Supply And Set Up Of Lyophilizer ( Year 2019-20 ), Meerut-Uttar Pradesh

Chaudhary Charan Singh University has published Supply And Set Up Of Lyophilizer ( Year 2019-20 ). Submission Date for this Tender is 08-01-2020. Laboratory Freeze Dryers and Lyopholizers Tenders in Meerut Uttar Pradesh. Bidders can get complete Tender details and download the document.




Tender Notice

22946762
Supply And Set Up Of Lyophilizer ( Year 2019-20 )
Open Tender
Indian
Uttar Pradesh
Meerut
08-01-2020

Tender Details

Supply And Set Up Of Lyophilizer ( Year 2019-20 ) -Lyophilizer (Freeze Dryer) (-840C)

Key Value

Document Fees
INR 1050 /-
EMD
INR 100000.0 /-
Tender Value
INR 15 Lakhs /-

BOQ Items

Name of Work: for Supply and Set Up of Lyophilizer (Year 2019-20)
Sl. No. Item Description Item Code / Make Quantity Units
1 1 Lyophilizer (Freeze Dryer) (-840C) BI01010001010000000000000515BI0100001112 1 Set
2 1.01 ·             Upright stainless steel collector coil capable of holding 2.5L of ice before defrosting. Stainless steel collector lid [Two 1/3 hp HCFC/CFC-free refrigeration systems cool the collector to -84°C(-119°F)], 3/4" thick, has neoprene gasket and 3" diameter port for connection of drying accessories. BI01010001010000000000000515BI0100001113
3 1.02 ·             Compact benchtop design with a small footprint. BI01010001010000000000000515BI0100001114
4 1.03 ·             Brushed stainless steel and glacier white, powder-coated steel exterior with blue accents. BI01010001010000000000000515BI0100001115
5 1.04 ·             Full color, 5" capacitive touch screen with OS: BI01010001010000000000000515BI0100001116
6 1.05 ·             Real time display of collector temperature, vacuum level, and End-Zone end point detection system. BI01010001010000000000000515BI0100001117
7 1.06 ·             Start–up mode for collector cool-down and vacuum pull-down (automatic or manual). BI01010001010000000000000515BI0100001118
8 1.07 ·             Operating mode (standby or run). BI01010001010000000000000515BI0100001119
9 1.08 ·             Vacuum set point, alerts, diagnostic tests, user manual, defrost control, and optional drying accessory parameters. BI01010001010000000000000515BI0100001120
10 1.09 ·             Data logging stores and displays data in table or chart format, and may be exported via USB or Ethernet. BI01010001010000000000000515BI0100001121
11 1.1 ·             Settings include language (English, Spanish, French, German, or Italian), time, time zone, temperature (°C or °F), and vacuum (mBar, Torr, or Pa). BI01010001010000000000000515BI0100001122
12 1.11 ·             Alerts are programmable to send via Ethernet connection to user-entered email addresses. BI01010001010000000000000515BI0100001123
13 1.12 ·             Display settings include brightness, touch sounds (on/off), and screen time out in minutes (on/off). BI01010001010000000000000515BI0100001124
14 1.13 ·             Maintenance alert settings include vacuum pump oil change, refrigeration cleaning, total operating hours and line voltage offset. BI01010001010000000000000515BI0100001125
15 1.14 ·             Level-specific security lock-out prevents changes to programs and other system parameters. BI01010001010000000000000515BI0100001126
16 1.15 ·             Hot gas defrost with auto shut off at +65°C (+149°F). BI01010001010000000000000515BI0100001127
17 1.16 ·             On/standby switch. BI01010001010000000000000515BI0100001128
18 1.17 ·             Moisture sensor that prevents refrigeration or vacuum start-up when moisture is detected in the drain line. BI01010001010000000000000515BI0100001129
19 1.18 ·             Vacuum control valve that maintains vacuum level set point. BI01010001010000000000000515BI0100001130
20 1.19 ·             Vacuum break valve that bleeds air into the system when power to the freeze dryer or vacuum pump is shut off. If a power outage less than approximately 5 minutes occurs, the freeze dryer will restart and the refrigeration and vacuum system will resume operation once power is restored. If the power failure is more than approximately 5 minutes and the collector warms above safe limits, the freeze dryer will not automatically restart. BI01010001010000000000000515BI0100001131
21 1.2 ·             Side-mounted USB port and rear-mounted Ethernet connection. BI01010001010000000000000515BI0100001132
22 1.21 ·             Side-mounted, quick-disconnect drain hose fitting and lanyard with hose connector. 18", 1/4" ID silicone tubing and clamp included. BI01010001010000000000000515BI0100001133
23 1.22 ·             Rear-mounted electrical receptacle for vacuum pump connection. BI01010001010000000000000515BI0100001134
24 1.23 ·             3-wire electrical cord with plug. BI01010001010000000000000515BI0100001135
25 1.24 ·             3/4" OD vacuum connection, 50" of 3/4" ID vacuum hose and two clamps. BI01010001010000000000000515BI0100001136
26 1.25 PTFE-coated collector coil and chamber for processing involving corrosive compounds. BI01010001010000000000000515BI0100001137
27 1.26 ·             Region-specific electrical plug configurations. BI01010001010000000000000515BI0100001138
28 1.27 Vacuum pump with a displacement of at least 98 L/min, 0.002 mBar ultimate pressure and fitting suitable for 3/4" ID vacuum hose., Oil mist filter BI01010001010000000000000515BI0100001139
29 1.28 ·             Small Tray Dryer and other drying accessories (Ampules and Flasks with adaptors) BI01010001010000000000000515BI0100001140
30 1.29 ·             All Freeze drying glass wares with different capacity (40ml, 100ml). BI01010001010000000000000515BI0100001141