaline detergents. antiseptics and furnigation agents without any degradation to the surface finish or performance. BI01010001010000000000000515BI0100001115
4 2 Laminar Air Flow Ceiling\nEach modular OT shall have Plan Air Ceiling. The Plan Air Ceiling shall be constructed out of 2mm each extruded aluminum profiles, overall size 2400 x 2400min or larger as per actual layout, having six no.s HEPA fitters with MPPS efficiency of 99,995% and spot efficiency of 99.99% @ 0.3 micron. The Plenum should provide the required air volume to achieve the temperature . pressure (15-30 Pa) and humidity specified, and terminal velocity 90 to 120 FPM at the diffuser face. Air and light diffuser made of two layer of mono filament precision woven polyester for plan air ceiling to give a laminar flow of filtered air, it shall also provide a shadow less lighting system with control of the intensity of luminance by using high efficiency electronic fluorescent tubes AM HEPA filters shall be factory tested and certified in accordance with international standards. All ceiling snail include integral lighting and composite air/light diffuser.:Air shall be diffused into the theater uniforrnaily over ,,the total area. The leminar flow ceiling should also have illumination across its total area, BI01010001010000000000000515BI0100001117 2 No.
5 3 Automatic Hermetically Sealing Door BI01010001010000000000000515BI0100001119
6 3.1 2100 x 1500mm BI01010001010000000000000515BI0100001120 2 No.
7 3.2 2100 x 1000mm BI01010001010000000000000515BI0100001121 2 No.
8 3.3 Automatic door shall be provided at the main entry of the OT Doorframes shall be made of aluminum extrusion profile to carry any door weight. The core door size is of 50mm size built of high density PUF (density 40kg/m3) of 52mm and 4mm consist of Solid Stainless Steel (SS-304) on both side of the door. Door blade is made with 3- tier specially designed 3 slide heavy duty replaceable EPDM gasket against wall frame. Bottom seakng with 2-tier heavy duty EPDM gasket to flush with finished floor. Stainless Steel lever handle on both side of the door blade for smooth and easy door opening and closing. The thickness of, aluminum extrusion for door frame & door panel surfaces shall not less than I.5mm There shall not be any visible screws or similar other uneven surfaces on the door panel. A lock shall be provided. Door shall have vision panels (viewing window) at least 300mm x 300mm, Automation of the door has Digital Conroller , Motor and suitable switches to open the door thru hands, foots etc. The door size shall be as per BOO. BI01010001010000000000000515BI0100001122
9 4 Surgeon Control Panel\nThe control panel will be user triendly and operation and maintenance will be easy. The control panel shall be "Tiled Membrane" type: can be configuired to incorporate all the services that operating theatre staff required The fascia shall be made with superior quality UV resistant, sterlisable membrane The membrane backing is made of 25mm thick aluminum that is rust resistance. The connection between the surgeon control and the distribution board will be by a multiplexed system. The panel will contain & services files for.\n(a) Time day clock\n(b) Time elapsed clock\n(c) Temperature / humidity indicator\n(d) General Lighting System\n(e) Medical gas alarm panel\n(f) Hand free phone.\n BI01010001010000000000000515BI0100001115 2 No.
10 5 LED TWIN Plate X-ray Viewer\nThe operation tneatre shall be equipped with a X-ray viewing LED Screen, designed to provide a high level of control luminance without flicker. It shall be equipped with loaded dips to secure the X-ray negative when in use. BI01010001010000000000000515BI0100001116 2 No.
11 6 Pressure Relief Dampers\nPressure relief dampers to be provided in theater to prevent cross contamination of air from clean and dirty areas. Suitably sized air pressure relief damper will be strategically placed, enable differential room pressure to be maintained and ensure that wnen doors are opened between clean and dirty area. Counter weight balancing system will be provided in the PRD to maintain positive pressure inside the operation theatre. Air pressure stabilizers will have unique capability of controlling differential pressure to close tolerance. The PRD will remain closed at pressure below the set pressure and will open fully at pressure only fractionally above the threshold pressure. The body will be epoxy pawder coated as per standard BS colors and meets international quality and safety requirements. First class electrolyzed steel plate will be used for body and with high -grade SS 304 Stainless steel for blades. BI01010001010000000000000515BI0100001117 2 No.
12 7 Anti-Static Flooring\nFlooring (Static conductive tiles): The floor finish in the operating room will be imported 2 mm statics conductive PVC tiles, laid on a semi conductive adhesive base The floor finish will terminate at the room perimeter passing over a corcealed cove former and continuing up the wall for 100mm. Flooring will fulfill EN it will have chemical resistance as per EN 423. it will have electrical resistance conforming to EN 1061 AEC 613404-1. All joints will be welded and the plastic wall finish will overlap the floor coving by 25mm , to provide a continuous seated surface. The self leveling compound will be of the type that does not promote bacteriological growth. Copper grounding strips (0.05mm thick, 50mm wide) will be laid flat on the floor in the conductive adhesive and connect to copper wire of grounding BI01010001010000000000000515BI0100001118 130 Sqmr.
13 8 Storage Cabinets\nProviding , Fixing, Supply and Installation of storage unit will tie made with the wall cladding material i.e SS-304,The unit will be divided into 2 equal parts and each part will have individual glass doors with locking system. Each part will have glass racks to keep surgica medicines etc BI01010001010000000000000515BI0100001119 2 No.
14 9 2-Bay Scrub Sink\nCompact Surgical Scrub sink will be designed for use in Operation theatre complex providing surgeons with a convenient sink for pre up scrub up. Each fixture will be fabricated from heavy gauge type 304 stainless steel & will be welded construction polished to a satin finish. The scrub will be provided with a front access panels, which will be easlly removed for access to the water control valve, waste connections stoppers & strainers. Hands free operation will include infrared sensor with built in range of adjustment. Thermostatic Mixing Valve Control will be located behind the access panel & maintain constant water temperature. User defined settings of 1 to 3 min are available. This timing will be adjustable to meet individual application requirement, provided with infrared sensor thermostatic controlled laps with fail-safe Temperature controls. BI01010001010000000000000515BI0100001120 2 No.
