The Agilent 1260 Infinity Ii Vialsampler Is An Hplc Instrument Module That Automatically Dispenses And Introduces Samples Into The Chromatograph. Вес 19 Кг. Dimensions (H × W × D) 320×396×468 Mm. Supply Voltage 100–240 V~, ±10%. Network Frequency Range 50-60 Hz, ±5%. Power Consumption 350 W. Operating Temperature +4 – +40 °C, Without Sample Thermostat Up To +55 °C. Storage Temperature (Non-Operating State) - 40 – +70 °C. Humidity <95% Rh Humidity At +40 °C, Non-Condensing. Working Altitude Up To 3000 M. Permitted Solvents: Boiling Point ≥ +56 °C. Auto-Ignition Temperature ≥ +200 °C. Safety Standards Iec, En, Csa, Ul: Overvoltage Category Ii Pollution Degree 2. Delivery, Installation, Commissioning, Installation Qualification And Operation Qualification Of The Autosampler, Training Of Quality Control Laboratory Personnel To Work With The Autosampler. The Autosampler Must Undergo Metrological Certification And Be Registered In The State Survey Register. Autosampler With A 12-Month Warranty From The Date Of Commissioning. The Autosampler Must Be Compatible With The Agilent 1260 Infinity Ii Hplc Instrument And Gina X Version 10.4 Software.