The Intra Vascular Ultra Sound Catheters With Mechanical Array Having Minimum Frequency Of 40 Mhz. Should Have Axial Resolution Of 38 Microns And Compatible With 5F Guide Catheter. The Entry Profile Should Be 2F With 0.014 Guide Wire Compatibility. The Ti-Tamil Nadu

Southern Railway-SR has published The Intra Vascular Ultra Sound Catheters With Mechanical Array Having Minimum Frequency Of 40 Mhz. Should Have Axial Resolution Of 38 Microns And Compatible With 5F Guide Catheter. The Entry Profile Should Be 2F With 0.014 Guide Wire Compatibility. The Ti. Submission Date for this Tender is 20-01-2020. Medical Diagnostic x ray Products Tenders in Tamil Nadu. Bidders can get complete Tender details and download the document.




Tender Notice

23076013
The Intra Vascular Ultra Sound Catheters With Mechanical Array Having Minimum Frequency Of 40 Mhz. Should Have Axial Resolution Of 38 Microns And Compatible With 5F Guide Catheter. The Entry Profile Should Be 2F With 0.014 Guide Wire Compatibility. The Ti
Open Tender
Indian
Tamil Nadu
20-01-2020

Tender Details

Supply Of The Intra Vascular Ultra Sound Catheters With Mechanical Array Having Minimum Frequency Of 40 Mhz. Should Have Axial Resolution Of 38 Microns And Compatible With 5F Guide Catheter. The Entry Profile Should Be 2F With 0.014 Guide Wire Compatibility. The Tip To Transducer Length Should Be 20Mm And Monorail Length Should Be 16Mm. Should Have Markers On The Catheter For Measuring Pullback Length. Should Be Compatible With Motor Drive System. Unit:No, The Intra Vascular Ultra Sound Catheters With Mechanical Array Having Minimum Frequency Of 40 Mhz. Should Have Axial Resolution Of 38 Microns And Compatible With 5F Guide Catheter. The Entry Profile Should Be 2F With 0.014 Guide Wire Compatibility. The Tip To Transducer Length Should Be 20Mm And Monorail Length Should Be 16Mm. Should Have Markers On The Catheter For Measuring Pullback Length. Should Be Compatible With Motor Drive System. Unit:No

Key Value

Document Fees
Refer document
EMD
Refer document
Tender Value
Refer document

BOQ Items

0 1 2 3 4 5 6 7 8 9
0 S.No. PL Code\r(Group) Item Type\rGST(Y/N) Stock /\rNonStock Ordering Consider\rFor Eval Approving\rAgency Inspection\rAgency Currency\rAllowed Estimated\rRate
1 1 161104 Goods (Y) Non Stock --- Yes CONSIGNEE INR
2 Description :THE INTRA VASCULAR ULTRA SOUND CATHETERS WITH MECHANICAL ARRAY HAVING\rMINIMUM FREQUENCY OF 40 MHz. SHOULD HAVE AXIAL RESOLUTION OF 38 MICRONS AND\rCOMPATIBLE WITH 5F GUIDE CATHETER. THE ENTRY PROFILE SHOULD BE 2F WITH 0.014 GUIDE\rWIRE COMPATIBILITY. THE TIP TO TRANSDUCER LENGTH SHOULD BE 20MM AND MONORAIL\rLENGTH SHOULD BE 16MM. SHOULD HAVE MARKERS ON THE CATHETER FOR MEASURING\rPULLBACK LENGTH. SHOULD BE COMPATIBLE WITH MOTOR DRIVE SYSTEM. Unit:No
3 Consignee SR DMO/S/RH/PERC Tamil Nadu 4.00 Numbers