15 10 Writing List Board\nThe writing unit shall comprise of flush mounted 1.5mm thick, white laminate board, bonded to a 40mm high density fireboard sheet tor additional rigidity. The unil shall be opened to create a wall mounted writing surface within operating room. The while board shall be constructed from 1.6mm thick, white laminate board. BI01010001010000000000000515BI0100001121 2 No.
16 11 Peripheral lights (LED)\nEight sets of double peripheral lights shall be with stepped finish aluminum with reflectors. Peripheral lights shall be clean room compatible. Rating of peripheral lighting small be iP 54, Control equipment for the peripheral lighting shall be provied in the theatre control panel to allow independent control at the lighting level by the surgical team Recessed mounted. BI01010001010000000000000515BI0100001114 16 No.
17 12 Hatch Box\nThe Hatch Box should be provided in each operation theatre to remove waste materials from the operation thetare to dirty Linen area just adjacent to operation Theatre. Each Hatch Should be equipped with two doors and the door should be opened one at a time, The Hatch should be designed in such a way that only one door should be opened at one time, The UV light should be so installed that it is kept on while both the doors are closed , this UV light has to be automatically turned off in case of opening of either of the doors There shall be indicators on both side of OT so that door open/close status can be monitored from both ends, BI01010001010000000000000515BI0100001115 2 No.
18 13 Surgeon's Double Arm Pendant BI01010001010000000000000515BI0100001116 2 No.
19 13.1 Pendant Should be Double arm (1000mm + 800mm) with load carrying capacity of 80 Kg. The arm should rotated upto 330°-340° with adjustable stopper. Dampened stoppers and very low hand forces for horizontal movements should contribute to an ergonomically optimized workplace. The large interior cross section for supply lines offers should completely new applications with 120mm diameter. The stoppers Should be infinitely variable from 0-330°- 340' BI01010001010000000000000515BI0100001117
20 13.2 The service head should be of modular design in achieve maximum location efficiency with minimum required space. Service head should designed to host. Base , Gas Module (Up to 8 Gas outlets), Electrical Module (10 Electrical outlets) and two shelves and one Infusion pump rack.,\na. The total length of the manager should be 800mm Surgeon pendant\nb. should have 2 arms with shelves as per following details'\nc. Horizontal arm system- 2\nd. Weight carrying capacity- 80 Kg\ne. 5/15 Amp. Electrical sockets without switches- 10 Nos\nf. Shelves with slide ralls- 2 Nos.\ng. Provision to fix Gas Outlets (i.e.)Oxygen-2, Vacuum-2 Air 4 Bar-1,\nh. Air 7 Bar-1, CO2.1 AGSS-1\ni. Ceiling mounting system up to 1000mm-1\nj. Ceiling cover for false ceiling-1\n\n BI01010001010000000000000515BI0100001118
21 14 Anaesthetist's Single Arm Pendant BI01010001010000000000000515BI0100001119 2 No.
22 14.1 Pendant should be Single arm (900mm) with load carrying capacity of 80 Kg. Dampened stoppers and very low hand forces for horizontal movements should contribute to an ergonomically optimized workplace. The interior cross section avaliable for supply lines shall be minimum 120mm diameter, The stoppers should be infinitely variable from 0-330'--340°. BI01010001010000000000000515BI0100001120
23 14.2 The service head should be provided with the modular design to achieve maximum supply with minimum required space Service head should be designed to host . Gas Module. Electric Module and\nshelves, Up to 8 Gas outlets & 10 Electrical Outlets. Racks & shelves should be provided to mount the equipments like monitor etc The total length of the manager should be 800mm. Anaesthesia pendent\nshould be have 1 arms with shelves as per following details.\na. Horizontal arm system-1\nb. Weight carrying capacity. 50 Kg\nc. 5/15 Amp. Electrical sockets without switches 10 Nos,\nd. Shelves with slide rails- 2 Nos.\ne. Provision to fix Gas Outlets (8 Nos.) Oxygen-2, Vaccum-2, Air 4 Bar-\nf. 1, N20- 1 , AGES-1. Spare-1\ng. Ceiling mounting system up to 1000mm-1\nh. Ceiling cover for false ceiling- 1\n BI01010001010000000000000515BI0100001121
24 15 Electrical Work inside the OT \nAir high voltage equipment is installed in a separate enclosure The remote cabinet houses the operating lamp transformers main failure relays electrical for all circuit within the operating theatre. All internal wiring is terminated in connectors with screw and clamp & spring connections of the clip on type mounted, on a DIN rail labeled with indelible proprietary labels. Individual miniature circuit breakers shall be provided to protect all internal circuits with 5/16 amp switch Sockets for the Modular Operation Theatre MCB for Circuits for critical equipments and outlets shall be clearly marked for UPS connection-OPERATING LIGHT. MONITOR, OT PANEL. OT GENERAL LIGHT and other critical equipments (4 out of 10 on the pendant) BI01010001010000000000000515BI0100001122 2 Job
25 16 Medical Gas Pipeline work Inside the OT BI01010001010000000000000515BI0100001115 2 Job
26 16.1 The Medical Gas Pipeline inside the OT shall provide Valve Box for 5 gases (02. CO2, N20, Vacuum, Surgical Air-7 Bar) outside each OT and shall bnng the Medical Gas Pipeline inside the OT till ceiling pendants and provide NIST connections for each gas pipeline as per EN 737/HTM standards or equivalent. Pendants shall connect flexible hose piping with Medical Gas Piping by means of NIST connections. The gas piping used for transportation should be as EN 13348/EN 1057 standards with copper fittings as per EN 1254 standards.The outlets fitted for all the 5 gases in the pendants shall in conformity of HTM standards. Other Accessories of Medical Gas Pipeline like Ward Vacuum Unit having 600ml autoclave able at 121 degree centigrade collection jar and Theatre Suction Trolley having 2000ml (2 Nos) autoclave able at 121 degree centigrade to be provided in each OT BI01010001010000000000000515BI0100001115
27 17 OT Light with provision of camera BI01010001010000000000000515BI0100001116 2 No.
