Reagent Strip For Checking Capillary Blood Glucose Technical Specification: Strip With Reagent Area For Checking Capillary Blood Glucose From Newborns To Adults, With Enzymatic Chemistry Of Glucose Dehydrogenase (Gdh-Fad Or Gdh-Nad) And Reading Method Through Amperometry, Without Allowing A Blood Sample In Contact With The Device, Accumulation Of Residues And Infections Crusades. The Reagent Area Of The Strip Must Be External To The Monitor. Measuring Range Should Be Between 10Mg/Dl To 600Mag/Di. Values Lower Than 10Mg/Dl And Higher Than 600Mg/Dl Are Accepted. Hematocrit Range From 15 To 65%. The Glucometers Must Have A Strip Ejector Button, In Compliance With Nr32 And Results Only In Mg/Dl. The Expiration Date Of The Strips Must Be Equal To The Expiration Date Described On The Label. The Winning Company Of This Item, Must Provide The Glucometer, Protective Bag, Usb Cable, Manuals In Portuguese And Stack(S) Or Battery(S), Software With Anvisa Registration On Lending Regime, In Quantity As Requested In The Terms Of Reference. Box With 50 Units.