28 17.1 Description Three Dome LED Surgical Lighting System with one dedicated. Spring -Arm Suspension flat pane and Progressive Scan HD Plat Panel with an Integrated In-Light Camera System (OR Only provision for installation of camera, flat panel and suspension arm in future, as may be specified in the BOQ) The OT light should be European CE/USFDA OT Light.\nOperating Room Surgical Lighting System will provide an ideal combination of brightness, maneuverability, and shadow resolution without sacrificing color accuracy through a consistent LED technology with a faceted reflector design technology . such Lighting system will have the following technical specifications'\nColor Temperature range- 4000-5000K adjustable. minimum range\n BI01010001010000000000000515BI0100001117
29 17.2 Field Size Diameter (Variable)• minimum 140rnm at minimum\nmaximum 300mm at maximum\nDepth of Field- minimum 650mm for all field diameters.\nIllumination Level at no dimming- minimum 160000 Lux\nColour rendition index- General Ra 94 or more, Enhanced red (R9)\n97 or more\nControl- Wall Control, Touch panel Rotation- Around 'X-360 degrees, around 'Y-290 degrees Location Positioning (C/I of light)- Within a minimum radius of 15 Mis from mount (both lights)\nMinimum floor clearance- 1.5 Mt for ceiling height of 2.75 Mt. Vertical Adjustment Range- +450mm to 600mm w r t main arm position (minimum)\nSecondary Arm Tilt Angie (Z) range +/- 30 degree minimum\nStarilizable Handle- Yes\nMounting Type. Ceiling\nSupply Voltage range. 140-270 VAC 50/60 Hz\nBulb Type- LEO\nProtection- IP54\nMaximum Power consumption - Dimming Range (Minimum)•10%.100%\nOperating/ Storage Humidity. 10-95%\nLife of Light Source- 040.000 Hrs\n BI01010001010000000000000515BI0100001118
30 18 Camera System WITH MONITOR (COMPATIBLE FOR O.T. LIGHT)\nDescription' Integrated in-Light Camera System will be integrated at the centre of one of the domes of this lighting system in order to capture images a video sequences of the open cases The camera and mount should be such that the camera can be removed and fixed to different lights as per need, without any special tools or training Sucn a camera will have the foliwing specifications)\nSignal to Noise Ratio (S/ N Ratio)- <50dB\nMinimum illumination- <3lx\nOptical Zoom- 25-30x\nDigital Zoom- 21-15x\nPower Supply. max 12W\nRelative Humidity- upto 90%\nVideo Output- DVl. S-Video & Composite Video\nWhite Balance & Gain- Automatic I Manual\nSuch Surgical Light system will be complaint with relevant European/US standards Such Light and tmegrated Camera will have control through. Touch Panel of the control equipment placed inside the operating room and at the documentation station / nurse works station.\n BI01010001010000000000000515BI0100001119 2 No.
31 18.1 Flat Panel Monitor\nWill be 26" High Defination Progressive Scan Fiat-panel Monitors with ceiling mounted spring arm suspension to support high- defination/HDTV progressive Scan images and will be able to support and display DVI/HDTV, RGENV, S-Video, Composite video signals.\nThe flat Panel suspension will be ready with the cables for integration of High Defination Digital (DVI/HDTV) , RGBHV (High Resolution) SVHS (S-Vicleo), Composite video signals to travel from the various sources of video like endoscopic camera room camera , in light camera . high defination flat panel monitors, while assuring native resolution /signal.\nSuch Monitor will at least meet the following technical criteria:\nResolution- 1600 dots a 1200 dots, Progressive Scan Display Colors-16 million Colors Inputs. DVI, RGBHV, S-Video, Composite Video Synchronization- 2.5-5.0 Vpp separated sync Response Time- <25 ms\nCable kit for Integration- DVI, Fiber Optic, RGBHV S-Videc, Composite\n BI01010001010000000000000515BI0100001120
32 19 UPS System with isolation transformer {10 KVA) AS PER SPECIFICATION ATTACHED BI01010001010000000000000515BI0100001121 2 Each
33 20 MOTORIZED VIEW WINDOW \nThe view window should be of Double insulated fixed glazing with not less than 5mm thick toughened safe glass. flush mounted with wall paneling. It should be aligned with present window and made of suitable size 1000mmx750mm. Maximum dimensions possible should be provided if asked for keeping in mind present civil setup. Motorised blinds should be provided which will be able to fully open and close such that full view from window is available or it can be closed fully. Also it should allow angulation of blinds view outside.AS PER SPECIFICATION ATTACHED.It shall be completed in all respect as per direction of I/E \n\n\n\n\n\n\n BI01010001010000000000000515BI0100001122 1 No.
34 21 AIR HANDLING UNIT (AHU) With AC OUTER UNIT \nSupply of Air handling unit- constructed in EGP sheet and the system is designed to give 30 Air Changes per hour and to create sufficent Positive Pressure, Temprature is 20-22 degree centigrade with double sckin ,0.3 , 5, 10 micron filter , 4000cfm , three phase electrical conection require . with 11 TR coolong coil It should essentially have temperature and humidity control inside OT. \n The AHU for MOT OT may be ceiling suspended or Floor mounted by creating mezzanine floor with necessary arrangements of horizontal structural members, etc. This also involves modifications to the existing AHUs ducting that may foul with the proposed location and it shall be done by the contractors by following similar specifications. The AHU shall be compliant with the latest guidelines in NABH (National Accreditation Board for Hospitals and Healthcare Providers) Standards for super specialty O.T. or equivalent international standards. The AHU shall supply fresh and re-circulated air. The fresh air shall be minimum 20%.The system shall have provision to keep the blower on during non-functional hours, without temperature control. The AHU shall be comply as per EU Medical Device Directive 93/42/EEC.AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of I/E \n BI01010001010000000000000515BI0100001115 2 No.
35 22 O.T SIGNAGE\nProviding and fixing of glowing signsge out side OT\nAS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of I/E BI01010001010000000000000515BI0100001116 2 No.
36 23 CENTRAL CONTROL ROOM (OT INTEGRATION SYSTEM)\nTrolley mounted audio/video recording & transmission system. Shift\nto the OT from when live transmission is to be displayed but could record only one operation at a time not all the 05 OTs.\nCommunication and Data archiving systems shall be placed on Mobile cart, Mobile cart may be moved OT to OT when streaming is required from OT, All the inputs connected with communication system can be streamed from OT, Output (HD-SDI) from HD OT light camera shall be provided on surgical pendant in each OT, Mobile cart may be moved near to surgical pendant in order to connect to Output of OT light camera, Surgical pendants in each OT shall be having Hospital LAN connection (Hospital Scope) Streaming shall be done using Hospital LAN network hence Multicast enabled LAN network with 3 layer switch will be required for uninterrupted streaming (Hospital Scope) Up to 5 computers (Including Auditorium) may be connected at the same time to get streaming from OT Streaming is one way Video and two way Audio.AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of I/E BI01010001010000000000000515BI0100001117 1 No.
37 24 DUCTING FOR IN SIDE O.T and fresh air outside O.T. \naluminum ducting in side O.T 22/24 gause with all accessories .AS PER SPECIFICATION ATTACHED.It shall be completed in all respect as per direction of I/E BI01010001010000000000000515BI0100001118 2 No.
38 25 EXHAUST FRESH AIR UNIT \nhe treated air temperature will be constantly monitored and regulated by a dedicated digital controller. It is also possible to modulate the air fl ow by operating onto the fan speed selector to assure the best agreement between air renewal and noisy.AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of I/E\n BI01010001010000000000000515BI0100001119 2 Lot
39 26 Pre-fabricated Walls & Ceiling Panel\nPre - fabricated modular construction shall be designed and constructed for exact size The prefabricated Operating Room will be free standing structure from Compasite free standing insulated steel wall panels. The wall will be constructed using 1.0 mm SS 304 sheet panel in the front panels & 0.8mm. SS 304 at the back panel The core between two panels will consist of rigid polyurethane foam. which has been injected under high pressure with a minimum density of 40 kg/m3. The individual wall Panels shall use the tongue and groove technology for joining two panels, no welding will be allowed . The gaps between panels be suitably filed with metal filter/epoxy and sanded flush. BI01010001010000000000000515BI0100001120 90 Sqmr.
40 26.1 The external walls of the room shall be constructed with solid bricks and mortar by the hospital as part of its scope of work. The inner surface walls shall be fixed to the brick wall with essential supports. There Natal be minimum possible cavity/gap in between the masonry and steel walls. The total distance between inside and outside surface of OR shall be variable to Suit architects's layout, out shall be sufficient for flush mounting of the equipments. The cavity between inner and outer walls shell be left with minimum obstructions for possible addition of equipment at a later date and to enable services, pipes, conduits etc; to be run within the cavity. All wall mounted equipment shall be flush mounted and sealed into theatre. The wail panel design and construction shall allow tor installation and support of all equipment and provision of openings required for the installations, without affecting rigidity and strength. Access boxes shall be fitted to the rear of all the wail -mounted equipments to enable maintenance to be carried out from outside the operating room. All the sharp edges and corners shall be smoothened to avoid tactrial contamination. BI01010001010000000000000515BI0100001121
41 26.2 Anti Bacterial/ Anti Fungal Paint\nThe internal surfaces of the SS-304 OT Walls should be sprayed with water based, non-reflective liquid plastic, to a color approved be the architect a minimum dry film thickness of 300 micronas, These plastic coatings should overlap the floor covering Ceiling system and-door frames by 25mm to provide a continuous sealed surface. Shade Coating applied should be wafer resistant. does not support bacteriological or fungal growth and is resistant. to most chemicals commonly used in hospital departments. The sterile coating should remains unaffected by radiation and other ionizing radiation at levels in excess of 1000 mred and is classified to class I when tested in accordance with the requirements specified under BS 478. part 7 1971. Surface spread of flame Test for Materials. The coating system should be easily maintained and can withsland-repeated cleaning with alkaline detergents. antiseptics and furnigation agents without any degradation to the surface finish or performance, BI01010001010000000000000515BI0100001114
42 27 Laminar Air Flow Ceiling\nEach modular OT shall have Plan Alr Ceiling. The Plan Air Ceiling shall be constructed out of 2mm each extruded aluminum profiles, overall size 2400 x 2400min or larger as per actual layout, having six no.s HEPA fitters with MPPS efficiency of 99,995% and spot efficiency of 99.99% @ 0.3 micron. The Plenum should provide the required air volume to achieve the temperature . pressure (15-30 Pa) and humidity specified, and terminal velocity 90 to 120 FPM at the diffuser face. Air and light diffuser made of two layer of mono filament precision woven polyester for plan air ceiling to give a laminar flow of filtered air, it shall also provide a shadow less lighting system with control of the intensity of luminance by using high efficiency electronic fluorescent tubes AM HEPA filters shall be factory tested and certified in accordance with international standards. All ceiling snail include integral lighting and composite air/light diffuser.:Air shall be diffused into the theater uniforrnaily over ,,the total area. The leminar flow ceiling should also have illumination across its total area, BI01010001010000000000000515BI0100001115 1 No.
43 28 Automatic Hermetically Sealing Door BI01010001010000000000000515BI0100001116
44 28.1 2100 x 1500mm BI01010001010000000000000515BI0100001117 1 No.
45 28.2 Automatic door shall be provided at the main entry of the OT Doorframes shall be made of aluminum extrusion profile to carry any door weight. The core door size is of 50mm size built of high density PUF (density 40kg/m3) of 52mm and 4mm consist of Solid Stainless Steel (SS-304) on both side of the door. Door blade is made with 3- tier specially designed 3 slide heavy duty replaceable EPDM gasket against wall frame. Bottom seakng with 2-tier heavy duty EPDM gasket to flush with finished floor. Stainless Steel lever handle on both side of the door blade for smooth and easy door opening and closing. The thickness of, aluminum extrusion for door frame & door panel surfaces shall not less than I.5mm There shall not be any visible screws or similar other uneven surfaces on the door panel. A lock shall be provided. Door shall have vision panels (viewing window) at least 300mm x 300mm. Automation of the door has Digital Conroller , Motor and suitable switches to open the door thru hands, foots etc. The door size shall be as per BOO. BI01010001010000000000000515BI0100001118
46 29 Surgeon Control Panel \nThe control panel will be user triendly and operation and maintenance will be easy. The control panel shall be "Tiled Membrane" type: can be configuired to incorporate all the services that operating theatre staff required The fascia shall be made with superior quality UV resistant, sterlisable membrane The membrane backing is made of 25mm thick aluminum that is rust resistance. The connection between the surgeon control and the distribution board will be by a multiplexed system. The panel will contain & services files for\n(a) Time day clock\n(b) Time elapsed clock\n(c) Temperature / humidity indicator\n(d) General Lighting System\n(e) Medical gas alarm panel\n(f) Hand free phone.\n BI01010001010000000000000515BI0100001119 1 No.
47 30 LED TWIN Plate X•ray Viewer\nThe operation tneatre shall be equipped with a X-ray viewing LED Screen, designed to provide a high level of control luminance without flicker. It shall be equipped with loaded dips to secure the X-ray negative when in use. BI01010001010000000000000515BI0100001120 1 No.
48 31 Pressure Relief Dampers\nPressure relief dampers to be provided in theater to prevent cross contamination of air from clean and dirty areas. Suitably sized air pressure relief damper will be strategically placed, enable differential room pressure to be maintained and ensure that wnen doors are opened between clean and dirty area. Counter weight balancing system will be provided in the PRD to maintain positive pressure inside the operation theatre. Air pressure stabilizers will have unique capability of controlling differential pressure to close tolerance. The PRD will remain closed at pressure below the set pressure and will open fully at pressure only fractionally above the threshold pressure. The body will be epoxy pawder coated as per standard BS colors and meets international quality and safety requirements. First class electrolyzed steel plate will be used for body and with high -grade SS 304 Stainless steel for blades. BI01010001010000000000000515BI0100001121 1 No.
49 32 Anti-Static Flooring\nFlooring (Static conductive tiles): The floor finish in the operating room will be imported 2 mm statics conductive PVC tiles, laid on a semi conductive adhesive base The floor finish will terminate at the room perimeter passing over a corcealed cove former and continuing up the wall for 100mm. Flooring will fulfill EN it will have chemical resistance as per EN 423. it will have electrical resistance conforming to EN 1061 AEC 613404-1. All joints will be welded and the plastic wall finish will overlap the floor coving by 25mm , to provide a continuous seated surface. The self leveling compound will be of the type that does not promote bacteriological growth. Copper grounding strips (0.05mm thick, 50mm wide) will be laid flat on the floor in the conductive adhesive and connect to copper wire of grounding BI01010001010000000000000515BI0100001122 40 Sqmr.
50 33 2-Bay Scrub Sink\nCompact Surgical Scrub sink will be designed for use in Operation theatre complex providing surgeons with a convenient sink for pre up scrub up. Each fixture will be fabricated from heavy gauge type 304 stainless steel & will be welded construction polished to a satin finish. The scrub will be provided with a front access panels, which will be easlly removed for access to the water control valve, waste connections stoppers & strainers. Hands free operation will include infrared sensor with built in range of adjustment. Thermostatic Mixing Valve Control will be located behind the access panel & maintain constant water temperature. User defined settings of 1 to 3 min are available. This timing will be adjustable to meet individual application requirement, provided with infrared sensor thermostatic controlled laps with fail-safe Temperature controls BI01010001010000000000000515BI0100001115 1 No.
51 34 Peripheral lights (LED)\nEight sets of double peripheral lights shall be with stepped finish aluminum with reflectors. Peripheral lights shall be clean room compatible. Rating of peripheral lighting small be iP 54, Control equipment for the peripheral lighting shall be provied in the theatre control panel to allow independent control at the lighting level by the surgical team Recessed mounted. BI01010001010000000000000515BI0100001116 8 No.
52 35 Surgeon's Double Arm Pendant\nPendant Should be Double arm (1000mm + 800mm) with load carrying capacity of 80 Kg. The arm should rotated upto 330°-340° with adjustable stopper. Dampened stoppers and very low hand forces for horizontal movements should contribute to an ergonomically optimized workplace. The large interior cross section for supply lines offers should completely new applications with 120mm diameter. The stoppers Should be infinitely variable from 0-330°- 340' BI01010001010000000000000515BI0100001117 1 No.
53 35.1 The service head should be of modular design in achieve maximum location efficiency with minimum required space. Service head should designed to host. Base , Gas Module (Up to 8 Gas outlets), Electrical Module (10 Electrical outlets) and two shelves and one Infusion pump rack.,\na. The total length of the manager should be 800mm Surgeon pendant\nb. should have 2 arms with shelves as per following details'\nc. Horizontal arm system- 2\nd. Weight carrying capacity- 80 Kg\ne. 5/15 Amp. Electrical sockets without switches- 10 Nos\nf. Shelves with slide ralls- 2 Nos.\ng. Provision to fix Gas Outlets (i.e.)Oxygen-2, Vacuum-2 Air 4 Bar-1,\nh. Air 7 Bar-1, CO2.1 AGSS-1\ni. Ceiling mounting system up to 1000mm-1\nj. Ceiling cover for false ceiling-1\n BI01010001010000000000000515BI0100001118
54 36 Anaesthetist's Single Arm Pendant\nPendant should be Single arm (900mm) with load carrying capacity of 80 Kg. Dampened stoppers and very low hand forces for horizontal movements should contribute to an ergonomically optimized workplace. The interior cross section avaliable for supply lines shall be minimum 120mm diameter, The stoppers should be infinitely variable from 0-330'--340°. BI01010001010000000000000515BI0100001119 1 No.
55 36.1 The service head should be provided with the modular design to achieve maximum supply with minimum required space Service head should be designed to host . Gas Module. Electric Module and\nshelves, Up to 8 Gas outlets & 10 Electrical Outlets. Racks & shelves should be provided to mount the equipments like monitor etc The total length of the manager should be 800mm. Anaesthesia pendent\nshould be have 1 arms with shelvs as per following details.\na. Horizontal arm system-1\nb. Weight carrying capacity. 50 Kg\nc. 5/15 Amp. Electrical sockets without switches 10 Nos,\nd. Shelves with slide rails- 2 Nos.\ne. Provision to fix Gas Outlets (8 Nos.) Oxygen-2, Vaccum-2, Air 4 Bar-\nf. 1, N20- 1 , AGES-1. Spare-1\ng. Ceiling mounting system up to 1000mm-1\nh. Ceiling cover for false ceiling- 1\n BI01010001010000000000000515BI0100001120
56 37 Electrical Work inside the OT\nAir high voltage equipment is installed in a separate enclosure The remote cabinet houses the operating lamp transformers main failure relays electrical for all circuit within the operating theatre. All internal wiring is terminated in connectors with screw and clamp & spring connections of the clip on type mounted, on a DIN rail labeled with indelible proprietary labels. Individual miniature circuit breakers shall be provided to protect all internal circuits with 5/16 amp switch Sockets for the Modular Operation Theatre MCB for Circuits for critical equipments and outlets shall be clearly marked for UPS connection-OPERATING LIGHT. MONITOR, OT PANEL. OT GENERAL LIGHT and other critical equipments (4 out of 10 on the pendant)\n BI01010001010000000000000515BI0100001121 1 Job
57 38 Medical Gas Pipeline work Inside the OT\nThe Medical Gas Pipeline inside the OT shall provide Valve Box for 5 gases (02. CO2, N20, Vacuum, Surgical Air-7 Bar) outside each OT and shall bnng the Medical Gas Pipeline inside the OT till ceiling pendants and provide NIST connections for each gas pipeline as per EN 737/HTM standards or equivalent. Pendants shall connect flexible hose piping with Medical Gas Piping by means of NIST connections. The gas piping used for transportation should be as EN 13348/EN 1057 standards with copper fittings as per EN 1254 standards.The outlets fitted for all the 5 gases in the pendants shall in conformity of HTM standards. Other Accessories of Medical Gas Pipeline like Ward Vacuum Unit having 600ml autoclave able at 121 degree centigrade collection jar and Theatre Suction Trolley having 2000ml (2 Nos) autoclave able at 121 degree centigrade to be provided in each OT BI01010001010000000000000515BI0100001114 1 Job
58 39 AIR HANDLING UNIT (AHU) With AC OUTER UNIT \nSupply of Air handling unit- constructed in EGP sheet and the system is designed to give 30 Air Changes per hour and to create sufficent Positive Pressure, Temprature is 20-22 degree centigrade with double sckin ,0.3 , 5, 10 micron filter , 4000cfm , three phase electrical conection require . with 11 TR coolong coil It should essentially have temperature and humidity control inside OT. \n The AHU for MOT OT may be ceiling suspended or Floor mounted by creating mezzanine floor with necessary arrangements of horizontal structural members, etc. This also involves modifications to the existing AHUs ducting that may foul with the proposed location and it shall be done by the contractors by following similar specifications. The AHU shall be compliant with the latest guidelines in NABH (National Accreditation Board for Hospitals and Healthcare Providers) Standards for super specialty O.T. or equivalent international standards. The AHU shall supply fresh and re-circulated air. The fresh air shall be minimum 20%.The system shall have provision to keep the blower on during non-functional hours, without temperature control. The AHU shall be comply as per EU Medical Device Directive 93/42/EEC.AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of I/E \nMake:-Waves/Edgetech/Zeco BI01010001010000000000000515BI0100001115 1 No.
59 40 DUCTING FOR IN SIDE O.T and fresh air outside O.T. \naluminum ducting in side O.T 22/24 gause with all accessories .AS PER SPECIFICATION ATTACHED.It shall be completed in all respect as per direction of I/E\n BI01010001010000000000000515BI0100001116 1 No.
60 41 Pre-fabricated Walls & Ceiling Panel\nPre - fabricated modular construction shall be designed and constructed for exact size The prefabricated Operating Room will be free standing structure from Compasite free standing insulated steel wall panels. The wall will be constructed using 1.0 mm SS 304 sheet panel in the front panels & 0.8mm. SS 304 at the back panel The core between two panels will consist of rigid polyurethane foam. which has been injected under high pressure with a minimum density of 40 kg/m3. The individual wall Panels shall use the tongue and groove technology for joining two panels, no welding will be allowed . The gaps between panels be suitably filed with metal filter/epoxy and sanded flush. BI01010001010000000000000515BI0100001117 920 Sqmr.
61 41.1 The external walls of the room shall be constructed with solid bricks and mortar by the hospital as part of its scope of work. The inner surface walls shall be fixed to the brick wall with essential supports. There Natal be minimum possible cavity/gap in between the masonry and steel walls. The total distance between inside and outside surface of OR shall be variable to Suit architects's layout, out shall be sufficient for flush mounting of the equipments. The cavity between inner and outer walls shell be left with minimum obstructions for possible addition of equipment at a later date and to enable services, pipes, conduits etc; to be run within the cavity. All wall mounted equipment shall be flush mounted and sealed into theatre. The wail panel design and construction shall allow tor installation and support of all equipment and provision of openings required for the installations, without affecting rigidity and strength. Access boxes shall be fitted to the rear of all the wail -mounted equipments to enable maintenance to be carried out from outside the operating room. All the sharp edges and corners shall be smoothened to avoid tactrial contamination. BI01010001010000000000000515BI0100001118
62 41.2 Anti Bacterial/ Anti Fungal Paint\nThe internal surfaces of the SS-304 OT Walls should be sprayed with water based, non-reflective liquid plastic, to a color approved be the architect a minimum dry film thickness of 300 micronas, These plastic coatings should overlap the floor covering Ceiling system and-door frames by 25mm to provide a continuous sealed surface. Shade Coating applied should be wafer resistant. does not support bacteriological or fungal growth and is resistant. to most chemicals commonly used in hospital departments. The sterile coating should remains unaffected by radiation and other ionizing radiation at levels in excess of 1000 mred and is classified to class I when tested in accordance with the requirements specified under BS 478. part 7 1971. Surface spread of flame Test for Materials. The coating system should be easily maintained and can withsland-repeated cleaning with alkaline detergents. antiseptics and furnigation agents without any degradation to the surface finish or performance. BI01010001010000000000000515BI0100001119
63 42 Anti-Static Flooring\nFlooring (Static conductive tiles): The floor finish in the operating room will be imported 2 mm statics conductive PVC tiles, laid on a semi conductive adhesive base The floor finish will terminate at the room perimeter passing over a corcealed cove former and continuing up the wall for 100mm. Flooring will fulfill EN it will have chemical resistance as per EN 423. it will have electrical resistance conforming to EN 1061 AEC 613404-1. All joints will be welded and the plastic wall finish will overlap the floor coving by 25mm , to provide a continuous seated surface. The self leveling compound will be of the type that does not promote bacteriological growth. Copper grounding strips (0.05mm thick, 50mm wide) will be laid flat on the floor in the conductive adhesive and connect to copper wire of grounding. BI01010001010000000000000515BI0100001120 450 Sqmr.
64 43 HVAC (AHU WITH OUTER UNIT) IN ICU & CORRIDORS\nProviding, fixing, supply, Installation commissioning of air handling unit AHU-(DX UNIT) constructed in EGP SHEET and The system is designed to give 15 TO 20 Air Changes Per Hour and to create sufficient Positive Pressure, Temp. range is 20-22 degree cent COIL CAPACITY 8TR 2no each AHU with fresh air system , centigrade with double sckin ,0.3, 5 to 10 micron filter ,8000cfm , with 16 TR. three phase electrical conection require . AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of E/I . BI01010001010000000000000515BI0100001121 5 No.
65 44 DUCTING IN SIDE & OUTSIDE IN ICU & CORRIDORS\nSupply of The internal ducting as per ISI-655 duly fabricated out of22/24 swg EGP sheet complete with flanges and accessories such as GI suspenders and GI supportscompletely sealed with Silicon sealant duly insulated with Aluminum foil Nitride rubber self adhesive typeinsulation.AS PER SPECIFICATION ATTACHED.It shall be completed in all respect as per direction of E/I BI01010001010000000000000515BI0100001122 5 LOT
66 45 EXHAUST AIR GRILL IN ICU & CORRIDORS\nThe openable and cleanable return-air exhaust GRILL should be provided in the ICU AREA . Designed flow rate should not be less than 1000 m3/hr. Distribution of exhaust air .AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of E/I BI01010001010000000000000515BI0100001115 48 No.
67 46 ELECTRICAL WIRING, INSIDE ICU & CORRIDORS\n Laying of PVC conduits, Modular Switches-sockets, Power and Light wiring including Earthing wire for all the lighting controls, Pendant and other equipment fixtures and fittings inside the ICU Wiring with low leakage current wires of FRLS wires as per requirements. Wiring for 250 volts single phase and neutral 6/16 Amps switched socket outlet with 3 sq.mm and 2.5 sq.mm PVC insulated copper conduct with cat6 cable for networking.AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of E/I \n BI01010001010000000000000515BI0100001116 4 LOT
68 47 GENERAL LIGHT WITH DEMING IN ICU & CORRIDORS\nThe fluorescent lamps 54 W 16mm Ø- 1 nos / Non-hygroscopic high glow low power LED based /digital blast peripheral lights (1’x2’) having high quality low wattage LED lighting system with highly spectacular anodized Aluminum reflectors and optical antiglare system for adjustable light distribution. Luminaire cover made of highly resistant, disinfectant proof laminated safety glass with fine grained surface, glass pane with white powder coated steel frame. Luminaire body made of sheet steel, white, powder coated supplied ready for connection. AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of E/I BI01010001010000000000000515BI0100001117 95 No.
69 48 HORIZONTAL WALL MOUNTED BED HEAD PANEL in SNCU \nThe modular medical Bed head panel should be fabricated of Aluminous die pressed in such a way to provide 100% safe and efficient means to deliver all the three gases as well as electrical to the patient/staff in intensive care conditions. To enclosed all piping, electrical wires, etc it should be fitted continuously on wall The design of modular should consist of aluminium section to meet the equirement of HTM 0201, NFPA and other international standard.There should be no sharp edges having smooth finish and without any visible screws etc.It should be have ability fix medical gas outlets, electrical sockets -. nurse’s call bell, IV fluid’s saline & Data socket. It should also have color code around the specific gas outlet and a slide rail for focus lamp. It should be polyester powder coating finish on all external surfaces and should be fitted with wall with the help of 8mm fasters at specific distance.1500mm Long Bedhead Panel It should be fixed in such a manner that it could be remove easily for servicing and\nrepair. It should be long and continuous to fix for any distance.\nProvision for accessories per bed: -\nOxygen outlet 2\nVacuum outlet 2\nCompressed air outlet 1\n Bracket for vacuum jar 1\nFocus light 1- optional\n15 Amp socket 2\n5 Amp socket 3\nIV saline 1- optional\nAS PER SPECIFICATION ATTACHED.It shall be completed in all respect as per direction of E/I BI01010001010000000000000515BI0100001118 12 Nos.
70 49 Intravenous Track I.V Track In ICU \nIt should be made from heavy Duty Aluminum Intravenous track 'U' configuration size : 35mm (W) x 19.2mm(H) make of aluminum natural anodized to 15 microns complete with wall brackets with matching screws all strictly in accordance with the manufacturer's instruction 1 no. IV carrier to a set of IV support track complete with 5 points Telescopic Bottle Holder adjustable .AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of E/I BI01010001010000000000000515BI0100001119 30 Nos.
71 50 Curtain Track partition In ICU \nCubicle Curtain Track : It should be made from heavy Duty Aluminium cubicle track size 20mm wide x 30mm high, made of aluminium natural anodized to 15 microns complete with continuous PVC liner, nylon gliders and hooks, plastic end cap, connecting bridge, overlapping joint connector, wall brackets with matching screws to make up cubicle height to 2100mm clearance from floor level at 1000mm spacing and securely fixed to above slab.AS PER SPECIFICATION ATTACHED ).It shall be completed in all respect as per direction of E/I BI01010001010000000000000515BI0100001120 30 Nos.
72 51 DOORS AND FRAMES (HERMETICALLY SEALED) Double Leaf, Size (1800mmx2100mm) SENSOR OPERATED \nAutomatic with manual over-ride sliding doors having durability, ease of control, versatility for clean environments.To maintain sterility and the correct air pressure in the room, all doors into and out of OT should be of this similar type. Automation with 2 sensors - foot operation and hand sensors (magic switch) Opening, closing and stay-open times should be programmable. Electrical motor drive should offer various control types such as energy saving feature (partial opening adjustable for distance and time); automatic closing (full opening, adjustable for time); sluice function and permanent opening. Lead protection on glass cutout is required. The door and control should comply current IEE regulations and BS 7971 standard. Double leaf Hermetically sealed Sliding Automatic Door shall be with Vision Panels 300 mm x 300 mm with double glazed panels with size of 1800mmx2100mm.AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of E/I BI01010001010000000000000515BI0100001121 4 Nos.
73 52 Control Panel in ICU\nthe control panel will be user triendly and operation and maintenance will be easy. The control panel shall be "Tiled Membrane" type: can be configuired to incorporate all the services that operating theatre staff required The fascia shall be made with superior quality UV resistant, sterlisable membrane The membrane backing is made of 25mm thick aluminum that is rust resistance. The connection between the surgeon control and the distribution board will be by a multiplexed system. The panel will contain & services files for.\n(a) Time day clock\n(b) Time elapsed clock\n(c) Temperature / humidity indicator\n(d) General Lighting System\n(e) Medical gas alarm panel\n(f) Hand free phone.\n BI01010001010000000000000515BI0100001114 4 No.
74 53 Distribution Board In ICU & Corridors\nElectrical Distribution Board will have all high voltage equipment should be installed in a separate enclosure. The remote cabinet should house the operating lamp transformers, mains failure relays, electrical distribution equipment and circuit protection equipment for all circuits within the operating theatre. All internal wiring should terminate in connectors with screw and clamp spring connections of the Clip-on type mounted, on a DIN rail. Individual fuses or miniature circuit breakers should protect all internal circuits. Should have CE marking. CE certificate must be submitted.AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of E/I BI01010001010000000000000515BI0100001115 5 Nos.
75 54 ICU PENDANT\nProviding, fixing, testing and commissioning of The pendant should be high degree of flexibility due to expandability in terms of equipment options. The pneumatic break should be used to secure system and prevent any accidental movement of the ceiling pendant. Variable placement of shelves to ensure adaptability of various requirements in OT Integration ICU equipment with built-in life system Double arm pendant should be installed depending on the given conditions and the individual requirements in the operating theatre. A ceiling supply system with two-part arm can not only move along a circular curve and rotate on its own axis, but may also be flexibly positioned at any point of a circular area. Configuration Rated for maximum 150 kg load,as per specifications CE/FDA/UL.AS PER SPECIFICATION ATTACHED .It shall be completed in all respect as per direction of E/I BI01010001010000000000000515BI0100001116 18 Nos.
76 55 VIEW WINDOW with Motorised Blinds In ICU \nThe view window should be of Double insulated fixed glazing with not less than 5mm thick toughened safe glass. flush mounted with wall paneling. It should be aligned with present window and made of suitable size 1000mmx750mm. Maximum dimensions possible should be provided if asked for keeping in mind present civil setup. Motorised blinds should be provided which will be able to fully open and close such that full view from window is available or it can be closed fully. Also it should allow angulation of blinds view outside.AS PER SPECIFICATION ATTACHED. It shall be completed in all respect as per direction of E/I BI01010001010000000000000515BI0100001117 6 Nos.